Published clinical literature (Compend Contin Educ Dent. 2008)
The sponsor used a published clinical study of 100 patients treated with the predicate device to establish the clinical timing (8 weeks) for implant placement in sockets grafted with the proposed device.
The healing socket and socket regeneration (Compend Contin Educ Dent. 2008); Retrospective clinical cohort; Follow-up/Duration: 3 years
Patients undergoing dental extraction and socket grafting; Sample Size: 100
Not applicable for this study
Implant integration and function
Indications for Use
SOCKET GRAFT is a bone grafting material indicated for use in dental extraction sockets that have all walls remaining. Socket Seal is a temporary implantable material intended for use in the oral cavity to be used as a space maintaining barrier over bone and other tissue. Socket Seal is indicated for use over bone grafts. Socket Seal in indicated for use over extraction sockets in the maxilla and mandible.
Device Story
Socket Graft is a dual-phase calcium phosphate biocement kit containing Socket Seal barrier membrane. Clinician wets powder with sterile water to form putty; fills dental extraction sockets with all walls remaining; covers with Socket Seal. Socket Seal is a nonporous, semi-rigid polyethylene foam sheet. Device stabilizes bone graft, provides space maintenance for regenerative healing. Socket Graft resorbs in 12-14 weeks. Used in dental clinics by dentists/oral surgeons. Output is physical bone regeneration scaffold. Benefits patient by facilitating bone growth in extraction sites, enabling subsequent implant placement.
Clinical Evidence
Bench testing only. Comparative analysis provided for trace impurities (USP 34), particle presence, Ca/P ratio, pH, volumetric porosity, and solubility. No clinical trials performed; resorption timing and clinical efficacy supported by literature review of predicate device (K052493) performance in 100-patient study.
Technological Characteristics
Socket Graft: Dual-phase calcium phosphate biocement with sodium carboxymethylcellulose (CMC). Socket Seal: Nonporous, closed-cell polyethylene foam. Sterilization: Gamma radiation. Form factor: 1.5 cc single-use syringes (graft) and semi-rigid sheets (seal).
Indications for Use
Indicated for dental extraction sockets with all walls remaining (Socket Graft) and as a space-maintaining barrier over bone grafts or extraction sockets in the maxilla and mandible (Socket Seal).
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
510(k) Summary
K113049
# SUMMARY REPORT
JAN 2 7 2012
SUBMITTED BY
Gregory Gene Steiner CEO STEINER LABORATORIES Kapolei, HI 96707 808 689 1710
01/15/2012
This summary of 510k substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807.92. Socket Graft is intended to be marketed as a kit containing Socket Graft bone grafting material and a barrier membrane titled Socket Seal.
#### CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME for Socket Graft
| Classification Name: | Bone grafting material, synthetic |
|-------------------------|-----------------------------------|
| Common/Usual Name: | Bone grafting material |
| Product Classification: | LYC |
| Proprietary Name: | SOCKET GRAFT |
## CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME for Socket Seal
| Classification Name: | Barrier, Synthetic, Oral |
|-------------------------|--------------------------|
| Regulation description: | Bone grafting material |
| Common/Usual Name: | Barrier, Synthetic, Oral |
| Product Classification: | NPK |
| Proprietary Name: | SOCKET SEAL |
#### SOCKET GRAFT grafting material is substantially equivalent to:
- 1) Socket Graft K052493
- 2) EquivaBone Osteoinductive Bone Graft K090855
- 3) CalMatrix Calcium Sulfate Bone Graft K041324
#### SOCKET SEAL is substantially equivalent to:
- 1) IMTEC BioBarrier Membrane K974752 2) Tefgen-LS K965205
{2}------------------------------------------------
#### DEVICE DESCRIPTION
SOCKET GRAFT is intended to regenerate bone in dental extraction sockets. SOCKET GRAFT is dual phase calcium phosphate biocement that is wetted by sterile water to a putty consistency. Extraction sockets are filled with SOCKET GRAFT and covered by the enclosed Socket Seal. SOCKET GRAFT is fully resorbed in 12-14 weeks resulting the extraction socket. SOCKET GRAFT is supplied in sterile, single use 1.5 cc syringes.
Socket Seal is a closed cell, medical grade, polyethylene foam material that is a nonporous, biocompatible semi-rigid sheet. Socket Seal is designed to stabilize, support and protect bone graft material and provide space maintenance for regenerative healing.
