K112475 · Xtract Solutions · FMF · Nov 10, 2011 · General Hospital
Device Facts
Record ID
K112475
Device Name
XTRACT SYRINGE KEY SYSTEM
Applicant
Xtract Solutions
Product Code
FMF · General Hospital
Decision Date
Nov 10, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The product family of Xtract Solutions Syringe Keys are indicated as accessories provided for specified commercially available piston hypodermic syringes (for medical purposes) which are designed to control the distance the syringe plunger can be drawn through the syringe barrel and to facilitate the user in obtaining accurate, consistent and precise volumetric draws from vials containing liquids such as vaccinations, allergenic extracts and other medications. The Syringe Keys (available for 1mL and/or 3 mL hypodermic syringes) are designed as an accessory only for the following syringes: 1 mL syringes: 1 mL Terumo SurSaver Mixing Syringes (SS01A2313T) and/or 1 mL Greer Optimix (GROM-23) SurSaver Mixing Syringes. 3 mL syringes: 3 mL VanishPoint (RTI10311) Syringes, 23g x 1" . 3 mL BD SafetyGlide (BD305905) Syringes, 23g x 1" . 3 mL BD Integra (BD305271) Syringes 23g x 1" . 3 mL BD Precision Glide (BD309578) Syringes 20g x 1" . The device assists the user in withdrawing precise and accurate liquid volume withdraws from a vial. The Syringe Keys do not facilitate or assist the user with injecting the contents of the hypodermic syringe. The Syringe Keys can be used by medical personnel as well as individuals who are not medical professionals; but, have been instructed in their use. Syringe Keys will be available for specified piston hypodermic syringe models manufactured by designated syringe manufacturers.
Device Story
Hand-held passive mechanical accessory; fits between plunger thumb rest and finger grip of specific 1mL/3mL piston hypodermic syringes. Features teeth, groove, and notch to physically limit plunger retraction distance; ensures consistent, accurate volumetric withdrawal from vials. Used by clinicians or trained individuals; not for injection. Benefits include improved accuracy and reproducibility of dose measurement compared to manual visualization. Device is non-electronic, purely mechanical.
Clinical Evidence
Bench testing only. Performance verified by measuring dispensed liquid weight across multiple trials for each compatible 3mL syringe model. Results demonstrated accurate, precise, and reproducible volume draws compared to standard manual methods.
Technological Characteristics
Mechanical spacer/accessory. Materials: Polycarbonate/ABS blend and/or Delrin. Dimensions: 5 inches long, curved for ergonomic grip. Features: Teeth, groove, and notch for plunger positioning. No energy source, connectivity, or software.
Indications for Use
Indicated for medical personnel or instructed non-professionals requiring precise volumetric withdrawal of liquids (vaccines, allergenic extracts, medications) from vials using specified 1mL or 3mL piston hypodermic syringes. Not for use in injecting syringe contents.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
K112475
NOV 1 0 2011
Image /page/0/Picture/2 description: The image shows a logo for Xtract Solutions. The word "Xtract" is written in a bold, sans-serif font, with a syringe replacing the "I" in the word. Below the word "Xtract" is the word "Solutions" written in a smaller, sans-serif font. The logo is simple and modern, and the syringe suggests that the company is involved in the medical or pharmaceutical industry.
#### 4.0 510(k) Summary (Prepared August 20, 2011):
In accordance with 21 CFR Section 807.92. Xtract Solutions. Is submitting the following 510(k) Summary:
#### 4.1 Submitter Information -
Xtract Solutions 9495 SW Locust, Suite E Portland, OR 97223, USA FDA Registration No.: 3008292087 Owner / Operator No .: EIN 931264310
#### 4.2 Preparer of Submission and Contact for Information -
Keith Lowrey, Vice President of Regulatory Affairs & Quality Assurance Xtract Solutions 9495 SW Locust Street, Suite E Portland, OR 97223 Date of Preparation: August 20, 2011
Contact Information for 510(k) Correspondence:
611 South Schoolhouse Creek Rd. Grants Pass, OR 97526 Office: 541-476-1628 Cell: 805-403-4676 E-mail: Lowrey_RA_QA_Soutions@yahoo.com
