← Product Code [HIF](/productcode/HIF) · K111955

# MEDLINE PNEUMOPERITONEUM NEEDLE (K111955)

_Medline Industries, Inc. · HIF · Nov 16, 2011 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K111955

## Device Facts

- **Applicant:** Medline Industries, Inc.
- **Product Code:** [HIF](/productcode/HIF.md)
- **Decision Date:** Nov 16, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1730
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Medline Pneumoperitoneum Needle has applications in gynecologic and abdominal endoscopic procedures for establishment of a pneumoperitoneum.

## Device Story

Medline Pneumoperitoneum Needle is a manual surgical instrument used during gynecologic and abdominal endoscopic procedures. Device facilitates establishment of pneumoperitoneum by providing access to the abdominal cavity for insufflation. Operated by physicians in clinical or surgical settings. Device design and function are consistent with standard pneumoperitoneum needles used to create space for laparoscopic visualization and instrumentation.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Manual surgical needle for insufflation. Design and materials are consistent with standard laparoscopic pneumoperitoneum needles. No electronic components, software, or energy sources.

## Regulatory Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Mr. Matt Clausen Regulatory Affairs Specialist Medline Industries, Inc. One Medline Place Mundelein, IL 60060

JUL 2 7 2015

Re: K11955

> Trade/Device Name: Medline Pneumoperitoneum Needle Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic insufflator Regulatory Class: II Product Code: HIF, FHO Dated (Date on orig SE ltr): October 27, 2011 Received (Date on orig SE ltr): November 2, 2011

Dear Mr. Clausen.

This letter corrects our substantially equivalent letter of November 16, 2011.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{1}------------------------------------------------

Page 2 -

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known):

K III 955

Device Name:

Medline Pneumoperitoneum Needle

Indications For Use:

The Medline Pneumoperitoneum Needle has applications in gynecologic and abdominal endoscopic procedures for establishment of a pneumoperitoneum.

| Prescription Use            | X  |
|-----------------------------|----|
| (Part 21 CFR 801 Subpart D) |    |
|                             | OR |
| Over-the-Counter Use        |    |
| (21 CFR 801 Subpart C)      |    |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of GDRH, Office of Device Evaluation (ODE)

۳۰

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K111955

---

**Source:** [https://fda.innolitics.com/device/K111955](https://fda.innolitics.com/device/K111955)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
