K112293 · Time Medical Systems, Inc. · MOS · Sep 30, 2011 · Radiology
Device Facts
Record ID
K112293
Device Name
HTS EXTREMITY COIL, HTS SURFACE COIL
Applicant
Time Medical Systems, Inc.
Product Code
MOS · Radiology
Decision Date
Sep 30, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
HTS Surface Coil is used to image peripheral anatomies, such as, wrist, ankle, Temporo-mandibular Joints (TMJ), Eye, finger and other parts of the body, close to the surface of skin. It is compatible to be used with Time Medical's 0.35T PICA Whole Body MRI System. HTS Extremity Coil is predominantly used for imaging anatomies, such as, knee, ankle, and wrist. It is compatible to be used with Time Medical's 0.35T PICA Whole Body MRI System.
Device Story
HTS Surface and Extremity Coils are receive-only radiofrequency (RF) coils for MRI imaging; designed for peripheral anatomy (wrist, ankle, TMJ, eye, finger, knee). Coils utilize liquid nitrogen (LN2) cooling to achieve high-temperature superconductivity (HTS). Devices connect to Time Medical 0.35T PICA Whole Body MRI System. Coils capture RF signals from patient anatomy; signals are processed by the MRI system to generate diagnostic images. Used in clinical settings by trained MRI technologists/radiologists. Output provides anatomical visualization for clinical diagnosis. Benefit includes high-sensitivity imaging of peripheral structures.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on technological comparison, design, and material similarity to predicate devices.
Technological Characteristics
High-temperature superconducting (HTS) receive-only RF coils. Cooling: Liquid Nitrogen (LN2). Construction: ABS plastic. Configuration: Single-channel (Surface coil for HTS Surface Coil; Volume coil for HTS Extremity Coil). Energy: Passive RF receive-only. Connectivity: Proprietary interface for Time Medical 0.35T PICA MRI system.
Indications for Use
Indicated for imaging peripheral anatomies (wrist, ankle, TMJ, eye, finger, knee) in patients requiring MRI examination. Compatible with Time Medical 0.35T PICA Whole Body MRI System. No specific contraindications listed.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
HTS Coil as Accessory of the MONA MRI System (K092230)
{1}------------------------------------------------
#### TIME MEDICAL SYSTEMS
510(k) Report for HTS Surface Coil, HTS Extremity Coil
| | | Attachment 2: 510(k) Summary | | |
|-----|----------------------------------|------------------------------|--|---|
| | Attachment 2 | K112293 | | |
| | 510 (k) Summary | SEP 30 2011 | | |
| 2.1 | summary | | | 1 |
| 2.2 | General Safety and Effectiveness | | | 2 |
| 2.3 | Substantial Equivalence | | | 2 |
# 2.1 SUMMARY
، ﺗﺴ
| Comparison<br>Element | Filing Device | Predicate Device | | | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | TBD | K092230<br>(Mona<br>Orthopedic<br>MRI System)<br>(an accessory<br>of LNH) | K022863 | | |
| Product Code | MOS | MOS | | | |
| Applicant | Time Medical | Time Medical | University of Hong Kong | | |
| Regulation Number | 892.1000 | 892.1000 | 892.1000 | | |
| Panel | Radiology | Radiology | Radiology | | |
| Class | Class II | Class II | Class II | | |
| Device Name (Model<br>Number) | HTS Surface Coil<br>(TM-HTS-SF001-0R35),<br>HTS Extremity Coil<br>(TM-HTS-KN001-0R35) | HTS Coil as<br>Accessory of<br>the MONA MRI<br>System | HTS Coil<br>(HTS Coil) | | |
| Characteristics | HTS Surface Coil | | HTS Extremity<br>Coil | K092230 | K022863 |
| Intended Use | HTS Surface Coil is<br>used to image<br>peripheral<br>anatomies, such as,<br>wrist, ankle,<br>Temporo-mandibular<br>Joints (TMJ), Eye,<br>finger and other<br>parts of the body,<br>close to the surface<br>of skin. It is<br>compatible to be<br>used with Time<br>Medical's 0.35T<br>PICA Whole Body<br>MRI System. | | HTS Extremity<br>Coil is<br>predominantly<br>used for<br>imaging<br>anatomies,<br>such as, knee,<br>ankle, and<br>wrist. It is<br>compatible to<br>be used with<br>Time Medical's<br>0.35T PICA<br>Whole Body<br>MRI System. | HTS Surface<br>Coil is used to<br>image<br>peripheral<br>anatomies<br>such as, Wrist,<br>Ankle, TMJ,<br>Eye, Finger<br>and other parts<br>of the body,<br>close to the<br>surface of the<br>skin. | MRI surface coil for<br>peripheral<br>anatomical imaging<br><br>Anatomical region:<br>Temporo-mandibular<br>joint (TMJ), wrist<br>and other anatomies<br>that are no deeper<br>than 1.5 inches from<br>the skin. |
