K112096 · Computer Programs and Systems, Inc. · LLZ · Sep 30, 2011 · Radiology
Device Facts
Record ID
K112096
Device Name
IMAGELINK
Applicant
Computer Programs and Systems, Inc.
Product Code
LLZ · Radiology
Decision Date
Sep 30, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ImageLink™ is an integrated medical imaging solution that allows for the storage, manipulation, and annotation of high resolution digital radiological images from multiple source modalities. It enhances the efficiency of diagnostic decision making and includes a feature-rich viewer with customizable worklists and hanging protocols that can be organized to meet the unique requirements and style of the individual radiologist.
Device Story
ImageLink is a PACS for hospital IT infrastructure; integrates with HIS/RIS systems. Inputs: digital radiological images (CR, CT, MR, NM, RF, US, DX, PT, XA, OT, MG) via DICOM. Server stores images and manages DICOM Modality Worklists. Viewer allows radiologists to manipulate, annotate, and compare images; supports hanging protocols. Web interface (ChartLink) enables remote viewing (not for primary diagnosis). System enhances diagnostic efficiency by providing quick access to studies across the enterprise. Used in clinical settings by radiologists and clinicians. Output: displayed images and diagnostic data. Benefits: streamlined workflow, centralized image management, and customizable viewing environments.
Clinical Evidence
Bench testing only. Compliance with design and development standards (21 CFR 820.30) and risk analysis (ISO 14971) documented. No clinical data provided.
Technological Characteristics
PACS system utilizing DICOM communications protocols. Supports CR, CT, MR, NM, RF, US, DX, PT, XA, OT, MG modalities. Server-based architecture with web interface. Requires diagnostic workstations with minimum 3-megapixel display capability. Conforms to DICOM, JPEG, and ACR standards.
Indications for Use
Indicated for use by radiologists and clinicians for the storage, manipulation, annotation, and viewing of high-resolution digital radiological images from multiple modalities to assist in diagnostic decision-making. Prescription use only.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
K112096
p. 1 of 3
510(k) Summary
| Section F | 510(k) Summary | | | | SEP 30 2011 |
|------------|----------------------------------------------------|---------------------------------------|--|--|-------------|
| F.1 | Submission Date | | | | |
| 07/19/2011 | | | | | |
| F.2 | Submitter/Owner | | | | |
| | Computer Programs and Systems, Incorporated (CPSI) | | | | |
| F.2.A | Owner / Operator Number | | | | |
| | 905588 | | | | |
| F.2.B | Establishment Registration Number | | | | |
| | 3004120479 | | | | |
| F.2.C | Contact Person | | | | |
| | Michael Hunt, Software Services, CPSI | | | | |
| F.2.D | Address | | | | |
| | 6600 Wall Street<br>Mobile, AL 36695 | | | | |
| F.2.E | Phone | | | | |
| | 1.800.711.2774 | | | | |
| F.2.F | Fax | | | | |
| | 251.639.8214 | | | | |
| F.3 | Device Trade Name | | | | |
| | ImageLink | | | | |
| F.4 | Common Name | | | | |
| | Picture Archiving Communications System (PACS) | | | | |
| F.5 | Classification Names | | | | |
| | Product Code: | LLZ | | | |
| | Device: | System, Image Processing Radiological | | | |
| | Regulation: | 21 CFR 892.2050 | | | |
## F.6 Predicate Device
The legally marketed device to which CPSI claims equivalence is found in 510(k) number K043415 and is called Centricity PACS System, manufactured by GE Medical Systems, Information Technologies.
{2}------------------------------------------------
K112096
P. 2 of 3
510(k) Summary
## F.7 Device Description
ImageLink is designed to fit seamlessly into the hospital's existing IT infrastructure, integrating with the CPSI HIS/RIS systems. Patient and order information is shared with CPSI clinical applications providing real-time update of the PACS worklist and DICOM Modality Worklist files, notification of completed orders and signed transcriptions, and quick access to the study throughout the enterprise.
This product conforms to recognized industry standards such as DICOM, JPEG, ACR, and others as applicable.
Support for the following modalities will be provided:
- · CR
- · CT
- · MR
- · NM
- RF
- US
- DX
- PT
- XA
- · OT
- · MG
The CPSI ImageLink application software includes the following major components: ImageLink Server, ImageLink Viewer, and ImageLink/ChartLink Interface for web viewing. These software applications, combined with the hardware platform and networking infrastructure, make up the ImageLink System.
#### F.8 Intended Use
ImageLink™ is an integrated medical imaging solution that allows for the storage, manipulation, and annotation of high resolution digital radiological images from multiple source modalities. It enhances the efficiency of diagnostic decision making and includes a feature-rich viewer with customizable worklists and hanging protocols that can be organized to meet the unique requirements and style of the individual radiologist.
#### F.9 Technology
The ImageLink Server provides a scalable, storage solution for medical images. Utilizing standard DICOM communications protocols, the server accepts images stored to it from authorized modalities, and also provides these modalities with worklists. The server also provides medical image studies to the ImageLink viewer when requested.
All archived image data is preserved in the format in which is it received by the ImageLink Server. The ImageLink server employs presentation states to store any changes such as annotations, editing, or image manipulations performed on the images. All manipulations and annotations can be reversed in the system to provide the user with the original images unchanged.
The ImageLink Viewer is used to view, analyze, edit, compare, annotate, store, and
{3}------------------------------------------------
K112096
P. 3 of 3
510(k) Summary
distribute images and data stored in the ImageLink server.
The ImageLink/ChartLink web interface allows the viewing of medical images from the archive, over secure web connections through the CPSI ChartLink application. The application can be accessed internally by hospital uses or with authentications. externally over the World Wide Web. The web interface is not intended for primary diagnosis.
Diagnostic workstations running the viewer must meet minimum requirements for safety, performance and effectiveness including display capabilities of no less than 3 mega pixels.
### F.10 Test Summary
The design and development process of ImageLink™ complies with standards listed in section I of this submission. Further ImageLink was developed and modified with change control principles based upon 21 CFR 820.30 as specified in section Q.1. Overall Quality Assurance activities of ImageLink is furthered by adherence to the principles of Risk Analysis in ISO 14971.
## F.11 Summary of Characteristics to that of the Predicate
CPSI ImageLinkTM is substantially equivalent to its predicate device, GE Medical Systems, Information Technologies Centricity PACS System (K043415). Both devices are similar in intended use, device description, materials used, the target population and intended environment for use. Both devices consist of a hardware platform, server application software, image viewing software, and web tools. There are minor differences in terms of technology employed as well as minor functional differences with configurable hotkeys, integration with historical reports and voice clip. These differences do not detract from the equivalence claim.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Michael Hunt Software Analysits, Software Services Computer Programs and Systems, Inc. (CPSI) 6600 Wall Street MOBILE AL 36695
SEP 3 0 2011
Re: K112096 Trade/Device Name: ImageLink™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 19. 2011 Received: July 22, 2011
Dear Mr. Hunt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 1 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Section E Indications for Use Statement
## E.1 Statement
# Indications for Use
510(k) Number (if known):
Device Name: ImageLinkTM
Indications For Use:
ImageLink™ is an integrated medical imaging solution that allows for the storage, manipulation, and annotation of high resolution digital radiological images from multiple source modalities. It enhances the efficiency of diagnostic decision making and includes a feature-rich viewer with customizable worklists and hanging protocols that can be organized to meet the unique requirements and style of the individual radiologist.
× _ AND/OR Over-The-Counter Use_ Prescription Use
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary S. Patel
(Division Sign-Off)
Office of In Vitro Di
510K K112096
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.