#### INDICATIONS FOR USE
SOCKET GRAFT is a bone grafting material indicated for use in dental extraction sockets that have all walls remaining.
Socket Seal is a temporary implantable material intended for use in the oral cavity to be used as a space maintaining barrier over bone and other tissue. Socket Seal is indicated for use over bone grafts. Socket Seal in indicated for use over extraction sockets in the maxilla and mandible.
#### PERFORMANCE TESTING:
In comparison to the predicate device Socket Graft K052493comparison summary: Non clinical testing.
| | Socket Graft predicate K052493 | Socket Graft K113049 |
|---------------------------------------------------------------------------------------|--------------------------------|----------------------|
| Trace impurities analysis<br>Reference US pharmacopeia 34<br>specification NMT 20 PPM | pass | pass |
| Particles | none | none |
| Ca/P ratio | equivalent | equivalent |
| Ph | 7.2 | 7.4 |
| volumetric<br>porosity | 25% | non porous |
| Solubility @ 20C | non soluble | non soluble |
{3}------------------------------------------------
The proposed device Socket Graft K113049 differs in composition only by the addition of sodium salt of carboxymethylcellulose.
The sodium salt of carboxymethylcellulose used in Socket Graft conforms to the GRAS classification 21 CFR 182.1745 with the following analysis:
| reference | test results |
|----------------------------------|------------------|
| 99.5% by dry weight | 99.8 |
| max substitution 0.95 | 0.77 |
| minimum viscosity 25 centipoises | 3100 centipoises |
Predicate devices EquivaBone Osteoinductive Bone Graft and CalMatrix Calcium Sulfate Bone Graft both contain comparable amounts of carboxymethylcellulose and both of these predicate devices have used carboxymethylcellulose to increase the viscosity of their graft material in the same manner as the proposed device Socket Graft K113049 . Carboxymethylcellulose has been shown to resorb in 12 weeks in vivo in the following reference "Carboxymethylcellulose-stabilized collagenous rhOP-1 device-a novel carrier biomaterial for the repair of mandibular continuity defects. Wang H et al".
The predicate device Socket Graft K052493 was shown to fully resorb in 12-14 weeks. The timing of implant placement for Socket Graft K113049 is based on published research for the predicate device Socket Graft K052493. In the published research article "The healing socket and socket regeneration. Compend Contin Educ Dent. 2008 Mar;29(2):114-6, 118, 120-4". One Hundred implants were placed on a protocol of 6 weeks for incisors and bicuspids and 8 weeks for molars. All one hundreds implants integrated and all implants were in function 3 years after placement. Therefore, the earliest time point for implant placement in sockets grafted with Socket Graft K113049 is 8 weeks.
#### Substantial Equivalence:
Socket graft grafting material and Socket Seal shares indications and design principles with the following predicate devices:
| Device name | Proposed Device | Predicate device #1 | Predicate Device #2 | Predicate Device #3 |
|--------------|---------------------------------------------|---------------------------------------------|--------------------------------------------|--------------------------------------------------|
| | Socket Graft | Socket Graft | EquivaBone<br>Osteoinductive Bone<br>Graft | CalMatrix Calcium<br>Sulfate Bone Graft |
| 510 K number | K113049 | K052493 | K090855 | K041324 |
| Intended use | SOCKET GRAFT is a<br>bone grafting material | SOCKET GRAFT is a<br>bone grafting material | EquivaBone is an<br>osteoinductive bone | used to treat multiple<br>types of maxillary and |
{4}------------------------------------------------
| | indicated for use in<br>dental extraction<br>sockets that have all<br>walls remaining. | indicated for use in<br>dental extraction<br>sockets. | graft substitute that<br>is resorbed and<br>replaced with new<br>bone during the healing<br>process. It is intended<br>for use to fill bony<br>voids or gaps of the<br>skeletal system of the<br>extremities, spine<br>(i.e.posterolateral<br>spine) and pelvis that<br>are not intrinsic to the<br>stability of the bony<br>structure. These voids<br>or gaps may result from<br>natural occurring bone<br>disease, traumatic<br>injury<br>or surgical intervention. | mandibular osseous<br>and periodontal defects<br>such as:<br>Intrabony/infrabony<br>defects, furcation<br>defects, recession<br>defects,<br>dehiscence/fenestratio<br>n defects (natural teeth<br>and prosthetic root<br>form implants),<br>extraction socket (ridge<br>preservation) defects,<br>ridge augmentation<br>defects, sinus lift<br>defects, endodontic<br>bony defects |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | A dual phase calcium<br>phosphate biocement<br>plus carboxymethyl<br>cellulose<br>(CMC) that is wetted by<br>sterile water to a putty<br>consistency. | A dual phase calcium<br>phosphate biocement<br>that is wetted by<br>sterile water to a paste<br>consistency | Synthetic calcium<br>phosphate,<br>carboxymethyl<br>cellulose<br>(CMC) and<br>demineralized bone<br>matrix (DBM) that is<br>wetted by sterile water<br>to a putty consistency. | A calcium sulfate<br>material that contains<br>approximately 10% of<br>pharmaceutical grade<br>sodium<br>carboxymethylcellulose<br>(CMC) that is wetted by<br>sterile water to a putty<br>consistency. |
