T2 FEMORAL NAIL MODEL 1825-XXXXCP, 1828-XXXXCP
K112059 · Stryker · HSB · Aug 8, 2011 · Orthopedic
Device Facts
| Record ID | K112059 |
| Device Name | T2 FEMORAL NAIL MODEL 1825-XXXXCP, 1828-XXXXCP |
| Applicant | Stryker |
| Product Code | HSB · Orthopedic |
| Decision Date | Aug 8, 2011 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3020 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The T2 Femoral Nail System is a fixation device comprised of femoral nails and the related locking screws, compression screws, and end caps intended to provide strong and stable fixation with minimal soft tissue irritation. This device is not as a substitute for normal intact bone and tissue.
Device Story
T2 Femoral Nail is a cylindrical, cannulated titanium alloy fixation rod; bowed to accommodate femoral anatomy. Used for femoral fracture fixation via antegrade or retrograde surgical approach. Available in diameters 8-17mm and lengths 140-480mm. Device provides stable internal fixation for various femoral fractures and nonunions. Operated by orthopedic surgeons in a clinical/OR setting. Output is mechanical stabilization of bone fragments to facilitate healing. Benefits include strong fixation with minimal soft tissue irritation.
Clinical Evidence
Bench testing only. Mechanical testing performed on worst-case predicate and subject devices to evaluate component strength and moment of inertia. Engineering analysis confirmed that additional sizes do not negatively impact device performance.
Technological Characteristics
Cylindrical, cannulated titanium alloy tube; bowed geometry. Diameters: 8-17mm; Lengths: 140-480mm. System includes locking screws, compression screws, and end caps. Mechanical fixation principle.
Indications for Use
Indicated for long bone fracture fixation, specifically femoral fracture fixation, including: open/closed femoral fractures; pseudoarthrosis and correction osteotomy; pathologic fractures, impending pathologic fractures, and tumor resections; supracondylar fractures (including intra-articular extension); ipsilateral femur fractures; fractures proximal to total knee arthroplasty; fractures distal to hip joint; nonunions and malunions.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
- T2 Femoral Nail System (K081152)
- T2 Femoral Nail System (K021744)
- T2 Femoral Nail System (K014220)
- T2 Femoral Nail System (K010801)
Related Devices
- K021744 — T2 FEMORAL NAIL · Howmedica Osteonics Corp. · Jun 26, 2002
- K982601 — OSTEO IC RETROGRADE/ANTEGRADE FEMORAL NAIL · Osteonics Corp. · Sep 14, 1998
- K253566 — Affixus Retrograde Femoral Nailing System · Zimmer, Inc. · Mar 12, 2026
- K021026 — S2 FEMORAL NAIL · Howmedica Osteonics Corp. · Apr 25, 2002
- K201147 — Elos® Intramedullary Nailing System · Intrauma S.P.A · Jul 19, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
AUG - 8 2011
K 112059
## 510(k) Summary of Safety and Effectiveness: T2 Nail
T2 Femoral Nail Proprietary Name: Intramedullary Nail, Femoral Nail Common Name: Intramedullary Fixation Rod, 21 CFR §888.3020 Classification Name and Reference: Class II Proposed Regulatory Class: HSB Product Codes: Valerie Giambanco For Information contact:
Regulatory Affairs Specialist Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 Manwall, NJ 671-6275 Fax: (201) 831-3275
Date Prepared:
7/15/2011
Legally Marketed Device to Which Substantial Equivilance is Claimed:
T2 Femoral Nail System: K081152, K021744, K014220, K010801
### Description:
Description:
The T2 Femoral nail is a cylindrical, cannulated titanium alloy tube, slightly bowed to aconnocated a The T2 Femoral nail is a cylindrial, cambian and the femoral canal using either a retrograde of shape of the femu: The T2 Femoral Nail is currently available in diameters ranging from 8 to 15
antegrade surgical approach. The T2 Femoral Nail is currently available in d mm and lengths ranging from 140 to 480 mm.
Additional sizes will offer 16 and 17 mm diameters.
