FEMTO LDV

K112154 · Sie Ag,Surgical Instument Engineering · HQF · Mar 16, 2012 · Ophthalmic

Device Facts

Record IDK112154
Device NameFEMTO LDV
ApplicantSie Ag,Surgical Instument Engineering
Product CodeHQF · Ophthalmic
Decision DateMar 16, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4390
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The FEMTO LDV Z2 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery or other treatment requiring lamellar resection of the cornea at constant depth relative to the corneal surface. The FEMTO LDV Z4 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface that does not enclose a volume of the cornea. The FEMTO LDV Z6 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants, lamellar keratoplasty, penetrating keratoplasty or other treatment requiring lamellar resection of the cornea at a varying depth with respect the corneal surface that may enclose a volume of the cornea.

Device Story

Femto LDV Z-Generation (Z2, Z4, Z6) are ophthalmic surgical lasers; utilize femtosecond laser technology for corneal tissue resection. Devices create precise corneal incisions, tunnels, or pockets for refractive surgery and keratoplasty. Operated by ophthalmologists in clinical settings. Z2 performs constant-depth lamellar resection; Z4 performs varying-depth resection (non-enclosed); Z6 performs varying-depth resection (including enclosed volumes). Laser energy facilitates photodisruption of corneal tissue; enables surgeon-controlled, computer-guided incisions. Benefits include high-precision tissue separation, improved surgical outcomes, and reduced manual instrumentation. Output is physical corneal incision/resection; surgeon visualizes procedure via integrated microscope/display.

Clinical Evidence

No clinical data provided in the document; substantial equivalence based on technological characteristics and intended use.

Technological Characteristics

Femtosecond surgical laser system; ophthalmic application. Models Z2, Z4, Z6 differ in depth control and volume resection capabilities. Laser-tissue interaction via photodisruption. System includes surgical laser, delivery optics, and control interface.

Indications for Use

Indicated for patients undergoing LASIK surgery, corneal ring tunnel creation, corneal implant pocket creation, lamellar keratoplasty, or penetrating keratoplasty requiring lamellar resection of the cornea.

Regulatory Classification

Identification

An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.

Submission Summary (Full Text)

{0}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes extending upwards and to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird-like figure. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SIE AG, Surgical Instrument Engineering c/o Kevin Walls, RAC Regulatory Insight, Inc. 5401 S. Cottonwood Ct. Greenwood Village, CO 80121 MAR 1 5 2012 . #### Re: K112154 Trade/Device Name: Femto LDVTM Z-Generation Femtosecond Surgical Laser models FEMTO LDV Z2, FEMTO LDV Z4 and FEMTO LDV Z6 Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: 2 Product Code: HQF Dated: February 25, 2012 Received: February 27, 2012 Dear Mr. Walls: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Page 2 - Mr. Kevin Walls If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Image /page/1/Picture/5 description: The image shows a signature in black ink on a white background. The signature appears to be cursive and stylized, with a prominent loop extending downwards. The letters are connected, and the overall impression is that of a quick, practiced autograph. Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use # 510(k) Number (if known): K112154 Device Name: FEMTO LDV Z-Generation Femtosecond Surgical Laser (FEMTO LDV Z2) Indications for Use: The FEMTO LDV Z2 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery or other treatment requiring lamellar resection of the cornea at constant depth relative to the corneal surface. × Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Aesa (Division Sign-Off) (Division of Ophthalinity, Neurological and Ear, Nose and Throut Devicess AND/OR 510(k) Number K112154 Page 1 of 3 {3}------------------------------------------------ #### Indications for Use ### 510(k) Number (if known): K112154 Device Name: FEMTO LDV Z-Generation Femtosecond Surgical Laser (FEMTO LDV ನ್ನೂ ನಾರ್ #### Indications for Use: The FEMTO LDV Z4 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants or other treatment requiring lamellar resection of the cornea at a varying depth with respect to the corneal surface that does not enclose a volume of the cornea. Prescription Use _ X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) Aer (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K112154 Page 2 of 3 {4}------------------------------------------------ ### Indications for Use ## 510(k) Number (if known): K112154 Device Name: FEMTO LDV Z-Generation Femtosecond Surgical Laser (FEMTO LDV 26) #### Indications for Use: The FEMTO LDV Z6 Femtosecond Surgical Laser is an ophthalmic surgical laser indicated for use in the creation of corneal incisions in patients undergoing LASIK surgery, tunnel creation for implantation of rings, pocket creation for implantation of corneal implants, lamellar keratoplasty, penetrating keratoplasty or other treatment requiring lamellar resection of the cornea at a varying depth with respect the corneal surface that may enclose a volume of the cornea. Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) Fesm (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K112154 Page 3 of 3
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...