ARK METHOTREXATE ASSAY, ARK METHOTREXATE CALIBRATOR, ARK METHOTREXATE CONTROL, ARK METHOTREXATE CONTROL (CALIBRATION RAN

K111904 · Ark Diagnostics, Inc. · LAO · Oct 18, 2011 · TX

Device Facts

Record IDK111904
Device NameARK METHOTREXATE ASSAY, ARK METHOTREXATE CALIBRATOR, ARK METHOTREXATE CONTROL, ARK METHOTREXATE CONTROL (CALIBRATION RAN
ApplicantArk Diagnostics, Inc.
Product CodeLAO · TX
Decision DateOct 18, 2011
DecisionSESE
Submission TypeTraditional
Device ClassClass U

Indications for Use

The ARK™ Methotrexate Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of Methotrexate in human serum or automated clinical chemistry analyzers. The measurements obtained are used in monitoring levels of methotrexate to help ensure appropriate therapy. The ARK™ Methotrexate Calibrator is intended for use in calibration of the ARK Methotrexate Assay. The ARKTM Methotrexate Control is intended for use in quality control of the ARK Methotrexate Assay. Specimens from patients who have received glucarpidase (carboxypeptidase G2) as a high dose methotrexate rescue therapy should not be tested with the ARK Methotrexate Assay.

Device Story

ARK Methotrexate Assay is a homogeneous enzyme immunoassay (EIA) for quantitative measurement of methotrexate in human serum or plasma. It utilizes a two-reagent system: R1 (anti-methotrexate rabbit polyclonal antibody with substrate) and R2 (methotrexate labeled with bacterial G6PDH enzyme). The assay operates on automated clinical chemistry analyzers (validated on Roche/Hitachi 917). Principle: competition between drug in specimen and enzyme-labeled methotrexate for antibody binding sites; enzyme activity is directly proportional to drug concentration. Active enzyme converts NAD to NADH, measured spectrophotometrically as rate of absorbance change. Used by clinicians to monitor methotrexate levels during therapy, particularly high-dose regimens, to guide leucovorin rescue and assess toxicity risk. Benefits include timely adjustment of therapeutic dosing to prevent renal or systemic toxicity. Includes calibrators and controls in synthetic protein matrix.

Clinical Evidence

Bench testing only. Performance validated on Roche/Hitachi 917. Precision (total CV 3.8-11.7%), linearity (R²=0.9978, range 0.04-1.20 μmol/L), and accuracy-by-recovery were established. Method comparison with predicate (N=102) showed slope 1.00, intercept 0.01, r²=0.978. Interference testing confirmed no significant cross-reactivity with common endogenous substances or folate analogs, though substantial cross-reactivity with the metabolite DAMPA was noted, precluding use in patients treated with glucarpidase.

Technological Characteristics

Homogeneous enzyme immunoassay (EIA). Reagents: anti-methotrexate rabbit polyclonal antibody, methotrexate-labeled bacterial G6PDH. Calibrators/Controls: synthetic protein matrix. Measuring range: 0.04-1.20 μmol/L. Analyzed via spectrophotometric rate of change in absorbance (NADH production). Validated on Roche/Hitachi 917 automated analyzer. Standards referenced: CLSI EP5-A2 (precision), EP6-A (linearity), EP9-A2 (method comparison), EP7-A2 (interference), EP17-A (LoD/LoQ).

