The Focus Touch Conception System is indicated for assisted insemination in instances where low sperm count, sperm immobility, or hostile vaginal environment has been diagnosed. The system (cervical cap in a condom-like silicone sheath) is used to collect semen into a cervical cap, and then deliver the cap to the outside of the cervix as an aid to conception. It is to be used at home following physician instruction. The Focus Touch Conception System should not be left in place for longer than 6 hours.
Device Story
Focus Touch Conception System assists conception via cervical cap insemination. System includes Conceptacle (cervical cap pre-assembled in silicone condom-like sheath) and plastic applicator. During intercourse, semen is collected in Conceptacle; male removes sheath. User loads cap onto applicator; applicator delivers cap to cervix. Applicator features removal cord mechanism for withdrawal. Device remains in place up to 6 hours. Used at home by couples under physician instruction. Output is physical placement of semen at cervical os. Benefits include aid to conception for patients with low sperm count, sperm immobility, or hostile vaginal environment.
Clinical Evidence
No clinical trials; evidence based on bench testing and human factors/usability studies. Bench testing: biocompatibility (non-cytotoxic, non-irritant, non-sensitizing), Human Sperm Survival Assay (HSSA) (≥80% forward progressive motility after 6 hours), burst pressure, and pull strength. Usability studies (Study 1 and 2): validated user ability to understand instructions, manipulate, insert, and remove device without supervision; confirmed comfort, acceptability, and safety during intercourse.
Technological Characteristics
Silicone cervical cap and condom-like sheath; plastic applicator. Non-sterile, single-use. Mechanical delivery and retention. Biocompatible materials. No energy source. No software.
Indications for Use
Indicated for assisted insemination in patients diagnosed with low sperm count, sperm immobility, or hostile vaginal environment. Used at home per physician instruction. Contraindicated for use longer than 6 hours.
Regulatory Classification
Identification
A cervical cap is a flexible cuplike receptacle that fits over the cervix to collect menstrual flow or to aid artificial insemination. This generic type of device is not for contraceptive use.
{0}------------------------------------------------
# K | | 2200
#### 510(k) Summary
[Refer to 21 CFR 807.92]
SEP 6 2012
| Owner: | Intimate Bridge 2 Conception, Inc<br>2425 Sidney Street,<br>Pittsburgh, PA 15203 | | |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--------------------------|
| Applicant: | Stephen Bollinger<br>Chief Executive Officer<br>412.770.1641 Office | | |
| Application<br>Correspondent: | Niki Caporali Spaniel<br>Senior, Regulatory & Quality Specialist<br>Regulatory & Quality Solutions, LLC<br>Phone: (724) 327-0230 x207 | | |
| Date Prepared: | August 24, 2012 | | |
| Proprietary Name: | Focus TouchTM Conception System | | |
| Common Name: | Conception Assistance Kit | | |
| Classification Name: | Kit, Conception-Assist, Home Use (Product Code: OBB) | | |
| Classification Panel: | Obstetrical/Gynecological | | |
| Regulation: | 21 CFR § 884.5250, Cervical Cap | | |
| Predicate Devices: | Manufacturer<br>Conceivex, Inc | Device Name<br>Conception Kit | 510(k) Number<br>K063227 |
| Reason for Submission | New Device | | |
| 510(k) Number: | K112200 | | |
### Description of Device
The Focus Touch™ Conception System includes detailed Instructions for Use, one Conceptacle and one plastic applicator. The system can be purchased packaged in a quantity of one (1) or three (3) individually packaged systems.
Once the couple has decided to attempt to become pregnant, the Focus Touch™ Conception System Instruction for Use recommends using the device during female's most fertile days in a given month. During sexual intercourse, semen will be collected using the Conceptacle, a cervical cap that is pre-inserted into a condom-like sheath. The male removes the Conceptacle
{1}------------------------------------------------
from the penis by rolling it down and separating it from the penis. After removal, the rim of the cervical cao is lightly pinched so that the Condom-like Sheath can be separated. (The Cervical Cap and Condom-like Sheath are separated in two separate parts). The semen is contained in the cervical cap and its reservoir. The second component, the Applicator, is used for ease of transition of the Cervical Cap to the cervix. The applicator seals the Cervical Cap and contains the semen for transition through the vaginal tract, assisting the female in guiding the cap to the cervix and releasing the semen and the cap in front of the cervical cap is loaded onto the applicator (while it is still outside of the Retainer locks onto the Cervical Cap and is delivered with the Cervical Cap intravaginally. The Retainer also has a removal cord that is attached for withdrawal. The Cervical Cap and Retainer then remain in place for up to 6 hours.
#### Indication For Use
The Focus Touch Conception System is indicated for assisted insemination in instances where low sperm count, sperm immobility, or hostile vaginal environment has been diagnosed. The system (cervical cap in a condom-like silicone sheath) is used to collect semen into a cervical cap, and then deliver the cap to the outside of the cervix as an aid to conception. It is to be used at home following physician instruction. The Focus Touch Conception System should not be left in place for longer than 6 hours.
