PROSPERA PRO-I, PRO-II AND PRO-III NEGATIVE PRESSURE WOUND THERAPY SYSTEM
K112458 · Medica-Rents Co., Ltd. · OMP · Jan 5, 2012 · General, Plastic Surgery
Device Facts
Record ID
K112458
Device Name
PROSPERA PRO-I, PRO-II AND PRO-III NEGATIVE PRESSURE WOUND THERAPY SYSTEM
Applicant
Medica-Rents Co., Ltd.
Product Code
OMP · General, Plastic Surgery
Decision Date
Jan 5, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Prospera Negative Pressure Wound Therapy System is indicated for patients that would benefit from a suction device particularly as the device may promote wound healing by removal of wound exudate, debris, and infectious material or for the aspiration and removal of surgical fluids, tissue (including bone), gases, bodily fluids or infectious material from the patient's airway or respiratory support system. The Prospera Negative Pressure Wound Therapy may be used during surgery or at the patient's bedside and is indicated for home use.
Device Story
Prospera PRO-I, PRO-II, and PRO-III are portable, AC/battery-powered suction pumps for negative pressure wound therapy (NPWT). Devices apply continuous or intermittent negative pressure (max -200 mmHg) to remove wound exudate, debris, and infectious material into a disposable canister via tubing and microbial filters. Operated by clinicians or patients (home use) via software-controlled interface with alarm functions. PRO-II and PRO-III feature integrated canisters and lithium-ion batteries; PRO-I uses NiMh battery and includes a relief valve. Healthcare providers use output to manage wound healing or clear respiratory fluids. Benefits include promoted wound healing and fluid aspiration. Devices are compatible with gauze or foam dressings and adhesive drapes to maintain a sealed environment.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified through design control testing, electrical safety testing per IEC 60601-1-2, and comparison of technical specifications to predicate device.
Technological Characteristics
Powered suction pump; max vacuum -200 mmHg; flow rate 9 L/min. Materials: rigid/collapsible canisters, hoses, microbial filters. Energy: AC, 12V automotive, or internal battery (Li-ion or NiMh). Dimensions: up to 290x359x130mm. Software-controlled continuous/intermittent modes. Manufactured per ISO 13485 and EEC Directive 93/42/EEC.
Indications for Use
Indicated for patients requiring suction for wound healing (chronic, acute, traumatic, subacute, dehisced wounds, burns, ulcers, flaps, grafts) or aspiration of fluids/tissues/gases from airway/respiratory systems. Suitable for hospital, surgical, or home use.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
K112458
JAN - 5 2012
10F7
### 510(k) Summary of Safety and Effectiveness
| Date prepared: | October 21, 2011 |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | Medica-Rents Co LTD<br>2831 Bledsoe Street, Fort Worth TX 76107<br>Telephone: 817-336-5536, Fax: 817-332-9140<br>Owner/Operator #: 1640798 |
| 510(k) Contact: | Dina L. Weissman, J.D.<br>P.O. Box 205, Derby CT 06418<br>Telephone: 203-954-9109, Email: dina.weissman@comcast.net |
| Contract Manufacturer: | Simex Medizintechnik, GmbH<br>Bruckstr. 30/1<br>Deisslingen-Lauffen, Baden-Wurtetmberg, 78652 Germany |
| Trade Name: | Prospera PRO-I, PRO-II and PRO-III Negative Pressure Wound<br>Therapy System |
| Common Name: | Negative Pressure Wound Therapy Powered Suction Pump |
| Classification: | Powered Suction Pump, 21 CFR 878.4780<br>Class II, General & Plastic Surgery |
| Device Product Code: | OMP - Pump, Portable, Aspiration, (Manual or Powered) |
| Substantial Equivalency: | Medica Rents Powered Suction Pump PRO-I K062456<br>(cleared as the NovaSpine Powered Suction Pump PRO-I) |
#### Indications for Use:
The Prospera Negative Pressure Wound Therapy System is indicated for patients that would benefit from a suction device particularly as the device may promote wound healing by removal of wound exudate, debris, and infectious material or for the aspiration and removal of surgical fluids, tissue (including bone), gases, bodily fluids or infectious material from the patient's airway or respiratory support system. The Prospera Negative Pressure Wound Therapy may be used during surgery or at the patient's bedside and is indicated for home use.