#### 4.3 Name of Device -
- 4.3.1 Trade / Proprietary Name: Xtract 3mL Syringe Keys
- 4.3.2 Common / Usual Name: Hypodermic Syringe Accessory Spacer or Holder
### 4.3.3 Classification Name: Accessory Device to Piston Hypodermic Syringe
- 4.4.4 Requlation Number: 21 CFR 880.5860
## Product Code: 4.4.5 The product code for the Syringe Keys is FMF.
- 4.4.6 Class: Class II (performance standards)
{1}------------------------------------------------
#### 4.4 Substantial Equivalence -
This submission establishes the substantial equivalence of the Xtract Solutions 3mL Syringe Keys as an accessory to piston hypodermic syringes such as the following predicate device:
- 4.4.1 Xtract Solutions Syringe Keys, K091200, SE 08/21/2009, manufactured by Xtract Solutions, Portland, OR.
- Xtract Solutions has added to the "family of Xtract Solutions Syringe 4.4.2 Keys" four (4) additional Syringe Keys to accommodate 3 mL piston hypodermic syringes.
#### 4.5 Description of the Device -
- The Xtract Solutions 3 mL Syringe Key is a hand held passive tool 4.5.1 and an accessory designed to conveniently and ergonomically fit between the plunger thumb rest and the finger grip of piston hypodermic syringes.
Image /page/1/Figure/7 description: The image shows two close-up shots of a white object with labels pointing to specific features. In the left shot, arrows point to the "TEETH" of the object. In the right shot, arrows point to the "GROOVE" and "NOTCH" of the object.
Figure 1 Teeth hold the finger grip of syringe. of the plunger.
Figure 2 The notch holds the end
- The 3 mL Syringe Key facilitates the user in obtaining rapid and 4.5.2 accurate volumetric with-drawls from vials containing vaccines and The Syringe Key has been re-designed for other medications. ergonomic improvement for the user and to fit four (4) additional piston hypodermic syringes:
- 3 mL VanishPoint (RTI10311) syringes, 23g x 1" .
- 3 mL BD SafetyGlide (BD305905) syringes, 23g x 1" .
- 3 mL BD Integra (BD305271) syringes 23g x 1" .
- 3 mL BD Precision Glide (BD309578) syringes 20g x 1" .
SECTION 4 PAGE 3
{2}------------------------------------------------
- The 3 mL Syringe Key is 5 inches long and is curved to fit comfortably 4.5.3 in the user's hand.
4.5.4
Image /page/2/Picture/3 description: In the image, a hand is holding a white rectangular object. The hand is gripping the object firmly, with the fingers curled around it. The object appears to be made of a solid material, and it is reflecting light. The background is a plain surface.
Figure 3 Key is curved to provide ergonomic grip by user
- 4.5.5 At the top of the Key on the underside are three features:
Teeth: These raised points allow the user to pull out the syringe plunger easily.
Groove: The width of the groove corresponds to the type of syringe designated for the specific Key. * The length of the groove equates to the dose the Key will withdraw.
Notch: The notch holds the end of the plunger secure when pulling back and pushing forward to measure the dose.
Image /page/2/Picture/9 description: The image shows a three-step process of connecting a syringe to a device. Step 1 shows the syringe and device separated with labels pointing to the teeth, groove, and notch. Step 2 shows the syringe partially connected to the device. Step 3 shows the syringe fully connected to the device.
e 4 Key controls the distance the syringe plunger can be drawn through the syringe barrel
Figur
{3}------------------------------------------------
- 4.5.5 The plunger is pulled slightly from the syringe barrel (Step 1) and the plunger thumb rest is then nested into a precise slot located at the proximal end of the Syringe Key (Step 2). After seating the thumb rest in the Syringe Key, the syringe's barrel is moved backward against the proximal flat face of the Syringe Key (Step 3).
{4}------------------------------------------------
#### 4.6 Indications for Use -
- The product family of Xtract Solutions Syringe Keys are indicated as accessories 4.6.1 provided for specified commercially available piston hypodermic syringes (for medical purposes) which are designed to control the distance the syringe plunger can be drawn through the syringe barrel and to facilitate the user in obtaining accurate, consistent and precise volumetric draws from vials containing liquids such as vaccinations, allergenic extracts and other medications. The Syringe Keys (available for 1mL and/or 3 mL hypodermic syringes) are designed as an accessory only for the following syringes:
#### 4.6.1.1 1 mL syringes:
- . 1 mL Terumo SurSaver Mixing Syringes (SS01A2313T) and/or
- 1 mL Greer Optimix (GROM-23) SurSaver Mixing Syringes. .
