The LITH method is an in vitro diagnostic test for the quantitative measurement of lithium in human serum and plasma on the Dimension Vista® System. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder). The Drug 4 Calibrator is an in vitro diagnostic product for the calibration of the LOCI Digoxin (DIGXN) and Lithium (LITH) methods on the Dimension Vista® System.
Device Story
The Dimension Vista Lithium (LITH) assay is an in vitro diagnostic test for quantitative lithium measurement in human serum and plasma. The device uses a lithium-specific chromoionophore that forms a complex with Li+ ions in an alkaline solution. The reaction is measured via a bichromatic (510 and 700 nm) endpoint technique on the Dimension Vista System. The concentration of lithium is proportional to the increase in absorbance resulting from the dye-lithium complex. The system is intended for use in clinical laboratory settings by trained personnel. Results assist clinicians in adjusting lithium dosage to maintain therapeutic levels (0.60–1.20 mmol/L) and avoid toxicity. The system includes a liquid, ready-to-use calibrator (Drug 4 Calibrator) derived from human serum, traceable to NIST SRM 924.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP5-A2 over 20 days (CVs 1.1–4.9%). Linearity established per CLSI EP6-A (0.20–3.00 mmol/L). LoB/LoD/LoQ determined per CLSI EP17-A (LoQ 0.158 mmol/L). Method comparison with predicate (N=144) showed Passing-Bablock slope 0.957, r=0.997. Interference testing per CLSI EP7-A2 confirmed no significant bias from common endogenous/exogenous substances at therapeutic levels, though hemolyzed specimens are contraindicated.
Technological Characteristics
Quantitative colorimetric assay using a lithium-specific chromoionophore. Reagents: dye, diethylene glycol monoethyl ether, KOH, surfactant, preservative. Bichromatic endpoint detection (510/700 nm). Calibrator: liquid human serum matrix, traceable to NIST SRM 924. System: Dimension Vista. Measuring range: 0.20–3.00 mmol/L. Sample volume: 2 uL.
Indications for Use
Indicated for quantitative measurement of lithium in human serum and plasma to monitor drug dosage in patients with mental disturbances, including bipolar disorder. Drug 4 Calibrator is indicated for calibration of LOCI Digoxin and Lithium methods on the Dimension Vista System.
Regulatory Classification
Identification
A lithium test system is a device intended to measure lithium (from the drug lithium carbonate) in serum or plasma. Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k112142
B. Purpose for Submission:
New device
C. Measurand:
Lithium
D. Type of Test:
Quantitative colorimetry
E. Applicant:
Siemens Healthcare Diagnostics
F. Proprietary and Established Names:
Dimension Vista® Lithium (LITH) Flex® Reagent Cartridge
Dimension Vista® Drug 4 Calibrator
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NDW | II | 21 CFR §862.3560 Lithium test system | Toxicology (81) |
| JIX | II | 21 CFR §862.1150 Calibrator | Chemistry (75) |
H. Intended Use:
1. Intended use(s):
See Indications for use below.
2. Indication(s) for use:
The LITH method is an in vitro diagnostic test for the quantitative measurement of lithium in human serum and plasma on the Dimension Vista® System.
Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder).
The Drug 4 Calibrator is an in vitro diagnostic product for the calibration of the LOCI Digoxin (DIGXN) and Lithium (LITH) methods on the Dimension Vista® System.
{1}
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
Dimension Vista System. The 510(k) demonstrates assay performance on the Dimension Vista 1500.
I. Device Description:
Reagent components of the LITH method are shown in the table below:
| Reagents wells | Form | Ingredients |
| --- | --- | --- |
| 1-6 | Liquid | Dye, Diethylene glycol monoethyl ether
2(3)-tert-Butyl-4-hydroxyanisole, Surfactant Preservative |
| 7-12 | Liquid | KOH, Diethylene glycol monoethyl ether, Surfactant |
The Drug 4 Calibrator is a five level liquid calibrator. The matrix is drug free normal human serum. Targets for the lithium levels are 0.0, 0.9, 1.8, 3.6, and 5.5 mmol/L. The calibrator was previously cleared for Digoxin (k093441).
Each donor unit used in the preparation of this product was tested by FDA-approved methods for the presence of antibodies to Human Immunodeficiency Virus Type 1 (HIV-1) and Type 2 (HIV-2), as well as for Hepatitis B Surface Antigen (HBsAg) and antibody to Hepatitis C Virus (HCV), and found to be negative (not repeatedly reactive).
