EPIGRAY

K112723 · Dosisoft SA · MUJ · Feb 8, 2012 · Radiology

Device Facts

Record IDK112723
Device NameEPIGRAY
ApplicantDosisoft SA
Product CodeMUJ · Radiology
Decision DateFeb 8, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

EPIgray is a software to be used by radiation oncologist and medical physicist to detect errors in the delivery of high energy X-rays during the course of patient treatment. This product verifies if the reconstructed dose, computed by the system, is in agreement with the planned dose given by the treatment planning system. This product uses the measurements performed by an Electronic Portal Imaging Device (EPID). This product uses the prescription computed by the treatment planning svstem. This product in not used to give a prescription of the radiation therapy.

Device Story

EPIgray is software for in-vivo dosimetry in X-ray high-energy radiation therapy. It inputs images from an Electronic Portal Imaging Device (EPID) and treatment plans from a planning system. The software reconstructs the dose delivered to the patient and compares it to the planned dose to detect delivery or positioning errors. It triggers warnings if discrepancies occur, allowing clinicians to monitor the actual dose given. The system includes an EPIgray workstation (extending the ISOgray planning system) and an In-Vivo Manager tool for web-based result retrieval. Used in clinical settings by radiation oncologists and medical physicists to improve treatment accuracy and patient safety.

Clinical Evidence

No clinical trials were performed. Substantial equivalence was demonstrated through non-clinical verification and validation testing, including system-level pass/fail criteria. Algorithm evaluation was conducted by medical physicists using measured data in a clinical facility.

Technological Characteristics

Software-based in-vivo dosimetry system. Integrates with existing EPID hardware and treatment planning systems. Utilizes modules from the ISOgray planning system (Information and Exacor modules). Operates as a workstation application with a web-based management tool for dose reconstruction and comparison.

Indications for Use

Indicated for use by radiation oncologists and medical physicists to detect errors in high-energy X-ray radiation therapy delivery by verifying reconstructed dose against planned dose using EPID measurements.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DOSIsoft 2002 LAUREAT CREATION D'ENTREPRISES DE TECHNOLOGIES INNOVANTES Ministère jeunesse éducation recherche ministère délégué recherche et nouvelles technologies # FEB - 8 2012 ### Premarket Notification [510(k)] Summary Tab 4 K112723 Tuesday, September 6th, 2011 Trade Name: EPIgray Classification Name: Accelerator, Linear, Medical, IYE (per 21 CFR section 892.5050) Manufacturer's Name: DOSIsoft SA Address: 45-47 avenue Carnot 94230 CACHAN - France | Corresponding Official: | Hanna Kafrouni, Ph.D. | |-------------------------|-----------------------| | Title: | CEO | | Telephone: | 33 1 41 24 26 26 | | Fax: | 33 1 41 24 26 28 | Predicate: Dosimetry Check with Exit Dose from Math Resolutions, LLC (K 101503) Device Description: EPIgray is comprehensive software that allows the user to perform an in-vivo dosimetry by means of imaging device such as an Electronic Portal Imaging Device (EPID). The product is composed of: EPIgray workstation and In-Vivo Manager tool. Epigray workstation is an extension of an already cleared product: ISOgray planning system (K103146). It uses only two modules of the previously approved system: Information module and Exacor module. In Vivo manager software is a year application intended for in-vivo measurements management. In particular, it allows to retrieve, on a web browser, the result of dose reconstruction by EPIgray workstation based on EPID. Intended Use: EPIgray is a software dedicated for error detection in dose delivery and patient positioning during X-ray high energy radiation therapy. EPIgray uses images acquired from a portal imaging system (EPID) to reconstruct the dose in the patient. The comparison between the reconstructed and the planned dose, activates warnings and allows a follow-up of the dose actually given to the patient. > DOSISOFT SA 45/47, avenue Carnot 94230 CACHAN S.A. capital 500.000 € .. . Tél. : 01 41 24 26 26 Fax : 01 41 24 26 28 info@dosisoft.com www.dosisoft.com {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "DOSI" in bold, followed by a butterfly-like shape with arrows pointing towards it. The word "soft" is written after the butterfly shape. The text and butterfly shape appear to be part of a logo or design. Image /page/1/Picture/1 description: The image shows a logo with the word "LAUREAT" at the top. Below that is the text "CREATION D'ENTREPRISES ET TECHNOLOGIES INNOVANTES". There is also text that says "Ministere" and "Jeunesse Education recherche". The logo also includes the text "ministre delegue recherche et nouvelles technologies" and the number "2002". Clinical tests: Clinical trials were not performed as part of the development of this product. Clinical testing is not advantageous in demonstrating substantial equivalence or safety and effectiveness of the device since testing can be performed such that no human subjects are exposed to risk. However, algorithm evaluation was performed by Medical Physicists team using measured data in a clinical facility. Evaluation summary is available in tab 13 of this submission. Non-Clinical tests: Verification tests were written and executed internally to validate that the system works according to the requirement. The documentation includes system level test pass/fail criteria and a summary of the test results. It is available in the Verification and Validation section from Tab 12, Sub Tab 6. > DOSISOFT SA 45/47, avenue Carnot 94230 CACHAN SA canital 500 000 € Tél. : 01 41 24 26 26 Fax : 01 41 24 26 28 info@dosisoft.com www.dosisoft.com {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 DOSIsoft SA % Mr. Robert J. Morton President Quality and Regulatory Services, Inc. 1244 Fairway Valley Court LINCOLN CA 95648-8489 FEB - 8 2012 Re: K112723 Trade/Device Name: EPIgray Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: January 18, 2012 Received: January 27, 2012 #### Dear Mr. Morton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {3}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Mary S Patel Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for Use Statement 510(k) Number (if known): Ki l 2723 Device Name: EPIgray Indications for Use: EPIgray is a software to be used by radiation oncologist and medical physicist to detect errors in the delivery of high energy X-rays during the course of patient treatment. This product verifies if the reconstructed dose, computed by the system, is in agreement with the planned dose given by the treatment planning system. This product uses the measurements performed by an Electronic Portal Imaging Device (EPID). This product uses the prescription computed by the treatment planning svstem. This product in not used to give a prescription of the radiation therapy. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...