← Product Code [JDR](/productcode/JDR) · K112837

# VILEX EZ-STAPLE (MEMORY BONE FIXATION STAPLE) (K112837)

_Vilex, Inc. · JDR · Oct 9, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K112837

## Device Facts

- **Applicant:** Vilex, Inc.
- **Product Code:** [JDR](/productcode/JDR.md)
- **Decision Date:** Oct 9, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Vilex eZ-Staple (Superelastic Bone Fixation Staple) is indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis.

## Device Story

Vilex eZ-Staple is a single-use, permanently implanted superelastic bone fixation staple; constructed from shape memory nickel-titanium alloy (Nitinol). Device provides compression across bone fragments or joints to facilitate healing. Available in monocortical and biocortical configurations with varying widths and leg lengths. Used by surgeons in clinical settings for osteotomy fixation and joint arthrodesis in hand and foot procedures. Device functions via shape memory properties of Nitinol to maintain constant compression on bone segments. Benefits include stable fixation of bone fragments during the healing process.

## Clinical Evidence

Bench testing only. Samples tested per ISO 10993 for biocompatibility (cytotoxicity, genotoxicity, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity) and per ASTM standards for mechanical performance (corrosion, pullout strength, bending stiffness, fatigue life) compared to predicate device.

## Technological Characteristics

Material: Nitinol (Titanium-Nickel alloy) per ASTM F2063-05. Principle: Superelastic shape memory alloy for bone compression. Form factor: Monocortical (10-20mm width, 10-20mm leg length) and Biocortical (10-18mm width, 15x13 to 19x17 leg lengths). Sterilization: Not specified. Connectivity: None.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Memometal Memory Staples ([K070031](/device/K070031.md)) (Trade name: MM EasyClip® SI)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 9 2012

1/1

## 510(k) Summary, K112837

| Sponsor                    | Vilex in Tennessee, Inc., 111 Moffitt St., McMinnville, TN 37110,<br>Phone: 931-474-7550, Fax: 931-474-7551, Email: sylvias@vilex.com                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact                    | Sylvia Southard                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Date                       | September 27, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Device Name                | Vilex eZ-Staple Superelastic Bone Fixation Staple                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification             | 21 CFR 888.3030 - Single/multiple component metallic bone fixation appliances and<br>accessories<br>Product Code: JDR<br>Regulatory Class: II                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Devices          | Memometal Technologies' Memometal Memory Staples - K070031 - (Trade name: MM<br>EasyClip® SI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Description of<br>Device   | The Vilex eZ-Staple Superelastic Bone Fixation Staple is a single-use bone fixation<br>appliance intended to be permanently implanted. Super Elastic staples are<br>compression staples made of shape memory nickel titanium alloy, Nitinol. Vilex will<br>offer Monocortical Staples ranging in width from 10mm to 20mm with leg lengths<br>ranging from 10mm to 20mm and Biocortical Staples ranging in width from 10mm to<br>18mm with leg lengths from 15x13 to 19x17.                                                                                                                       |
| Material                   | Titanium-Nickel alloy, Nitinol, for human implanting. ASTM F2063-05 Standard<br>Specification for Wrought Nickel-Titanium Shape Memory Alloys for Medical Devices<br>and Surgical Implants                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for Use        | The Vilex eZ-Staple (Superelastic Bone Fixation Staple) is indicated for hand and foot<br>bone fragments osteotomy fixation and joint arthrodesis.                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Non-Clinical Test<br>Data  | Samples of the Vilex eZ-Staple have been tested according to ISO 10993 and applicable<br>FDA regulations for:<br>• Cytotoxicity (MEM Elution)<br>• Genotoxicity (Ames Assay)<br>• Intracutaneous Reactivity<br>• Acute Systemic Toxicity<br>• Material-Mediated Pyrogenicity<br>Additionally, samples of the Vilex eZ-Staple have also been tested to relevant ASTM<br>standards versus samples of the predicate device for:<br>• Corrosion Susceptibility<br>• Auger Electron Spectroscopy of Oxide Layer<br>• Pullout Strength<br>• Static Bending Stiffness<br>• Dynamic Bending Fatigue Life |
| Substantial<br>Equivalence | Documentation is provided which demonstrates that the Vilex eZ-Staple is substantially<br>equivalent to other legally marketed devices in terms of intended use and indications<br>for use, material, design and function. Any minor differences between these two<br>devices do not raise new questions of safety and effectiveness.                                                                                                                                                                                                                                                            |
| Establishment<br>Reg. No.  | 1051526                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

1

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three wing-like shapes.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

OCT
9
2012

Vilex, Inc. % Ms. Sylvia Southard 111 Moffitt Street McMinnville, Tennessee 37110

Re: K112837 Trade/Device Name: Vilex eZ-Staple (Memory Bone Fixation Staple) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories ・ Regulatory Class: II Product Code: JDR Dated: September 27, 2012 Received: September 28, 2012

Dear Ms. Southard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{2}------------------------------------------------

Page 2 - Ms. Sylvia Southard

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Mellekerson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## INDICATIONS FOR USE

510(k) Number: K12837

Device Name: Vilex eZ-Staple (memory bone fixation staple).

Indications for Use:

The Vilex memory bone fixation staples (eZ-Staple) are indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K. Asahi.

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K112837

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K112837](https://fda.innolitics.com/device/K112837)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
