SMR RESURFACING SHOULDER SYSTEM- HUMERAL HEADS, CTA HUMERAL HEADS

K112900 · Lima Corporate S.P.A. · HSD · Jun 1, 2012 · Orthopedic

Device Facts

Record IDK112900
Device NameSMR RESURFACING SHOULDER SYSTEM- HUMERAL HEADS, CTA HUMERAL HEADS
ApplicantLima Corporate S.P.A.
Product CodeHSD · Orthopedic
Decision DateJun 1, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3690
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SMR Resurfacing Shoulder Prosthesis system is indicated for partial or total primary shoulder replacement in patients where the humeral head and neck have sufficient bone stock to support the prosthesis. The Standard Resurfacing heads are indicated for partial or total shoulder replacement when the rotator cuff is intact or reconstructable. Specific indications include: · non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; · Post-traumatic osteoarthritis; · Rheumatoid arthritis. The CTA Resurfacing Heads are indicated for partial shoulder replacement in patients with rotator cuff tears and arthritis. Specific indications include: · Rotator cuff tear arthropathy. The humeral heads and stems are intended for uncemented use only. The glenoid components, used in total shoulder replacement, are intended for cemented use only.

Device Story

Modular shoulder resurfacing system; includes humeral head, humeral stem, and optional glenoid component. Humeral heads (standard or CTA) feature spherical polished articulating surface; plasma-sprayed titanium inner surface; Morse taper coupling to stem. CTA version features extended articulating surface for acromial contact in rotator cuff tear arthropathy. Humeral stems feature fluted cross-section; sand-blasted exterior; two sizes. Glenoid components feature single central peg with grooves for cement fixation. Used in orthopedic surgery; implanted by surgeons. Provides joint replacement; restores function; alleviates pain from degenerative joint disease or rotator cuff arthropathy. Modular design allows intraoperative sizing/selection; uncemented humeral fixation; cemented glenoid fixation.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including static disassembly, fatigue testing of taper assembly, and torsional stability testing of stems.

Technological Characteristics

Modular humeral head (CoCrMo, ASTM F1537) with plasma-sprayed titanium coating (ASTM F1580). Humeral stem (Ti6Al4V, ISO 5832-3/ASTM F1472). Glenoid (UHMWPE, ISO 5834-2/ASTM F648). Morse taper coupling. Sterile, single-use. 5-year shelf life.

Indications for Use

Indicated for partial or total primary shoulder replacement in patients with sufficient humeral bone stock. Standard heads for patients with intact/reconstructable rotator cuff (osteoarthritis, avascular necrosis, post-traumatic osteoarthritis, rheumatoid arthritis). CTA heads for patients with rotator cuff tears and arthritis (rotator cuff tear arthropathy).

