← Product Code [LLZ](/productcode/LLZ) · K112570

# CENTRICITY CARDIO IMAGING (K112570)

_Ge Healthcare Integrated IT Solutions · LLZ · Nov 30, 2011 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K112570

## Device Facts

- **Applicant:** Ge Healthcare Integrated IT Solutions
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Nov 30, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Centricity* Cardio Imaging is a software only Picture Archiving and Communication System (PACS). It will be sold as a software only device to operate on general purpose computing hardware. Centricity* Cardio Imaging receives medical images and other information from various data sources. Information can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Centricity* Cardio Imaging is intended to assist trained professionals in the viewing, analysis, and diagnostic interpretation of images and other information for the diagnosis and treatment of cardiac and vascular disease. These trained professionals include but are not limited to physicians, cardiologists, radiologists, nurses, medical technicians, and assistants.

## Device Story

Centricity Cardio Imaging is a web-enabled, software-only PACS platform for cardiology. It ingests medical images and data from various sources, providing tools for storage, communication, processing, and display across distributed networks. The system comprises a web-accessed client for worklists, viewing, and reporting, and a server for database management, authentication, and hanging protocols. It operates on off-the-shelf general-purpose computing hardware. Used by clinicians in clinical settings (not patient vicinity) to facilitate diagnostic interpretation of cardiac and vascular images. It leverages clinical toolsets from predicate devices to assist in clinical decision-making, potentially improving diagnostic efficiency and patient management.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical testing, including risk analysis, requirements/design reviews, unit-level verification, integration testing, performance testing, safety testing, and simulated use validation.

## Technological Characteristics

Software-only PACS; operates on general-purpose off-the-shelf computing hardware. Web-enabled client-server architecture. Provides tools for ECHO/NIPV image analysis and data management. Connectivity via computer networks. Not intended for patient vicinity use.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Centricity PACS-IW
- TomTec Imaging System's Image Arena

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV 3 0 2011

K112570

Image /page/0/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The border has decorative flourishes around the circle.

### 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date: August 31, 2011

Submitter: GE Healthcare 540 West Northwest Highway Barrington, IL 60010

John Manarik Primary Contact Person: Requlatory Affairs Manager GE Healthcare Phone: 847-227-5504 Fax: 847-277-5240

- Secondary Contact Jeme Wallace Person: Requlatory Affairs Director GE Healthcare Phone: 847-277-4468 Fax: 847-939-1446
- Centricity* Cardio Imaging Device: Trade Name:

Common/Usual Name: Picture Archiving and Communication System

Classification Names: 21 CFR 892.2050, System, Image Processing, Radiological

> Product Code: LLZ

Predicate Device(s): GE Healthcare's Centricity* PACS-IW and TomTec Imaging System's Image Arena

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular border. The letters and the border are in white, set against a black background, creating a high-contrast and recognizable emblem.

Device Description: Centricity* Cardio Imaging is a web enabled cardiology Picture Archiving and Communication System that offers highly integrated imaging, workflow and information management in a single platform. Centricity* Cardio Imaging leverages a rich ECHO/NIPV clinical toolset and the data management capability from its predicate devices.

> Centricity* Cardio Imaging is a software only medical device comprised of a client and server. The client is web-accessed and provides the user-facing functions such as the work list, viewing, and reporting. The server provides background functions such as data storage, data transfer, database management, application deployment, user authentication, user profiles, licensing, and hanging protocols.

> Centricity* Cardio Imaaina is a software only medical device intended for use with commercially available off the shelf hardware.

- Intended Use: Centricity* Cardio Imaging is a software only Picture Archiving and Communication System (PACS). It will be sold as a software only device to operate on general purpose computing hardware. Centricity* Cardio Imaging receives medical images and other information from various data sources. Information can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations.
Centricity* Cardio Imagina is intended to assist trained professionals in the viewing, analysis, and diagnostic interpretation of images and other information for the diagnosis and treatment of cardiac and vascular disease. These trained professionals include but are not limited to physicians, cardiologists, radiologists, nurses, medical technicians, and assistants.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular frame. The frame has four stylized protrusions, one at the top, bottom, left, and right, giving it a slightly decorative appearance.

Centricity* Cardio Imaging is not intended for use in the patient vicinity.

Technology: Centricity* Cardio Imaging employs the same fundamental scientific technology as its predicate devices, Centricity* PACS IW and Tomtec Imaging Systems Image Arena.

> Centricity* Cardio Imaging receives medical images and other information from various data sources, and provides the capability to analyze echocardiography and NIPV images the same as its predicate devices.

> Information can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations the same as its predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular shape. The letters and the circle are in black, set against a white background. 

#### Summary of Non-Clinical Tests Determination of

Substantial Equivalence:

Centricity* Cardio Imaging and its applications comply with voluntary standards as detailed in this premarket notification submission. The following quality assurance measures were applied to the development of the system:

- . Risk Analysis
- Requirements Reviews ●
- Design Reviews .
- Testing on unit level (Verification) ●
- Integration testing (Verification) �
- � Performance testing (Verification)
- Safety testing (Verification) .
- . Simulated use testing (Validation)

## Summary of Clinical Tests:

The subject of this premarket notification submission, Centricity* Cardio Imaging, did not require clinical studies to support substantial equivalence.

GE Healthcare considers the Centricity* Cardio Imaging Conclusion: product to be as safe, as effective and substantially equivalent in performance to its predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human profiles facing to the right, stacked on top of each other.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. John Manarik Regulatory Affairs Manager GE Healthcare HCIT 540 West Northwest Highway BARRINGTON IL 60010

NOV 3 0 2011

Re: K112570

Trade/Device Name: Centricity Cardio Imaging Regulation Number: 21 CFR 872.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 31, 2011 Received: September 6, 2011

Dear Mr. Manarik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{5}------------------------------------------------

Page 2

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Pastel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known):

Centricity Cardio Imaging Device Name:

Indications for Use:

Centricity* Cardio Imaging is a software only Picture Archiving and Communication System (PACS). It will be sold as a software only device to operate on general purpose computing hardware. Centricity* Cardio Imaging receives medical images and other information from various data sources. Information can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations.

Centricity* Cardio Imaging is intended to assist trained professionals in the viewing. analysis, and diagnostic interpretation of images and other information. These trained professionals include but are not limited to physicians, cardiologists, radiologists, nurses, medical technicians, and assistants.

Prescription Use Yes (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use No (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K 112576

Page 1 of

---

**Source:** [https://fda.innolitics.com/device/K112570](https://fda.innolitics.com/device/K112570)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