J. Substantial Equivalence Information:
1. Predicate device name(s):
Dimension® Lithium (LI) Flex® Reagent Cartridge
Dimension® Drug Calibrator
2. Predicate K number(s):
k011033, k011035
{2}
3. Comparison with predicate:
Assay:
| Similarities | | |
| --- | --- | --- |
| Item | Proposed Device | Predicate Device (k011033) |
| Intended use | Is an in vitro diagnostic test for the quantitative measurement of lithium in human serum and plasma. | Same |
| Indications for use | Measurements of lithium are used to assure that the proper drug dosage is administered in the treatment of patients with mental disturbances, such as manic-depressive illness (bipolar disorder). | Same |
| Sample Types | Serum and Sodium Heparin Plasma | Same |
| Differences | | |
| --- | --- | --- |
| Item | Proposed Device | Predicate Device (k011033) |
| Measuring Range | 0.20 – 3.00 mmol/L | 0.20 - 5.00 mmol/L |
| Sample size | 2 uL | 10 uL |
| Bichromatic endpoints | 510 and 700 nm | 540 and 700 nm |
Calibrator:
| Similarities | | |
| --- | --- | --- |
| Item | Proposed Device | Predicate Device (k011035) |
| Intended use/ Indications for use | An in vitro diagnostic product for the calibration of the Lithium (LITH) and other methods on the Dimension Vista® System. | Same |
| Matrix | Human serum | Same |
| Preparation | Liquid; ready to use | Same |
| Storage | 2 - 8°C | Same |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Proposed Device | Predicate Device (k011035) |
| Analytes | Lithium and Digoxin | Lithium, Digoxin, Phenobarbital, Phenytoin, and Theophylline |
K. Standard/Guidance Document Referenced (if applicable):
- CLSI Guideline EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods
- CLSI Guideline EP 17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation
- CLSI Guideline EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach
- CLSI Guideline EP9-A2: Method Comparison and Bias Estimation Using Patient Samples
- CLSI Protocol EP7-A2: Interference Testing in Clinical Chemistry
L. Test Principle:
The LITH method employs a lithium-specific chromoionophore that forms a complex with the $\mathrm{Li^{+}}$ ion in an alkaline solution.
$$
\mathrm{Li}^{+} + \text{Lithium Dye} \quad \rightarrow \quad \text{Dye-Li Complex (absorbs at 510 nm)}
$$
The concentration of lithium in the sample is proportional to the increase in absorbance, which is due to the formation of the dye-lithium complex. The reaction is measured using a bichromatic (510 and 700 nm) endpoint technique.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision testing for the LITH method was performed over twenty days according to CLSI EP5-A2. Test samples consisted of two levels of a commercially available control, two serum pools and one plasma (Na+ heparin) pool. The two serum pools and one plasma pool were spiked with lithium carbonate. On each day of testing, each sample was run in duplicate, in two separate runs. Testing was performed over 20 days, two separate runs
{4}
with two test samples for each test material. The studies included one Flex lot, one instrument and one calibrator lot.
The range of samples spans the analytical measurement range. Analysis of variance (ANOVA) was used to evaluate the data consistent with recommendations of EP5-A2.
| Sample | Mean (mmol /L) | Repeatability | | Within-lab | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | SD | % CV | SD | % CV | |
| Controls | Level 1 | 0.86 | 0.01 | 1.7 | 0.02 | 2.5 |
| | Level 2 | 1.67 | 0.02 | 1.1 | 0.03 | 1.6 |
| Serum pool | Level 1 | 0.67 | 0.01 | 1.9 | 0.03 | 4.9 |
| | Level 2 | 2.48 | 0.03 | 1.2 | 0.07 | 2.7 |
| Plasma pool | Level 1 | 1.24 | 0.02 | 1.4 | 0.03 | 2.7 |
# b. Linearity/assay reportable range:
The linear range was determined according to CLSI EP6-A. Based on the results of this study and that from the Limit of Detection study, the analytical measuring range was established. The LITH measuring range is 0.20 - 3.00 mmol/L.