| Applicable Systems | Time Medical,<br>Pica Whole-body<br>MRI System | | Time Medical,<br>Pica Whole-body<br>MRI System | Time Medical,<br>Mona<br>Orthopedic<br>MRI System | GE<br>0.2T Signa Profile<br>MRI System |
| Mode of Operation | Single-Channel | | Single-Channel | Single-Channel | Single-Channel |
| Coil Configuration (Linear,<br>Quad, array, etc) | Single Channel<br>surface coil | | Single Channel<br>Volume coil | Single Channel<br>surface coil | Single Channel<br>surface coil |
{2}------------------------------------------------
| Safety | | | | |
|------------------------------------|---------------------------|---------------------------|---------------------------|---------------------------|
| SAR (T/R Coils) | N/A, Receive<br>Only Coil | N/A, Receive<br>Only Coil | N/A, Receive<br>Only Coil | N/A, Receive Only<br>Coil |
| Power Input Protection (T/R Coils) | N/A, Receive<br>Only Coil | N/A, Receive<br>Only Coil | N/A, Receive<br>Only Coil | N/A, Receive Only<br>Coil |
| Material | ABS | ABS | ABS | PVC |
| Technological<br>Parameter | Filing Device:<br>HTS Surface Coil | HTS Extremity Coil | Predicate Device:<br>HTS Coil as Accessory of<br>K092230 (Mona Orthopedic MRI System) | HTS Coil (K022863) |
|----------------------------|------------------------------------|-------------------------------|---------------------------------------------------------------------------------------|--------------------------------|
| T/R | Receive-Only | Receive-Only | Receive-Only | Receive-Only |
| Coil Type | Single Channel<br>Surface Coil | Single Channel<br>Volume Coil | Single Channel<br>Surface Coil | Single Channel<br>Surface Coil |
| Cooling liquid | LN2 (Liquid<br>Nitrogen) | LN2 (Liquid<br>Nitrogen) | LN2 (Liquid<br>Nitrogen) | LN2 (Liquid<br>Nitrogen) |
## 2.2 GENERAL SAFETY AND EFFECTIVENESS
The technological characteristics of Time Medical's HTS Surface Coil and HTS Extremity Coil are similar to the predicate device. They work on the same principle, have similar design, are constructed of similar materials and are of similar safety and effectiveness.
lt does not induce other safety issues and warning than already valid for the current cleared RF external coils.
## 2.3 SUBSTANTIAL EQUIVALENCE
The HTS Surface coil and the HTS Extremity Coil are substantially equivalent to the other HTS coil(s) which has been cleared for commercial distribution as part of Time Medical Limited's MONA Orthopedic MRI System (ref. K092230), and the HTS coil of University of Hong Kong (K022863).
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines representing its wings.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Time Medical Systems, Inc. % Mr. John Baby Director-Regulatory Affairs Time Medical Limited, G/F Bio-Informatics Centre No. 2 Science Park West Avenue Hong Kong Science Park, Shatin, New Territories. Hong Kong CHINA
SEP 3 0 2011
Re: KI12293
Trade/Device Name: HTS Surface Coil, HTS Extremity Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: August 1. 2011 Received: August 10. 2011
Dear Mr. Baby:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809): medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
510(k) Report for HTS Surface Coil, HTS Extremity Coil
TIME MEDICAL
Attachment 3: Indications of Use
### Indications for Use
510(k) Number (if known):
Device Name: HTS Surface Coil, HTS Extremity Coil
Indications for Use:
HTS Surface Coil is used to image peripheral anatomies, such as, wrist, ankle, Temporomandibular Joints (TMJ), Eye, finger and other parts of the body, close to the surface of skin. It is compatible to be used with Time Medical's 0.35T PICA Whole Body MRI System.
HTS Extremity Coil is predominantly used for imaging anatomies, such as, knee, ankle, and wrist. It is compatible to be used with Time Medical's 0.35T PICA Whole Body MRI System.
Prescription Use _____x______x
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Pastel
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
**510(k) Number** K112293
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.