SNORERX 9.0

K112205 · Consumer Health Products, Inc. · LRK · Nov 16, 2011 · Dental

Device Facts

Record IDK112205
Device NameSNORERX 9.0
ApplicantConsumer Health Products, Inc.
Product CodeLRK · Dental
Decision DateNov 16, 2011
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.5570
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Consumer Health Products "SnoreRx NS 9.0" is intended for use on adult patients 18 years of age or older as an aid for the reduction of snoring.

Device Story

SnoreRx NS 9.0 is an intraoral mandibular anterior repositioning device; used at night to reduce snoring. Device consists of two custom-fabricated acrylic trays fitting over upper and lower dental arches; trays engage in anterior area. Mechanism advances lower jaw to increase pharyngeal space; minimizes air obstruction and turbulence during sleep. Device is patient-worn; intended for adult use. Benefits include improved air exchange during sleep. No clinical data provided; substantial equivalence based on design, intended use, and technological characteristics compared to predicate devices.

Clinical Evidence

No clinical data. Substantial equivalence determined via non-clinical performance testing: dimensional conformance, visual inspection, design verification (airway passage), and biocompatibility (ISO 10993-5, ISO 10993-10). Materials confirmed identical to predicate devices.

Technological Characteristics

Intraoral mandibular repositioning device; custom-fitted acrylic trays; adjustable mechanism for jaw advancement. Biocompatibility per ISO 10993-1 (MEM Elution, Intracutaneous Irritation, Guinea Pig Maximization).

Indications for Use

Indicated for adult patients 18 years of age or older as an aid for the reduction of snoring.