The linearity study covered the range of 0.00 to $3.16\mathrm{mmol / L}$ for serum and 0.02 to $3.22\mathrm{mmol / L}$ for plasma $(\mathrm{Na + }$ heparin). Linearity testing was performed using human serum and plasma pools spiked with lithium carbonate and then diluted for intermediate levels by sequential mixing with drug-free serum and/or plasma. For each sample type, nine samples were prepared by sequential mixing and analyzed in $\mathbf{N} = 5$ replicates.
The expected values listed in the tables below are based on dilutions of a standard solution of known concentration.
The predicted values from the most statistically significant polynomial regressions were compared to the predicted values from the linear regressions and the bias was calculated for each level.
{5}
Plasma:
| Sample | Expected value | Observed value | Predicted linear | Predicted Quadratic | Bias | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | mmol/L | % |
| L1 | 0.00 | 0.02 | 0.02 | 0.01 | -0.01 | n/a |
| L2 | 0.41 | 0.41 | 0.43 | 0.42 | 0.00 | -0.6 |
| L3 | 0.82 | 0.83 | 0.83 | 0.83 | 0.00 | 0.4 |
| L4 | 1.23 | 1.24 | 1.23 | 1.24 | 0.01 | 0.5 |
| L5 | 1.64 | 1.65 | 1.63 | 1.64 | 0.01 | 0.5 |
| L6 | 2.04 | 2.03 | 2.02 | 2.03 | 0.01 | 0.3 |
| L7 | 2.45 | 2.43 | 2.43 | 2.43 | 0.00 | 0.1 |
| L8 | 2.86 | 2.82 | 2.83 | 2.82 | 0.00 | -0.1 |
| L9 | 3.27 | 3.22 | 3.23 | 3.22 | -0.01 | -0.3 |
Linear regression analysis:
$$
y = 0.9795x + 0.0257
$$
$$
R^2 = 0.9999
$$
Serum:
| Sample | Expected value | Observed value | Predicted linear | Predicted Cubic | Bias | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | | mmol/L | % |
| L1 | 0.00 | -0.03 | 0.00 | -0.03 | -0.02 | n/a |
| L2 | 0.40 | 0.39 | 0.40 | 0.40 | 0.00 | 0.0 |
| L3 | 0.80 | 0.81 | 0.79 | 0.81 | 0.01 | 1.4 |
| L4 | 1.19 | 1.19 | 1.18 | 1.20 | 0.01 | 1.2 |
| L5 | 1.59 | 1.61 | 1.58 | 1.59 | 0.01 | 0.7 |
| L6 | 1.99 | 1.98 | 1.98 | 1.99 | 0.00 | 0.2 |
| L7 | 2.39 | 2.38 | 2.38 | 2.38 | 0.00 | -0.1 |
| L8 | 2.78 | 2.79 | 2.77 | 2.76 | -0.01 | -0.3 |
| L9 | 3.18 | 3.16 | 3.17 | 3.16 | -0.01 | -0.2 |
Linear regression analysis:
$$
y = 0.998x - 0.0031
$$
$$
R^2 = 0.9997
$$
Recovery: A recovery study was also performed. Samples were each independently volumetrically prepared using a highly concentrated lithium stock solution, and expected values are calculated based upon the amount of lithium carbonate added to each sample. In addition, a control is prepared for each sample by spiking the diluent used to prepare the lithium carbonate stock solution (in this case, water) into the sample matrix. Recovery is determined relative to control for each sample.
{6}
Serum results:
| Expected (mmol/L) | Observed (mmol/L) | | % Recovery |
| --- | --- | --- | --- |
| | Test | Control | |
| 0.00 | -0.03 | n/a | n/a |
| 0.60 | 0.58 | -0.03 | 101% |
| 1.20 | 1.14 | -0.05 | 99% |
| 1.80 | 1.75 | -0.04 | 99% |
| 2.40 | 2.34 | -0.03 | 99% |
| 3.00 | 2.93 | -0.04 | 99% |
Plasma results:
| Expected (mmol/L) | Observed (mmol/L) | | % Recovery |
| --- | --- | --- | --- |
| | Test | Control | |
| 0.00 | 0.02 | n/a | n/a |
| 0.60 | 0.66 | 0.03 | 105% |
| 1.20 | 1.20 | 0.02 | 98% |
| 1.80 | 1.78 | 0.02 | 98% |
| 2.40 | 2.36 | 0.02 | 97% |
| 3.00 | 2.96 | 0.02 | 98% |
Out of range specimens (with results in excess of 3.00 mmol/L) are to be manually diluted. Autodilution is not recommended.