#### Device Description:
The Prospera PRO Series of Powered Suction Pump is a portable, AC-powered device to be used to remove fluids, including wound exudate, irrigation fluids, and infectious materials from a wound. The pump is operated through computer software, having help and alarm functions. The software settings permit continuous or intermittent negative pressure to be applied for proscribed time periods. The device is indicated for management of chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts.
The pumps may be used at a patient's bedside. Some models are portable. The Prospera series of pumps are sutiable for use in either a hospital or home, with proper training and supervision.
{1}------------------------------------------------
V112458
20fr
The Prospera Negative Pressure Wound Therapy Powered Suction Pump is available in three models. The PRO-I was previously cleared in K062456 and is again included here because a relief valve was added. The PRO-II is a smaller, more portable unit having an integrated disposable canister. Both the PRO-II and PRO-III have a lithium ion battery while the PRO-I has a Nickel-metal Hydride (NiMh) battery. All three models are rechargeable and include a port for an external power supply and/or a battery charger. The pump can also be externally powered from a 12V automotive battery source.
All models provide suction at the rate of 9 liters per minute; have a maximum vacuum of -200 mmHg (negative pressure); weigh between 1.8 and 2.8 kg; and operate in continuous or intermittent modes. Maximum dimensions of the unit are 290mmX359mmX130mm (PRO-1).
Consumable accessory components include collection canisters (rigid or collapsible), hoses, connectors and drains (tubing), and foam or gauze dressings.
The device provides negative pressure wound therapy at a range of pressure settings and removes exudates from the wound site through tubing and microbial filters to a disposable canister. The pumps are compatible with a variety of wound dressing kits, using either gauze or foam dressings.
| Feature | Predicate: | This submission: |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| | K062456, PRO-I NPWT pump | PRO-II, PRO-III |
| Dressing system | Foam or gauze based dressing with an adhesive film drape to<br>create a sealed wound environment | Same as predicate |
| Pressure sensing | Sensors in unit | Same as predicate |
| Therapy unit | Computer software controlled, battery and AC powered negative<br>pressure pumps using continuous or intermittent modes to<br>remove exudates from the wound to a collection canister | Same as predicate |
## Summary of Technological Characteristics
### Non-Clinical Test (Bench):
The Prospera PRO series of pumps is manufactured in accordance with EEC Directive 93/42/EEC Annex IX. Testing for electrical safety was conducted to ensure it meets the requirements for IEC 60601-1-2 including the National Differences for the U.S. The pump's manufacture and quality systems management are in accordance with certified international standards, including ISO 13485. Testing was also conducted to verify the changes in the design specifications and demonstrated substantial equivalence to the predicate device, the PRO-I.
## Basis For Substantial Equivalence:
In order to demonstrate substantial equivalence, Medica-Rents evaluated the indications for use (unchanged), materials (unchanged), technology (unchanged), product specifications (design control verification and validation conducted), and energy requirements of the device (unchanged). Peformance testing and electrical safety has been successfully completed and documented to demonstrate that the modified PRO-I, along with the PRO-II, are substantially equivalent to the previously cleared PRO-1.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medica-Rents Co., Ltd % Dina L. Weissman, J.D. Legal and Regulatory Affairs Consultant P.O. Box 205 Derby. Connecticut 06418
JAN - 5 2012
Re: K112458
Trade/Device Name: Prospera PRO-I, PRO-II and PRO-III Negative Pressure Wound Therapy Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: December 12, 2011 Received: December 13, 2011
Dear Ms. Weissman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Dina L. Weissman, J.D.
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Sincerely yours,
J. N. Mull
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement
## 510(k) Number: K112458
## Device Name:
Prospera PRO-1, PRO-11 and PRO-111 Negative Pressure Wound Therapy System
# Indication for Use:
The Prospera Negative Pressure Wound Therapy System is indicated for patients that would benefit from a suction device particularly as the device may promote wound healing by removal of wound exudate, debris, and infectious material or for the aspiration and removal of surgical fluids, tissue (including bone), gases, bodily fluids or infectious material from the patient's airway or respiratory support system. The Prospera Negative Pressure Wound Therapy may be used during surgery or at the patient's bedside and is indicated for home use.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Knaufu MXM
Division Sign Off
gical. Orthoned
510(k) Number K112458
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.