K112995 · Instrumentation Laboratory CO · CEM · May 3, 2012 · Clinical Chemistry
Device Facts
Record ID
K112995
Device Name
GEM PREMIER 4000
Applicant
Instrumentation Laboratory CO
Product Code
CEM · Clinical Chemistry
Decision Date
May 3, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1600
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K112995 · May 3, 2012
GEM PREMIER 4000
Instrumentation Laboratory CO
Clinical field site testing
Field site testing was conducted to compare the performance of the modified GEM Premier 4000 system against the predicate GEM Premier 3000/3500 in a clinical setting.
Field site testing; Clinical setting; Method comparison
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Field Site Testing; Method comparison in a clinical setting
Clinical patients (samples tested at three field sites); Sample Size: 454 (Syringe), 304 (Full Capillary), 304 (Micro Capillary); Number of Sites: 3
GEM Premier 3000/3500
Potassium (K+) performance comparability
Indications for Use
The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2, pO2, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, O2Hb, COHb, MetHb, HHb) parameters. Total bilirubin can also be quantitated from heparinized plasma samples when analyzed in the tBili/CO-Ox mode. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity. Total bilirubin measurements are used in the diagnosis and management of biliary tract obstructions, liver disease and various hemolytic diseases and disorders involving the metabolism of bilirubin. In neonates, the level of total bilirubin is used to aid in assessing the risk of kernicterus. Intelligent Quality Management (iQM) is used as the quality control and assessment system for the GEM Premier 4000 system. iQM is an active quality process control program designed to provide continuous monitoring of the analytical process with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls. Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed. As part of this program, GEM CVP (Calibration Valuation Product) with CO-Ox, GEM CVP tBili and GEM CVP Hematocrit are external solutions intended to complete the calibration process and final accuracy assessment of the iQM cartridge calibration following warm-up. The reported values for GEM CVP (two levels for pH, blood gases, electrolytes, metabolites, total bilirubin, CO-Oximetry and hematocrit) must meet IL's specifications before the iQM cartridge can be used for patient sample measurements. Once the cartridge calibration is verified, the internal iQM program monitors the status of the system during the cartridge use life.
Device Story
Portable critical care system for rapid whole blood analysis at point-of-care or central lab. Inputs: whole blood (syringe, full capillary, micro capillary) or heparinized plasma. Principle: potentiometric measurement for electrolytes (K+ sensor). iQM (Intelligent Quality Management) provides continuous monitoring, real-time error detection, automatic system correction, and documentation, replacing traditional external QC. Output: quantitative measurements of pH, blood gases, electrolytes, metabolites, hematocrit, total bilirubin, and CO-Oximetry. Used by clinicians to aid diagnosis of acid/base, electrolyte, and oxygenation status. Benefits: rapid clinical decision-making via automated quality control and point-of-care testing.
Clinical Evidence
Bench testing only. Precision studies (N=160 per level) across three sample modes showed total imprecision %CV ranging from 0.86% to 3.55%. Linearity confirmed across 0.2–19.0 mmol/L range (R^2 > 0.999). Method comparison vs. GEM Premier 3000 (N=220 per mode) and field site testing (N=304-454) demonstrated comparable performance (slopes 0.978–1.050). Sodium citrate identified as an interferent at ≥ 7.3 mmol/L.
Indicated for use by healthcare professionals in clinical settings or central laboratories to analyze whole blood samples for pH, blood gases, electrolytes, metabolites, hematocrit, total bilirubin, and CO-Oximetry. Used for diagnosis/management of acid/base status, electrolyte/metabolite balance, oxygen delivery capacity, biliary/liver disease, hemolytic disorders, and kernicterus risk in neonates. Prescription use only.
Regulatory Classification
Identification
A potassium test system is a device intended to measure potassium in serum, plasma, and urine. Measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Predicate Devices
GEM Premier 4000 with iQM and CVP (k061974)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k112995
B. Purpose for Submission:
Modification of a cleared device (k061974 and k093623) -Potassium sensor change
C. Measurand:
Potassium
D. Type of Test:
Quantitative, potentiometric method
E. Applicant:
Instrumentation Laboratory Co.
