← Product Code [DRG](/productcode/DRG) · K112559

# MEDAPPS 2.0 REMOTE PATIENT MONITORING SYSTEM (K112559)

_Medapps, Inc. · DRG · Dec 2, 2011 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K112559

## Device Facts

- **Applicant:** Medapps, Inc.
- **Product Code:** [DRG](/productcode/DRG.md)
- **Decision Date:** Dec 2, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2910
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The MedApps 2.0 - Remote Patient Monitoring System consists of a patient device, MedApps HealthPAL, a mobile over-the-counter wireless communication hub, or MedApps HealthAIR, a portable over-the-counter wireless communication hub, which connects to commercially available glucose meters, scales, blood pressure monitors and pulse oximeters and HealthCOM, MedApps' secure host server system. MedApps Remote Patient Monitoring devices receive and store measurements collected from the described monitors, either wirelessly (HealthPAL) or tethered (HealthPAL or HealthAIR). MedApps devices do not alter the indicated use of the peripheral monitors that they integrate with. MedApps devices indicate successful or failed reception and transmission of data with visual and audio cues (HealthPAL via OLED display screen, verbal message and audio tones; HealthAIR via LED lights and audio tones). MedApps devices store collected data and transmit to HealthCOM using commercially available, FCC compliant, wireless telecommunication protocols (including but not limited to cellular GSM, CDMA and WiMax). Healthcare professionals can review the transmitted information within the MedApps HealthCOM system, set thresholds to flag readings based on specific thresholds being exceeded. In addition, the MedApps Interactive Voice Response (IVR) has the ability to contact the patient remotely and use pre-approved ("canned") educational or reminder messages. ("Your nurse would like to talk to you, can I connect you now", "We haven't received a reading from you today, please send us your readings"). The MedApps 2.0 - Remote Patient Monitoring System is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. The data is made available to the patients when time-critical care is not required. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.

## Device Story

System acts as communication hub for home-use medical devices (glucose meters, scales, BP monitors, pulse oximeters). Inputs: biometric data via wired (USB/Smart Cable) or wireless (Bluetooth) connections. Transformation: firmware captures, stores, and transmits data via embedded M2M cellular module to secure HealthCOM server. Outputs: visual/audio status cues on device; data availability on HealthCOM web portal for healthcare providers. IVR system provides automated voice/text reminders and educational content. Used in home/remote settings by patients; reviewed by clinicians. Benefits: facilitates remote data collection for chronic condition management; improves adherence via reminders. Not for real-time monitoring or diagnosis.

## Clinical Evidence

No clinical data. Substantial equivalence based on bench testing, design control verification, and validation of software functionality, risk management, and adherence to consensus standards (e.g., 60601-1, 60601-1-2).

## Technological Characteristics

Plastic enclosure; OLED (HealthPAL) or LED (HealthAIR) display; embedded M2M cellular module; Bluetooth V2.0; wired USB/Smart Cable interfaces. Power: wall plug (both) and rechargeable battery (HealthPAL only). Connectivity: cellular (GSM/CDMA/WiMax). Standards: 60601-1 (Safety), 60601-1-2 (EMC/EMI), FCC, PTCRB, CTIA, ETSI.

## Regulatory Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

## Predicate Devices

- Intel Health Guide PHS6000 ([K080798](/device/K080798.md))
- Confidant 2.5 ([K072698](/device/K072698.md))
- MedApps Remote Patient Monitoring System (D-PAL) ([K062377](/device/K062377.md))
- MedApps 2.0 - Remote Patient Monitoring System ([K083862](/device/K083862.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