| Design | Powder, water | Powder, water | Powder, water | Powder, water |
| Sterile/<br>non sterile | Sterile | sterile | Sterile | Sterile |
| Sterilization<br>method | gamma | gamma | gamma | gamma |
| Biocompatible | Yes | yes | yes | yes |
| Resorbable/<br>Nonresorbable | resorbable | resorbable | resorbable | resorbable |
## PREDICATE DEVICES for Socket Seal
:
| Device name | Proposed Device | Predicate Device #1 | Predicate Device#2 |
|--------------|---------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| | Socket Seal | IMTEC BioBarrier<br>Membrane | Tefgen-LS |
| 510 K number | K113049 | K974752 | K965205 |
| Intended use | A temporary<br>implantable material<br>intended for use in the<br>oral cavity to be used as | A temporary implantable<br>material intended to be<br>used in the oral cavity as a<br>space maintaining barrier | A implant material which is<br>intended to be used in the<br>oral cavity as a temporary<br>space making barrier over |
{5}------------------------------------------------
| | a space maintaining<br>barrier over bone and<br>connective tissue.<br>Socket Seal is indicated<br>for use over bone<br>grafts. Socket Seal in<br>indicated for use over<br>extraction sockets in<br>the maxilla and<br>mandible. | over bone. Use in<br>augmentation around<br>implants placed in<br>immediate extraction<br>sockets and around<br>implants placed in delayed<br>extraction sockets and<br>should be used in<br>combination with space-<br>making bone graft<br>material. | bone or other tissue. |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Material | polyethylene | polyethylene | polyethylene |
| Design | sheets | sheets | sheets |
| Sterile/non<br>sterile | Sterile | Sterile | Sterile |
| Sterilization<br>method | gamma | Not specified | Not specified |
| Biocompatible | Yes | yes | yes |
| Resorbable/<br>Nonresorbable | nonresorbable | nonresorbable | nonresorbable |
| porosity | nonporous | nonporous | nonporous |
## Conclusion (statement of equivalence)
The data submitted in this 510(k) supports substantial equivalence of Socket Graft and Socket Seal to the following aforementioned commercially marketed devices.
Substantial equivalence is based on the indications for use, product design and configuration, and materials used. The intended use of Socket Graft K113049 is the same as Socket Graft K052493, and the materials used are the same as those found in Socket Graft K052493 with the exception of the addition of carboxymethyl cellulose(CMC). Socket The comparative analysis demonstrates the substantial equivalence of Socket Graft and Socket Seal to the predicate devices.
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Gregory G. Steiner CEO STEINER LABORATORIES 590 Farrington Highway, #524 Suite 132 Kapolei, Hawaii 96707
Re: K113049
Trade/Device Name: Socket Graft Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: January 15, 2012 . Received: January 23, 2012
Dear Mr. Steiner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
JAN 2 7 2012
{7}------------------------------------------------
#### Page 2 - Mr. Steiner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
G for Anthony Watson
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{8}------------------------------------------------
### Indications for Use Statement
| 510(k) Number (if known) | K113049, Socket Graft, January 15 2012 |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Socket Graft |
| Indications for Use | SOCKET GRAFT is a bone grafting material indicated for use in dental extraction sockets that have all walls remaining. Socket Seal is a temporary implantable material intended for use in the oral cavity to be used as a space maintaining barrier over bone and other tissue. Socket Seal is indicated for use over bone grafts. Socket Seal in indicated for use over extraction sockets in the maxilla and mandible." |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X | OR | Over-The-Counter Use |
|------------------|---|----|----------------------|
|------------------|---|----|----------------------|
Prescription Use_ X (Per 21 CFR 801. 109)
Over-The-Counter Use_
Susan Kunne
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
KII13049 510(k) Number: __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.