Samples with results less than 0.20 mmol/L will be reported as "less than 0.20 mmol/L" by the instrument.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: The calibrator is traceable to NIST SRM1 924. It is a liquid human serum base product containing lithium packaged as two vials for each of five levels. Level A is drug free normal human serum and Levels B - E contain lithium carbonate material in human serum.
Value assignment: The four levels of Lithium master pool are manufactured by weighing in the appropriate amount of NIST SRM 924 Lithium reference material into drug free normal human serum. (Level 1 does not contain analyte and has a concentration of 0.0 mmol/L.) Each Masterpool level is tested on Dimension for content verification against a previous masterpool and current calibrator lot. The estimated uncertainties in the assigned values of calibrator material range from 0.4 to 0.7%.
Commercial calibrator lots are produced by adding lithium gravimetrically to drug free normal human serum. Values are assigned to each lot of calibrator from the Lithium master pool using multiple instruments and reagent lots.
{7}
Stability: The calibrators are stable until the expiration date printed on the vial when stored unopened and opened at 2-8 °C.
Real time stability studies are ongoing for unopened, opened and punctured calibrators. The sponsor also evaluated three freeze-thaws. Stability testing protocols were presented in the 510(k) and sponsor's acceptance criteria were reviewed and found to be acceptable.
d. Detection limit:
The sponsor followed CLSI Guideline EP 17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation to determine the Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantitation (LoQ).
Five lithium free samples (five independent lots of Calibrator Level A (zero level)) and five low level samples (serum samples spiked with lithium carbonate) were tested for three days, one run per day, two replicates per run, with two reagent lots, on two instruments. Calibration of both instruments was performed using the same lot of calibrator and all testing was performed by the same operator.
The Limit of Blank (LoB) for LITH was determined to be 0.05 mmol/L and the Limit of Detection (LoD) for LITH was determined to be 0.08 mmol/L.
In order to establish the Limit of Quantitation (LoQ) four unique patient samples were gravimetrically prepared to 0.15 mmol/L lithium using NIST SRM 924a Lithium Carbonate. Two instruments were used and two lots of reagents (a different reagent lot per instrument). Each of the four samples were processed in triplicate during each run. Three runs were performed on each instrument system (one run per day). A total of 72 measurements were obtained. The mean observed sample concentration was 0.158 mmol/L, with overall bias (relative to the concentration based on the standard) of 0.008 mmol/L. The %CV at 0.158 mmol/L was 8.5%, and a Total Analytical Error (TAE) of 0.035 mmol/L was determined. Consistent with this, the sponsor has set the lower limit of the claimed measuring range to be 0.20 mmol/L. Samples with results less than 0.20 mmol/L will be reported as "less than 0.20 mmol/L" by the instrument.
e. Analytical specificity:
Interference testing was performed according to CLSI Guideline EP7-A2 to determine the effect of various endogenous and exogenous substances on the LITH assay. For all interferents the percent bias was determined by testing a control sample without the interferent and comparing it to the value obtained from a test sample to which the potential interferent had been added.
Test samples were prepared by spiking the potential interferent into serum spiked with lithium carbonate. Testing was performed at LITH concentrations
8
{8}
of approximately $0.60\mathrm{mmol / L}$ and $1.50\mathrm{mmol / L}$ . The low level of 0.60 mmol/L was chosen based upon its medical relevance as the lower limit of the therapeutic range. The upper level of $1.50\mathrm{mmol / L}$ was chosen based upon CLSI Guideline EP7-A2.
Interference testing was performed with one lot of reagents. Five replicates were tested for each substance.
Significant interference was defined as a bias of $>0.10 \, \text{mmol/L}$ for the low LITH sample and $>10\%$ bias for the high LITH sample relative to the control sample.
