MAGNESIUM ASSAY

K111915 · Sekisui Diagnostics P.E.I., Inc. · JGJ · Dec 2, 2011 · Clinical Chemistry

Device Facts

Record IDK111915
Device NameMAGNESIUM ASSAY
ApplicantSekisui Diagnostics P.E.I., Inc.
Product CodeJGJ · Clinical Chemistry
Decision DateDec 2, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1495
Device ClassClass 1

Intended Use

The Sekisui Magnesium Assay is intended for the quantitative measurement of magnesium in human serum and plasma (Lithium Heparin) on automated clinical chemistry analyzers. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia and hypermagnesemia. This device is intended for professional use and IN VITRO diagnostic use only.

Device Story

The Sekisui Magnesium Assay is a liquid, colorimetric reagent kit for the quantitative measurement of magnesium in human serum and lithium heparin plasma. It is designed for use on automated clinical chemistry analyzers by healthcare professionals in clinical laboratory settings. The assay utilizes xylidyl blue-1, which forms a red complex with magnesium under alkaline conditions. The concentration of magnesium is determined spectrophotometrically by measuring the decrease in absorbance at 660 nm, which is directly proportional to the amount of magnesium-xylidyl blue complex formed. Results are used by clinicians to diagnose and manage magnesium-related metabolic disorders, specifically hypomagnesemia and hypermagnesemia.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (Total CV 3.2-6.1%). Linearity confirmed from 0.3 to 8.0 mg/dL (y=1.01x+0.013). LoQ established at 0.3 mg/dL. Interference testing per CLSI EP7-A2 showed no significant interference (<10%) for hemoglobin (1000 mg/dL), conjugated/unconjugated bilirubin (40 mg/dL), and ascorbic acid (3 mg/dL). Method comparison with predicate (n=100) yielded Deming regression slope 0.981, r=0.98. Matrix comparison (serum vs LiHeparin plasma, n=54) showed r=0.99.

Technological Characteristics

Liquid ready-to-use reagent containing buffer (pH 11.2), xylidyl blue-1, EGTA, and surfactant. Principle: Colorimetric spectrophotometry (660 nm). Measuring range: 0.3–8.0 mg/dL. Designed for automated clinical chemistry analyzers. Complies with CLSI guidelines EP5-A2, EP6-A2, EP7-A2, EP9-A2, EP17-A2, and C28-A.

Indications for Use

Indicated for the quantitative measurement of magnesium in human serum and Lithium Heparin plasma to aid in the diagnosis and treatment of hypomagnesemia and hypermagnesemia in professional clinical laboratory settings.

