RENASYS AB ABOMINAL DRESSING KIT

K112784 · Smith & Nephew, Inc. · FTL · Nov 22, 2011 · General, Plastic Surgery

Device Facts

Record IDK112784
Device NameRENASYS AB ABOMINAL DRESSING KIT
ApplicantSmith & Nephew, Inc.
Product CodeFTL · General, Plastic Surgery
Decision DateNov 22, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The RENASYS AB Abdominal Kit with Soft Port is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The RENASYS AB Abdominal Kit with Soft Port is intended for use in acute hospital settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre.

Device Story

NPWT abdominal dressing kit; includes hydrophobic reticulated polyurethane foam, polyurethane organ protection layer, transparent film drapes, and Soft Port suction assembly. Used with RENASYS EZ/EZ PLUS pumps to provide negative pressure to abdominal wounds. Soft Port replaces previous suction port; features 'quick-click' connector for secure canister tubing attachment. Applied in operating theatre by clinicians; used in acute hospital settings. Facilitates exudate removal from open abdominal wounds with exposed viscera; supports temporary bridging of abdominal wall openings. Benefits patient by managing exudate and maintaining wound environment in complex abdominal cases.

Clinical Evidence

Bench testing only. Testing verified Soft Port performance against clottable blood, resistance to blockage/compression, flow rate maintenance (minimum 48 hours), and connector security. Biocompatibility testing (cytotoxicity, irritation, sensitization, implantation, sub-acute toxicity, genotoxicity) performed on all kit components per ISO 10993.

Technological Characteristics

Materials: hydrophobic reticulated polyurethane foam, polyurethane organ protection layer, transparent film drapes. Soft Port suction assembly. Biocompatibility: ISO 10993 compliant. Energy: passive dressing kit used with external NPWT pump. Sterilization: sterile, single-use.

Indications for Use

Indicated for temporary bridging of abdominal wall openings where primary closure is not possible or repeat entries are necessary; for open abdominal wounds with exposed viscera, including abdominal compartment syndrome. Intended for acute hospital settings (trauma, general, plastic surgery wards).

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ NOV 2 2 2011 K112784 # 10f3 # 510(k) Summary | General Information | | |------------------------------------|------------------------------------------------------------------------------------------| | Submitters Name/Address: | Smith & Nephew, Inc.<br>970 Lake Carillon Drive<br>Suite 110<br>St. Petersburg, FL 33716 | | Establishment Registration Number: | 3006760724 | | Contact Person: | Laura D. Reynolds<br>Director, Regulatory Affairs | | Phone Number: | (727) 329-7702 | | Date Prepared: | November 1, 2011 | | Trade Name: | RENASYS™™ AB Abdominal Dressing Kit with Soft<br>Port | | Generic/Common Name: | NPWT Abdominal Wound Dressing Kit | | Classification Name: | Mesh, Surgical, Polymeric, 21 CFR 878.3300 | | Product Classification/Code: | Class II, FTL | | Predicate Device Information | | | 510(k) # | Device | Manufacturer | Clearance Date | |----------|-------------------------------------------------------------------------|----------------------|----------------| | K100787 | RENASYS™ F/AB<br>Abdominal Dressing Kit | Smith & Nephew, Inc. | 9/17/2010 | | K110647 | RENASYS Foam and<br>Gauze NPWT Wound<br>Dressing Kits with Soft<br>Port | Smith & Nephew, Inc. | 6/22/2011 | # Device Description The RENASYS™ AB Abdominal Dressing Kit with Soft Port consists of two large hydrophobic reticulated polyurethane foam dressings that incorporate several cuts to facilitate custom sizing if needed. Also included in the kit are a polyurethane organ protection layer, six transparent film drapes and a Soft Port suction port assembly with tubing that attaches to the exudate canister. The kit is designed specifically for abdominal wounds and is supplied sterile, single use. The modification to this kit is the inclusion of a Soft Port suction port, to replace the existing suction port. The RENASYS AB Abdominal Dressing Kit is used in conjunction with Smith & Nephew RENASYS EZ and RENSASYS EZ PLUS negative pressure wound therapy pumps and canister kits, which have been cleared under 510(k) numbers K082426 and K091470. {2}------------------------------------------------ 2 of 2 ## Indications for Use The RENASYS AB Abdominal Dressing Kit with Soft Port is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The RENASYS AB Abdominal Dressing Kit with Soft Port is intended for use in acute hospital settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre. # Non-Clinical Tests (Bench) Design verification testing has been conducted to verify the Soft Port suction device meets the required specifications and functions equivalent to the existing suction port component in the kit. Testing verified that the Soft Port meets the design specifications and demonstrated substantial equivalence to the predicate device. Summary of testing conducted: - Laboratory testing was completed to confirm the Soft Port and the existing suction port demonstrate comparable performance patterns when tested with clottable blood. - Testing to verify that the Soft Port performs to specification when it is placed under excessive ・ weight, becomes blocked by compression, folding, or is subjected to particulate challenge. - -Testing to verify the Soft Port assembly will effectively remove exudate from the abdomen at the predetermined flow rate for a minimum of 48 hours. - Testing to verify that the new "quick-click" connector establishes a secure connection to the exudate canister tubing. The following biocompatibility testing for all kit components has been successfully completed per applicable parts of ISO 10993: | Kit Component | Tests Completed | |----------------------------|--------------------| | Foam | Cytotoxicity | | | Irritation | | | Sensitization | | Organ Protection Layer | Cytotoxicity | | | Irritation | | | Sensitization | | | Implantation | | | Sub-acute Toxicity | | | Genotoxicity | | Transparent Film Drape | Cytotoxicity | | | Irritation | | | Sensitization | | Soft Port Suction Assembly | Genotoxicity | | | Cytotoxicity | | | Irritation | | | Sensitization | | Sub-acute Toxicity | | {3}------------------------------------------------ 4112784 3.83 # Conclusion In establishing substantial equivalence to the currently marketed devices, Smith & Nephew, Inc evaluated the indications for use, materials, technology and product specifications of the device. Performance testing has been successfully completed to demonstrate that the RENASYS AB Abdominal Dressing Kit with Soft Port is substantially equivalent to the marketed device and is appropriate for the intended use. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002 NOV 2 2 2011 Smith & Nephew, Inc. % Ms. Laura D. Reynolds Director, Regulatory Affairs 970 Lake Carillon Drive, Suite 110 St. Petersburg, Florida 33716 Re: K112784 Trade/Device Name: RENASYS™ AB Abdominal Kit with Soft Port Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: November 01, 2011 Received: November 02, 2011 Dear Ms. Reynolds: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {5}------------------------------------------------ Page 2 -- Ms. Laura D. Reynolds or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use 510(k) Number (if known): K1|2784 Device Name: RENASYS™ AB Abdominal Kit with Soft Port Indications for Use: The RENASYS AB Abdominal Kit with Soft Port is indicated for temporary bridging of abdominal wall openings where primary closure is not possible and/or repeat abdominal entries are necessary. It is intended to be used in open abdominal wounds with exposed viscera, including but not limited to abdominal compartment syndrome. The RENASYS AB Abdominal Kit with Soft Port is intended for use in acute hospital settings (trauma, general and plastic surgery wards) and should ideally be applied in the operating theatre. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Kurech MXIM al. Orthopedic, 510(k) Number K112784
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...