Interference was evaluated at several levels of potential interferent for the following compounds:
| Interferent | Amount | Amount (SI units) | Bias at 0.60 mmol/L LITH | Bias at 1.5 mmol/L LITH |
| --- | --- | --- | --- | --- |
| Bilirubin (conjugated) | 20 mg/dL | 342 μmol/L | 0.13 mmol/L | 11% |
| | 15 mg/dL | 256 μmol/L | 0.10 mmol/L | 8% |
| | 10 mg/dL | 171 μmol/L | 0.07 mmol/L | 5% |
| | 5 mg/dL | 86 μmol/L | 0.03 mmol/L | 3% |
| | 2 mg/dL | 34 μmol/L | 0.00 mmol/L | -1% |
| | 1 mg/dL | 17 μmol/L | 0.01 mmol/L | -1% |
| Bilirubin (unconjugated) | 20 mg/dL | 342 μmol/L | 0.18 mmol/L | 14% |
| | 15 mg/dL | 256 μmol/L | 0.14 mmol/L | 10% |
| | 10 mg/dL | 171 μmol/L | 0.10 mmol/L | 6% |
| | 5 mg/dL | 86 μmol/L | 0.04 mmol/L | 2% |
| | 2 mg/dL | 34 μmol/L | 0.02 mmol/L | 4% |
| | 1 mg/dL | 17 μmol/L | -0.01 mmol/L | 0% |
| Hemoglobin | 500 mg/dL | 0.32 mmol/L | 0.21 mmol/L | 18% |
| | 300 mg/dL | 0.19 mmol/L | 0.11 mmol/L | 11% |
| | 200 mg/dL | 0.13 mmol/L | 0.10 mmol/L | 7% |
| | 100 mg/dL | 0.06 mmol/L | 0.07 mmol/L | 4% |
| | 50 mg/dL | 0.03 mmol/L | 0.01 mmol/L | 1% |
| | 25 mg/dL | 0.02 mmol/L | 0.01 mmol/L | 3% |
The package insert indicates the following levels of endogenous compounds do not cause interference (defined as bias less than $0.10\mathrm{mmol / L}$ or $10\%$ at lithium concentrations 0.60 and $1.50\mathrm{mmol / L}$ , respectively):
| Substance tested | Concentration |
| --- | --- |
| Hemoglobin(hemolysate) | 200 mg/dL |
| Bilirubin, unconjugated | 10 mg/dL |
| Bilirubin, conjugated | 15 mg/dL |
| Lipemia (Intralipid) | 3000 mg/dL |
{9}
The sponsor indicates in their package insert that "Hemolyzed specimens should not be used with this assay."
Interference was not detected for the following compounds:
| Substance | Test Concentration | SI Units |
| --- | --- | --- |
| Acetaminophen | 20 mg/dL | 1324 μmol/L |
| Amikacin | 8.0 mg/dL | 137 μmol/L |
| Ampicillin | 5.3 mg/dL | 152 μmol/L |
| Ascorbic Acid | 6.0 mg/dL | 342 μmol/L |
| Caffeine | 6.0 mg/dL | 308 μmol/L |
| Calcium | 25 mg/dL | 6.3 mmol/L |
| Carbamazepine | 3.0 mg/dL | 127 μmol/L |
| Chloramphenicol | 5.0 mg/dL | 155 μmol/L |
| Chlordiazepoxide | 1.0 mg/dL | 33.3 μmol/L |
| Chlorpromazine | 0.2 mg/dL | 6.27 μmol/L |
| Cholesterol | 503 mg/dL | 13 mmol/L |
| Cimetidine | 2.0 mg/dL | 79.2 μmol/L |
| Copper | 1588 μg/dL | 250 μmol/L |
| Creatinine | 30 mg/dL | 2.65 mmol/L |
| Dextran 40 | 6000 mg/dL | 1500 μmol/L |
| Diazepam | 0.51 mg/dL | 18 μmol/L |
| Digoxin | 6.1 ng/mL | 7.8 nmol/L |
| Erythromycin | 6.0 mg/dL | 81.6 μmol/L |
| Ethanol | 400 mg/dL | 86.8 mmol/L |
| Ethosuximide | 25.0 mg/dL | 1770 μmol/L |
| Furosemide | 6.0 mg/dL | 181 μmol/L |
| Gentamicin | 1.0 mg/dL | 21 μmol/L |
| Heparin | 3.0 U/mL | 3000 U/L |
| Ibuprofen | 50.0 mg/dL | 2425 μmol/L |
| Immunoglobulin G (IgG) | 5000 mg/dL | 50 g/L |
| Intralipid | 3000 mg/dL | 30 g/L |
| Iron | 1117 μg/dL | 200 μmol/L |
| Lidocaine | 1.2 mg/dL | 51.2 μmol/L |
| Magnesium | 10.5 mg/dL | 4.3 mmol/L |
| Nicotine | 0.10 mg/dL | 6.2 μmol/L |
| Penicillin G | 25 U/mL | 25,000 U/L |
| Pentobarbital | 8.0 mg/dL | 354 μmol/L |
| Phenobarbital | 10.0 mg/dL | 431 μmol/L |
| Phenytoin | 5.0 mg/dL | 198 μmol/L |
| Potassium | 8.0 mmol/L | 8.0 mmol/L |
| Primidone | 4.0 mg/dL | 183 μmol/L |
| Propoxyphene | 0.16 mg/dL | 4.91 μmol/L |
{10}
| Protein: Albumin | 6000 mg/dL | 60 g/L |
| --- | --- | --- |
| Protein: Total | 12,000 mg/dL | 120 g/L |