#### 4.6.1.2 3 mL syringes:
- . 3 mL VanishPoint (RTI10311) Syringes, 23g x 1"
- 3 mL BD SafetyGlide (BD305905) Syringes, 23g x 1" .
- 3 mL BD Integra (BD305271) Syringes 23g x 1" .
- . 3 mL BD Precision Glide (BD309578) Syringes 20g x 1"
- The device assists the user in withdrawing precise and accurate liguid volume 4.6.2 withdraws from a vial. The Syringe Keys do not facilitate or assist the user with injecting the contents of the hypodermic syringe.
- The Syringe Keys can be used by medical personnel as well as individuals who 4.6.3 are not medical professionals: but, have been instructed in their use,
- Svringe Keys are available for specified piston hypodermic syringe models 4.6.4 manufactured by designated syringe manufacturers.
#### 4.7 Specifications -
- 4.7.1 The 3mL Syringe Keys are designed and fabricated for use with the following specified piston hypodermic syringes and specified volume withdrawals:
- mL VanishPoint (RTI10311) Syringes, 23g x 1" .
- 3 mL BD SafetyGlide (BD305905) Syringes, 23g x 1" .
- 3 mL BD Integra (BD305271) Syringes 23g x 1 .
- . 3 mL BD Precision Glide (BD309578) Syringes 20g x 1"
{5}------------------------------------------------
XTRACT Solutions, Syringe Keys 510(k) Submission, August 2011
Image /page/5/Picture/1 description: The image shows four different types of syringes. The first syringe is a B-D PrecisionGlide 3mL syringe with a reference number of 309578 and a volume of 1.50 mL. The second syringe is a B-D Integra 3mL syringe with a reference number of 305271 and a volume of 0.5 mL. The third syringe is a SafetyGlide 3 mL syringe with a volume of 0.5 mL, and the fourth syringe is a VanishPoint 3 mL syringe with a volume of 0.5 mL.
Figure 5 Examples of 3 mL Syringe Keys
- 4.7.2 The new 3mL Syringe Keys are designed for the following volume withdrawals:
| Safety Glide<br>P/N 120-0003 | Vanish Point<br>P/N 120-0002 | BD Precision Glide<br>P/N 190-0101 | BD Integra<br>P/N 190-0110 |
|------------------------------|------------------------------|------------------------------------|----------------------------|
| 0.5 mL | 0.5 mL | 0.5 mL | 0.5 mL |
- The Xtract Solutions 3mL Syringe Keys will be sold by prescription 4.7.4 only and labeling will bear the statement, "Caution: Federal Law restricts this device to sale by or on the order of a physician."
#### Technological Characteristics in Comparison to the Predicates -4.8
#### 4.8.1 Predicate Device -
The Xtract Solutions 3mL Syringe Key is substantially equivalent to the Xtract Solutions 1mL Syringe Keys, K091200, SE 08/21/2009
Manufacturer: Xtract Solutions, Portland, OR 97223
#### 4.8.2 Materials -
Materials used to fabricate the Xtract 3mL Syringe Keys are polycarbonate/ABS blend and/or Delrin® that have a long use in medical devices.
{6}------------------------------------------------
#### 4.8.3 Performance Comparison-
As an accessory for the hypodermic svringe, the 3mL Svringe Keys and the 1 mL Svringe Keys are designed to control the distance that the syringe plunger can be pulled through the syringe barrel by functioning as a "spacer" fitting on the plunger between the thumb rest and the barrel's finger rest/flange. The devices assist the user in withdrawing precise and accurate liquid volume withdraws from a vial. The devices do not facilitate or assist the user with injecting the contents of the hypodermic syringe.
#### Performance Testing -4.9
- The 3mL Syringe Keys were designed by first establishing an 4.9.1 empirical (not just mathematic) relationship of plunger travel to dispensed volume (weight). This data was then used to establish the ideal physical size of the syringe keys (the "Syringe Key Design" worksheet). This process was used with each of the volume sizes for each 3 mL Syringe Key.
