K112496 · Pioneer Surgical Technology, Inc. · MAX · Sep 28, 2011 · Orthopedic
Device Facts
Record ID
K112496
Device Name
PIONEER CROSSFUSE II IBF/VBR SYSTEM
Applicant
Pioneer Surgical Technology, Inc.
Product Code
MAX · Orthopedic
Decision Date
Sep 28, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pioneer IBF/VBR System, when used as an IBF implant, is indicated for intervertebral body fusion of the spine in skeletally mature patients. Pioneer IBFs are designed for use with autogenous bone graft to facilitate fusion. One device may be used per intervertebral space. Pioneer IBFs are intended to be used with supplemental spinal fixation cleared for the implanted level, such as the Quantum, Streamline, Contact ALP or SlimFuse systems. The Cervical IBF device in intended for use at one level in the cervical spine, from C3 to T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment. Lumbar IBFs are also intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Lumbar IBFs are to be used in patients who have had six months of non-operative treatment. The Pioneer IBF/VBR System, when used as a VBR implant, is intended for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. Pioneer VBRs are also indicated for treating fractures of the thoracic and lumbar spine. Pioneer VBRs are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period of time. The system must be used with the Pioneer® Quantum Spinal Fixation System or supplemental internal fixation systems cleared for the conditions listed above (i.e., tumor or trauma of T1-L5). Additionally, the Pioneer Vertebral Spacer implant is intended to be used with bone graft.
Device Story
Pioneer IBF/VBR System consists of spinal implants (interbody fusion/vertebral body replacement) used by surgeons in clinical settings; implants restore vertebral height and provide biomechanical stability; used for cervical/lumbar disc disease, tumors, or fractures; requires supplemental internal fixation and autogenous bone graft; implants available in various shapes/sizes to match patient anatomy; device provides structural support to spinal column; facilitates fusion or maintains integrity post-resection.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing only, including static compression, static torsion, subsidence, and expulsion testing.
Technological Characteristics
Implants manufactured from PEEK with tantalum or titanium alloy markers. System includes various shapes and sizes. Mechanical performance validated via static compression, static torsion, subsidence, and expulsion testing.
Indications for Use
Indicated for skeletally mature patients requiring intervertebral body fusion (cervical C3-T1 for disc disease; lumbar L2-S1 for DDD with Grade 1 spondylolisthesis) or vertebral body replacement (thoracolumbar T1-L5 for tumors or fractures). Requires supplemental fixation and autogenous bone graft.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
## 510(k) SUMMARY
# Pioneer IBF/VBR System
SEP 2 8 2011
| Sponsor: | Manufacturer | Pioneer Surgical Technology<br>375 River Park Circle<br>Marquette, MI 49855 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| | Official Contact<br>Phone:<br>Fax: | Emily M. Downs<br>(906) 225-5602<br>(906) 226-4459 |
| | Date prepared: | September 28, 2011 |
| Device Name: | Pioneer Interbody Fusion/Vertebral Body Replacement System | |
| Classification<br>Name: | Intervertebral Body Fusion Device / Vertebral Spacer Device | |
| Classification<br>Number: | 21 CFR 888.3080 - product codes MAX and ODP, class II | |
| | 21 CFR 888.3060 - product code MQP, class II | |
| Description: | The Pioneer IBF/VBR System is an implantable device manufactured from<br>PEEK and tantalum or titanium alloy that is available in a variety of different<br>shapes and sizes to suit the individual pathology and anatomical conditions<br>of the patient. | |
| | The purpose of the submission was to add the CrossFuse II device to the<br>system. | |
| Intended Use: | The Pioneer IBF/VBR System, when used as an IBF implant, is indicated for<br>intervertebral body fusion of the spine in skeletally mature patients. Pioneer<br>IBFs are designed for use with autogenous bone graft to facilitate fusion.<br>One device may be used per intervertebral space. Pioneer IBFs are intended<br>to be used with supplemental spinal fixation cleared for the implanted level,<br>such as the Quantum, Streamline, Contact ALP or SlimFuse systems. | |
| | The Cervical IBF device in intended for use at one level in the cervical spine,<br>from C3 to T1, for the treatment of cervical disc disease (defined as neck pain<br>of discogenic origin with degeneration of the disc confirmed by history and<br>radiographic studies). The cervical device is to be used in patients who have<br>had six weeks of non-operative treatment. | |
Lumbar 1BFs are also intended for use at either one level or two contiguous
{1}------------------------------------------------
levels in the lumbar spine, from L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Lumbar IBFs are to be used in patients who have had six months of non-operative treatment.