F. Proprietary and Established Names:
GEM Premier 4000 with iQM (Intelligent Quality Management)
GEM CVP 1 and 2 (Calibration Valuation Product) with CO-Ox
GEM CVP 3 and 4 (Calibration Valuation Product) Hematocrit
GEM CVP 5 (Calibration Valuation Product) tBili
G. Regulatory Information:
- GEM Premier 4000 with iQM (Intelligent Quality Management):
| Description | CFR Section | Device Class | Product Code |
| --- | --- | --- | --- |
| Blood gases and blood pH | 862.1120 | Class II | CHL |
| Sodium test system | 862.1665 | Class II | JGS |
| Potassium test system | 862.1600 | Class II | CEM |
| Calcium test system | 862.1145 | Class II | JFP |
{1}
2
| Chloride test system | 862.1170 | Class II | CGZ |
| --- | --- | --- | --- |
| Glucose test system | 862.1345 | Class II | CGA |
| Lactic acid test system | 862.1450 | Class I, meets limitations of exemptions per 21 CFR 862.9 (c)(9) | KHP |
| Automated hematocrit instrument | 864.5600 | Class II | GKF |
| Carboxyhemoglobin assay | 864.7425 | Class II | GHS |
| Automated hemoglobin system | 864.5620 | Class II | GKR |
| Whole blood hemoglobin assays | 864.7500 | Class II | GLY |
| Bilirubin (Total or Direct) Test System | 862.1110 | Class II | CIG |
| (Total and Unbound) in the Neonate Test System | 862.1113 | Class I, Reserved | MQM |
- GEM CVP 1 and 2 (Calibration Valuation Product) with CO-Ox:
- GEM CVP 5 (Calibration Valuation Product) tBili:
| Description | CFR Section | Device Class | Product Code |
| --- | --- | --- | --- |
| Quality Control Material | 862.1660 | Class I, reserved | JJY |
- GEM CVP 3 and 4 (Calibration Valuation Product) Hematocrit:
| Description | CFR Section | Device Class | Product Code |
| --- | --- | --- | --- |
| Hematocrit Control | 864.8625 | Class II | GLK |
## H. Intended Use:
1. Intended use(s):
Refer to indications for use below.
2. Indication(s) for use:
The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, $p\mathrm{CO}_2$, $p\mathrm{O}_2$, sodium, potassium, chloride, ionized calcium, glucose, lactate, hematocrit, total bilirubin and CO-Oximetry (tHb, $\mathrm{O}_2\mathrm{Hb}$, COHb, MetHb, HHb)
{2}
parameters. Total bilirubin can also be quantitated from heparinized plasma samples when analyzed in the tBili/CO-Ox mode. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity. Total bilirubin measurements are used in the diagnosis and management of biliary tract obstructions, liver disease and various hemolytic diseases and disorders involving the metabolism of bilirubin. In neonates, the level of total bilirubin is used to aid in assessing the risk of kernicterus.
Intelligent Quality Management (iQM) is used as the quality control and assessment system for the GEM Premier 4000 system. iQM is an active quality process control program designed to provide continuous monitoring of the analytical process with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls. Facilities should follow local, state and federal regulatory guidelines to ensure that a total quality management system is followed.
As part of this program, GEM CVP (Calibration Valuation Product) with CO-Ox, GEM CVP tBili and GEM CVP Hematocrit are external solutions intended to complete the calibration process and final accuracy assessment of the iQM cartridge calibration following warm-up. The reported values for GEM CVP (two levels for pH, blood gases, electrolytes, metabolites, total bilirubin, CO-Oximetry and hematocrit) must meet IL's specifications before the iQM cartridge can be used for patient sample measurements. Once the cartridge calibration is verified, the internal iQM program monitors the status of the system during the cartridge use life.
3. Special conditions for use statement(s):
For prescription use only
At point-of-care or central laboratory settings.
4. Special instrument requirements:
GEM Premier 4000 Analyzer
I. Device Description:
The GEM Premier 4000 analyzer utitilizes an ion-selective electrode for the measurement of potassium. The potassium (K+) electrode has a sensor membrane on the surface of the electrode. The GEM Premier 4000 K sensor membrane is being modified to lower the Valinomycin concentration, along with a proportional decrease in the amount of counterion. This modification is to the Potassium sensor for the K+ assay on the GEM Premier 4000. There are no changes to other analyte measurements and device components.