EXHIBIT 02

# PREMARKET NOTIFICATION 510(k) SUMMARY As required by 21 CFR §807.92(c)

#### Submitter

| 510(k) Owner:         | MedApps, Inc.                                             |
|-----------------------|-----------------------------------------------------------|
| Owner / Operator:     | 10027842                                                  |
| Registration:         | 3005916763                                                |
| Address:              | 7975 North Hayden Road, Suite A-203, Scottsdale, AZ 85258 |
| Telephone:            | 480-305-6323                                              |
| Fax Number:           | 480-393-1892                                              |
| Contact Person:       | Kent Dicks                                                |
| Contact Person Title: | Founder / CEO                                             |
| Date Prepared:        | August 31, 2011                                           |

### Device Information

MedApps 2.0 - Remote Patient Monitoring System Trade Name: Remote Patient Monitoring System Common Name: Classification Status: Class II per regulations 870.2910 Classification Name: Transmitters and Receivers, Physiological Signal, Radiofrequency (21 CFR 870.2910, Product Code DRG)

#### LEGALLY MARKETED PREDICATE DEVICE A.

Legally marketed predicate devices are: K080798 Intel Health Guide PHS6000 K072698 Confidant 2.5 MedApps Remote Patient Monitoring System (D-PAL) K062377 MedApps 2.0 - Remote Patient Monitoring System K083862

#### B. INDICATIONS FOR USE

The MedApps 2.0 - Remote Patient Monitoring System consists of a patient device, MedApps HealthPAL, a mobile over-the-counter wireless communication hub, or MedApps HealthAIR, a portable over-the-counter wireless communication hub, which connects to commercially available glucose meters, scales, blood pressure monitors and pulse oximeters and HealthCOM, MedApps' secure host server system.

Remote Patient Monitoring devices receive and store MedApps measurements collected from the described monitors, either wirelessly (HealthPAL) or tethered (HealthPAL or HealthAIR). MedApps devices do not alter the indicated use of the peripheral monitors that they integrate with. MedApps devices indicate successful or failed reception and transmission of data with visual and audio cues (HealthPAL via OLED display screen, verbal message and audio tones; HealthAIR via LED lights and audio tones). MedApps devices store collected data and transmit to

Page 1 of 8

{1}------------------------------------------------

HealthCOM using commercially available, FCC compliant, wireless telecommunication protocols (including but not limited to cellular GSM, CDMA and WiMax).

Healthcare professionals can review the transmitted information within the MedApps HealthCOM system, set thresholds to flag readings based on specific thresholds being exceeded. In addition, the MedApps Interactive Voice Response (IVR) has the ability to contact the patient remotely and use pre-approved ("canned") educational or reminder messages. ("Your nurse would like to talk to you, can I connect you now", "We haven't received a reading from you today, please send us your readings").

The MedApps 2.0 - Remote Patient Monitoring System is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. The data is made available to the patients when time-critical care is not required. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.

#### MedApps 2.0 SYSTEM DESCRITPION C.

The MedApps 2.0 - Remote Patient Monitoring System consists of:

- (1) MedApps HealthPAL hardware:
The physical component of the MedApps HealthPAL is an electronic device contained in a plastic enclosure with an OLED screen, built-in M2M cellular chip, speaker, smart cable connection, smart cables, wireless module, LED lights to indicate activity, timer button to assist patients with their reading schedule (i.e. remind them to take their reading in X minutes), last reading button, volume up and down buttons.

- MedApps HealthPAL firmware / software: (2)
The firmware captures data from commercially available health monitors, and stores and transmits the information to the MedApps HealthCOM server, via the embedded communication chip / platform.

The firmware allows HealthPAL to receive information via wire or via embedded wireless module from accessory medical devices that are compatibly wireless enabled, which have been paired to the MedApps HealthPAL.

The firmware has many additional functions including:

Download of user profiles from the server to configure HealthPAL . remotely.

HealthPAL has audio capability to deliver verbal announcement of . readings and acknowledgment of data transmission from all connected accessory medical devices, time settings, volume control, educational content and reminders, in any language that is loaded to the device.

Timer capability, activated by the user to provide assistance with . adhering to a reading schedule (reminders to take readings within a set timeframe).

