K112308 · C.R. Bard, Inc. · PDU · Aug 17, 2011 · Cardiovascular
Device Facts
Record ID
K112308
Device Name
CROSSER RECANALIZATION SYSTEM
Applicant
C.R. Bard, Inc.
Product Code
PDU · Cardiovascular
Decision Date
Aug 17, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The CROSSER™ Recanalization System is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions via atherectomy. The device is contraindicated for use in the carotid arteries. The CROSSER™ Catheter is only intended for use with the CROSSER™ Generator. Refer to the CROSSER™ Generator Manual of Operations for proper use.
Device Story
System uses high-frequency mechanical vibration to recanalize peripheral artery chronic total occlusions (CTO). Components: electronic generator, foot switch, transducer, single-use catheter, sterile drape. Generator/transducer convert AC power to high-frequency mechanical vibrations; energy propagates through Nitinol core wire to titanium or stainless steel catheter tip. Vibrational energy ablates plaque via mechanical impact and cavitation. Used in clinical settings by physicians. Requires power injector for saline delivery. Output is mechanical vibration at catheter tip; enables guidewire passage through occlusions. Benefits patient by facilitating recanalization of otherwise difficult-to-cross arterial blockages.
Clinical Evidence
Bench testing only. No clinical data presented. Substantial equivalence supported by prior bench testing and risk assessment confirming the device meets design requirements for the clarified indications.
Technological Characteristics
System consists of electronic generator, foot switch, transducer, and single-use catheter. Catheter materials: Nitinol core wire, titanium or stainless steel tip, Pebax shaft with hydrophilic coating. Energy source: AC power converted to high-frequency mechanical vibration. Catheter configurations vary by diameter (0.6mm to 1.5mm) and length (125cm to 154cm). Sterilization: Not specified. Connectivity: Standalone system requiring manual operation via foot switch.
Indications for Use
Indicated for patients with peripheral artery chronic total occlusions to facilitate intra-luminal guidewire placement via atherectomy. Contraindicated for use in carotid arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
K112308 page 1 of 4
AUG 17 2011
### CROSSER™ Recanalization System
#### 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### Submitter Information:
| Applicant: | FlowCardia, Inc<br>745 N. Pastoria Ave.<br>Sunnyvale, CA 94085 |
|------------|----------------------------------------------------------------|
| Phone: | 480-638-2939 |
| Fax: | 480-449-2546 |
| Contact: | Timothy Wade, Regulatory Affairs Associate |
| Date | July 28, 2011 |
| Device Trade Name: | CROSSER TM Recanalization System |
|-----------------------|------------------------------------------------------------------------|
| Common or Usual Name: | Peripheral Atherectomy Catheter (21 CFR<br>870.4875, Product Code MCW) |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
#### Predicate Devices:
- CROSSER™ System .
- Models: GEN200, CRU14S, CRU14P, CRU18 (K072776; cleared o December 7, 2007)
- Models: GEN200, CRUO14S, CRUO14P (K091119; cleared May o 15, 2009)
- Models: GEN200, CRUS6 (K092175; cleared February 17, 2010) O
- Diamondback 360 Orbital Atherectomy System (K090521; cleared March ● 20, 2009)
- . Jetstream System (K110626; cleared April 20, 2011)
Image /page/0/Picture/17 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, with a white fill. The letters are evenly spaced and appear to be of uniform size. The overall impression is clean and professional.
{1}------------------------------------------------
#### Device Description:
The CROSSER™ Recanalization System is a high frequency mechanical vibration device designed for placement of guidewires through obstructed peripheral arteries by the removal or reduction of occlusive material. The system consists of an electronic CROSSER™ Generator, Foot Switch, high frequency Transducer, single-use CROSSER™ Catheter and sterile drape. The catheters are offered in a variety of configurations as shown in the table below. The system is required to be used with a power injector system for the delivery of saline. The CROSSER™ Catheter, which is intended for one procedure only, is connected to the CROSSER™ Generator through the high frequency The Foot Switch is used to activate the CROSSER™ System. The Transducer. CROSSER™ Generator and Transducer convert AC power into high frequency mechanical vibrations, which are propagated through a Nitinol core wire to the titanium or stainless steel tip of the CROSSER™ Catheter. The main body of the catheter is constructed from Pebax and a hydrophilic coating which covers the distal end of the catheter. The vibrational energy ablates intravascular plaque by mechanical impact and cavitational effects.