#### Comparison between predicate and proposed device
The tables below compare the two devices:
| Device | Predicate - Conceivex Inc.<br>Conception Kit | Proposed - Focus Touch<br>Conception System | Conclusio<br>n |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Indication<br>for Use | The Conceivex Conception<br>Kit is indicated for assisted<br>insemination in situations in<br>which low sperm count, sperm<br>immobility, or hostile vaginal<br>environment have been<br>diagnosed. The kit is used for<br>semen collection and<br>placement into the bowl of a<br>cervical cap as an aid to<br>conception. It is to be used at<br>home following physician<br>instruction. The Cap should<br>not be left in place for longer<br>than 6 hours. | The Focus Touch Conception<br>System is indicated for assisted<br>insemination in instances where<br>low sperm count, sperm<br>immobility, or hostile vaginal<br>environment has been<br>diagnosed. The system (cervical<br>cap in a condom-like silicone<br>sheath) is used to collect semen<br>into a cervical cap, and then<br>deliver the cap to the outside of<br>the cervix as an aid to<br>conception. It is to be used at<br>home following physician<br>instruction. The Focus Touch<br>Conception System should not<br>be left in place for longer than 6<br>hours. | Same |
| Usage | Prescription Use. For use at | Prescription Use. For use at | Same |
#### Device Similarities:
{2}------------------------------------------------
| | home following physician<br>instruction. | home following physician<br>instruction. | |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| User Group | Couples who are trying to<br>conceive and have been<br>diagnosed with low sperm<br>count, sperm immobility, or<br>hostile vaginal environment. | Couples who are trying to<br>conceive and have been<br>diagnosed with low sperm count,<br>sperm immobility, or hostile<br>vaginal environment. | Same |
| Material<br>Composition of<br>Cervical Cap | Silicone Cervical Cap | Silicone Cervical Cap | Same |
| Insemination<br>Technology | Cervical Cap Insemination | Cervical Cap Insemination | Same |
| Biocompatibility | Materials tested and shown to<br>be biocompatible. | Materials tested and shown to be<br>biocompatible. | Same |
| Sperm<br>Compatibility | Human Sperm Survival Assay<br>Tested to demonstrate that<br>device materials have no<br>deleterious effects on human<br>sperm. | Conceptacle Human Sperm<br>Survival Assay (HSSA) Tested.<br>To pass the HSSA test, the<br>human sperm must demonstrate<br>forward progressive motility rate<br>of ≥80% of the initial forward<br>progressive motility sperm swim<br>up after 6 hours of sample<br>preparation. | Same |
| Sterility | Nonsterile | Nonsterile | Same |
| Single use | Yes | Yes | Same |
### Device Differences:
・・・ #
ﮮ ﮧ۔
| Device | Predicate - Conceivex Inc.<br>Conception Kit | Proposed - Focus Touch Conception<br>System |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Semen<br>Collection<br>Method | A collection condom is used during<br>intercourse to collect semen. | The Focus Touch™ Conception System<br>does not contain a separate collection<br>condom, rather the collection cervical<br>cap and condom-like sheath are pre-<br>assembled into one unit. |
| Discussion: | | |
| Collection is still performed in the same manner as it is with the predicate device. | | |
| Usability and acceptability testing was successfully performed to validate that the | | |
| use of the <i>Conceptacle</i> for semen collection was comfortable (slip, breakage, | | |
| ability to ejaculate) and adequate (able to collect and contain semen). Therefore, | | |
| the semen collection method utilizing the <i>Conceptacle</i> (cervical cap and condom-<br>like sheath assembly) is substantially equivalent to and the collection method of | | |
| the predicate device. | | |
| Collection<br>Condom<br>Material | Polyurethane | The condom-like sheath component of the<br><i>Conceptacle</i> is made from silicone. |
| Discussion: | | |
| Silicone has been used historically for medical devices used in the vaginal tract. | | |
·
{3}------------------------------------------------
| | The Condom-like Sheath of the Conceptacle is very similar in malleability to<br>standard polyurethane condoms with a durometer of 30-40 on the Shore A scale.<br>The device has been shown to be biocompatible and acceptable as a semen<br>collection device with regard to comfort (slip, breakage, ability to ejaculate) and<br>adequacy (ability to collect and contain semen) in validation testing. Therefore,<br>the material used for the condom-like sheath of the proposed device is<br>substantially equivalent to the material used for the condom for the predicate<br>device. | |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Placement<br>and<br>Delivery<br>Method | The device is placed into the<br>vagina and in front of the cervical<br>os by hand. | The Focus Touch™ Conception System<br>uses an applicator for placement and<br>delivery of the cervical cap to the cervical<br>os. |
| | Discussion: | |
| | The applicator component, for which biocompatibility and usability have been<br>demonstrated during testing, allows the user to insert the cervical cap into the<br>vaginal tract without the use of the user's hand. It is performed in a similar<br>manner as inserting a tampon with a plastic applicator. It has smoothness<br>requirements for comfort and safety and allows for insertion and placement.<br>Therefore, the placement and delivery method of the proposed device is<br>substantially equivalent to that of the predicate device. | |
| Removal<br>Method | The proposed device contains a<br>loop that is molded with the<br>cervical cap. The user reaches<br>into the vagina and grasps the<br>loop to remove the device. | The Focus Touch™ Conception System<br>applicator features a mechanism which<br>applies a removal cord to the cervical cap<br>(during the process of loading the applicator)<br>to aid in removal of the device. |
| | Discussion: | |
| | The removal means of the looped removal device to grip and pull to withdraw is<br>the same for the predicate and proposed device. The difference is that the remove<br>cord on the proposed device remains outside of the vagina and provides for<br>removal of the device without insertion of the hands into the vaginal opening, as<br>was demonstrated during Usability testing. Testing was also performed to verify<br>that the pull strength of the collet/string attached to the cervical cap is sufficient<br>to ensure that the cap will not be separated from the collet. Therefore the<br>removal cord method of the proposed device is substantially equivalent to that of<br>the predicate device. | |
#### Testing Summary
.