Page 2 of 8

{2}------------------------------------------------

OLED screen displays information regarding the HealthPAL's status . including battery level, volume level, data transmission status, transmission pending indicator, activity icons / messages and other information to provide ease of use and promote patient adherence; as well as information received from accessory medical devices, such as the type of device, measurement, date and time of the last reading collected.

Battery charging, isolation circuits, and interfaces to individual . accessory medical devices / protocols via the smart cable.

MedApps HealthAIR hardware / software: (3)

MA020 HealthAIR is a modified MA105 HealthPAL device. The physical component of the MedApps HealthAIR is an electronic device contained in a plastic enclosure with built-in M2M cellular chip, speaker, standard USB cable and USB Smart Cable connection, and LED lights to indicate activity regarding the receiving and transmitting of collected data.

Like the HealthPAL, HealthAIR's firmware / software captures, data from commercially available retail health monitors, and stores and transmits information to the MedApps HealthCOM server, via the embedded communication chip / platform.

The firmware allows HealthAIR to receive information via wire, either standard USB or with a MedApps USB Smart Cable, from accessory medical devices.

The firmware has many additional functions including:

Download of user profiles from the server to configure HealthAIR . remotely.

HealthAIR's Audio feature uses audio tones to indicate . acknowledgment of collected readings from all connected accessory medical devices as well as reading transmission via the cellular network.

HealthAIR's visual user interface utilizes LED lights of collected . readings from all attached medical devices as well as reading transmission acknowledgements (via).

(5) MedApps HealthCOM software application:

The HealthCOM software application allows caregivers access to review patient data collected from accessory medical devices using MedApps hardware on the secure HealthCOM website. HealthCOM software allows professional caregivers to set patient readings.

HealthCOM software also allows the patient to establish an account and to direct / authorize their data to be directed to an outside, validated Personal Health Record (PHR), Electronic Health Record or Medical Record (EHR or EMR).

Page 3 of 8

{3}------------------------------------------------

- MedApps IVR software application: (6)
The IVR (Interactive Voice Response) software application provides the ability to contact the patient remotely, by phone (designated in the user profile), and executes an pre-approved ("canned") scripts to deliver preapproved ("canned") reminder messages ("Your nurse would like to talk to you, can I connect you now", "We haven't received a reading from you today, please send us your readings"), educational content and gather survey information.

In addition, the MedApps IVR application will send out Email, SMS / Text Messages, Paging, IM and other forms of communications in order to contact patients or caregivers. This will include reminders and alerts, based on clinically defined parameters / thresholds established in HealthCOM by the professional care provider.

The MedApps 2.0 - Remote Patient Monitoring System is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. The data is made available to the patients when time-critical care is not required. - The device is contraindicated for patients requiring direct medical supervision or emergency intervention.

