K112626 · Medicis Technologies Corporation · OHV · Oct 20, 2011 · General, Plastic Surgery
Device Facts
Record ID
K112626
Device Name
LIPSONIX(R) SYSTEM MODEL 2
Applicant
Medicis Technologies Corporation
Product Code
OHV · General, Plastic Surgery
Decision Date
Oct 20, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4590
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LipoSonix system delivers high intensity focused ultrasound (HIFU) energy that can disrupt subcutaneous adipose tissue (SAT) to provide a non-invasive approach to achieve a desired aesthetic effect. The LipoSonix system is specifically indicated for non-invasive waist circumference reduction.
Device Story
LipoSonix system Model 2 is a cart-based device for non-invasive disruption of subcutaneous adipose tissue (SAT). It utilizes high intensity focused ultrasound (HIFU) energy to induce thermal cellular disruption and collagen contraction in soft tissue. The device consists of a cart, a detachable handpiece, and a user interface display. It is operated by a clinician in a clinical setting. The thermal energy targets adipose tissue, which is subsequently cleared by the body's natural inflammatory and lymphatic processes. The device output is controlled energy delivery to the target area. Clinical benefits include waist circumference reduction, potentially equivalent to one dress or pant size. The system is designed to provide a non-invasive aesthetic alternative to surgical procedures.
Clinical Evidence
Multicenter, randomized study evaluated waist circumference reduction. Primary endpoint met: 59 J/cm² treatment group showed 1.01 cm greater reduction compared to sham control. Mean change from baseline to 12 weeks in the 59 J/cm² group was -2.44 cm. Safety assessed through 24 weeks; adverse events were mild, short-lived, and resolved without incident. No serious adverse events or unanticipated adverse device effects reported.
Technological Characteristics
Cart-based system with detachable handpiece and user interface. Employs high intensity focused ultrasound (HIFU) for thermal tissue disruption. Conforms to FDA consensus and international harmonized standards for thermal, electrical, electromagnetic, and mechanical safety. Addresses risks including thermal/mechanical injury, ocular injury, electrical shock, and inflammation.
Indications for Use
Indicated for non-invasive waist circumference reduction in patients with at least 1.0 cm of adipose tissue thickness beyond the selected treatment focal depth. Contraindicated in pregnant or lactating females, patients with a hernia in the treatment area, or patients with insufficient adipose tissue thickness.
Regulatory Classification
Identification
A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
K112626
Pg 1 of 4
:
LipoSonix® system Model 2 Special 510(k) Premarket Notification Medicis Technologies Corporation
## 510(k) Summary
In accordance with 21 CFR 807.92(c) the following summary of information is provided:
| Date Prepared: | September 8, 2011 |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Medicis Technologies Corporation<br>11818 North Creek Parkway North<br>Bothell, WA 98011 USA |
| Contact Person: | Michael A. Hoffman, MPH<br>Vice President - Regulatory Affairs and Quality Systems<br>Medicis Technologies Corp.<br>11818 North Creek Parkway North<br>Bothell, WA 98011 USA |
| Telephone: | 425.420.2135 |
| Facsimile: | 425.420.2335 |
| FR Number: | 21 CFR 878.4590 |
| Product Code: | OHV |
| Classification Name | Focused Ultrasound Stimulator System for Aesthetic Use |
| Common Name: | Focused Ultrasound Stimulator System for Aesthetic Use |
| Trade Name: | LipoSonix® system Model 2 |
| Regulatory Class: | Class II |
| Classification<br>Advisory<br>Committee: | Division of General, Restorative and Neurological<br>Devices; General Surgery Devices Branch |
| Special Controls: | Class II Special Controls Guidance Document: Focused<br>Ultrasound Stimulator System for Aesthetic Use (July 20,<br>2011) |
| Predicate Device(s): | K100874, LipoSonix® system Model 1 |
| Intended<br>Use/Indications for<br>Use: | The LipoSonix system delivers high intensity focused<br>ultrasound (HIFU) energy that can disrupt subcutaneous<br>adipose tissue (SAT) to provide a non-invasive approach<br>to achieve a desired aesthetic effect.<br>The LipoSonix system is specifically indicated for non-invasive waist circumference reduction |
| Contraindications: | A patient is ineligible for treatment with the LipoSonix system if any of the following criteria are met:<br>The patient is a female who is pregnant, may be pregnant, or is lactating. Female patients of child bearing age should have a negative pregnancy test before being treated with the LipoSonix system. The patient has less than 1.0 cm of adipose tissue thickness beyond the selected treatment focal depth in the area to be treated. Hernia in the area to be treated. |
| Device Description: | The LipoSonix system is a cart-based medical device intended for non-invasive disruption of subcutaneous adipose tissue (SAT) by employing ultrasound energy. The system consists of a cart, a detachable handpiece and a user interface display in one unit. |
| Technology: | The LipoSonix system accomplishes its intended use through the use of precisely targeted high intensity focused ultrasound energy that produces a thermal cellular disruption in the subcutaneous adipose tissue (soft tissue). This thermal energy also causes collagen to contract. Thus creating a dual tissue response. The destroyed adipose tissue is subsequently cleared via an |
{2}------------------------------------------------
K112626 pg 2 of 4
LipoSonix® system Model 2 Special 510(k) Premarket Notification Medicis Technologies Corporation
lymphatic system.