Regulatory Classification

Identification

A shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis is a device made of alloys, such as cobalt-chromium-molybdenum. It has an intramedullary stem and is intended to be implanted to replace the articular surface of the proximal end of the humerus and to be fixed without bone cement (§ 888.3027). This device is not intended for biological fixation.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Summary of Safety and Effectiveness Date: September 26, 2011 Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 – Villanova di San Daniele Udine - Italy U.S. Contact Person: Cheryl Hastings Principal Consultant Phone: 574-527-4220 | Product | Product Code | Regulation and Classification Name | |------------------------------------|--------------|-----------------------------------------------------------------------------------------------------------| | SMR Resurfacing<br>Shoulder System | HSD | Shoulder joint humeral (hemi-shoulder)<br>metallic uncemented prosthesis; Class II<br>per 21 CFR 888.3690 | | | KWS | Shoulder joint metal/polymer semi-<br>constrained cemented prosthesis; Class II<br>per 21 CFR 888.3660 | #### Description: The SMR Resurfacing Shoulder System consists of a humeral head, a humeral stem and an optional glenoid. The SMR Resurfacing Shoulder Prosthesis is intended for partial or total shoulder replacement. The heads and the stems are intended for uncemented use only. The glenoid components, used in total shoulder replacement, are intended for cemented use only. The glenoid components were previously cleared for use in the SMR Shoulder System in K100858. Two versions of the SMR Resurfacing humeral heads are available: standard and CTA. Both versions have a spherical polished articulating surface with a plasma spray titanium coated inner surface. In the inner part of the head a male taper allows coupling with the humeral stem. The CTA version of the head is characterized by an extended articulating surface to keep the implant surface in contact with the acromion in patients with irreparable rotator cuff tear arthropathy. Eight sizes are available for the SMR Resurfacing standard heads (from 40 to 54mm diameter, increasing 2mm for each greater size): seven sizes are available for the SMR Resurfacing CTA heads (from 42 to 54mm diameter, increasing 2mm for each greater size). Heads are made from CoCrMo (ASTM F1537); the inner surface is plasma spray titanium coated (ASTM F1580). The SMR Resurfacing humeral stem is characterized by a female taper for coupling with the resurfacing head. The external geometry of the stem has two taper angles and a fluted cross-section. The stem is externally sand-blasted. Two sizes of stems are available: 11 Traditional 510(k) -SMR Resurfacing Shoulder System May 4, 2012 Section 5 - Page 17 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the text "K112400 2 of 3" in a handwritten style. The text appears to be a code or identifier, possibly related to document management or tracking. The "2 of 3" likely indicates that this is the second page of a three-page document. mm diameter (length 32 mm) and 13 mm diameter (length 36 mm). Stems are made from Ti6Al4V (ISO 5832-3 -- ASTM F1472). The glenoid components are manufactured from Ultra-High Molecular Weight Polyethylene (UHMWPE ISO 5834-2, ASTM F648). The articulating surface has a radius of curvature greater than the corresponding humeral head, which allows translation in the superior/inferior and anterior/posterior directions. The back surface of the component is spherical in geometry and has a single central peg which is inserted in the hole drilled in the glenoid cavity during surgery. The peg surface has three grooves to provide enhanced cement fixation. #### Intended Use/Indications: The SMR Resurfacing Shoulder Prosthesis system is indicated for partial or total primary shoulder replacement in patients where the humeral head and neck have sufficient bone stock to support the prosthesis. The Standard Resurfacing heads are indicated for partial or total shoulder replacement when the rotator cuff is intact or reconstructable. Specific indications include: - · non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis: - · Post-traumatic osteoarthritis; - · Rheumatoid arthritis. The CTA Resurfacing Heads are indicated for partial shoulder replacement in patients with rotator cuff tears and arthritis. Specific indications include: · Rotator cuff tear arthropathy. The humeral heads and stems are intended for uncemented use only. The glenoid components, used in total shoulder replacement, are intended for cemented use only. #### Predicate Devices: - · Copeland MB Resurfacing Humeral Heads (Biomet, K010657); - · Copeland EAS Humeral Resurfacing Heads (Biomet, K051843); - · Global CAP Resurfacing Replacement Shoulder(DePuy Orthopaedics, K031971); - · Global CAP CTA Resurfacing Shoulder (DePuy Orthopaedics, K080990); - · Bio-Modular shoulder system (Biomet, K992119); - · DePuy Global Shoulder Glenoid (DePuy K981487). #### Comparable Features to Predicate Devices: The intended use and indications for the Limacorporate SMR Resurfacing Shoulder System are similar to those of the referenced predicate devices. Like the predicate t Traditional 510(k) -SMR Resurfacing Shoulder