POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, YELLOW COLOR
K112266 · Shandong Zhiyuan Medical Products Co., Ltd. · LYZ · May 8, 2012 · General Hospital
Device Facts
Record ID
K112266
Device Name
POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, YELLOW COLOR
Applicant
Shandong Zhiyuan Medical Products Co., Ltd.
Product Code
LYZ · General Hospital
Decision Date
May 8, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Intended Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
Device Story
Disposable synthetic vinyl patient examination gloves; yellow color; non-sterile. Worn by healthcare personnel on hands/fingers during patient examinations. Acts as physical barrier to prevent cross-contamination of body fluids, waste, or environmental contaminants between patient and examiner. Used in clinical settings. Benefits include infection control and protection for both patient and provider.
Clinical Evidence
No clinical data. Bench testing only. Testing included physical dimensions, water fill test (AQL 2.5), and biocompatibility (primary skin irritation and sensitization) per ASTM D5250-06.
Technological Characteristics
Synthetic vinyl material; powder-free (max 2 mg powder/glove); yellow color; non-sterile. Dimensions and physical properties conform to ASTM D5250-06. Biocompatible per skin irritation/sensitization testing.
Indications for Use
Indicated for use by healthcare and similar personnel as a disposable barrier worn on hands or fingers to prevent cross-contamination between the examiner and the patient.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
K070038 — POWDER FREE VINYL PATIENT EXAMINATION GLOVE, YELLOW · Super Safe Plastic Products Co., Ltd. · Mar 23, 2007
K122656 — POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, YELLOW COLOR · Shandong Yifan Plastic Products Co., Ltd. · May 6, 2014
K070188 — POWDER FREE VINYL PATIENT EXAMINATION GLOVE, YELLOW · Hong DI Plastic Products Co., Ltd. · Mar 19, 2007
K070044 — POWDER FREE (YELLOW) VINYL PATIENT EXAMINATION · Shijiazhuang Lianhe Shunda Plastic Products Co., L · Jan 29, 2007
K110857 — POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, YELLOW COLOR · Shandong Zhanpeng Medical Products Co., Ltd. · Jul 15, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
K112266
# 510 (K) SUMMARY
MAY - 8 2012
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.
## 1. Submitter's Identification:
Shandong Zhiyuan Medical Products Co., Ltd. No. 20 Shengshui Road Mihe Industry Park Qingzhou, Shandong, China
Date summary prepared: August 03, 2011
### 2. Name of the Device:
Shandong Zhiyuan Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves -- Powder Free, Yellow Color
## 3. Predicate Device Information:
Shijiazhuang Hongxiang Plastic Products Ltd. Synthetic Vinyl Patient Examination Gloves - Powder Free (K992821)
### 4. Device Description:
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirement of ASTM Standard D5250-06.
## ડ. Intended Use:
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
#### 6. Comparison to Predicate Devices:
Shandong Zhiyuan Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free, yellow color are substantially equivalent in safety and effectiveness to the Shijiazhuang Hongxiang Plastic Products Co., Ltd.(K992821) and Shanghai Jiawang Industrial Co., Ltd. (K031854)
{1}------------------------------------------------
## Discussion of Non-Clinical tests performed for Determination of 7. Substantial Equivalence are as follows:
The standards used for Shandong Zhiyuan Medical Products Co., Ltd. glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AOL 2.5.
The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5, Inspection Level I, meeting these requirements, Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.
There are no special labeling claims and we do not claim our gloves as hypoallergenic is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).
### Discussion of Clinical Tests Performed: 8:
Not Applicable - There is no hypoallergenic claim.
## 9. Conclusions:
Shandong Zhiyuan Medical Products Co., Ltd. Synthetic Vinyl Patient Examination Gloves, Powder-Free, yellow color conform fully to ASTM-D-5250-06 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The text is in all capital letters and is evenly spaced around the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shandong Zhiyuan Medical Products Co., Ltd C/O Mr. Frank Liu Official Correspondent Basic Medical Industries, Inc. 12390 East End Avenue Chino, California 91710
MAY - 8 2012
Re: K112266
Trade/Device Name: Patient Vinyl Examination Gloves, Powder Free, Non-Sterile, Yellow Color
Regulation Number: 21 CFR 880.6250
Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYZ
Dated: April 16, 2012
Received: April 19, 2012
Dear Mr. Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 - Mr. Liu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
hh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{4}------------------------------------------------
Shandong Zhiyuan Medical Products Co., Ltd.
No. 20, shengshui Road, Mihe Industry Park Qingzhou, Shandong, China
# INDICATIONS FOR USE
Shandong Zhiyuan Medical Products Co., Ltd. Applicant:
510(k) Number:
Device Name:
Patient Vinyl Examination Gloves, Powder free, Non-Sterile, Yellow Color
Indications of Use:
A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)
Prescription Use _______________
x Over the Counter Use ___
Factory Initials
E. Elliott S., Clavens-Will
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
11226 510(k) Number:
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.