The Pioneer IBF/VBR System, when used as a VBR implant, is intended for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. Pioneer VBRs are also indicated for treating fractures of the thoracic and lumbar spine. Pioneer VBRs are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period of time. The system must be used with the Pioneer® Quantum Spinal Fixation System or supplemental internal fixation systems cleared for the conditions listed above (i.e., tumor or trauma of T1-L5). Additionally, the Pioneer Vertebral Spacer implant is intended to be used with bone graft.
- The Pioneer IBF/VBR System are composed of Polyether ether ketone (PEEK) Material: with tantalum or titanium alloy markers. The predicate device is composed of the same materials.
The indication for use and material composition of the Pioneer IBF/VBR Comparison to Predicate System are the same as the currently cleared predicate device, Pioneer Devices IBF/VBR System (K043206/K061151/K073177). The differences in dimensional specifications between the subject devices and the predicate were not considered significant based on mechanical testing.
For a determination of substantial equivalence, the following non-clinical Performance mechanical tests were performed: Data:
- Static Compression �
- Static Torsion ●
- Subsidence .
- Expulsion ●
Based on the supporting documentation within this premarket notification, Performance the subject device demonstrates substantial equivalence to the listed and SE predicate devices. Determination:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three arms, representing the department's mission to protect the health of all Americans and provide essential human services. The figure is positioned to the right of the department's name, which is written in a circular pattern around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
SEP 2 8 2011
Pioneer Surgical Technology % Ms. Emily M. Downs 375 River Park Circle Marquette, Michigan 49855
Re: K112496
Trade/Device Name: Pioneer Interbody Fusion / Vertebral Body Replacement System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX. ODP. MOP Dated: August 24, 2011 Received: August 29, 2011
Dear Ms. Downs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Emily M. Downs
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutfDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Euna M. Keith
So Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known): K112496
Device Name:
### Pioneer Interbody Fusion /Vertebral Body Replacement System
Indications:
The Pioneer IBF/VBR System, when used as an IBF implant, is indicated for intervertebral body fusion of the spine in skeletally mature patients. Pioner 1BFs are designed for use with autogenous bone graft to facilitate fusion. One device may be used per intervertebral space. Pioneer IBFs are intended to be used with supplemental spinal fixation cleared for the implanted level, such as the Quantum, Streamline, Contact ALP or SlimFuse systems.
The Cervical IBF device in intended for use at one level in the cervical spine, from C3 to T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment
Lumbar IBFs are also intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to St, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Lumbar IBFs are to be used in patients who have had six months of non-operative treatment.
The Pioneer IBF/VBR System, when used as a VBR implant, is intended for use in the thoracolumbar spine (T1-L.S) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of turnors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. Pioneer VBRs are also indicated for treating fractures of the thoracic and lumbar spine. Pioneer VBRs are designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a proloneed period of time. The system must be used with the Pioneer® Quantum Spinal Fixation System or supplemental internal fixation systems cleared for the conditions listed above (i.e., tumor or trauma of TI-L5). Additionally, the Pioneer Vertebral Spacer implant is intended to be used with bone graft.
> イ Prescription Use (Per 21 CFR 801.109)
Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
KI12496 510(k) Number_
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.