ELECSYS T4 CALCHECK 5

K112528 · Roche Diagnostics Corporation · JJX · Oct 18, 2011 · Clinical Chemistry

Device Facts

Record IDK112528
Device NameELECSYS T4 CALCHECK 5
ApplicantRoche Diagnostics Corporation
Product CodeJJX · Clinical Chemistry
Decision DateOct 18, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

The Elecsys T4 CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys T4 reagent on the indicated Elecsys and cobas e immunoassay analyzers, for in vitro diagnostic use only.

Device Story

Elecsys T4 CalCheck 5 is a lyophilized, multi-level control material consisting of human serum matrix spiked with synthetic L-Thyroxine (T4). It is used in clinical laboratory settings on Elecsys and cobas e immunoassay analyzers. The device is reconstituted with distilled or deionized water by laboratory technicians. It serves as a tool for calibration verification and assay range verification; it is not a primary calibrator or routine control. By comparing analyzer results against the assigned target values provided for each lot, clinicians verify the accuracy and reportable range of the T4 assay, ensuring reliable thyroid function testing for patient diagnosis and management.

Clinical Evidence

No clinical data. Bench testing only. Value assignment was performed using a minimum of three Elecsys 2010/cobas e411 and MODULAR ANALYTICS E170/cobas e601 analyzers, with each sample tested in duplicate for at least 6 measurements. Target values are the median of these determinations. Stability testing supports a 36-month shelf-life at 2-8°C and 4-hour open-vial stability at 20-25°C.

Technological Characteristics

Lyophilized human serum matrix spiked with synthetic L-Thyroxine. Five levels of concentration. Reconstituted with 1.0 mL distilled/deionized water. Standardized against ID-GC/MS. Compatible with Elecsys 2010, cobas e411, MODULAR ANALYTICS E170, cobas e601, and cobas e602 analyzers.

Indications for Use

Indicated for use as an assayed control for calibration verification and assay range verification of the Elecsys T4 reagent on Elecsys and cobas e immunoassay analyzers. No specific patient population or contraindications described.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K11a528 OCT 1 8 2011 | 510(k) Summary | OCT 1 8 2011 | |----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Introduction | According to the requirements of 21 CFR 807.92, the following information<br>provides sufficient detail to understand the basis for a determination of<br>substantial equivalence. | | Submitter<br>name, address,<br>contact | Roche Diagnostics<br>9115 Hague Road, P.O. Box 50416<br>Indianapolis, IN 46250-0416<br>317-521-3577 | | | Contact Person: Michael Leuther, Ph.D.<br>Phone: 317-521-3930<br>Fax: 317-521-2324<br>Email: michael.leuther@roche.com | | | Secondary Contact: Stephanie Greeman<br>Phone: 317-521-2458<br>Fax: 317-521-2324<br>Email: stephanie.greeman@roche.com | | | Date Prepared: August 29, 2011 | | Device Name | Proprietary name: Elecsys T4 CalCheck 5<br>Common name: T4 CalCheck 5<br>Classification name: Single (specified) analyte controls (assayed and<br>unassayed) | | Predicate<br>device | The Elecsys T4 CalCheck 5 is substantially equivalent to other products in<br>commercial distribution intended for similar use. We claim equivalency to the<br>currently marketed Elecsys T3 CalCheck 5 (K111552). | | Device<br>Description | The Elecsys T4 CalCheck 5 is a lyophilized product consisting of T4 in<br>human serum matrix. During manufacture, the analyte is spiked into the<br>matrix at the desired concentration levels. | | Intended use | The Elecsys T4 CalCheck 5 is an assayed control for use in calibration<br>verification and for use in the verification of the assay range established by<br>the Elecsys T4 reagent on the indicated Elecsys and cobas e immunoassay<br>analyzers, for in vitro diagnostic use only. | Continued on next page {1}------------------------------------------------ # 510(k) Summary, Continued #### The table below compares Elecsys T4 CalCheck 5 with the predicate device, Comparison Table Elecsys T3 CalCheck 5 (K111552). The predicate shows that T4 CalCheck 5 is substantially equivalent to T3 CalCheck 5, with several key similarities. The shaded fields indicate similar characteristics between the candidate device and the predicate device. | Characteristic | Elecsys T3 CalCheck 5<br>(Predicate Device, K111552) | Elecsys T4 CalCheck 5<br>(Candidate Device) | |-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The Elecsys T3 Calcheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys T3 quantitative assay reagent on the indicated Elecsys and cobas e immunoassay analyzers. | The Elecsys T4 Calcheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys T4 reagent on the indicated Elecsys and cobas e immunoassay analyzers. | | Analyte | Triiodothyronine (T3) | Thyroxine (T4) | | Levels | Five | Five | | Assay Measuring Range | 0.300 - 10.0 nmol/L | 5.40 - 320 nmol/L | | Check Target Values | Check 1: ≤ 0.2 nmol/L<br>Check 2: 2.0 nmol/L<br>Check 3: 5 nmol/L<br>Check 4: 8 nmol/L<br>Check 5: 10 nmol/L | Check 1: <0.78ug/dL, <10 nmol/L<br>Check 2: 7.77ug/dL, 100 nmol/L<br>Check 3: 12.43ug/dL, 160 nmol/L<br>Check 4: 19.42ug/dL, 250 nmol/L<br>Check 5: 24.86ug/dL, 320 nmol/L | | Format | Lyophilized | Lyophilized | | Handling | Reconstitute Check 1, Check 2, Check 3, Check 4, and Check 5 with exactly 1.0 mL distilled or deionized water. Allow to stand closed for 15 minutes, then mix gently by inversion. | Reconstitute Check 1, Check 2, Check 3, Check 4, and Check 5 with exactly 1.0 mL distilled or deionized water. Allow to stand closed for 15 minutes, then mix gently by inversion. | | Stability | Unopened:<br>• Store at 2-8°C until expiration date<br>Reconstituted:<br>• 20-25°C: 4 hours | Unopened:<br>• Store at 2-8°C until expiration date<br>Reconstituted:<br>• 20-25°C: 4 hours | | Matrix | Human Serum matrix | Check 1: BSA/Buffer matrix<br>Checks 2-5: Human serum | | Traceability | The assayed value of each CalCheck level was standardized against reference standards by weighing T3 into an analyte free | Thyroxine is traceable to the reference method ID-GC/MS using serum/plasma samples as reference material | ### Performance Characteristics The Elecsys T4 CalCheck 5 was evaluated for value assignment and stability. . . {2}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Silver Spring. MD 20993 OCT 1 8 2011 Roche Diagnostics c/o Dr. Michael Leuther Manager, Regulatory Affairs 9115 Hague Road, PO Box 50416 Indianapolis, IN 46250-0416 > k112528 Trade Name: Elecsys T4 CalCheck 5 Regulation Number: 21 CFR §862.1660 Regulation Name: Quality control material (assayed and unassayed). Regulatory Class: Class I, reserved Product Codes: JJX Dated: August 30, 2011 Received: August 31, 2011 Dear Dr. Leuther: Re: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ #### Page 2 - If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric' s (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Sincerely yours, signature Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for Use 510(k) Number (if known): Kll 2528 Device Name: Elecsys T4 CalCheck 5 Indication For Use: The Elecsys T4 CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys T4 reagent on the indicated Elecsys and cobas e immunoassay analyzers. Prescription Use __ X (21 CFR Part 801 Subpart D) Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) And/Or Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Qus Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K112528 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...