| Feature                                                                                    | Intel Health<br>Guide PHS6000<br>K080798                                                                     | Confidant 2.5<br>K072698                                                               | MedApps<br>Submission<br>(HealthPAL &<br>HealthCOM)<br>K083862                                                                    | MedApps<br>Submission<br>(HealthAIR &<br>HealthCOM)<br>K112559                                                                                  |
|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications of<br>Use                                                                      | Enables<br>healthcare<br>providers to<br>monitor and<br>manage chronic<br>conditions of<br>patients remotely | Same                                                                                   | Same                                                                                                                              | Same                                                                                                                                            |
| Intended Use                                                                               | Telemedicine<br>System                                                                                       | Same                                                                                   | Same                                                                                                                              | Same                                                                                                                                            |
| Intended Users                                                                             | Home users and<br>Healthcare<br>providers                                                                    | Same                                                                                   | Same                                                                                                                              | Same                                                                                                                                            |
| Site of Use                                                                                | Home, Clinic                                                                                                 | Same                                                                                   | Same                                                                                                                              | Same                                                                                                                                            |
| Data Collection<br>Software                                                                | Intel Care<br>Management<br>Suite Software                                                                   | The Hermes<br>Proprietary<br>Software                                                  | MedApps<br>Proprietary<br>Software                                                                                                | MedApps<br>Proprietary<br>Software                                                                                                              |
| Data Collection<br>Software<br>Functionality                                               | Transmit data<br>from Sensor<br>devices to Central<br>Database                                               | Same                                                                                   | Same                                                                                                                              | Same                                                                                                                                            |
| Communication<br>method of hub<br>with Central<br>Server                                   | Via DSL or Phone<br>Line Connection                                                                          | Via Cellular<br>Phone                                                                  | Via Embedded<br>Cellular<br>Technology                                                                                            | Via Embedded<br>Cellular<br>Technology                                                                                                          |
| Feature                                                                                    | Intel Health<br>Guide PH\$6000<br>K080798                                                                    | Confidant 2.5<br>K072698                                                               | MedApps<br>Submission<br>(HealthPAL &<br>HealthCOM)<br>K083862                                                                    | MedApps<br>Submission<br>(HealthAIR &<br>HealthCOM)<br>K112559                                                                                  |
| Types of sensors<br>which can be<br>interfaced (wired<br>or wirelessly) to<br>receiver hub | Medical Devices<br>designed for<br>Home use:<br>Glucose<br>Scale<br>Blood Pressure<br>Pulse Ox<br>Peak Flow  | Medical<br>Devices<br>designed for<br>Home use:<br>Glucose<br>Scale<br>Blood Pressure  | Medical Devices<br>designed for Home<br>use:<br>Glucose<br>Scale<br>Blood Pressure<br>Pulse Ox                                    | Medical Devices<br>designed for Home<br>use:<br>Glucose<br>Scale<br>Blood Pressure<br>Pulse Ox                                                  |
| Implementation<br>method of<br>collecting data<br>from sensors                             | Short range radio<br>system using<br>Wireless<br>(Bluetooth) and<br>Wired (tethered)<br>cables.              | Short range<br>radio system<br>using<br>Bluetooth                                      | Short range radio<br>system using<br>Wireless<br>(Bluetooth) and<br>Wired (tethered)<br>cables.                                   | Currently using<br>Wired (tethered)<br>cables (USB),<br>Smart Cables.                                                                           |
| Sensor Software                                                                            | Sensor Software<br>unchanged                                                                                 | Same                                                                                   | Same                                                                                                                              | Same                                                                                                                                            |
| Connectivity                                                                               | Short range radio<br>system using<br>Bluetooth and<br>Wired (tethered)<br>cables.                            | Short range<br>radio system<br>using<br>Bluetooth                                      | Short range radio<br>system using<br>Bluetooth and<br>Wired (tethered)<br>cables.                                                 | Currently using<br>Wired (tethered)<br>cables<br>Future capability to<br>use Bluetooth<br>dongles.                                              |
| Communication<br>method of hub<br>with devices                                             | Short range radio<br>system using<br>Wireless<br>(Bluetooth) and<br>Wired (tethered)<br>cables.              | Short range<br>radio system<br>using Wireless<br>(Bluetooth)                           | Short range radio<br>system using<br>Wireless<br>(Bluetooth) and<br>Wired (tethered)<br>cables.                                   | Currently using<br>Wired (tethered)<br>cables.                                                                                                  |
| Communications<br>Protocol                                                                 | Wireless<br>(Bluetooth) V2.0<br>& Wired<br>(Tethered)                                                        | Wireless<br>(Bluetooth)<br>V2.0                                                        | Wireless<br>(Bluetooth) V2.0<br>and Wired<br>(Tethered)                                                                           | Wired (Tethered)                                                                                                                                |
| Communication<br>Frequency                                                                 | Bluetooth : 2.402<br>to 2.480 GHz                                                                            | Bluetooth :<br>2.402 to 2.480<br>GHz<br>GSM: 850/<br>900 / 1800 /<br>1950 Mhz          | Bluetooth : 2.402<br>to 2.480 GHz<br>GSM: 850 / 900 /<br>1800 / 1950 Mhz                                                          | GSM: 850 / 900 /<br>1800 / 1950                                                                                                                 |
| Power Source                                                                               | Wall power plug<br>(120 VAC/50-60)                                                                           | Wall power<br>plug (120<br>VAC/50-60)<br>and<br>Rechargeable<br>Batteries in<br>Device | Wall power plug<br>(120 VAC/50-60)<br>or Rechargeable<br>Batteries in<br>HealthPAL                                                | Wall power plug<br>(120 VAC/50-60)                                                                                                              |
| Visual Feedback<br>/ Display                                                               | On devices and<br>hub, and monitors<br>connected to<br>central server                                        | Same                                                                                   | OLED for<br>HealthPAL                                                                                                             | HealthAIR uses<br>LED light indicators                                                                                                          |
| Communication<br>with Patients                                                             | On screen display                                                                                            | Same                                                                                   | Audio/visual<br>reading feedback<br>on screen and by<br>speaker; and<br>Interactive Voice<br>Response (IVR)<br>System for patient | Audio/visual<br>reading feedback<br>from LED light<br>indicators & audio<br>tones; Interactive<br>Voice Response<br>(IVR) system for<br>patient |

