K112366 · Toshiba Medical Systems Coporation, Japan · MOS · Sep 15, 2011 · Radiology
Device Facts
Record ID
K112366
Device Name
4CH FLEX SPEEDER
Applicant
Toshiba Medical Systems Coporation, Japan
Product Code
MOS · Radiology
Decision Date
Sep 15, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Imaging of the extremities (Foot and Hand), joints (knee, shoulder, ankie, wrist and 1. elbow), and body (chest, abdomen and pelvis). MJAJ-192A is intended to be used on the Toshiba Vantage Titan 3T MR System.
Device Story
4ch Flex SPEEDER (MJAJ-192A) is a flexible, phased-array receive-only coil for MRI systems. Device consists of four single-loop coil elements arranged in a row on a PC board. Coil wraps around patient anatomy (extremities, joints, trunk) to receive NMR signals. Operates with Toshiba Vantage Titan 3T MR System; coil decouples during transmission via PIN diodes. Used by clinicians for diagnostic imaging, including 2D/3D imaging, MR angiography, MR fluoroscopy, and MRS. Provides signal input to the MR system for image reconstruction; aids clinical decision-making through visualization of anatomical structures.
Clinical Evidence
Bench testing only. Sample phantom images and clinical images provided to demonstrate imaging performance.
Technological Characteristics
Phased-array receive-only coil; 4 coil elements; flexible form factor; PIN diode decoupling; 3T field strength compatibility; biocompatible materials.
Indications for Use
Indicated for diagnostic imaging of extremities (foot, hand), joints (knee, shoulder, ankle, wrist, elbow), and body (chest, abdomen, pelvis) using the Toshiba Vantage Titan 3T MR System.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K112 366 p½" in bold black font. Below this text is the text "Appendix A" in a smaller, regular black font. The text is centered in the image and the background is white.
SEP 1 5 2011
## 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
#### Classification and Device Name 1.
### Classification Name
Magnetic Resonance Diagnostic Device
### Model Number
MJAJ-192A
### Trade/Propriety Name
4ch Flex SPEEDER
#### 2. Establishment Registration
2020563
#### 3. U.S. Agent Name and Address
### Agent Name ·
Toshiba America Medical Systems, Inc. (TAMS) 2441 Michelle Drive Tustin, CA 92780
### Contact Person
Paul Biggins, Director Requlatory Affairs T: (714) 730-5000 F: (714) 730-1310 pbiggins@tams.com
#### 4. Manufacturing Site
Toshiba Medical Systems Corporation (TMSC) 1385, Shimoishigami, Otawara-Shi, Tochigi 324-8550, Japan
#### 5. Date of Submission
August 31, 2011
#### 6. Device Intended Use
### Field Strength
3 T
### Resonant Nucleus
Hydrogen
### Anatomical Region of Interest
.
The extremities, joints, and trunk of the human body
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "K112366 p2/2 Appendix A". The text is in bold, and the font is sans-serif. The text is arranged in two lines, with "K112366 p2/2" on the first line and "Appendix A" on the second line. The text is black, and the background is white.
### Diagnostic Use
Diagnostic imaging of the human body, fluid visualization, 2D and 3D imaging, MR angiography, MR fluoroscopy and MRS
#### 7. Device Description
The 4ch Flex SPEEDER (MJAJ-192A), is a phased array coil that can receive NMR signal from the extremities, joints, and trunk of the human body. The 4ch Flex SPEEDER is mechanically flexible, and can wrap around various regions mentioned above.
The 4ch Flex SPEEDER (MAJA-192A) consists of four coil elements and PC board. The coil elements shape single loop and the four single loop coil elements are arranged to the row. The coil is decoupled during transmission from QD whole body by means of activating the PIN diodes.
MJAJ-192A is intended to be used on the Toshiba Vantage Titan 3T MR System.
#### 8. Safety Parameter
#### Maximum static field strength
3 T
#### Maximum dB/dt
1st operation mode specified in IEC60601-2-33 (2002)
#### Maximum SAR
1st operation mode specified in IEC60601-2-33 (2002)
### Peak and A-weighted Acoustic Noise Level
Not applicable
### Biocompatibility
All materials used in contact with the patient have a history of use or test data that demonstrates its biocompatibility, i.e., non-toxic, non-irritating.
#### Imaging Performance Parameter 9.
Sample phantom images and clinical images are presented in Appendix F & G.
#### 10. Equivalency Information
Toshiba Medical Systems Corporation believes that this 4ch Flex SPEEDER (Model Number MJAJ-192A is substantially equivalent to the current 4ch Flex SPEEDER (Model number MJAJ-197A, K111251).
#### 11. Software
There is no software required for this coil.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes extending upwards and to the right. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Toshiba Medical Systems Corporation, JAPAN % Mr. Paul Biggins Director Regulatory Affairs/US Agent 2441 Michelle Dr. TUSTIN CA 92780
SEP 15 2011
Re: K112366
Trade/Device Name: 4ch Flex Speeder, MJAJ-192A Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: August 16, 2011 Received: August 17, 2011
Dear Mr. Biggins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (2) CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Statel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K 117 366
Device Name:
4ch Flex SPEEDER, MJAJ-192A
### Indications for Use:
- Imaging of the extremities (Foot and Hand), joints (knee, shoulder, ankie, wrist and 1. elbow), and body (chest, abdomen and pelvis).
MJAJ-192A is intended to be used on the Toshiba Vantage Titan 3T MR System.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Pastet
(Division Sign Off)
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
**510(k) Number** K112366
Page 1 of
Indication for Use Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.