TRI AUTO MINI - ENDODEONTIC TREATMENT MOTORIZED HANDPIECE
K112665 · J. Morita USA, Inc. · EKX · May 4, 2012 · Dental
Device Facts
Record ID
K112665
Device Name
TRI AUTO MINI - ENDODEONTIC TREATMENT MOTORIZED HANDPIECE
Applicant
J. Morita USA, Inc.
Product Code
EKX · Dental
Decision Date
May 4, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Tri Auto mini motorized handpiece can be used to enlarge and prepare root canals, remove gutta-percha points, and for professional tooth cleaning. When connected to Apex locator, the Tri Auto mini can be used to measure the length of root canals.
Device Story
Battery-driven motorized handpiece; includes chuck for rotary instruments (dental files, reamers). Used for root canal enlargement/preparation, gutta-percha removal, and professional tooth cleaning. Operates by rotating attached instruments. When connected to an external Apex locator, provides indication of file tip position within root canal. Used in dental clinics by dental professionals. Output allows clinicians to perform endodontic procedures and measure canal length; facilitates efficient root canal preparation and cleaning.
Clinical Evidence
Bench testing only. Testing conducted in accordance with IEC 60601-1, IEC 60601-1-4, UL 60601-1, CAN/CSA C 22.2 No.601.1-M90, and IEC 60601-1-2. Biocompatibility, sterilization, and software validation were performed.
Technological Characteristics
Battery-driven motorized handpiece; rotary instrument chuck. Standards met: IEC 60601-1, IEC 60601-1-4, UL 60601-1, CAN/CSA C 22.2 No.601.1-M90, IEC 60601-1-2. Includes software for motor control and integration with external Apex locator.
Indications for Use
Indicated for patients requiring endodontic root canal enlargement and preparation, gutta-percha point removal, professional tooth cleaning, and root canal length measurement when used with an Apex locator.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Kllalds
### 510(k) SUMMARY
---. ...
MAY - 4 2012
# J. MORITA USA Inc.'s Tri Auto mini
# 1. NAME OF DEVICE
Common/Usual Name: Device trade or proprietary name: Tri Auto mini Product Model :
Endodontic Treatment Motorized Handpiece TR-CM
#### 2. SUBMITTER NAME AND ADDRESS WITH PHONE/FAX
Registration No. 2081055 Initial Distributor: J. Morita USA, Inc. 9 Mason Irvine, CA 92618 USA Telephone: 949-581-9600 Facsimile: 949-581-9688 Registration No. 3002807636 Manufacturer: J. MORITA MFG. CORP. 680 Higashihama Minami-cho Fushimi-ku, Kyoto Japan 612-8533 +81-75-611-2141 +81-75-605-2354
### 3. CONTACT PERSON
Keith A. Barritt Fish & Richardson P.C. 1425 K Street, N.W. Suite 1100 Washington, DC 20005 Phone: (202) 783-5070 Facsimile: (202) 783-2331
#### 4. DATE SUMMARY PREPARED: August 5, 2011
#### 5. DEVICE CLASSIFICATION/CLASSIFICATION PANEL
Device: Endodontic Treatment Motorized Handpiece Review Panel: 872 Dental Product Code: EKX Device Class: Class I
{1}------------------------------------------------
# 6. DEVICE DESCRIPTION/SUBSTANTIAL EQUIVALENCE
#### DEVICE DESCRIPTION
. 1;
The Tri Auto mini is an battery-driven handpiece with a motor, equipped with the chuck for holding rotary instrument such as a dental file and a reamer. Tri Auto mini can be used for enlargement and preparation of root canals. When connected to Apex Locator (which is not included in this application), the Tri Auto mini indicates the position of the file tip inside the root canal.
#### SUBSTANTIAL EQUIVALENCE
The Tri Auto mini is substantially equivalent to Tri Auto ZX (K921979 / K953867) and ROOT ZX II (K071190).
#### 1) Predicate device I : Tri Auto ZX (K921979 / K953867)
The Tri Auto mini is substantially equivalent to Tri Auto ZX from J.MORITA MFG.CORP. The Tri Auto mini has similar general intended uses, similar principles of operation, and similar technological characteristics to the predicate device Tri Auto ZX (K970339).
#### 2) Predicate device II : ROOT ZX II (K071190)
The Tri Auto mini is substantially equivalent to ROOT ZX II (K071190) from J.MORITA MFG.CORP. The Tri Auto mini has similar general intended uses, similar principles of operation, and similar technological characteristics to the predicate device, ROOT ZX II (K071190).
#### PERFORMANCE TESTING
The Tri Auto mini was tested in accordance with IEC 60601-1, IEC 60601-1-4, UL 60601-1, CAN/CSA C 22.2 No.601.1-M90, and IEC 60601-1-2. Also, Biocompatibility, sterilization, and software validation were conducted.
Test results shows that Tri Auto mini is substantially equivalent to predicate devices.
{2}------------------------------------------------
| | Predicate devices | |
|------------------------------------------------------|--------------------------|-------------------------|
| TECHNOLOGICAL<br>CHARACTERISTICS of<br>Tri Auto mini | Tri Auto ZX<br>(K970339) | ROOT ZX II<br>(K071190) |
| Indication for use<br>Canal Enlargement Function | Identical | Identical |
| Target population | Identical | Identical |
| Design | Similar | Different |
| Materials | Similar | Similar |
| Performance<br>Canal Enlargement Function | Identical | Identical |
| Sterility | Similar | Similar |
| Biocompatibility | Similar | Similar |
| Mechanical safety | Identical | Identical |
| Chemical safety | Identical | Identical |
| Anatomical sites | Similar | Similar |
| Human factors | Similar | Similar |
| Energy used and/or delivered | Different | Different |
| Compatibility with environment<br>and other devices | Similar | Similar |
| Where used | Identical | Identical |
| Standards met | Similar | Identical |
| Electrical safety | Similar | Identical |
| Thermal safety | Identical | Identical |
#### Table- 1 Comparison summary table
#### 7. INDICATIONS FOR USE
The Tri Auto mini motorized handpiece can be used to enlarge and prepare root canals, remove gutta-percha points, and for professional tooth cleaning. When connected to Apex locator, the Tri Auto mini can be used to measure the length of root canals.
#### 8. Conclusion
Based on the information provided in this premarket notification, the Tri Auto mini is substantially equivalent to the Tri Auto ZX (K970339) and the ROOT ZX II (K071190).
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized eagle-like symbol with three curved lines representing the body and wings.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
J. Morita USA, Inc. C/O Mr. Keith A. Barritt Attorney Fish & Richardson P.C. 1425 K Street NW Suite 1100 Washington. District of Columbia 20005
MAY - 4 2012
Re: K112665
Trade/Device Name: Tri Auto Mini (Endodontic Treatment Motorized Handpiece) Regulation Number: 21 CFR 872.4200 Regulation Name: Dental handpiece and accessories. Regulatory Class: I Product Code: EKX Dated: April 30, 2012 Received: May 1, 2012
Dear Mr. Barritt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2- Mr. Barritt
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): unknown
Device Name: Tri Auto mini (Endodontic Treatment Motorized Handpiece)
Indications For Use:
-------
- The Tri Auto mini motorized handpiece can be used to enlarge and prepare root canals,
- . remove gutta-percha points, and for professional tooth cleaning.
- When connected to Apex locator, the Tri Auto mini can be used to measure the length of root canals.
Susan Runge
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112665
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.