#### TECHNOLOGICAL CHARACTERISTICS SUMMARY - as required by D. 807.92(a)(6)

Page 4 of 8

{4}------------------------------------------------

# MedApps, Inc.
510(k) SUMMARY

# EXHIBIT 02 KII2559

Page 5 of 8

{5}------------------------------------------------

## Data Collection:

The 2 predicate devices and the MedApps solution connect to medical devices (designed for home use) by either wired (cable) connection or wireless (HealthPAL- Bluetooth). The data is collected from the devices and sent to a secure central server using various communication methods.

## Telecommunication Platform to Central Server:

Intel Health uses DSL connectivity (wired point of care), Confidant uses an off-the-shelf Cellular Phone; MedApps uses embedded Machine to Machine (M2M) module to transmit data via cellular connectivity.

## Patient Feedback Technology:

The 2 predicate devices and the MedApps solution allow data and messages to be displayed on a screen (for the HealthPAL) for the patient to read and acknowledge. HealthAIR uses audio and visual acknowledgement / feedback. The MedApps solution also uses an Interactive Voice Response (IVR) system in order to call the patient and ask questions, gather survey information, or issue reminders.

## Backend Data Storage:

All systems (both 2 predicate devices and the MedApps solution), provide a backend system that allows data to be stored, and healthcare professionals to access and monitor collected patient data.

#### NON-CLINICAL PERFORMANCE DATA TESTING AND REVIEW - as E. required by 807.92(b)(1)

#### Non-Clinical Testing

The submitted 2.0 System has undergone MedApps' design control verification and validation testing. MedApps 2.0 validation testing include testing of all executable code and functionality and confirmation that all identified risks have been adequately addressed by software functionality, the user interface, documentation or user SOP.

MedApps 2.0 System verification and validation activities as part of the design control process include testing of all Design Specifications (Design Control Inputs) based on risk analysis, certification standards, and Verification plans. MedApps Product Verification and Release Plan execution on both HealthPAL and HealthAIR ensures both medical devices work with each type of user accessory medical device (glucose, blood pressure monitor, scale, and pulse oximeter) as part of the MedApps 2.0 System including integration to HealthCOM backend software application. The output of these design control verification analysis documents MedApps 2.0 - Remote Patient Monitoring System shall meet its requirements and design specifications as intended.

Lastly, MedApps has used its Risk Management Plan to perform risk analysis comparing the current MA105 HealthPAL device to the modified MA020 HealthAIR device regarding residual risks, control analysis risks, and human management factors for usability to determine that no significant risks were added by allowing either the MA105 HealthPAL or the MA020 HealthAIR to be functionally used as part of the MedApps 2.0 System.

Page 6 of 8

{6}------------------------------------------------

Lastly, all relevant certification testing such as EMC (60601-1-2) and Safety (60601-1) are described in MedApps' Declaration of Conformity.

#### SUBSTANTIAL EQUIVALENT F.

The MedApps 2.0 Remote Patient Monitoring System is substantially equivalent to the predicate devices in terms of data collection software functionality, operating system for the patient device, communication method of patient device with central server, types of sensors which can be integrated to the patient medical device, implementation methods of collecting data from sensors, sensor software, connectivity, communication protocol, power source and general display method.

The HealthAIR communication hub device is substantially equivalent to the HealthPAL (described in 510k K083862) as both devices, as part of the MedApps 2.0 System, connect to commercially available Glucose Meters, Scales, Blood Pressure Monitors and Pulse Oximeters; data is collected, stored and transmitted using off-the-shelf FCC approved wireless / cellular connectivity. Both provide audio and visual feedback / acknowledgement that readings have been collected and transmitted to MedApps' secure host server called

| Feature                                                                                 | MA105 HealthPAL                                                                                    | MA020 HealthAIR                                                      |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Indications of Use                                                                      | Enables healthcare providers to<br>monitor and manage biometric<br>patient data collected remotely | Same                                                                 |
| Intended Use                                                                            | Telemedicine System                                                                                | Same                                                                 |
| Intended Users                                                                          | Home users and patients outside<br>of the clinical setting, as well as<br>Healthcare providers     | Same                                                                 |
| Site of Use                                                                             | Remote setting (e.g. Home /<br>Work), Clinic                                                       | Same                                                                 |
| Data Collection<br>Software & firmware                                                  | MedApps Proprietary Software                                                                       | Same                                                                 |
| Data Collection<br>Software Functionality                                               | Transmit data from Sensor<br>devices to Central Database                                           | Same                                                                 |
| Communication<br>method of device hub<br>with Central Server                            | Via Embedded Cellular<br>Technology (GSM or CDMA)                                                  | Same                                                                 |
| Types of sensors which<br>can be interfaced<br>(wired or wirelessly)<br>to receiver hub | Medical Devices designed for<br>Home use:<br>Glucose, Scale, Blood Pressure<br>Pulse Ox            | Same                                                                 |
| Transmission                                                                            | Transmits information to the<br>MedApps secure host server<br>called "HealthCOM"                   | Same                                                                 |
| Implementation<br>method of collecting<br>data from sensors and<br>general Connectivity | Short range radio system using<br>Wireless (Bluetooth) and Wired /<br>tethered (Smart Cables).     | HealthAIR uses wired /<br>tethered connection<br>(USB, Smart Cables) |

Below is a Technological Characteristics Summary between the HealthPAL and the HealthAIR medical devices:

Page 7 of 8

{7}------------------------------------------------

## MedApps, Inc. 510(k) SUMMARY

## EXHIBIT 02 ।। ১২২২৪

| Communication<br>method of hub with<br>devices | Short range radio system using<br>Wireless (Bluetooth) and Wired /<br>tethered (Smart Cables).                          | HealthAIR uses wired /<br>tethered connection<br>(USB, Smart Cables)                                                            |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Communication<br>Frequency                     | Bluetooth : 2.402 to 2.480 GHz<br>GSM: 850 / 900 / 1800 / 1950<br>Mhz                                                   | GSM: 850 / 900 / 1800 /<br>1950 Mhz                                                                                             |
| Power Source                                   | Wall power plug (120 VAC/50-<br>60) and Rechargeable Batteries<br>in Device                                             | Same Wall power plug but<br>HealthAIR does not have<br>a rechargeable battery                                                   |
| Device Communication<br>with Patients          | On screen display and audio<br>voice feedback                                                                           | LEDs lights for visual<br>feedback and audio tones<br>(beeps).                                                                  |
| Certification Testing                          | Safety 60601-1, EMC/EMI/FCC<br>(60601-1-2), ESD & Radiated<br>Immunity, FCC Bluetooth,<br>(PTCRB), CTIA (battery), ETSI | Safety 60601-1,<br>EMC/EMI/FCC (60601-1-<br>2), ESD & Radiated<br>Immunity, (PTCRB), ETSI<br>(See Declaration of<br>Conformity) |

