ACCUMED WRIST AUTOMATIC BLOOD PRESSURE MONITOR

K112270 · Accumed Healthcare (Shanghai), Inc. · DXN · Nov 22, 2011 · Cardiovascular

Device Facts

Record IDK112270
Device NameACCUMED WRIST AUTOMATIC BLOOD PRESSURE MONITOR
ApplicantAccumed Healthcare (Shanghai), Inc.
Product CodeDXN · Cardiovascular
Decision DateNov 22, 2011
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

The Accumed Wrist Automatic Blood Pressure Monitor, Model ABA-701 is a noninvasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home. The device has an irregular heartbeat (IHB) indicator. The device detects the appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected.

Device Story

Wrist-worn automatic blood pressure monitor; uses oscillometric method to measure systolic/diastolic blood pressure and pulse rate in adults. Electronic pressure sensor converts cuff pressure variations into electrical signals; microprocessor analyzes signals to determine blood pressure and pulse. Includes irregular heartbeat (IHB) detection feature, triggering an LCD symbol if pulse rate deviation exceeds 25%. Intended for home use by patients. Provides digital readings to the user to assist in monitoring cardiovascular health.

Clinical Evidence

Bench testing only. Device performance verified against ANSI/AAMI SP10, IEC/EN 60601-1, IEC/EN 60601-1-2, and IEC/EN 60601-1-4. Biocompatibility testing for cytotoxicity, sensitivity, and irritation performed per ISO 10993 with passing results.

Technological Characteristics

Oscillometric measurement principle; semiconductor pressure sensor; LCD display; battery-powered. Biocompatible cuff materials tested per ISO 10993. Software-based signal processing for pressure-to-electrical signal conversion and IHB detection.