{3}
J. Substantial Equivalence Information:
1. Predicate device name:
GEM Premier 4000 with iQM and CVP
2. Predicate 510(k) number:
k061974
3. Comparison with predicate:
| Comparison Table: Potassium | | |
| --- | --- | --- |
| Item | New Device (k112995) | Predicate (k061974) |
| Intended Use | Quantitative measurement of potassium | Same |
| Test Principle | Potentiometric ion-selective electrode | Same |
| Specimen Type | Whole blood | Same |
| Assay Range | 0.2 to 19.0 mmol/L | Same |
| Membrane Formulation | Modified | Unmodified |
K. Standard/Guidance Document Referenced (if applicable):
1. CLSI EP5-A2. Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline Second edition
2. CLSI EP6-A. Evaluation of the Linearity of Quantitative Analytical Methods; Approved Guideline
3. CLSI EP7-A2. Interference Testing in Clinical Chemistry; Approved Guideline- Second edition
4. CLSI EP9-A2. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline Second edition
5. CLSI EP17-A. Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline.
6. CLSI C29-A2. Standardization of Sodium and Potassium Ion-selective electrode systems to the flame photometric reference method; Approved Guideline, Second Edition
{4}
L. Test Principle:
The GEM Premier 4000 potassium sensor technology is based on ion selective electrode methodology. Valinomycin is an ionophore that functions as the potassium recognition component in the sensor. When the sensor is exposed to solutions containing potassium, an electrical signal response (change in electrode potential in millivolts) proportional to the concentration of potassium in the sample is measured. The measured response is converted into potassium concentration based on the slope from the sensor calibration.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies were performed to evaluate the performance of the modified K+ sensor using three levels of whole blood samples under the syringe mode (150 μL), the full capillary mode (150 μL) and the micro capillary (65 μL) mode. Due to the instability of whole blood, fresh levels of whole blood were prepared (using 0.1M KCL as the spiking solution) for each run, each day, with a target concentration of 3.0, 4.0 and 7.0 mmol/L of potassium. Testing was completed in replicates of 8 per run, twice a day for 5 days on 2 GEM Premier 4000 instruments. The mean, standard deviation (SD), and coefficients of variation (CV) were determined for each level and each mode as summarized below:
| Precision Data Summary | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Whole Blood- Syringe Mode | | | | | | |
| Sample | N | Within Run | | | Total | |
| | | Mean (mmol/L) | SD (mmol/L) | CV (%) | SD (mmol/L) | CV (%) |
| 1 | 160 | 2.90 | 0.04 | 1.51 | 0.04 | 1.51 |
| 2 | 160 | 3.88 | 0.05 | 1.32 | 0.05 | 1.32 |
| 3 | 160 | 7.41 | 0.06 | 0.86 | 0.06 | 0.86 |
| Whole Blood- Full Capillary Mode | | | | | | |
| Sample | N | Within Run | | | Total | |
| | | Mean (mmol/L) | SD (mmol/L) | CV (%) | SD (mmol/L) | CV (%) |
| 1 | 160 | 3.06 | 0.11 | 3.55 | 0.11 | 3.55 |
| 2 | 160 | 4.06 | 0.11 | 2.71 | 0.11 | 2.71 |
| 3 | 160 | 7.44 | 0.12 | 1.58 | 0.16 | 2.16 |
| Whole Blood- Micro Capillary Mode | | | | | | |
| Sample | N | Within Run | | | Total | |
| | | Mean (mmol/L) | SD (mmol/L) | CV (%) | SD (mmol/L) | CV (%) |
| 1 | 160 | 3.34 | 0.11 | 3.31 | 0.11 | 3.31 |
| 2 | 158 | 4.25 | 0.11 | 2.66 | 0.11 | 2.66 |
| 3 | 160 | 7.83 | 0.10 | 1.27 | 0.10 | 1.27 |
{5}
# b. Linearity/assay reportable range:
A linearity study was performed using whole blood samples according to the CLSI EP6-A guideline. 7 different concentrations of potassium were measured in triplicate on 3 different GEM Premier 4000 analyzers using 3 different sample modes. The measured values were plotted against the expected values measured by the reference method (Flame Photometer). The linear regressions analysis is summarized in the table below:
| Linearity Data Summary | | | |
| --- | --- | --- | --- |
| Testing Mode | Linear regression equation | R2 | Sample range tested |
| Syringe mode | Y= 1.0144X - 0.0612 | 0.9999 | 0.18-24.17 mmol/L |
| Full capillary | Y= 1.0115X + 0.0251 | 0.9997 | 0.13-24.57 mmol/L |
| Micro capillary | Y= 1.0791X - 0.2189 | 0.9998 | 0.17-26.14 mmol/L |
The results of the study support the sponsor's claim that the assay is linear from 0.2 to $19.0\mathrm{mmol / L}$
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
Potassium is traceable to a reference material (NIST standard). See k061974
GEM CVP 1 and 2 (Calibration Valuation Product) with CO-Ox, GEM CVP 3 and 4 (Calibration Valuation Product) Hematocrit, and GEM CVP 5 (Calibration Valuation Product) tBili are part of the iQM quality control and assessment system for the GEM Premier 4000 system. They have been previously cleared in k061974 and k093623.
Potassium sensor stability:
The shelf life stability study was performed on three lots of sensors and the in-use (cartridge use life) stability study was performed on six lots of sensors. Real-time stability study is still on-going. Based on the results, the sponsor claims that the K sensor has a shelf life stability of 2 months at 15 to $25^{\circ}\mathrm{C}$ and an in-use stability of at least 600 samples. The stability study protocol and acceptance criteria were found to be adequate.
