INFINITI VISION SYSTEM EVERGREEN II AUTOSERT IOL INJECTOR HANDPIECE
Applicant
Alcon Manufacturing, Ltd.
Product Code
HQC · Ophthalmic
Decision Date
Nov 21, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4670
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Infinit® Vision System is indicated for emulsification, separation, and removal of cataracts, the removal of residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intra-ocular lens injection. The AutoSert™ IOL Injector Handpiece is intended to deliver qualified AcrySof® intraocular lenses into the eye following cataract removal. The following system modalities additionally support the described indications: - Ultrasound with UltraChopper Tip achieves the functionality of cataract separation. - AquaLase achieves the functionality for removal of residual cortical material and lens epithelial cells - The AutoSert ™ IOL Injector Handpiece achieves the functionality of injection of intraocular lenses. The AutoSert™ is indicated for use with ACRYSOF lenses SN60WF and SN6AD1, as well as approved AcrySof lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.
Device Story
Infiniti® Vision System is a modular phacofragmentation system for cataract surgery. It utilizes ultrasound (with UltraChopper tip), AquaLase technology, and the AutoSert™ IOL injector handpiece to emulsify, separate, and remove cataracts, aspirate vitreous, perform bipolar coagulation, and inject IOLs. Operated by surgeons in clinical settings, the system provides integrated control over surgical modalities. The 'Power Watch' feature monitors system performance. The device processes user inputs to manage fluidics and energy delivery for lens removal and IOL insertion. By automating key surgical steps, the system enhances surgical efficiency and precision, potentially improving patient outcomes during cataract extraction and lens replacement procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993 series), sterilization validation (ISO 11135-1), and electrical safety/EMC testing (IEC 60601 series).
Technological Characteristics
Modular phacofragmentation system. Components include ultrasound handpieces, AquaLase, and AutoSert IOL injector. Materials comply with ISO 10993. Sterilization via EtO for single-use accessories; reusable handpieces are reprocessable. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 80601-2-58 (lens removal/vitrectomy).
Indications for Use
Indicated for patients undergoing cataract surgery requiring emulsification, separation, and removal of cataracts, cortical material, and lens epithelial cells; anterior vitrectomy; bipolar coagulation; and intraocular lens (IOL) implantation. AutoSert™ handpiece is indicated for delivery of specific AcrySof® IOLs.
Regulatory Classification
Identification
A phacofragmentation system is an AC-powered device with a fragmenting needle intended for use in cataract surgery to disrupt a cataract with ultrasound and extract the cataract.
{0}------------------------------------------------
NOV 2 1 2011
k112425
# PREMARKET NOTIFICATION 510(k) SUMMARY
The submitter of the 510(k) is:
Martin A. Kaufman Director. Regulatory Affairs Alcon Research, Ltd. 18500 Alton Parkway Irvine, CA 92618 Phone: (949) 753-6250 Fax: (949) 753-6237
Device Subject to this 510(k):
Trade Name: Common Name: Classification Name: Infiniti® Vision System Phacofragmentation System Phacofragmentation System (per 21 CFR 886.4670)
## Predicate Devices:
The legally marketed devices(s) to which we are claiming equivalence to are:
| 510(k) Number | Device |
|---------------|---------------------------------------|
| K021566 | Infiniti® Vision System |
| K082845 | Infiniti® Vision System with OZil® IP |
| K091777 | Alcon UltraChopper Tip |
| K093210 | AquaLase Capsule Wash Tip |
| K063155 | Monarch III IOL Delivery System |
#### Device Description
The Infinit® Vision System is an enhanced version of the Infinit® Vision System with OZil® IP (K082845) that is modular in design and incorporates the Power Watch feature, along with the Alcon UltraChopper Tip, the AquaLase Capsule Wash Tip, the AutoSert™ IOL Injector, and its associated software.
{1}------------------------------------------------
# Indications for Use:
The Infinity Vision System is indicated for emulsification, separation, and removal of cataracts, the removal of residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intraocular lens injection. The AutoSert™ IOL Injector Handpiece is intended to deliver qualified AcrySof® intraocular lenses into the eye following cataract removal.
The following system modalities additionally support the described indications:
- Ultrasound with UltraChopper Tip achieves the functionality of cataract separation. -
- AquaLase achieves the functionality for removal of residual cortical material and lens epithelial cells
- The AutoSert™ IOL Iniector Handpiece achieves the functionality of injection of । intraocular lenses. The AutoSert™ is indicated for use with ACRYSOF lenses SN60WF and SN6AD1, as well as approved Acrysof lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.