System May 4, 2012 Section 5 - Page 18 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows handwritten text on a white background. The top line reads 'K112900', with the 'K' being a capital letter. Below this, the text '3 of 3' is written, indicating a page number or a sequence within a set. The handwriting appears to be casual and slightly uneven. devices, the humeral head and stem components of the SMR Resurfacing Shoulder System are intended for uncemented use only. Like the Copeland Resurfacing Heads, the SMR Resurfacing Shoulder System can be used in Partial or Total Shoulder Arthroplasty. The SMR and predicate glenoids are all intended for cemented use only. The heads of the SMR Resurfacing Shoulder Prosthesis are available in two versions (standard head and CTA head) similar to the designs offered in the predicate devices. The SMR Resurfacing Shoulder Prosthesis has a modular design (head and stem coupled by means of a Morse taper) while the predicates are monolithic. Apart from this difference, the design of the SMR Resurfacing Shoulder Prosthesis is similar to the predicate devices with a spherical polished articulating surface (extended for the CTA version) and a coated inner surface. The SMR glenoids feature a single central peg and cement pockets to achieve fixation while predicate glenoids are available with multiple pegs or a keeled design. The material used for the SMR Resurfacing Shoulder Prosthesis (CoCrMo) is identical to that used for the predicate devices, while the humeral stems of the SMR Resurfacing system are made from Ti6Al4V. The SMR and the Biomet Bio-modular allpoly glenoids are made from UHMWPE while DePuy Global Shoulder Glenoids are made from crosslinked polyethylene. The diameters of the standard heads of the SMR Resurfacing Shoulder Prosthesis are comparable to those of the Global CAP Resurfacing Replacement Shoulder. The diameters of the CTA heads of the SMR Resurfacing Shoulder Prosthesis are comparable to those of the Copeland EAS Humeral Resurfacing Heads. The heights of the heads and the lengths of the stems of the SMR Resurfacing Shoulder Prosthesis are comparable to those of the DePuy predicate devices. All systems utilize single use devices which are provided sterile. Shelf life for the components of the SMR Resurfacing System is 5 years. #### Non-Clinical Testing: The SMR Resurfacing Shoulder System has undergone static disassembly and fatigue testing with post-test evaluation of connection strength and examination for corrosion to verify the strength of the taper assembly between the stems and the heads. The SMR Resurfacing stems have undergone torsional tests to test primary stability. The testing results demonstrated the device's ability to perform under expected clinical conditions. Testing of the glenoid components according to ASTM F1829 and ASTM F2028 was submitted in K100858. No additional glenoid testing was performed for this application because the previous testing with SMR total shoulder humeral heads also supports glenoid performance with the SMR Resurfacing system. Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of the SMR Resurfacing Shoulder System to the predicate devices. Traditional 510(k) -SMR Resurfacing Shoulder System May 4. 2012 Section 5 - Page 19 {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Lima-Lto- S.P.A % Hastings Regulatory Consulting, LLC. Ms. Cheryl Hastings Principal Consultant P.O. Box 696 Winona Lake, Indiana 46590 JUN - 1 2012 Re: K112900 Trade/Device Name: SMR Resurfacing Shoulder Prosthesis Regulation Number: 21 CFR 888.3690 Regulation Name: Shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis Regulatory Class: Class II Product Code: HSD, KWS Dated: May 4, 2012 Received: May 8, 2012 Dear Ms. Hastings: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ #### Page 2 - Ms. Hastings forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation(21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, El Keith Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K112900 510(k) Number (if known): Unknown Device Name: SMR Resurfacing Shoulder System Indications for Use: ### SMR Resurfacing Shoulder System Indications for Use The SMR Resurfacing Shoulder Prosthesis system is indicated for partial or total primary shoulder replacement in patients where the humeral head and neck have sufficient bone stock to support the prosthesis. The Standard Resurfacing heads are indicated for partial or total shoulder replacement when the rotator cuff is intact or reconstructable. Specific indications include: - · non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; - · Post-traumatic osteoarthritis; - · Rheumatoid arthritis. The CTA Resurfacing Heads are indicated for partial shoulder replacement in patients with rotator cuff tears and arthritis. Specific indications include: · Rotator cuff tear arthropathy. The humeral heads and stems are intended for uncemented use only. The glenoid components, used in total shoulder replacement, are intended for cemented use only. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ars (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 4112900 510(k) Number Page 1 of 1 Traditional 510(k) - SMR Resurfacing Shoulder System May 4, 2012 83
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...