| Salicylic Acid | 60 mg/dL | 4.34 mmol/L |
| Sodium | 200 mmol/L | 200 mmol/L |
| Theophylline | 4.0 mg/dL | 222 μmol/L |
| Urea | 500 mg/dL | 83 mmol/L |
| Uric Acid | 20 mg/dL | 1.2 mmol/L |
| Valproic Acid | 50 mg/dL | 3467 μmol/L |
| Vancomycin | 10 mg/dL | 69 μmol/L |
| Zinc | 1625 μg/dL | 250 μmol/L |
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
The predicate device used to demonstrate substantial equivalence to the Dimension Vista LITH assay is the Dimension LI Assay for the Dimension Clinical Chemistry System previously cleared under k011033. The method comparison study included 104 excess, de-identified serum samples and 40 excess de-identified Na⁺ Heparin plasma samples containing measurable amounts of lithium within the analytical measurement range of the DV LITH assay. Testing was performed over 9 days by two operators at one site. One instrument and one lot of calibrators and reagents were used.
Passing Bablock Regression Statistics:
N = 144
Slope = 0.957
95% CI: 0.956, 0.957
Intercept = 0.04 mmol/L
95% CI: 0.04, 0.04
r = 0.997
Sy, x = 0.05 mmol/L
Range = 0.23 - 2.91 mmol/L
b. Matrix comparison:
Matched serum and Na⁺ heparin plasma samples were collected from patients and donors and tested on the Dimension Vista System with the Dimension Vista LITH assay. Both fresh and frozen samples were analyzed.
Spiked samples were prepared by adding aqueous lithium carbonate to freshly drawn matched sets of serum and Na⁺ heparin plasma from healthy donors. Sample spikes were made such that the lithium carbonate stock solution contributed to no more than 2% of the final sample matrix.
{11}
LITH values were determined with the Dimension Vista LITH assay. Samples were run in duplicate and the first replicate was used to compare LITH in plasma versus serum by Passing-Bablock regression analysis.
Comparison of 55 matched undiluted serum and Na+ heparin plasma samples on the Dimension Vista System gave the following statistics using Passing-Bablock regression:
Slope: 1.01
Intercept: -0.02 mmol/L
N: 55
r: 0.997
Range: 0.32 - 2.93 mmol/L
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The therapeutic range of 0.60 - 1.20 mmol/L is obtained from the literature (see below).
The following is included in the package insert:
Therapeutic lithium concentrations vary significantly, depending on the individual. A therapeutic range of 0.60 to 1.20 mmol/L includes effective serum/plasma concentrations for many patients; however, some individuals are best treated at concentrations outside this range. $^{1,2,3}$
Concentrations of 1.20 to 1.50 mmol/L signify a warning range, and a concentration in excess of 1.50 mmol/L in a specimen drawn 12 hours after the dose indicates a significant risk of intoxication. $^{1}$
Concentrations greater than 2.00 mmol/L are often associated with toxic symptoms. $^{1,2,3}$
{12}
Note: The physician must ultimately determine the most appropriate lithium therapeutic range for each patient.
1. Burtis CA, Ashwood ER. Tietz Fundamentals of Clinical Chemistry, Fifth Edition, W.B. Saunders Company, Philadelphia, PA 2001: pp. 1023.
2. Kaplan LA, Pesce AJ. Clinical Chemistry: Theory, analysis, and correlation, Fourth Edition, Mosby, Inc., St. Louis, MO 2003; pp. 1085.
3. Baselt RC, Cravey RH. Disposition of Toxic Drugs and Chemicals in Man, Fourth Edition, Chemical Toxicology Institute, Foster City, CA 1995; pp. 430.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.