## Brief Summary of Nonclinical test and Results:
Safety tests of the Infinity Vision System have demonstrated its compliance with applicable requirements of the following standards:
| Standard # | Title |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 11135-1:2007<br>ISO | Sterilization of health care products - Ethylene oxide - Requirements<br>for development, validation, and routine control of a sterilization<br>process for medical devices. |
| 11979-3:2006<br>BS EN ISO | Ophthalmic implants - Intraocular Lenses - Part 3: Mechanical<br>properties and test methods |
| 14971:2007<br>EN ISO | Medical Devices: Application of Risk Management to Medical Devices |
| IEC 60601-1:<br>1988, A1:1991,<br>A2:1995 | Medical Electrical Equipment - Part 1: General requirements for safety |
| IEC 60601-1-1:<br>2000 | Medical electrical equipment. Part 1-1: General requirement for safety -<br>Collateral standard: Safety requirements for medical electrical<br>systems |
| EN 60601-1-2:<br>2004 | Medical electrical equipment - Part 1: General requirements for safety 2.<br>Collateral Standard: Electromagnetic compatibility - Requirements and<br>tests |
| IEC 60601-1-4:<br>2000 | Medical electrical equipment - Part 1: General requirements for safety -<br>4. Collateral standard: Programmable electrical medical systems. |
| IEC 60601-1-6:<br>2006 | Medical electrical equipment - Part 1-6: General requirements for safety -<br>Collateral standard: Usability. |
| IEC 60601-2-2:<br>2006 | Medical electrical equipment - Part 2: Particular requirements for the<br>safety of high frequency surgical equipment. |
| IEC 80601-2-58:<br>2008 | Medical electrical equipment -- Part 2-58: Particular requirements for<br>basic safety and essential performance of lens removal devices and<br>vitrectomy devices for ophthalmic surgery |
{2}------------------------------------------------
Biocompatibility evaluations of materials coming in contact with the patient fluid path have been performed to the following standards:
| Standard # | Title |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------|
| ISO 10993-1:<br>2009 | Biological Evaluation of Medical Devices -- Part 1: Evaluation and<br>Testing within a Risk management Process |
| ISO 10993-5:<br>2009 | Biological Evaluation of medical devices -- Part 5: Tests for In Vitro<br>cytotoxicity |
| ISO 10993-10:<br>2002/A1:2006 | Biological Evaluation of Medical Devices -- Part 10: Tests for irritation<br>and delayed-type hypersensitivity |
| ISO 10993-11:<br>2006 | Biological Evaluation of Medical Devices -- Part 11: Tests for systemic<br>toxicity |
| ISO 10993-12:<br>2007 | Biological Evaluation of Medical Devices --Part 12: Sample Preparation<br>and Reference Materials |
Many Infiniti® accessories are provided sterile and are intended for single use only. These products are EtO sterilized and the process has been validated per ISO 11135-1:2007: Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization. Reusable handpieces are provided non-sterile. Validated reprocessing instructions for cleaning, sterilization, and re-use of the handpieces are provided in the Directions for Use of the product.
Technological characteristics affecting clinical performance are similar to that of predicate devices previously listed. The Infiniti® Vision System has been developed and will be manufactured in compliance with 21 CFR 820 and ISO 14971:2007. Non-clinical testing noted above has demonstrated that the functional requirements have been met and that the modified device is equivalent to the predicate device.
#### Accessories Provided with the System (subject to change)
| Configurations of tips and sleeves currently used with the Irrigation/Aspiration Handpiece.<br>Description |
|------------------------------------------------------------------------------------------------------------|
| ASSY, SHIP, INTREPID POLYMER I/A TIP, STRAIGHT |
| ASSY, SHIP, INTREPID POLYMER I/A TIP, ANGLED |
| ASSY, SHIP, INTREPID POLYMER I/A TIP, CURVED |
| I/A Tip 0.5 mm |
| I/A Tip 0.2 mm |
| I/A Tip 0.3 mm Small Bore |
| I/A Tip 0.3 mm Small Bore Mod |
| I/A Tip 0.3 mm Bent |
| I/A Tip 0.3 mm Bent & Sand Blast |
| I/A Tip .033 OD, .3mm, Intrepid |
{3}------------------------------------------------
| Configurations of tips and sleeves currently used with the Irrigation/Aspiration Handpiece. | Description |
|---------------------------------------------------------------------------------------------|-------------|
| I/A Tip .033 OD, .3mm Bent, Intrepid | |
| Silicone I/A Tip, Bent | |
| Silicone I/A Tip, Straight | |
| Ultraflow Tool/O-rings | |
| Ultraflow IA Handpiece Comp | |
| Ultraflow IA Handpiece Only | |
| Ultraflow IA Tip STR | |
| Ultraflow IA Tip CRVD | |
| Ultraflow IA 45 | |
| Ultraflow IA 90 | |
| Ultraflow IA 120 | |
| Ultraflow CNL STTL | |
| Ultraflow Luer | |
| Straight Tip, 3 mm | |
| Ultraflow Tip Protector, Standalone | |
| 45° Bent Tip, .3 mm | |
| 90° Bent Tip, .3mm | |
| Irrigation Only Luer | |
| Curved Tip, .3mm | |
| Threaded Tip - STTL | |
| Ultraflow I/A Box | |
| Replacement O-ring | |
Configurations of Paks currently used with the Infiniti Vision System.