Indications for Use

Indicated for adult patients for non-invasive measurement of systolic and diastolic blood pressure and pulse rate at home using the oscillometric method. Includes irregular heartbeat (IHB) detection.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 11 2270 1/3 NOV 2 2 2 2011 Accumed Healthcare(Shanghai) Inc. No. 6133, Huyi Rd., Waigang, Jiading District, Shanghai, China . 1 ---------------------------------------------------------------------- # Accumed 510(k) Summary | 510(k) owner's information: | Accumed Healthcare (Shanghai) Inc.<br>No. 6133, Huyi Rd., Waigang, Jiading<br>District, Shanghai, China.<br>Tel: +86-21-69519220<br>Fax: +86-21-69519221 | |-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------| | Official Contact: | Cindy Green, Authorized Representative<br>P.O. Box 1277; Maple Valley, WA 98038<br>Tel: (425) 432-8623 | | Proprietary or Trade Name: | Accumed Wrist Automatic Blood Pressure<br>Monitor, Model ABA-701 | | Common/Usual Name: | Non-Invasive Blood Pressure Monitor | | Classification Name: | System, Measurement, Blood-Pressure,<br>Non-Invasive | | Regulation | CFR 870.1130 | | Product Code | DXN | | Predicate Device: | Citizen CH-609, K100055 | | Date Summary Prepared: | November 7, 2011 | # Device Description: The Accumed Wrist Automatic Blood Pressure Monitor, Model ABA-701 measures systolic and diastolic blood pressure and pulse rate of an adult individual. The method used to define blood pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The device utilizes the oscillometric method whereby the electronic pressure sensor converts variation in cuff pressure to electrical signals. The electrical signals are analyzed and used to define the systolic and diastolic measurements and determine the pulse rate. The irregular heart beat (IHB) indicator is displayed when there is more than a 25% deviation from the averaged pulse rate measurements. {1}------------------------------------------------ Accumed Healthcare(Shanghai) Inc. No. 6133, Huyi Rd., Waigang, Jiading District, Shanghai, China ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . # Accumed ### Intended Use: The Accumed Wrist Automatic Blood Pressure Monitor, Model ABA-701 is a noninvasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home. The device has an irregular heartbeat (IHB) indicator. The device detects the appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected. ### Comparison to Predicate Device The Accumed Wrist Automatic Blood Pressure Monitor, Model ABA-701 is substantially equivalent to the Citizen CH-609 (K100055). The Accumed ABA-701 and the Citizen CH-609 predicate have the same: - · Intended use - · Intended use environment - · Intended patient population - · Display type - · Accuracy - · Measurement range - · Pulse range measurement - · Memory capacity - · Well known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse - · Same semi-conductor pressure sensor technology used to translate the pressure variations to electrical signals that can be interpreted by the microprocessor - · Irregular heart beat detection function The subject device and the predicate devices CH-609 are similar; however, not identical with regards to operating environment, storage environment, dimensions, and weight. The subject device is substantially equivalent to the Citizen CH-609 (K100055). There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate device. #### Performance Testing: Bench testing was conducted to demonstrate that the device meets its Requirements Specification. The following performance tests were completed: - ANSI/AAMI SP10 . - IEC/EN 60601-1 - IEC/EN 60601-1-2 - IEC/EN 60601-1-4 . {2}------------------------------------------------ r - 12 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - Accumed Healthcare(Shanghai) Inc. No. 6133, Huyi Rd., Waigang, Jiading District, Shanghai, China # Accumed The Accumed Wrist Automatic Blood Pressure Monitor, Model ABA-701 meets all applicable requirements. ... .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ## Biocompatibility Testing was performed for in vitro cytotoxicity, sensitivity, and irritation according to ISO 10993 requirements. Test results demonstrate that there is no cytotoxicity, sensitivity, or irritation caused by the Blood Pressure Monitor Cuff. ## Conclusion: Based on the evaluations completed there are no new safety or effectiveness issues introduced with this new medical device. Therefore, the Accumed Wrist Automatic Blood Pressure Monitor, Model ABA-701 is substantially equivalent to the predicate device; Citizen CH-609 (K100055). {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol appears to be an abstract representation of a human figure or a caduceus. The logo is presented in black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JAN 11 2012 Accumed Healthcare (Shanghai) Inc. c/o Cindy Green Authorized Representative P.O. Box 1277; Maple Valley, WA 98038 Re: K112270 KT12270 Trade/Device Name: Accumed Wrist Automatic Blood Pressure Monitor, Model ABA-701 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: November 8, 2011 Received: November 10, 2011 Dear Ms. Green: This letter corrects our letter of November 22, 2011 regarding the incorrect name identified on the Indications for Use statement form. We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bootion of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use stated in the onerosale) to regard cate of the Medical Device Amendments, or to commerce provision of May 20, 1977, and occordance with the provisions of the Federal Food, DNA devices that have occh roomsoried in actre approval of a premarket approval application (PMA). and Costine Act (710c) that Go not required to the general controls provisions of the Act. The I ou may, incretore, market the act include requirements for annual registration, listing of general controls provisions of the fice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (000 acove). Existing major regulations affecting your device can be may be subject to additional ochaons. "Bhoung and to any and to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean I least be advised that I Dri i securities your device complies with other requirements of the Act {4}------------------------------------------------ Page 2 - Ms. Cindy Green or any Federal statutes and regulations administered by other Federal agencies. You must or any I callul statures and reguirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical CITY art 607); adoling (21 CFR 803); good manufacturing practice requirements as set device-related adverse events) (2) OFF 003), good in and if applicable, the electronic forth in the quality by croms (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific advisor ser your Centers Offices/CDRH/CDRHOffices/ucm115809.htm for go to map. W. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . the Center 10 Devices and Parison ing by reference to premarket notification" (21CFR Part note the regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the I ou may obtain of Scheral memational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. M.g. Hilleber Bram D. Žuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Accumed Wrist Automatic Blood Pressure Monitor, Model ABA-701 Indications For Use: The Accumed Wrist Automatic Blood Pressure Monitor, Model ABA-701 is a non-invasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home. The device has an irregular heartbeat (IHB) indicator. The device detects the appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected. Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) A. Hillelson ion of Cardiovascular Devices Page 1 of 1 \$\frac{1}{1}\$ 5109) Number K112270
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...