Biocompatibility Testing performed on the device demonstrated that like the predicate device, the Focus Touch Conception System is considered non-cytotoxic, is not an irritant and did not elicit a sensitization response. Human Sperm Survival Assay (HSSA) testing was performed on the silicone Conceptacle (Cervical cap inserted into condom-like sheath) to demonstrate that like the predicate device, the Focus Touch collection device and cervical cap, Conceptacle, material has no deleterious effects on human semen. Bench testing was performed to verify that the condom-like sheath of the Focus Touch Conceptacle exceeded the minimum requirements for
{4}------------------------------------------------
burst pressure and volume for commercially available condoms. Bench testing demonstrated that the pull strength of the withdrawal cord, specifically Cervical Cap/Collet Separation Force for the Focus Touch System, was met for all systems tested. A shelf life study was performed to verify that all components of the Focus Touch System were shown to function procently and according to their design intent following a minimum real-time storage during of and or accelerated aging period equivalent to six (6) months. Clinical human factors/usability testing was performed to validate the usability of the device. Specifically, Study 1 testing was performed with the cervical cap and applicator to demonstrate that the end user can understand the Instructions for Use, and manipulate, insert and remove the cap cherectly without direct supervision. Study 2 testing was performed to demonstrate appropriate ease of yoe, confort and acceptability and safety when used during sexual intercourse as a collection device. During Study 1, a physical examination occurred to verify safety of the device to the vaginal tract and all success criteria were met.
#### Conclusion
The differences between the predicate device and the Focus Touch™ Conception System have been evaluated through testing to show that they are not critical to the intended therapeutic effect of the device, nor do they affect the safety or effectiveness of the device when used as labeled. Therefore, the Focus Touch Conception System is substantially equivalent to the Conceivex Conception Kit
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Intimate Bridge 2 Conception, Inc. % Ms. Niki Caporali Spaniel Senior Regulatory & Quality Specialist Regulatory and Quality Solutions, Inc. 3919 William Penn Highway, Suite 200 MURRYSVILLE PA 15668
Re: K112200
Trade/Device Name: Focus Touch™ Conception System Regulation Number: 21 CFR\$ 884.5250 Regulation Name: Cervical cap Regulatory Class: II Product Code: OBB Dated: April 18, 2012 Received: August 29, 2012
Dear Ms. Caporali Spaniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that 1 Drival statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
SEP 6 2012
{6}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set de rice-relation adverse or ones (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you uchte spoolite active to: your le ntersOffices/CDRH/CDRHOffices/ucm115809.htm for go to mtp.//www.tdai.gov/rtoodable/s/cal Health's (CDRH's) Office of Compliance. Also, please
the Center for Devices and Radiological Health's (CDRH's) Office of Compliance the Concertion be recording by reference to premarket notification" (21CFR Part note the regulation online and the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R. Ecker
Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
# 510(k) Number (if known): FBB K HZ2 DO
# Device Name: Device Name: Focus Touch™ Conception System
## Indications for Use:
The Focus Touch Conception System is indicated for assisted insemination in instances The Focus I ouch Conception by stom in mobility, or hostile vaginal environment has been where low spenn (cervical cap in a condom-like silicone sheath) is used to collect semen into a cervical cap, and then deliver the cap to the outside of the cervisa. The conception. It is to be used at home following physician instruction. The and all to conception. The is to be assume as work in place for longer than 6 hours.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Norm
Division Sign-Off vision of Reproductive, Gastro-Renal, and frological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.