Regulatory Classification

Identification

A magnesium test system is a device intended to measure magnesium levels in serum and plasma. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypermagnesemia (abnormally high plasma levels of magnesium).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Kll915 DEC - 2 2011 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION ## DECISION SUMMARY # DEVICE ONLY TEMPLATE # A. 510(k) Number: K111915 # B. Purpose for Submission: New Device - C. Measurand: Magnesium #### D. Type of Test: Quantitative, Photometric Method - E. Applicant: Sekisui Diagnostics P.E.I. Inc. #### F. Proprietary and Established Names: Magnesium Assay #### G. Regulatory Information: - 1. Requlation section: 21 CFR 862.1495 - 2. Classification: Class I - 3. Product code: JGJ 4. Panel: Chemistry (75) # H. Intended Use: - 1. Intended use(s): For the IN VITRO quantitative measurement of magnesium in serum and plasma. - 2. Indication(s) for use: The Sekisui Magnesium Assay is for the quantitative determination of magnesium in human serum and plasma (Lithium Heparin) on automated chemistry analyzers. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium). This device is intended for professional use and IN VITRO diagnostic use only. - 3. Special conditions for use statement(s): For prescription use only {1}------------------------------------------------ # 4. Special instrument requirements: Clinical chemistry analyzers (testing was performed on the Roche/Hitachi® 717 Automated Analvzer). ## I. Device Description: The Sekisui Magnesium assay kit consists of the following: Magnesium Reagent: A solution containing buffer (pH 11.2 at 25°C), 0.14 mmol/L xylidy/ blue-1, 0.1 mmol/L EGTÅ, and a surfactant. ## J. Substantial Equivalence Information: - 1. Predicate device name(s): Beckman Coulter Magnesium Reagent 2. Predicate 510(k) number(s): K944407 # 3. Comparison with predicate: (Similarities and Differences between the candidate device and the predicate device). | Parameter | Candidate Device | Predicate Device | |-----------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| | | Sekisui Diagnostics<br>Magnesium Assay | Beckman Coulter<br>Magnesium Reagent OSR6189 | | | | K944407 | | Analyte | Magnesium | Magnesium | | Intended Use | For the IN VITRO quantitative<br>measurement of magnesium in<br>serum and plasma. | System reagent for the quantitative<br>determination of Magnesium in<br>Human serum, plasma and urine on<br>Beckman Coulter AU analyzers. | | Sample Matrix | Serum or LiHeparin Plasma | Serum or Heparinized Plasma or<br>urine | | Reportable Range | 0.3 - 8.0 mg/dL | 0.5 -8.0 mg/dL | | Reference Interval | 1.6-2.6 mg/dL | 1.9-2.7 mg/dL | | Technology /<br>Methodology | Colorimetric endpoint method<br>using xylidyl blue. | Colorimetric endpoint method using<br>xylidyl blue. | | Format | Liquid | Liquid | #### K. Standard/Guidance Document Referenced (if applicable): CLSI EP5-A2, Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline CLSI EP7-A2, Interference Testing in Clinical Chemistry; Approved Guideline CLSI EP9-A2, Method Comparison and Bias Estimation Using Patient samples; Approved Guideline CLSI EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline CLSI C28-A2, How to define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline {2}------------------------------------------------ #### L. Test Principle: Xylidyl blue-1 + Mg** -> Mg-xylidyl blue complex (red) Under the alkaline conditions of the Magnesium reagent, magnesium present in the sample, when mixed with reagent, forms a red complex with the diazonium salt of xylidyl blue. The concentration of magnesium in the sample is directly proportional to the amount of xylidyl blue -Magnesium complex formed and can be measured spectrophotometrically by the decrease in absorbance measured at 660 nm. #### M. Performance Characteristics (if/when applicable): - 1. Analytical performance: - a. Precision/Reproducibility: Precision testing was performed according to CLSI guideline EP5-A2. Samples, including serum pools and control materials, were tested twice per day in duplicate for 20 days. Results are summarized below: Precision table for Sekisui Diagnostics Magnesium on Roche Hitachi 717 automated analyzer: | Concentration | | Total SD | | Total<br>CV % | Concentration | | Within Run SD | | Within<br>Run<br>CV % | |---------------|-------|----------|--------|---------------|---------------|--------|---------------|--------|-----------------------| | mg/dL | mmol/ | mg/dL | mmol/L | | mg/d | mmol/L | mg/d | mmol/L | | | 0.8 | 0.33 | 0.05 | 0.021 | 6.1 | 0.9 | 0.37 | 0.06 | 0.024 | 7.0 | | 2.0 | 0.82 | 0.07 | 0.029 | 3.7 | 2.0 | 0.82 | 0.04 | 0.016 | 2.2 | | 4.7 | 1.93 | 0.15 | 0.062 | 3.2 | 4.9 | 2.01 | 0.08 | 0.033 | 1.6 | #### b. Linearity/assay reportable range: A linearity study was performed in accordance with CLSI protocol EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures, a Statistical Approach, and the Sekisui Diagnostics Magnesium method was demonstrated to be linear across the reportable range from 0.2 mg/dL (0.08 mmol/L) to 8.0 mg/dL (3.29 mmo//L). A high concentration serum pool was diluted with saline to produce 11 concentrations of magnesium. Samples were assayed in replicates of four. The results were used to generate best fit regression lines using a linear equation. Nonlinearity is less than allowable nonlinearity. #### Summary of Linear Fit Slope 1.01 Intercept 0.013 mg/dL N 11 The linearity testing confirmed a reportable range of 0.3 - 8.0 mg/dL (0.12 - 3.29 mmol/L). - c. Detection limit: The determination of the method detection limits was performed according to the CLSI EP17-A Guideline, Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline. Limit of Detection (LoD) is defined as the smallest amount of an analyte that the method can reliably detect to determine presence or absence of the analyte. LoD was {3}------------------------------------------------ determined by measuring 60 replicates of a zero magnesium sample and 60 replicates of the low concentration serum control sample. The results were used following the EP17-A quidelines to calculate the Limit of Blank (LoB). Limit of Detection, and Limit of Quantitaion (LoQ). The Limit of Blank was determined as 0.15 mg/dL magnesium. The Pooled SD for the low values of the samples was determined as 0.06 mg/dL; therefore, the LoD is calculated to be 0.25 mg/L. The LoQ represents the magnesium concentration where the CV is less than or equal to 20%. The LoQ is therefore 0.30 mg/L. - d. Analytical specificity: Interference studies on the effects of endogenous and exogenous substances were designed according to the CLSI EP7-A2 guideline. Potentially cross-reactive or interfering substances that were evaluated included five endogenous substances (unconjugated bilirubin, conjugated bilirubin, hemoglobin, lipemia, and ascorbic acid). Three samples of human serum with different concentrations of magnesium were tested with the potential interference substances and analyzed in quadruplicate on the Roche Hitachi 717 automated analyzer. No significant interference was defined to be within +/- 10% of the corresponding control result for the high magnesium sample. | Concentration of<br>Analyte | | Substance Tested | Concentration of Interferent where<br>Interference is Insignificant | | |-----------------------------|--------|---------------------------|---------------------------------------------------------------------|--------------------------------------------------------| | mg/dL | mmol/L | | | | | 2.1 | 0.86 | | 1000 mg/dL | 155 µmol/L | | 4.2 | 1.73 | Hemoglobin | 1000 mg/dL | 155 µmol/L | | 7.9 | 3.25 | | 1000 mg/dL | 155 µmol/L | | 2.0 | 0.82 | | 40 mg/dL | 684 µmol/L | | 4.3 | 1.77 | Conjugated Bilirubin | 40 mg/dL | 684 µmol/L | | 6.8 | 2.79 | | 40 mg/dL | 684 µmol/L | | 2.0 | 0.82 | | 40 mg/dL | 684 µmol/L | | 4.2 | 1.73 | Unconjugated<br>Bilirubin | 40 mg/dL | 684 µmol/L | | 6.8 | 2.79 | | 40 mg/dL | 684 µmol/L | | 2.3 | 0.95 | | 3000 µg/dL | 170 µmol/L | | 4.4 | 1.81 | Ascorbic Acid | 3000 µg/dL | 170 µmol/L | | 6.8 | 2.79 | | 3000 µg/dL | 170 µmol/L | | 1.9 | 0.78 | | 800 mg/dL | 2400 mg/dL (27.1<br>mmol/L) Simulated<br>Triglycerides | | 4.1 | 1.69 | Intralipid | 1000<br>mg/dL | 3000 mg/dL (33.9<br>mmol/L) Simulated<br>Triglycerides | | 6.9 | 2.84 | | 1000<br>mg/dL | 3000 mg/dL (33.9<br>mmol/L) Simulated<br>Triglycerides | e. Assay cut-off: Not applicable - 2. Comparison studies: - a. Method comparison with predicate device: A method comparison study was performed based on the CLSI EP9-A2 guideline. Serum samples (n=100) were obtained for the study, of which 100 were determined to be within the reportable range of the candidate method and the Predicate method (sample recoveries ranged from 0.5 mg/dL to 7.3 mg/dL magnesium. The Beckman Coulter method on the AU640 was used to compare the results performed using the Sekisui Diagnostics Magnesium Assay on the Roche Hitachi 717 automated analyzer. The results demonstrate the Sekisui Diagnostics Magnesium test method on Roche Hitachi 717 is substantially comparable with the Beckman Coulter magnesium assay on AU640 analyzer. A summary of the regression statistics is provided below: Samples, n= 100 {4}------------------------------------------------ Slope: 0.981 Intercept: 0.00 mg/dL Correlation: 0.9848 b. Matrix comparison: A matrix comparison study was conducted using fresh samples from non-fasting donors drawn as serum, and Li Heparin plasma. Sekisui Diagnostics Magnesium method with serum was used to compare the results obtained using the Sekisui Diagnostics Magnesium Assay with plasma on the Roche Hitachi 717 automated analyzer. The results demonstrate the Sekisui Diagnostics Magnesium test method with plasma is substantially comparable with the Sekisui Diagnostics Magnesium test method with serum on Roche Hitachi 717 analyzer. A summary of the regression statistics is provided below: Samples, n= દ્વ Slope: 1.009 Intercept: - 0.12 mg/dL Correlation: 0.9903 - 3. Clinical studies: a. Clinical Sensitivity: - Not applicable - b. Clinical specificity: Not applicable - c. Other clinical supportive data (when a. and b. are not applicable): Not applicable - 4. Clinical cut-off: Not applicable - 5. Expected values/Reference range: Testing followed the CSLI C28-A guideline to validate the reference range using a study of 20 serum samples from apparently healthy donors. The reference range is validated if the samples result in no outliers. The testing did not result in outliers supporting the literature reports of adult reference ranges for magnesium. [Tietz, N.W., (Editor) Clinical Guide to Laboratory Tests, W.B. Saunders Company, Philadelphia (1983) p.338.]. #### N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. #### O. Conclusion: The information presented in the premarket notification demonstrates that the performance of the Sekisui Diagnostics Magnesium Reagent for use with human serum and lithium heparin plasma is substantially equivalent to the cleared predicate device. Equivalence was demonstrated using patient samples with magnesium values that span the assay range. The information presented in the premarket notification provides a reasonable assurance that the Sekisui Diagnostics Magnesium for use with human serum and lithium heparin plasma is safe and effective for the stated intended use. {5}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized image of a bird-like figure, possibly representing a human profile, with outstretched wings. 10903 New Hampshire Avenue Silver Spring, MD 20993 SEKISUI DIAGNOSTICS P.E.I. INC C/O Penny White 70 Watts Ave Charlottetown Prince Edward Island Canada C1E2B9 DEC - 2 2011 Re: k111915 > Trade Name: MAGNESIUM ASSAY Regulation Number: 21 CFR §862.1495 Regulation Name: Magnesium test system Regulatory Class: Class I,reserved Product Codes: JGJ Dated: October 19, 2011 Received: October 24, 2011 Dear Ms. White We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {6}------------------------------------------------ # Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ... You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm Sincerely yours, N Countney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use Form 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: ___ Magnesium Assay Indications for Use: The Sekisui Magnesium Assay is intended for the quantitative measurement of magnesium in human serum and plasma (Lithium Heparin) on automated clinical chemistry analyzers. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia and hypermagnesemia. This device is intended for professional use and IN VITRO diagnostic use only. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Signature Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) < 111915 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...