| Description |
|----------------------------------------------------------------------|
| Ultrasound FMS, Basic |
| Ultrasound FMS, Tipless, 0.9mm |
| Ultrasound FMS, Tipless, 1.1mm |
| Ultrasound FMS, Tipless, 0.9mm HIS |
| Ultrasound FMS, Tipless, 1.1mm HIS |
| Ultrasound FMS, 30° Round, 0.9 mm ABS |
| Ultrasound FMS, 45° Round, 0.9 mm ABS |
| Ultrasound FMS, 30° Kelman, 0.9 mm ABS |
| Ultrasound FMS, 45° Kelman, 0.9 mm ABS |
| Ultrasound FMS, 30° Round, 1.1 mm ABS |
| Ultrasound FMS, 45° Round, 1.1 mm ABS |
| Ultrasound FMS, 30° Kelman, 1.1 mm ABS |
| Ultrasound FMS, 45° Kelman, 1.1 mm ABS |
| Ultrasound FMS, 30° Round, 0.9 mm MicroTip Flared ABS |
| Ultrasound FMS, 45° Round, 0.9 mm MicroTip Flared ABS |
| Ultrasound FMS, 30° Kelman, 0.9 mm MicroTip Flared ABS |
| Ultrasound FMS, 45° Kelman, 0.9 mm MicroTip Flared ABS |
| Description |
| Ultrasound FMS, 30° Round, 1.1 mm Flared ABS |
| Ultrasound FMS, 45° Round, 1.1 mm Flared ABS |
| Ultrasound FMS, 30° Kelman, 1.1 mm Flared ABS |
| Ultrasound FMS, 45° Kelman, 1.1 mm Flared ABS |
| Ultrasound FMS, 30° Round, 0.9 mm Mackool |
| Ultrasound FMS, 45° Round, 0.9 mm Mackool |
| Ultrasound FMS, 30° Kelman, 0.9 mm Mackool |
| Ultrasound FMS, 45° Kelman, 0.9 mm Mackool |
| Ultrasound FMS, 30° Round, 1.1 mm Mackool Flared ABS |
| Ultrasound FMS, 45° Round, 1.1 mm Mackool Flared ABS |
| Ultrasound FMS, 30° Kelman, 1.1 mm Mackool Flared ABS |
| Ultrasound FMS, 45° Kelman, 1.1 mm Mackool Flared ABS |
| Ultrasound FMS, 30° Round, 0.9 mm Tapered ABS |
| Ultrasound FMS, 45° Round, 0.9 mm Tapered ABS |
| Ultrasound FMS, 30° Kelman, 0.9 mm Tapered ABS |
| Ultrasound FMS, 45° Kelman, 0.9 mm Tapered ABS |
| Ultrasound FMS, Tipless, 0.9 Ultra |
| Ultrasound FMS, 30° Kelman, 0.9 mm Mini Flared ABS, Ultra |
| Ultrasound FMS, 45° Kelman, 0.9 mm Mini Flared ABS, Ultra |
| Ultrasound FMS, 30° OZil 12, 0.9 mm Mini Flared ABS, Ultra |
| Ultrasound FMS, 45° OZil 12, 0.9 mm Mini Flared ABS, Ultra |
| Ultrasound FMS, 30° Kelman, 0.9 mm Mini Flared ABS |
| Ultrasound FMS, 45° Kelman, 0.9 mm Mini Flared ABS |
| Ultrasound FMS, 30° OZil 12, 0.9 mm Mini Flared ABS |
| Ultrasound FMS, 45° OZil 12, 0.9 mm Mini Flared ABS |
| Ultrasound FMS, Tipless, 0.9mm J |
| Ultrasound FMS, Tipless, 1.1mm J |
| Ultrasound FMS, Tipless, 0.9mm HIS J |
| Ultrasound FMS, Tipless, 1.1mm HIS J |
| Ultrasound FMS, Tipless, 0.9 Ultra J |
| Ultrasound FMS, Intrepid, Basic |
| Ultrasound FMS, Intrepid, Tipless, 0.9 Ultra |
| Ultrasound FMS, Intrepid, Tipless, 1.1 Ultra |
| Ultrasound FMS, Intrepid, Tipless, 0.9 Nano |
| Ultrasound FMS, Intrepid, Tipless, 1.1 Nano |
| Ultrasound FMS, Intrepid, 30° Kelman, 0.9 mm Mini-Flared ABS, Ultra |
| Ultrasound FMS, Intrepid, 45° Kelman, 0.9 mm Mini-Flared ABS, Ultra |
| Ultrasound FMS, Intrepid, Tipless, 0.9 |
| Ultrasound FMS, Intrepid, 30° Kelman, 0.9 mm Mini Flared ABS, Nano |
| Ultrasound FMS, Intrepid, 45° Kelman, 0.9 mm Mini Flared ABS, Nano |
| Ultrasound FMS, Intrepid, 30° OZil 12, 0.9 mm Mini Flared ABS, Nano |
| Ultrasound FMS, Intrepid, 45° OZil 12, 0.9 mm Mini Flared ABS, Nano |
| Ultrasound FMS, Intrepid, 30° OZil 12, 0.9 mm Mini Flared ABS, Ultra |
| Description |
| Ultrasound FMS, Intrepid, 45° OZil 12, 0.9 mm Mini Flared ABS, Ultra |
| Ultrasound FMS, Intrepid, 30° Kelman, 0.9 mm Mini Flared ABS, 0.9 |
| Ultrasound FMS, Intrepid, 45° Kelman, 0.9 mm Mini Flared ABS, 0.9 |
| Ultrasound FMS, Intrepid, 30° OZil 12, 0.9 mm Mini Flared ABS, 0.9 |
| Ultrasound FMS, Intrepid, 45° OZil 12, 0.9 mm Mini Flared ABS, 0.9 |
| Ultrasound FMS, Intrepid, Basic, J |
| Ultrasound FMS, Intrepid, Tipless, 0.9 Ultra, J |
| Ultrasound FMS, Intrepid, Tipless, 1.1 Ultra, J |
