K111924 · Calibra Medical, Inc. · OPP · Apr 10, 2012 · General Hospital
Device Facts
Record ID
K111924
Device Name
FINESSE PERSONAL INSULIN DELIVERY PATCH
Applicant
Calibra Medical, Inc.
Product Code
OPP · General Hospital
Decision Date
Apr 10, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Finesse Insulin Delivery System is indicated for the subcutaneous, bolus delivery of physician prescribed insulins, in adult persons requiring such medications for the management of diabetes mellitus.
Device Story
Finesse Personal Insulin Delivery Patch is a manual, non-electrically powered, disposable, ambulatory bolus dosing pump. System includes: insulin delivery device (IDD), fill syringe/needle, reusable inserter, and dose count card. Patient fills IDD with insulin (max 2ml) using syringe; inserter places IDD on skin and implants cannula into subcutaneous tissue. Device adheres to skin for up to 72 hours. User delivers discrete insulin doses by manually pressing buttons on IDD, which actuate a positive displacement piston. Patient tracks dosing via dose count card. Used by patients, caregivers, or clinicians. Benefits include precise, manual bolus delivery of insulin without external power source.
Clinical Evidence
No clinical performance data required. Evidence consists of bench testing, including chemical, physical, and microbiological stability studies, biocompatibility testing, and mechanical reliability verification. Human factors and simulated use studies were conducted to validate usability and device performance.
Technological Characteristics
Manual, non-electrically powered, positive displacement piston pump. Materials: biocompatible plastics, elastomers, stainless steel. Single lumen catheter/cannula. Wear duration: 72 hours. Reservoir capacity: 2ml. Form factor: wearable patch. No software or electronic components.
Indications for Use
Indicated for adult patients with diabetes mellitus requiring subcutaneous bolus insulin delivery.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Calibra Medical Finesse Insulin Delivery System (K093065)
Calibra Medical Finesse Insulin Delivery System (K100947)
Submission Summary (Full Text)
{0}------------------------------------------------
K111924
# 510(k) Summary of Safety and Effectiveness
APR 1 0 2012
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807,92(c),
### Owner's Name, Address, Telephone Number, Contact 1. Person
### Name, Address. Telephone Number
Calibra Medical. Inc. 220 Saginaw Drive Redwood City, CA 94063-4725
## Contact Person
Richard J. Meader Vice President Regulatory and Quality Affairs Calibra Medical, Inc. 220 Saginaw Drive Redwood City, CA 94063-4725 Direct: 1.650.298.4740 Fax: 1.650.587.8994 rmeader@calibra.com
## Date Prepared
- . July 5, 2011
#### 2. Trade Name
Finesse Personal Insulin Delivery Patch
#### 3. Common Name
Disposable Insulin Infusion Pump
#### 4. Classification Name
Pump, infusion, insulin bolus
{1}------------------------------------------------
#### 5. Classification
| Code of Federal Regulations Number | 880.5725 |
|------------------------------------|----------------------------------|
| Classification Panel | General Hospital |
| Product Code | OPP (primary)<br>LZG (secondary) |
| Regulatory Class | Class II |
| Review Category | Tier 2 |
#### 6. Identification of the Predicate or Legally Marketed Device
Calibra Medical, Inc. believes that the System described in this Submission is substantially equivalent to the Calibra Medical Finesse Insulin Delivery System cleared under 510(k)s K093065 and K100947.
#### 7. Device Description
The Finesse Insulin Delivery System (IDS) is comprised of a sterile. nonpyrogenic, single-use, external, disposable, ambulatory, liquid medication, bolus dosing device (IDD); a single use, non-pyrogenic, sterile. syringe and needle (Fill Syringe); a reusable Inserter; and a Dose Count Card. The device is intended for subcutaneous delivery of clinicianprescribed medications, and is adhered to the skin for up to 72 hours with a biocompatible adhesive.
The Insulin Delivery Device (IDD) is a manual, user filled, positive displacement, bolus dosing pump. The Inserter is used to place the IDD on the skin and simultaneously implant the cannula into the subcutaneous tissues. The syringe and needle are for patient filling of the IDD with liquid medication prior to deployment on the body, and have a maximum capacity of 2ml. The Dose Count Card is utilized by the patient during the dosing session and provides for a written record of date, time and amount of Insulin delivery.
The Finesse IDS materials are biocompatible plastics, elastomers, and stainless steel.
{2}------------------------------------------------
#### 8. Intended Use
The Finesse Insulin Delivery System is indicated for the subcutaneous, bolus delivery of physician prescribed insulins, in adult persons requiring such medications for the management of diabetes mellitus.
#### 9. Technological Characteristics
The Finesse Insulin Delivery System meets the description of a pump, infusion, insulin bolus as established in product code 80-OPP. The system is identical to other insulin delivery devices in that it uses a positive volume displacement type of manual piston to precisely deliver discrete doses of medication from the internal reservoir. The Finesse has no power source. It is non-electrically powered. The Finesse piston is actuated by the mechanical action of the user's fingers pressing on the buttons. The Finesse has a reservoir to hold the medicinal product (insulin). The Finesse has a single lumen catheter/cannula that delivers the drug to the subcutaneous tissues. The Finesse is used by a health care professional, a patient, or a patient care-giver to deliver the drug (i.e., drug delivery requires competent human interaction).
The Finesse is worn for up to 72 hours and has a flexible cannula placed in the subcutaneous tissues to deliver insulin. This is equivalent to the characteristics of the Calibra Medical Finesse Insulin Delivery System (product code 80-OPP) cleared under 510(k) K093065 and K100947.
#### Non-Clinical Performance Data 10.
Compatibility and stability studies have been completed that demonstrate the chemical, physical, microbiological stability and biocompatibility of insulin in the Finesse Insulin Delivery device for the period of time specified in the labeling. Design verification studies have been completed that demonstrate the wear and mechanical reliability of the device for the intended period of time. Design verification studies have been completed that demonstrate the usability and reliability of the Inserter accessory. Human factors and clinical evaluation studies have been completed that demonstrate device performance and usability.
#### Clinical Performance Data 11.
{3}------------------------------------------------
No clinical performance data is required to validate the intended uses and user needs of the system. Design validation is completed by human factors simulated use and clinical evaluation testing.
#### 12. Conclusions
The Finesse Insulin Delivery System is substantially equivalent in its intended use, performance, safety, effectiveness, underlying scientific and operating principles used and performs as well as or better than the predicate Finesse Insulin Delivery System (K093065 and K100947).
# END OF SUMMARY
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR 1 0 2012
Mr. Richard J. Meader Vice President Regulatory and Quality Affairs Calibra Medical, Inc 220 Saginaw Drive Redwood City, California 94063
Re: K111924
Trade/Device Name: Finesse Personal Insulin Delivery Patch Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: OPP, LZG Dated: April 5, 2012 Received: April 6, 2012
### Dear Mr. Meader:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Mr. Meader
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony V. Adutan
(Inthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 510(k) Page 0030 of 3660
## Indications for Use
510(k) Number (if known): Not known.
Device Name: Finesse Personal Insulin Delivery Patch
The Finesse Insulin Delivery System is intended for subcutaneous, bolus delivery of insulin for the management of diabetes mellitus in adult persons requiring insulin.
Prescription Use
AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Gil. for RZC Apr 11, 2012
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: _ K 111924
Section 2 - Indications for Use
510(k) Page 0030 of 3660
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.