{6}
d. Detection limit:
See linearity study in section M.1.b above.
e. Analytical specificity:
An interference study was performed based on the CLSI EP7-A guideline. A comprehensive interference screen was conducted, followed by a dose-response evaluation to characterize the interference according to the CLSI EP7-A. Two levels of human whole blood pooled samples (K approximately 3.0 and $5.0\mathrm{mmol / L}$ ) were spiked with various potential interference substances (48 drugs and endogenous compounds). All samples were analyzed in triplicates on the GEM Premier 4000 and the reference method (flame photometer). The sponsor defines non-significant interference as difference within $\pm 0.1\mathrm{mmol / L}$ when compared to the reference method. Results are summarized in the table below for non-significant interference when tested at the concentrations listed:
Potential Interfering Substances and the Concentrations Tested
| Substance | “Worst Case” Test Concentration | Substance | “Worst Case” Test Concentration |
| --- | --- | --- | --- |
| Acetaminophen | 20 mg/dL | Hydroxyurea | 0.8 mg/dL |
| Acetoacetate | 2 mmol/L | Ibuprofen | 2.425 mmol/L |
| Ammonium | 107 μmol/L | Icodextrin | 20 mg/dL |
| Aprotinin | 50 mg/L | Iodide | 3 mmol/L |
| Ascorbic acid | 6 mg/dL | Isoniazide | 4 mg/dL |
| Atracurium | 50 mg/L | Lactate | 6.6 mmol/L |
| Benzalkonium | 5 mg/L | Lithium | 3.2 mmol/L |
| Bromide | 37.5 mmol/L | Ionized Magnesium | 15 mmol/L |
| Ionized Calcium | 2.5 mmol/L | Maltose | 0.2 mg/dL |
| Chlorpromazine | 0.0063 mmol/L | Mannose | 20 mg/dL |
| Creatinine | 5 mg/dL | Oxalate | 500 mg/dL |
| Dobutamine | 2 mg/dL | Perchlorate | 20 mg/dL |
| Dopamine | 0.1 mg/dL | pH | 6.80 |
| Ethanol | 400 mg/dL | Pralidoxime iodide | 40 μg/mL |
| Etomidate | 50 mg/L | Pyruvate | 0.309 mmol/L |
| Flaxedil | 5 mg/dL | Salicylate | 4.34 mmol/L |
| Fluoride | 0.4 mg/dL | Sodium | 180 mmol/L |
| Fructose | 18 mg/dL | Thiocyanate | 56 mg/dL |
| Galactose | 15 mg/dL | Thiopental | 66 mg/L |
{7}
8
| Substance | “Worst Case” Test Concentration | Substance | “Worst Case” Test Concentration |
| --- | --- | --- | --- |
| Glucose | 1000 mg/dL | Triglycerides | 3250 mg/dL |
| Glycolic acid | 1 mmol/L | Urea | 42.9 mmol/L |
| Heparin | 300 IU/mL | Uric acid | 23.5 mg/dL |
| β-hydroxybutyrate | 2 mmol/L | Xylose | 20 mg/dL |
Because hemolysis and sodium citrate have shown significant interference, the sponsor stated the following limitations in the labeling:
“Hemolyzed samples may result in falsely elevated potassium levels.”
“Blood collection tubes containing sodium citrate as an additive will produce clinically significant interference on Potassium and should not be used.”
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study was performed for potassium according to the CLSI EP9-A guideline. Six GEM Premier 400 analyzers were tested with two in each sample mode, over a period of 18 days, against the results performed with a commercially available blood gases analyzer. A total of 109 samples were tested. Approximately 10% to 14% of the samples were altered in order to cover the hard-to-find sample range. The linear regressions analysis is summarized in below:
| Test mode | N | Slope (95%CI) | Intercept (95% CI) | R² | Sample range tested (mmol/L) |
| --- | --- | --- | --- | --- | --- |
| Syringe | 109 | 1.002 (0.994-1.011) | 0.052 (0.012-0.1092) | 0.998 | 0.2-18.6 |
| Full capillary | 109 | 1.001 (0.992-1.011) | 0.092 (0.047-0.137) | 0.997 | 0.2-18.2 |
| Micro capillary | 109 | 1.016 (1.006-1.027) | 0.148 (0.095-0.200) | 0.997 | 0.2-18.2 |
b. Matrix comparison:
Not applicable
{8}
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
K: 3.4 to 4.5 mmol/L
The reference range for potassium was taken from the literature.
References: Wu, A., Tietz Clinical Guide to Laboratory Tests, W.B. Saunders Co., St. Louis MO, 4th, 2006.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.