| Ultrasound FMS, Intrepid, Tipless, 0.9 Nano, J |
| Ultrasound FMS, Intrepid, Tipless, 1.1 Nano, J |
| Ultrasound FMS, Intrepid, Tipless, 0.9 J |
| INFI, Multipak, Basic Pak |
| INFI, Multipak, 0.9mm Tipless |
| INFI, Multipak, 1.1mm Tipless |
| INFI MP Procedural Pouch Basic |
| INFI MP Procedural Pouch 0.9mm |
| INFI MP Procedural Pouch 1.1mm |
| Infiniti MP Drain Bags |
| INFI Multi-Pak Basic Stand Alone |
| INFI Multi-Pak 0.9 Tipless Stand Alone |
| INFI Multi-Pak 1.1 Tipless Stand Alone |
| Ultrasound FMS, Intrepid Plus, Basic |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Tipless Ultra |
| Ultrasound FMS, Intrepid Plus, 1.1 MM Tipless Ultra |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Tipless Nano |
| Ultrasound FMS, Intrepid Plus, 1.1 MM Tipless Nano |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 30K, Ultra |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 45K, Ultra |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Tipless |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 30K, Nano |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 45K, Nano |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 30 OZil, Nano |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 45 OZil, Nano |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 30 OZil, Ultra |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 45 OZil, Ultra |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 30K |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 45K |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 30 OZil |
| Ultrasound FMS, Intrepid Plus, 0.9 MM Mini-Flared ABS 45 OZil |
{4}------------------------------------------------
·
{5}------------------------------------------------
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV 2 1 2011
Alcon Research, Ltd. c/o Mr. Martin Kaufman, RAC Director, Regulatory Affairs 15800 Alton Parkway Irvine, CA 92816
Re: K112425
Trade/Device Name: Infiniti Vision System (Evergreen II) Regulation Number: 21 CFR 886.4670 Regulation Name: Phacofragmentation system Regulatory Class: II Product Codes: HQC, HQR, MLZ, KYB Dated: August 19, 2011 Received: August 23, 2011
Dear Mr. Kaufman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{7}------------------------------------------------
CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kesia Alexander
Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): _ k1/2 425
Device Name: Infiniti® Vision System
Indications for Use:
The Infinit® Vision System is indicated for emulsification, separation, and removal of cataracts, the removal of residual cortical material and lens epithelial cells, vitreous aspiration and cutting associated with anterior vitrectomy, bipolar coagulation, and intra-ocular lens injection. The AutoSert™ IOL Injector Handpiece is intended to deliver qualified AcrySof® intraocular lenses into the eye following cataract removal.
The following system modalities additionally support the described indications:
- Ultrasound with UltraChopper Tip achieves the functionality of cataract separation.
- AquaLase achieves the functionality for removal of residual cortical material and lens epithelial cells
- The AutoSert ™ IOL Injector Handpiece achieves the functionality of injection of intraocular lenses. The AutoSert™ is indicated for use with ACRYSOF lenses SN60WF and SN6AD1, as well as approved AcrySof lenses that are specifically indicated for use with this inserter, as indicated in the approved labeling of those lenses.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear. Nose and Throat Devices
510(k) Number f/12425
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.