| | CROSSER Catheters<br>14P and 14S | CROSSER Catheter<br>18 | CROSSER Catheter<br>S6 |
|----------------------------------|----------------------------------------|--------------------------------|--------------------------------|
| Working Length | 146cm | 125cm | 154cm |
| Tip Outer Diameter | 1.1mm (0.044") | 1.5mm (0.061") | 0.6mm (0.025") |
| Catheter Shaft Outer<br>Diameter | Rapid<br>Exchange<br>1.3mm<br>(0.052") | 1.5mm (0.058") | 1.3mm (0.051") |
| | Over The<br>Wire<br>1.6mm<br>(0.063") | | |
| Guidewire Lumen<br>Length | Rapid<br>Exchange<br>20cm | Rapid Exchange<br>20cm | n/a* |
| | Over The<br>Wire<br>152 cm | | |
| Guidewire Compatibility | 0.014" (0.36mm) | 0.018" (0.46mm) | n/a* |
| Guide Catheter<br>Compatibility | 6Fr<br>(min ID 0.067" [1.7mm]) | 7Fr<br>(min ID 0.080" [2.0mm]) | 5Fr<br>(min ID 0.053") [1.3mm] |
| Sheath Compatibility | 5Fr<br>(min ID 0.067" [1.7mm]) | 6Fr<br>(min ID 0.080" [2.0mm]) | 5Fr<br>(min ID 0.053") [1.3mm] |
* CROSSER™ Catheter S6 does not contain a guidewire lumen but creates a channel large enough to accommodate a guidewire up to 0.018".
#### Indications for Use of Device:
The CROSSER™ Recanalization System is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions
Image /page/1/Picture/8 description: The image shows the word "BARD" in a stylized, bold font. Each letter is outlined in black, giving it a three-dimensional appearance. The letters are closely spaced, creating a compact and unified design.
{2}------------------------------------------------
via atherectomy. The device is contraindicated for use in the carotid arteries. The CROSSER™ Catheter is only intended for use with the CROSSER™ Generator. Refer to the CROSSER™ Generator Manual of Operations for proper use.
### Review of the Modified Indications for Use:
The modified indication for use statement for the CROSSER™ Recanalization System does not raise any new issues of safety and effectiveness as demonstrated through the risk analysis process based on the proposed indications for use statement as compared to the predicate devices, the CROSSER™ System, the Diamondback System, and the Jetstream System. The modified indication clarifies the mechanism of action by which the CROSSER™ Recanalization System recanalizes a Chronic Total Occlusion (CTO) as the removal or reduction of occlusive material from the artery (atherectomy). The Diamondback System and Jetstream Systems are also indicated for the atherectory of occlusive material in the peripheral arteries. Therefore, the subject device, the CROSSER™ Recanalization System, is substantially equivalent to the predicate devices.
### Technological Comparison to Predicate Devices:
The CROSSER™ Recanalization System has the following similarities to the predicate devices:
- Similar intended use (all predicates) .
- Similar indications for use (all predicates) .
- Same target population (all predicates) .
- Same fundamental scientific technology (all predicates) .
- Same operating principle (Crosser predicates) .
- Similar operating principle (Diamondback and Jetstream predicates) .
- Same mechanism of action (all predicates) .
- Same materials (Crosser predicates) .
- Same packaging materials and configuration (Crosser predicates) .
- Same sterility assurance level and method of sterilization (Crosser ● predicates)
#### Performance Data:
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a handwritten note that reads "K112308 page 41 of 4". The note appears to be a page number or document identifier. The handwriting is cursive and legible.
ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ
To demonstrate substantial equivalence of the subject device, the CROSSER™ Recanalization System to the predicate devices, the technological characteristics and performance criterion were evaluated through bench testing. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, no additional in vitro tests or in vivo studies were required as the Crosser System previously met all requirements. As this 510(k) is clarifying the system mechanism of action, prior bench testing is presented.
#### Conclusions:
The subject device, the Crosser™ Recanalization System, satisfied the design requirements associated with the modified indications for use statement, as specified by applicable standards, guidance, test protocols and/or customer inputs. Therefore, the Crosser™ Recanalization System is substantially equivalent to the legally marketed predicate devices, the Crosser™ System, the Diamondback System, and the Jetstream System.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, with three human profiles incorporated into the design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002
C.R. Bard, Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th St. NW Buffalo, MN 55313
SEP 1 8 2013
Re: K112308
Trade/Device Name: Crosser Recanalization System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: August 10, 2011 Received: August 11, 2011
Dear Mr. Job:
This letter corrects our substantially equivalent letter of August 17, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
ZZn.Z
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
# 510(k) Number (if known): K112308
..
Device Name: CROSSER™ Recanalization System
Indications for Use: The CROSSER™ Recanalization System is indicated · to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions via atherectomy. The device is contraindicated for use in carotid arteries.
The CROSSER™ Catheter is only intended for use with the CROSSER™ Generator. Refer to the CROSSER™ Generator Manual of Operations for proper use.
Prescription Use_X (Part21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image does not contain any text.
(Division Sign-Oit) Division of Cardiovascular Devices
510(k) Number K112309
Image /page/6/Picture/14 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The font is simple and easy to read.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.