Regulatory Classification

Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K112205/S2 Response to Al email 11-01-11 Page 20 of 23 #### Section 6 510(k) Summary #### 510(k) Summarv This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT: | Consumer Health Products, Inc.<br>17 Brownsbury Road #110<br>Laguna Niguel, CA 92677 | |------------------------|-------------------------------------------------------------------------------------------------------------| | CONTACT : | Gary Mocnik<br>49 Coastal Oak, Aliso Viejo, CA 92656<br>949.433.0413<br>949.831.9944 fax<br>gmocnik@cox.net | | DATE PREPARED | July 25, 2011 | | TRADE NAME: | SnoreRx NS 9.0 | | COMMON NAME: | Anti-Snoring Mouth Piece | | CLASSIFICATION NAME: | Anti-Snoring Device, 21 CFR, 872.5570 | | DEVICE CLASSIFICATION: | Class II | | PRODUCT CODE | LRK | | PREDICATE DEVICES: | SnoreGuard (K103004), Silencer (K954530), Snore | Control (K963591), SnoreMaster (K954128) #### Substantially Equivalent To: The Consumer Health Products SnoreRx NS 9.0 is substantially equivalent in intended use, principal of operation and technological characteristics to the SnoreGuard (K103004), the Silencer (K954530), the SnoreControl (K963591), and the SnoreMaster (K954128), as well as other predicate devices cleared with an LRK Product Code. #### Description of the Device Subject to Premarket Notification: The Consumer Health Products SnoreRx NS 9.0 is an intraoral device used at night to reduce snoring by advancing the lower jaw and thereby minimizing air obstruction and turbulence. The device consists of two custom fabricated trays that fit separately over the upper and lower dental arches and engage each other in the anterior area of the mouth. This interface, and thus this device, functions as a mandibular anterior repositioner, which acts to increase the patient's pharyngeal space, improving the ability to exchange air during sleep Consumer Health Products, Inc SnoreRx NS 9.0 Premarket Notification {1}------------------------------------------------ ## Indication for Use: The Consumer Health Products "SnoreRx NS 9.0" is intended for use on adult patients 18 years of age or older as an aid for the reduction of snoring. ## Discussion of Technological Characteristics: The Consumer Health Products SnoreRx NS 9.0 has similar physical and technical characteristics to the predicate devices. The Consumer Health Products SnoreRx NS 9.0 and the identified predicates all provide means for advancing the lower jaw in a predetermined manner. The technical designs and manufacture of the SnoreRx NS 9.0 and the predicate devices are very similar, being composed of custom fitted acrylic trays which fit onto the upper and lower teeth and which are positioned in relation to each other by an adjustable mechanism. ### Non-Clinical Performance Data: .. . . Performance testing was conducted to evaluate and characterize the performance of the Consumer Health Products SnoreRx NS 9.0. Preclinical testing conducted included dimensional conformance evaluation, visual inspections, design verification testing to confirm airway passage equivalency, and biocompatibility testing of device materials based on the applicable elements of ISO 10993-1 shown below. | Test Performed | Standard | Test Result/Conclusion | |------------------------------------------------------------|--------------|--------------------------------------------------| | ISO MEM Elution Assay with L-929<br>Mouse Fibroblast Cells | ISO 10993-5 | Passed.<br>Non-cytotoxic | | ISO Intracutaneous Irritation Test | ISO 10993-10 | Passed.<br>Non-irritant | | Sensitization: Guinea Pig Maximization | ISO 10993-10 | Passed/Negative for evidence of<br>sensitization | Additionally material characterization testing was performed and concluded that the materials used in the construction of the Consumer Health Products SnoreRx NS 9.0 are identical the listed predicate device. ### Clinical Data This submission does not rely on clinical data to determine substantial equivalency to the predicate devices. ### Basis for Determination of Substantial Equivalence: The following table displays the differences and similarities between the new SnoreRx NS 9.0 and other previously marketed devices. Consumer Health Products, Inc SnoreRx NS 9.0 Premarket Notification {2}------------------------------------------------ Section 6 | Products<br>SnoreRx NS<br>9.0 | or older as an aid for the reduction of snoring. | repositioning<br>to increase<br>pharyngeal<br>space | device inserted<br>over the upper<br>and lower dental<br>arches. | |-------------------------------|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------|------------------------------------------------------------------| | SnoreGuard<br>(K103004) | Intended to reduce night time snoring and mild to<br>moderate obstructive sleep apnea (OSA) in adults | SAME | SAME | | Silencer<br>(K954530) | Intended to reduce or eliminate night time snoring<br>in patients 18 years of age or older only. | SAME | SAME | | SnoreControl<br>(K963591) | The anti-snoring device is intended to alleviate or<br>correct snoring | SAME | SAME | | SnoreMaster<br>(9541285) | The anti-snoring device is intended to alleviate or<br>correct snoring | SAME | SAME | #### Conclusions Drawn As shown, the Consumer Health Products SnoreRx NS 9.0 has the following similarities to the predicate devices: - . Same intended use - . Same design characteristics - · Same operating principal- - Same mechanism of action . - . Same technological characteristics Upon reviewing the safety and efficacy information provided in this submission and comparing intended use, principle of operation and overall technological characteristics, the Consumer Health Products SnoreRx NS 9.0 is determined to be substantially equivalent to existing legally marketed devices, performs as well as the predicate devices, and is as safe and effective for its intended use. Premarket Notification {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 NOV 1 6 2011 Consumer Health Products, Incorporated C/O Mr. Gary Mocnik Regulatory Consultants Gary Mocnik and Associates 49 Coastal Oak Aliso Viejo, California 92656 Re: K112205 Trade/Device Name: SnoreRx NS 9.0 Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II Product Code: LRK Dated: November 4, 2011 Received: November 7, 2011 Dear Mr. Mocnik: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however. that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Mocnik Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 1 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. · Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Nh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K112205/S2 Response to Al email 11-01-11 Page 18 of 23 Image /page/5/Picture/1 description: The image shows a handwritten alphanumeric string. The string is "K112205". The characters are written in a cursive style, with some connections between the letters and numbers. The writing appears to be done with a pen or marker. Indications for Use Statement # INDICATIONS FOR USE STATEMENT 510(k) Number (if known): Device Name: SnoreRx NS 9:0 Indications for Use: The Consumer Health Products "SnoreRx NS 9.0" is intended for use on adult patients 18 years of age or older as an aid for the reduction of snoring. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) - - - OR Prescription Use _ X (Per 21 CFR 801.109) . Over-The-Counter Use (Optional Format 1-2-96) Page of Susan Roses (Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: k12205
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...