# VESSEL GUARDIAN (K112357)

_Lexion Medical, LLC · HIF · Nov 10, 2011 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K112357

## Device Facts

- **Applicant:** Lexion Medical, LLC
- **Product Code:** [HIF](/productcode/HIF.md)
- **Decision Date:** Nov 10, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1730
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

Vessel Vapor™ is a gas conditioner accessory device intended to heat, humidify and filter an insufflating CO2 gas stream being used in minimally invasive cardiovascular. thoracoscopic, and vessel harvesting surgical procedures.

## Device Story

Vessel Vapor™ is a single-use accessory for insufflators or regulated CO2 sources. It attaches to the gas outlet to condition CO2 before it enters the patient's surgical cavity. The system includes a disposable tubing set with an in-line filter and a heater/humidifier cassette, managed by a control module containing safety and control circuits. CO2 gas flows through the filter, then through the cassette where it is warmed and humidified, and finally through a Luer lock connector to an insufflation needle or trocar. Used in minimally invasive cardiovascular, thoracoscopic, and vessel harvesting procedures, the device aims to mitigate the effects of dry, cold insufflation gas on internal tissues. It is operated by surgical staff in an OR setting.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technical comparison to predicate devices.

## Technological Characteristics

Single-use disposable tubing set with in-line filter and heater/humidifier cassette. Includes a control module for system safety and heating regulation. Connects via Luer lock to standard insufflation needles/trocars. Operates by warming and humidifying CO2 gas stream.

## Regulatory Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

## Predicate Devices

- Vessel Guardian® ([K102136](/device/K102136.md))
- Insuflow® ([K090456](/device/K090456.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a sequence of handwritten characters and numbers. The sequence starts with the letter 'K', followed by the numbers '112357'. The handwriting style appears somewhat casual, with slight variations in the thickness and slant of the strokes.

NOV 1 0 2011

# SECTION 1. SUMMARY AND CERTIFICATION

## A. 510(K) SUMMARY

, ''

### Summary of Safety and Effectiveness

In accordance with 21 CFR 807.92, the following information constitutes the summary of safety and effectiveness for Vessel VaporTM

| SUBMITTER'S NAME:   | LEXION Medical LLC                       |
|---------------------|------------------------------------------|
| ADDRESS:            | 545 Atwater Circle<br>St. Paul, MN 55103 |
| CONTACT PERSON:     | Bernard (Bud) Horwath                    |
| TELEPHONE NUMBER:   | 651-361-8041                             |
| FAX NUMBER:         | 651-351-8001                             |
| DATE OF SUBMISSION: | 12 August 2011                           |

#### Identification of device 1.

Proprietary Name: Vessel Vapor™ Common Name: Laparoscopic Insufflator Gas Conditioner Classification Status: Class II per regulations 884.1730 Product Code: HIF

#### 2. Equivalent devices

LEXION Medical believes that Vessel Vapor™ is substantially equivalent to the following devices:

Vessel Guardian®, K102136 Insuflow®, K090456

Vessel Vapor™ is the same gas conditioner insufflator accessory device and has the same intended use as the predicate devices cleared under 510(k) K102136 and K090456.

#### 3. Description of the Device

The Vessel Vapor™ device is a single use device that attaches to the outlet port of an insufflator or other regulated CO2 source and is designed to warm and humidify the CO2 gas stream prior to insufflation into the surgical cavity. The Vessel Vapor™ device consists of a disposable tubing set with a filter and heater/humidifier cassette and a control module that houses the control and safety circuits for the system.

Regulated CO2 gas flows into the Vessel Vapor™ device, through the in-line filter, continues along the tube to enter the Vessel Vapor™ device cassette that contains the heating element and humidification media, through a tube that connects via a Luer lock connector to a gas entrance port or an insufflation needle/trocar and finally flows into the patient's surgical cavity.

{2}------------------------------------------------

#### 4. Intended use

. . .

Vessel Vapor™ is a gas conditioner accessory device intended to heat, humidify and filter an insufflating CO2 gas stream being used in minimally invasive cardiovascular. thoracoscopic, and vessel harvesting surgical procedures.

The indications for use for the Vessel Vapor™ are patterned after the predicate devices: however, they have been expanded to include minimally invasive cardiovascular surgical procedures. The addition of this specific indication does not affect the safety and effectiveness of the device when used as labeled.

#### ડ. Technological characteristics, comparison to predicate device.

Technically, the Vessel Vapor™ is identical to the Vessel Guardian® and Insuflow devices cleared for market in 510(k) K102136 and K090456, respectively.

#### Conclusion 6.

Based on a comparison to the predicate devices, it is the conclusion of LEXION Medical that Vessel Vapor™ is substantially equivalent to devices already on the market being used for this application (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness.

{3}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

NOV 1 0 2011

Mr. Bernard Horwath Regulatory Consultant LEXION Medical LLC 545 Atwater Circle ST PAUL MN 55103

Re: K112357 Trade/Device Name: Vessel Vapor™ Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic insufflator Regulatory Class: II Product Code: HIF. OSV Dated: August 12, 2011 Received: August 16, 2011

Dear Mr. Horwath:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{4}------------------------------------------------

Page 2

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default:htm.

Sincerely yours,

Hubert H. Humphrey

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# INDICATIONS FOR USE

K112357 510(k) Number

Device Name: Vessel Vapor™

### Indications for Use:

Vessel Vapor™ is a gas conditioner accessory device intended to heat, humidify and filter an insufflating CO2 gas stream being used in minimally invasive cardiovascular, thoracoscopic, and vessel harvesting surgical procedures. ・・

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use_X OR Over the Counter Use (Per 21 CFR 801.109) vision Signoductive, Gastro-Renal, and

0(k) Numbe

---

**Source:** [https://fda.innolitics.com/device/K112357](https://fda.innolitics.com/device/K112357)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).