← Product Code [KWS](/productcode/KWS) · K112438

# TITAN MODULAR TOTAL SHOULDER SYSTEM (K112438)

_Ascension Orthopedic · KWS · Apr 11, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K112438

## Device Facts

- **Applicant:** Ascension Orthopedic
- **Product Code:** [KWS](/productcode/KWS.md)
- **Decision Date:** Apr 11, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3660
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The Ascension TITAN™ Total Shoulder System is indicated for use as a hemi or total shoulder replacement for: Severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicated that alternativce methods of treatment are unsatisfactory. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. - revision of a failed primary component). Shoulder Hemiarthroplasty is also indicated for: Ununited humeral head fractures. Avascular necrosis of the humeral head. Rotator cuff arthropathy. Deformity and/or limited motion. The humeral component is intended for cemented or un-cemented use. The glenoid component is intended for cemented use only.

## Device Story

Modular shoulder prosthesis system; includes metaphyseal bodies, humeral stems, humeral heads, and polyethylene glenoid components. Used for hemiarthroplasty or total shoulder replacement. Humeral components allow cemented or uncemented fixation; glenoid components require cement. Designed to accommodate natural humeral geometry, provide proximal bone loading, and ensure proper head placement. Humeral heads available in concentric and eccentric articulating surfaces. Implanted by orthopedic surgeons in clinical/surgical settings to restore joint function, reduce pain, and address complex fractures or failed primary components.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Modular metal/polymer semi-constrained prosthesis. Components include metaphyseal bodies, humeral stems, humeral heads, and polyethylene glenoid. Conforms to ASTM F1378. Fixation: cemented or uncemented (humeral), cemented (glenoid).

## Regulatory Identification

A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Predicate Devices

- Ascension® TITAN™ Total Shoulder System ([K100448](/device/K100448.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# | | | 2438 ( | | 2 )

APR 1 1 2012

- -

## 510(k) Summary

:

·

:

.

| Sponsor            | Ascension Orthopedics, Inc.<br>8700 Cameron Road<br>Austin, TX 78754-3832                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person     | Susan Walton<br>susan.walton@integralife.com<br>Phone: (512) 836-5001 x1591<br>Alternate FAX (512) 836-6933                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date               | March 22, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Trade Name         | Ascension® TITAN™ Total Shoulder System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Common Name        | Hemi- or Total Shoulder                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification     | 21 CFR 888.3660 - Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis<br>21 CFR 888.3690 - Shoulder joint humeral (hemi-shoulder)<br>metallic uncemented prosthesis.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code       | HSD and KWS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Panel              | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Device   | K100448 – Ascension® TITAN™ Total Shoulder System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Description | The TITAN Modular Total Shoulder System consists of a line of<br>metaphyseal bodies, humeral stems, humeral heads and all<br>polyethylene glenoid components. The body, stem and humeral<br>head may be used by themselves, as a hemiarthroplasty, if the<br>natural glenoid provides a sufficient bearing surface, or in<br>conjunction with the glenoid, as a total replacement. The<br>metaphyseal bodies and humeral stems are shaped to provide<br>proximal fixation and optimal fixation area. Their variable length<br>and proximally-filling shape are designed to accommodate the<br>natural humeral geometry and provide stable fixation, proximal<br>bone loading and proper head placement. The humeral heads are<br>offered with both concentric and eccentric articulating surfaces. The<br>humeral head may articulate against the natural glenoid bone, if it is<br>of sufficient quality, or against the all polyethylene cemented<br>glenoid. The glenoid has two options: keeled or standard pegged (3<br>pegs). All glenoid options are designed to function with both the<br>concentric and eccentric heads.<br>The humeral components are intended for cemented or uncemented<br>use, while the glenoid component is for use with cement only. |
| Intended Use       | A shoulder joint metal/polymer semi-constrained cemented<br>prosthesis is a device intended to be implanted to replace a shoulder<br>joint.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Intended Use       | The Ascension TITAN™ Total Shoulder System is indicated for<br>use as a hemi or total shoulder replacement for:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{1}------------------------------------------------

- > Severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis.
- > Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicated that alternativce methods of treatment are unsatisfactory.
- > Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. - revision of a failed primary component).

Shoulder Hemiarthroplasty is also indicated for:

- > Ununited humeral head fractures.
- > Avascular necrosis of the humeral head.
- > Rotator cuff arthropathy.
- > Deformity and/or limited motion.

The humeral component is intended for cemented or un-cemented use. The glenoid component is intended for cemented use only.

Ascension Orthopedics believes that this system is substantially equivalent to the legally marketed predicate device based on similarities in design, materials and indications.

Testing to support the material change and additional lengths for the humeral stems has been completed. The testing conforms to ASTM F1378.

Clinical performance data are not required for the minor changes that are the subject of this Special 510(k).

Basis of Substantial Equivalence

Non-clinical Performance Data

Clinical Performance Data

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ascension Orthopaedics, Incorporated % Ms. Susan Walton Senior Director of Regulatory Affairs 8700 Cameron Road Austin, Texas 78754

APR 1 1 2012

Re: K112438

Trade/Device Name: Ascension® TITANTM Total Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS, HSD Dated: March 22, 2012 Received: March 23, 2012

Dear Ms. Walton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

## Page 2 - Ms. Susan Walton

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Molkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## 112438 (11)

## Indications for Use Statement

510(K) Number:

Ascension® TITAN™ Total Shoulder System Device Name:

Indications for Use:

The Ascension TITAN Total Shoulder System is indicated for use as a hemi or total shoulder replacement for:

- > Severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis.
- > Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicated that alternative methods of treatment are unsatisfactory.
- > Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. - revision of a failed primary component).

Shoulder Hemiarthroplasty is also indicated for:

- > Ununited humeral head fractures.
- > Avascular necrosis of the humeral head.
- > Rotator cuff arthropathy.
- Deformity and/or limited motion. s

The humeral component is intended for cemented or uncemented use. The glenoid component is intended for cemented use only.

OR

Prescription Use __ x (Part 21 CFR 801Subpart B)

Over-The-Counter Use (Part 21 CFR 801Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ro. Oulil
(Division Sign Off)

Division Sign-Off) ivision of Surgical, Orthopedic, and Restorative Devices

510(k) Number K112438

---

**Source:** [https://fda.innolitics.com/device/K112438](https://fda.innolitics.com/device/K112438)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