inflammatory response by macrophages that transported the destroyed cells and their contents to the liver via the
{3}------------------------------------------------
K11 2626 pg.3 of 4.
LipoSonix® system Model 2 Special 510(k) Premarket Notification Medicis Technologies Corporation
## Summary of Non-Clinical Tests: Determination of
The system has been found to conform to the system Substantial essential specifications, thermal. electrical. Equivalence: electromagnetic and mechanical safety, and has been found to conform to FDA consensus, medical device safety standards and international harmonized standards.
> The following risks as identified in the FDA special control quidance document entitled, "Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use" have been addressed.
- Thermal Injury from Focused Ultrasound Exposure . (Thermal Damage)
- Mechanical Injury from Focused Ultrasound Exposure .
- (Cavitation or other Mechanical Damage) .
- . Ocular Injury
- Electrical Shock
- Inflammation/Foreign Body Response ●
- Use Error ●
{4}------------------------------------------------
K112626 pg 4 of 4
LipoSonix® system Model 2 Special 510(k) Premarket Notification
The LipoSonix system was evaluated in a multicenter, Performance Data: randomized study. Treatment with the LipoSonix system was shown to be superior to a sham control in reducing waist circumference, meeting the pre-specified primary endpoint of the study. The 59 J/cm² treatment group demonstrated a 1.01 cm greater waist circumference reduction as compared to sham. The least square (LS) mean change from baseline to 12 weeks of the waist circumference in the same 59 J/cm2 group was -2.44 cm. On average, the mean change in waist circumference of approximately 2.55 cm (1 inch) compared to baseline that was demonstrated in the 59 J/cm2 treatment group. This could equal one dress or pant size.
> The safety of treatment with the LipoSonix system was assessed through 24 weeks post-treatment. The AEs resulting from treatment with the LipoSonix system during this study were mostly mild, short-lived in duration, and resolved without incident. There were no serious adverse events (SAEs) or unanticipated adverse device effects (UADEs) related to treatment with the investigational device.
The LipoSonix system is substantially equivalent to the Substantial LipoSonix system Model1 (K100874) device. Equivalence:
> The LipoSonix system Model 2 has the same intended use as the Model 1 system, namely performing a noninvasive treatment to achieve a desired aesthetic effect. Both systems use the same technological principles, namely HIFU, to thermally disrupt soft tissues which result in the aesthetic affect. The mechanism of action for both devices is essentially the same as well. Disrupted soft tissue is removed by normal healing processes of the body resulting in an aesthetic affect. Both devices can deliver more than one energy dose.
Therefore, the LipoSonix system Model 2 is substantially equivalent to the predicate device.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three stripes representing the three branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002
OCT 2 0 2011
Medicis Technologies Corporation % Michael A. Hoffman, MPH Vice President - Regulatory Affairs and Quality Systems 11818 North Creek Parkway North Bothell, Washington 98011
Re: K112626
Trade/Device Name: LipoSonix® System Model 2 Regulation Number: 21 CFR 878.4590 Regulation Name: Focused ultrasound stimulator system for aesthetic use Regulatory Class: Class II Product Code: OHV Dated: October 04, 2011 Received: October 04, 2011
Dear Mr. Hoffman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Orug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not musicading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device tan be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Michael A. Hoffman, MPH
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Evin L. Keith
For Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): This Submission
Device Name: LipoSonix® system Model 2
Indications For Use:
The LipoSonix system delivers high intensity focused ultrasound (HIFU) energy that can disrupt subcutaneous adipose tissue (SAT) to provide a non-invasive approach to achieve a desired aesthetic effect.
The LipoSonix system is specifically indicated for non-invasive waist circumference reduction.
Prescription Use
AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
NilkRedden for mkm
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of
510(k) Number K112626
Section 18 - Indications for Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.