- 4.9.2 Upon receipt of the 3mL Syringe Keys, a First Article Inspection was performed to verify that all dimensions met drawing specifications and requirements as noted in following graph:
## Example data: 3 mL Syringe Keys tested with BD 3 mL Integra Syringes:
| Dose Verification at Design<br>Length (0.578") | | | |
|------------------------------------------------|-----|----------|----------|
| Syringe | # 1 | #2 | |
| DRAW | 1 | 0.502 | 0.504 |
| | 2 | 0.503 | 0.506 |
| | 3 | 0.495 | 0.501 |
| | 4 | 0.499 | 0.499 |
| | 5 | 0.500 | 0.506 |
| | 6 | 0.505 | 0.502 |
| | 7 | 0.500 | 0.502 |
| | 8 | 0.493 | 0.500 |
| | 9 | 0.504 | 0.499 |
| | 10 | 0.503 | 0.501 |
| | AVG | 0.500 | 0.502 |
| | STD | 0.003893 | 0.002582 |
{7}------------------------------------------------
The channel length measurements of the actual syringe keys and the "draw volumes" were evaluated and compared. Using the 3mL Syringe Keys, piston hypodermic syringes with needles were introduced into the septum of vials containing H₂0.
The liquid was withdrawn from the vials and dispensed into tarred weighing dishes and subsequently weighed. Weight measurement analyses were conducted on each of the different 3 mL Syringe Keys multiple times.
- 4.9.3 Performance data demonstrated that the 3mL Syringe Keys facilitate the user in obtaining more accurate, precise and reproducible volume draws as compared to the standard method or visualization procedure.
{8}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV 1 0 2011
Xtract Solutions C/O Keith Lowrey Vice President of Regulatory Affairs & Quality Assurance 9495 SW Locust Street, Suite E Portland, Oregon 97223
Re: K112475
Trade/Device Name: Xtract Syringe Keys Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: August 26, 2011 Received: September 8, 2011
Dear Mr. Lowrey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
## Page 2 - Mr. Lowery
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony V. Anston
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{10}------------------------------------------------
Image /page/10/Figure/1 description: The image shows the logo for Xtract Solutions. The logo is black and white and features a stylized "X" with a syringe in place of the "T" in Xtract. Below the word Xtract is the word Solutions in a smaller font.
# INDICATIONS FOR USE STATEMENT
KI12475 510(k) Number (if known): To be determined by FDA .
Device Name: Xtract Svringe Kevs
# Indications for Use:
- 1.1 The product family of Xtract Solutions Syringe Keys are indicated as accessories provided for specified commercially available piston hypodermic syringes (for medical purposes) which are designed to control the distance the syringe plunger can be drawn through the svringe barrel and to facilitate the user in obtaining accurate, consistent and precise volumetric draws from vials containing liquids such as vaccinations, allergenic extracts and other medications. The Syringe Keys (available for 1mL and/or 3 mL hypodermic syringes) are designed as an accessory only for the following syringes:
- 1.1.1 1 mL svringes:
- 1 mL Terumo SurSaver Mixing Syringes (SS01A2313T) and/or .
- 1 mL Greer Optimix (GROM-23) SurSaver Mixing Syringes. .
#### 1.1.2 3 mL syringes:
- 3 mL VanishPoint (RTI10311) Syringes, 23g x 1" .
- 3 mL BD SafetyGlide (BD305905) Syringes, 23g x 1" .
- . 3 mL BD Integra (BD305271) Syringes 23g x 1"
- 3 mL BD Precision Glide (BD309578) Syringes 20g x 1" .
- 1.2 The device assists the user in withdrawing precise and accurate liquid volume withdraws from a vial. The Syringe Keys do not facilitate or assist the user with injecting the contents of the hypodermic syringe.
- 1.3 The Syringe Keys can be used by medical personnel as well as individuals who are not medical professionals; but, have been instructed in their use.
- 1.4 Syringe Keys will be available for specified piston hypodermic syringe models manufactured by designated syringe manufacturers.
| Prescription Use<br>(Part 21 CFR 801 Subpart D | X |
|------------------------------------------------|---|
|------------------------------------------------|---|
AND/OR
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
|------------------------------------------------|--|
|------------------------------------------------|--|
(Division Sign-GHFEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
| 510(k) Number: | K112475 | SECTION 3 PAGE 2 |
|----------------|-------------------------------------|------------------|
| | Kature Loftused for Richard Chapman | |
| | 11/10/11 | |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.