Intended Use:
The modifications do not alter the intended use of the predicate system as cleared in the referenced The modifications do not alter the nie presents a fixetion device compised of femoral
premarket notifications. The T2 Femoral Nail System is a fraction of Nail System is premarket notifications. The 12 Femoral Nati System is a The T2 Femoral Nail System is
nails and the related locking screws, compression screation with soft tissue irritatio nails and the related locking screws, compression with minimal soft tissue irritation. This intended to provide strong and stable meeting not as a substitute for normal intact bone and tissue.
2
{1}------------------------------------------------
Indications for Use:
The T2 Femoral Nail is indicated for long bone fracture fixation specifically femoral fracture fixation, which may include the following:
X112054
3
- · Open and closed femoral fractures
- · Pseudoarthrosis and correction osteotomy
- · Pseudoarthrosis and correction visioners and tumor resections
Pathologic fractures, impending pathologic fractures, exicular extension
- Painologic fractures, including those with intra-articular extension
Supracondylar fractures, including those with intra-articular extension
Supracondylar fractures, in
- · Ipsilateral femur fractures
- · Fractures proximal to a total knee arthroplasty
- Fractures distal to hip joint
- Nonunions and malunions
### Proposed Modification:
Proposed Modification:
The subject T2 Femoral Nails have the same intended use, material, and operational The subject T2 Femoral Nails have the same multiplikes of 0.0.152, K021744, K014220, and
principles as the previously cleared T2 Femoral Nation Colditions Larger dizes of 16 principles as the previously cleared it Femoral Natis Clear of Career izes of 16 and 17mm. The
K010801. The subject T2 Femoral device consists of the nails with a radius of S K010801. The subject T2 Femoral device consists of administrations with a radius of 3m and
lengths will continue to range from 140-480mm in 20mm increments for the nails with lengths will comm for the nails with a radius of 1.5m.
### Summary of Data:
:
Mechanical testing has been performed on the worst case predicate T2 Femoral Naily equivalent subject devices Mechanical testing has been performed on the more as a beantialy equivalent subject devices
demonstrates the increased diameter of the T2 femoral naily equivalent subject dev demonstrates the increased diameter of the life in the subs... This, the enginesing andysis of
based on component strength, moment of increases arber the safety and based on component strength, moment of the the additional sizes do not affec the safety and the new 16 and 17mm diameter sizes demonstral include anatomical fit of the new diameter. These risks
effectiveness of the device. Potential risks andres fFMFA and mechanical effectiveness of the device. Potential issues android (RMEA) and mechanical testing which have been investigated with use of failure mode cries and your was your affect device performance.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a series of flowing lines, possibly representing health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Howmedica Osteonics Corp. % Ms. Valerie Giambanco 325 Corporate Drive Mahwah, New Jersey 07430
AUG - 8 2011
Re: K112059
Trade/Device Name: T2 Femoral Nail Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Code: HSB Dated: July 15, 2011 Received: July 19, 2011
Dear Ms. Giambanco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 – Ms. Valerie Giambanco
device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours
e
Mark N. Melkerson 1 Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Special 510(k)
Indications for Use
# 510(k) Number (if known): K 11205 9
Device Name: T2 Femoral Nail
Indications for Use:
The T2 Femoral Nail is indicated for long bone fracture fixation specifically femoral fracture fixation, which may include the following:
- Open and closed femoral fractures
- Pseudoarthrosis and correction osteotomy
- Pathologic fractures, impending pathologic fractures, and tumor resections
- Supracondylar fractures, including those with intra-articular extension
- · Ipsilateral femur fractures
- · Fractures proximal to a total knee arthroplasty
- Fractures distal to hip joint
- Nonunions and malunions
| Prescription Use | X |
|-----------------------------|--------|
| (Part 21 CFR 801 Subpart D) | |
| | AND/OR |
| Over-The-Counter Use | |
| (21 CFR 807 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. Melleer
Page 1 of 1
(Division Sign-Of (Division Sign-Or), Orthopedic, Division of Restorative Devices
510(k) Number K112059