#### SAFETY AND EFFICACY G.

The MedApps 2.0 Remote Patient Monitoring System does not rely on an assessment of clinical performance data. The device conforms to FDA's recognized consensus standards and relies on its conformity to demonstrate its safety and efficacy. The device does not introduce any new questions concerning the safety or efficacy and is, therefore, substantially equivalent to the predicate devices.

Page 8 of 8

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of an eagle or other bird with outstretched wings.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC - 2 2011

MedApps, Inc. c/o Mr. Kent Dicks Founder / CEO 7975 North Hayden Road, Suite A-203 Scottsdale. AZ 85258

Re: K112559

Trade/Device Name: Modification to MedApps 2.0 – Remote Patient Monitoring System Regulation Number: 21 CFR 870.2910 Regulation Name: Transmitters and Receivers, Physiological Signal, Radiofrequency Regulatory Class: Class II (two) Product Code: DRG Dated: October 25, 2011 Received: November 3. 2011

Dear Mr. Dicks:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{9}------------------------------------------------

Page 2 - Mr. Kent Dicks

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

## MedApps, Inc. (FDA-IU-8014 Rev C) STATEMENT OF INDICATIONS FOR USE

510(k) Number: K112559

Preparation Date: August 31, 2011

Modification to MedApps 2.0 - Remote Patient Monitoring Device Name: System

Indications For Use:

The MedApps 2.0 - Remote Patient Monitoring System consists of a patient device, MedApps HealthPAL, a mobile over-the-counter wireless communication hub, or MedApps HealthAIR, a portable over-the-counter wireless communication hub, which connects to commercially available qlucose meters, scales, blood pressure monitors and pulse oximeters and HealthCOM, MedApps' secure host server system.

MedApps Remote Patient Monitoring devices receive and store measurements collected from the described monitors, either wirelessly (HealthPAL) or tethered (HealthPAL or HealthAIR). MedApps devices do not alter the indicated use of the peripheral monitors that they integrate with. MedApps devices indicate successful or failed reception and transmission of data with visual and audio cues (HealthPAL via OLED display screen, verbal message and audio tones; HealthAIR via LED lights and audio tones). MedApps devices store collected data and transmit to HealthCOM using commercially available, FCC compliant, wireless telecommunication protocols (including but not limited to cellular GSM, CDMA and WiMax),

Healthcare professionals can review the transmitted information within the MedApps HealthCOM system, set thresholds to flaq readings based on specific thresholds being exceeded. In addition, the MedApps Interactive Voice Response (IVR) has the ability to contact the patient remotely and use pre-approved ("canned") educational or reminder messages. ("Your nurse would like to talk to you, can I connect you now", "We haven"t received a reading from you today, please send us your readings").

The MedApps 2.0 - Remote Patient Monitoring System is not intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. The data is made available to the patients when time-critical care is not required. The device is contraindicated for patients requiring direct medical supervision or emergency intervention.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use<br>(Per 21 CFR 801.109) |   |
|------------------------------------------|---|
| OR Over-The-Counter Use                  | X |

(Division Sign-Off)
Division of Cardiovascular Devices

| 510(k) Number | K112559 |
|---------------|---------|
|---------------|---------|

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K112559](https://fda.innolitics.com/device/K112559)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