Indications for Use

Indicated for quantitative monitoring of methotrexate levels in human serum or plasma to ensure appropriate therapy. Contraindicated for patients who have received glucarpidase (carboxypeptidase G2) as high-dose methotrexate rescue therapy.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k111904 B. Purpose for Submission: New assay, calibrator and control C. Measurand: Methotrexate D. Type of Test: Quantitative Homogeneous Enzyme Immunoassay E. Applicant: ARK Diagnostics, Inc. F. Proprietary and Established Names: ARK™ Methotrexate Assay, ARK™ Methotrexate Calibrator, ARK™ Methotrexate Control G. Regulatory Information: 1. Regulation section: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | LAO - Methotrexate enzyme immunoassay | II | Unclassified | Toxicology (91) | | DLJ - Clinical toxicology calibrator | II | 21 CFR §862.3200 | Toxicology (91) | | LAS - Clinical toxicology control material | I, reserved | 21 CFR §862.3280 | Toxicology (91) | H. Intended Use: 1. Intended use(s): See indications for use below. 2. Indication(s) for use: The ARK Methotrexate Assay is a homogeneous enzyme immunoassay intended for the quantitative determination of methotrexate in human serum or plasma on automated clinical chemistry analyzers. The measurements obtained are used in {1} monitoring levels of methotrexate to help ensure appropriate therapy. The ARK Methotrexate Calibrator is intended for the calibration of the ARK Methotrexate Assay. The ARK Methotrexate Control is intended for the quality control of the ARK Methotrexate Assay. 3. Special conditions for use statement(s): For prescription use only. Specimens from patients who have received glucarpidase (carboxypeptidase G2) as a high dose methotrexate rescue therapy should not be tested with the ARK Methotrexate Assay. 4. Special instrument requirements: Performance characteristics were established on the Roche/Hitachi 917 automated clinical chemistry analyzer. I. Device Description: The ARK Methotrexate Assay consists of reagents R1 (anti-Methotrexate rabbit polyclonal antibody with substrate) and R2 (Methotrexate labeled with bacterial G6PDH enzyme.) The ARK Methotrexate Calibrator consists of a six-level set to calibrate the assay. The calibrators consist of a synthetic protein matrix and the levels are 0.00 umol/L, 0.05 umol/L, 0.15 umol/L, 0.25 umol/L, 0.50 umol/L and 1.2 umol/L. The ARK Methotrexate Control consists of a six-level set used for quality control of the assay (tri-level calibration range set and tri-level high range set). The controls consist of a synthetic protein matrix and the levels are 0.07 umol/L, 0.40 umol/L, 0.80umol/L, 5 umol/L, 50 umol/L and 500 umol/L. The ARK Methotrexate Dilution Buffer is equivalent to zero calibrator (Calibrator A). J. Substantial Equivalence Information: 1. Predicate device name(s): Abbott TDx®/TDxFLx® METHOTREXATE II 2. Predicate K number(s): k932615 {2} 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Proposed Device | Predicate Device (k932615) | | Intended Use/Indications for use | For the quantitative determination of methotrexate. The measurements obtained are used in monitoring levels of methotrexate to ensure appropriate therapy. | Same | | Specimen type | Serum and plasma | Same | | Number of calibrators and controls | 6 levels (each) | Same | | Reagent matrix and storage condition | Liquid, 2-8°C | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Methodology | Homogenous enzyme immunoassay (EIA) | Fluorescence polarization immunoassay (FPIA) | | Reagent Components | Two (2) reagent system: Anti-Methotrexate Antibody/Substrate Reagent (R1) (contains rabbit polyclonal antibodies to Methotrexate) Enzyme Reagent (R2) (contains Methotrexate labeled with bacterial G6PDH) | Reagent Pack: W (Wash solution) S (containing Methotrexate antibody (mouse monoclonal)) T (Methotrexate Fluorescein Tracer) P (Pretreatment Solution) | | Matrix for calibrators and controls | Synthetic matrix | Human serum | {3} K. Standard/Guidance Document Referenced (if applicable): - CLSI document EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - CLSI Guideline EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach - CLSI Guideline EP9-A2: Method Comparison and Bias Estimation Using Patient Samples - CLSI Protocol EP7-A2: Interference Testing in Clinical Chemistry - CLSI Protocol EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation L. Test Principle: The ARK Methotrexate Assay is a homogeneous immunoassay based on competition between drug in the specimen and Methotrexate labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for binding to the antibody reagent. As the latter binds antibody, enzyme activity decreases. In the presence of drug from the specimen, enzyme activity increases and is directly proportional to the drug concentration. Active enzyme converts the coenzyme nicotinamide adenine dinucleotide (NAD) to NADH that is measured spectrophotometrically as a rate of change in absorbance. Endogenous serum G6PDH does not interfere with the results because the coenzyme NAD functions only with the bacterial enzyme used in the assay. M. Performance Characteristics (if/when applicable): Performance was validated on the Hitachi 917 instrument. 1. Analytical performance: a. Precision/Reproducibility: Precision was determined as described in CLSI Guideline EP5-A2. Data for the precision evaluation studies (total and within-laboratory precision) were collected on a single Roche/Hitachi 917 automated clinical chemistry analyzer. Each level of control and patient sample was assayed in quadruplicate twice a day over twenty non-consecutive days. A total of 160 determinations were made for each sample. Four calibrations were performed during this interval. The ARK Methotrexate Controls (three levels) and three patient sample pools were tested in each run. Mean methotrexate concentration, standard deviation (SD) and coefficients of variation (% CVs) were calculated for within-run, between-day, and total precision. Patient specimens that contained methotrexate were pooled to create three levels with sufficient volume to complete the 20-day protocol. 4 {4} | Sample | N | Mean (μmol/L) | Within Run | | Between Day | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | | Controls | | | | | | | | | | Low | 160 | 0.06 | 0.005 | 8.2 | 0.005 | 7.3 | 0.007 | 10.7 | | Mid | 160 | 0.37 | 0.011 | 3.0 | 0.008 | 2.1 | 0.014 | 3.8 | | High | 160 | 0.76 | 0.032 | 4.3 | 0.030 | 4.0 | 0.045 | 5.9 | | Patient Pools | | | | | | | | | | Low | 160 | 0.07 | 0.006 | 9.1 | 0.005 | 7.5 | 0.008 | 11.7 | | Mid | 160 | 0.41 | 0.013 | 3.3 | 0.026 | 6.4 | 0.029 | 7.2 | | High | 160 | 0.82 | 0.037 | 4.5 | 0.043 | 5.2 | 0.057 | 7.0 | *Samples were diluted in ARK Methotrexate Dilution Buffer. Mean result and SD were multiplied by the dilution factor. The sponsor also performed a precision study to evaluate serum and plasma samples. The study was modified from the recommendations in CLSI Guideline EP5-A2; 5 days of precision studies were performed. The results showed that precision for serum and plasma was not significantly different. b. Linearity/assay reportable range: Linearity studies were performed as suggested in CLSI Guideline EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. Linearity studies were performed by testing concentrations of methotrexate across the claimed measuring range. Gravimetric addition of pure methotrexate (USP) to dimethylformamide and volumetric addition of this stock solution to human serum negative for methotrexate was made to achieve a concentration (HIGH Linearity Sample; 1.30 μmol/L) above the calibration range as suggested in CLSI Protocol EP-6A. Dilutions of this 1.30 μmol/L HIGH Linearity Sample were made proportionally with pooled human serum negative for methotrexate. Two separately calibrated analytical runs of three replicates of each sample per run were assayed (N=6). The results of the six replicates were averaged. Regression analyses were performed between the measured mean methotrexate and the theoretical values for each dilution, using first order and second order polynomial determinations. Predicted 1st and 2nd order regressed values were calculated from the regression equations and compared. {5} | Theoretical (μmol/L) | Observed Results (μmol/L) | 1^{st} Order Predicted Results | 2nd Order Predicted Results | Difference (μmol/L or %) | | --- | --- | --- | --- | --- | | 0.00 | 0.00 | 0.009 | -0.003 | na | | 0.02 | 0.02 | 0.026 | 0.016 | -0.010 μmol/L | | 0.04 | 0.04 | 0.042 | 0.034 | -0.008 μmol/L | | 0.05 | 0.06 | 0.059 | 0.053 | -0.006 μmol/L | | 0.07 | 0.08 | 0.076 | 0.072 | -0.004 μmol/L | | 0.11 | 0.11 | 0.110 | 0.109 | -0.7 % | | 0.18 | 0.17 | 0.178 | 0.183 | 3.1 % | | 0.36 | 0.34 | 0.347 | 0.364 | 4.8 % | | 0.65 | 0.63 | 0.618 | 0.639 | 3.4 % | | 0.72 | 0.72 | 0.686 | 0.705 | 2.9 % | | 0.86 | 0.84 | 0.821 | 0.835 | 1.7 % | | 1.01 | 0.99 | 0.957 | 0.960 | 0.4 % | | 1.15 | 1.06 | 1.092 | 1.082 | -1.0 % | | 1.30 | 1.19 | 1.228 | 1.199 | -2.3 % | Linear regression analysis yields the following result: $$ y = 0.9407x + 0.086 $$ $$ R^2 = 0.9978 $$ There are no concentrations at which the percent difference was more than 10% between the predicted 1st and 2nd order regressed values for concentrations &gt;0.10 μmol/L or ±0.01 μmol/L at concentrations ≤ 0.10 μmol/L. The claimed measuring range of this device is 0.04 - 1.20 μmol/L. ## Evaluation of recovery: An accuracy-by-recovery study was conducted to determine the trueness of quantitative measurements of methotrexate across the calibration range of the ARK™ Methotrexate Assay. Recovery of methotrexate was studied with two sources of methotrexate for comparison. Samples were prepared by volumetric addition of methotrexate (Cerilliant Certified Stock solution 99.8% purity) to human serum negative for methotrexate. Drug concentrations across the assay range (0.06, 0.10, 0.30, 0.60, and 1.00 μmol/L) were tested. Each sample was assayed in triplicate in each of two separately calibrated runs for a total of six replicates. The results were averaged and compared to the theoretical target concentration and percentage recovery calculated. $$ \% \text{Recovery} = 100 \times \frac{\text{Mean recovered concentration}}{} $$ {6} 7 # Target concentration For comparison, samples were prepared by gravimetric addition of methotrexate (USP &gt; 99.9% purity) to dimethylformamide (DMF) and volumetric addition of this stock solution to human serum negative for methotrexate, then assayed as described above. Recovery was acceptable if the mean methotrexate concentration measured was within ±10% of the target level for concentrations at &gt;0.1 μmol/L and ±0.01 at ≤0.1 μmol/L. | Target Concentration (μmol/L) | Cerilliant Methotrexate | | USP Methotrexate | | | --- | --- | --- | --- | --- | | | Mean Recovered Concentration (μmol/L) | Recovery (%) | Mean Recovered Concentration (μmol/L) | Recovery (%) | | 0.06 | 0.067 | +0.007 μmol/L (111.1%) | 0.060 | 100.0 | | 0.10 | 0.100 | 100.0 | 0.098 | 98.3 | | 0.30 | 0.295 | 98.3 | 0.282 | 93.9 | | 0.60 | 0.613 | 102.2 | 0.597 | 99.4 | | 1.00 | 0.988 | 98.8 | 1.008 | 100.8 | ## Evaluation of samples above the claimed measuring range: Frequently, patient samples may have methotrexate levels higher than the claimed measuring range of the device. They require dilution into the measuring range for evaluation. Evaluation of linearity, precision and accuracy after dilution of samples was evaluated. ## Evaluation of linearity for samples with elevated methotrexate levels: A sample containing approximately 1200 μmol/L of methotrexate in human serum was prepared to assess the proportional measurement of the drug upon dilution into the calibration range with ARK Methotrexate Dilution Buffer. This HIGH sample was diluted proportionally in pooled human serum to obtain concentrations ranging from 2 to 1200 μmol/L. As a pre-analytical step, these serum samples were diluted in ARK Methotrexate Dilution Buffer. Ten-fold dilution factors were used. Two separately calibrated analytical runs of three replicates of each sample per run were assayed (N=6). Linear regression analysis was performed to demonstrate by visual analysis that measurement of methotrexate was adequately proportional for concentrations exceeding the calibration range. Measurement of methotrexate is linear within the calibration range of the ARK Methotrexate Assay when the predicted 1st and 2nd order regressed values agree within 10% for concentrations &gt;0.10 μmol/L or within ±0.01 μmol/L at concentrations ≤0.10 μmol/L. {7} Measurement of elevated methotrexate above the calibration range was linear. Regression analysis demonstrated proportionality along a straight line by visual inspection. Reported values (after calculating for dilution) were plotted versus the expected concentration of methotrexate (2.00 to 1200.00 μmol/L). | Expected (μM) | 2.00 | 8.00 | 20.00 | 80.00 | 200.00 | 800.00 | 1200.00 | | --- | --- | --- | --- | --- | --- | --- | --- | | Dilution | 1/10 | 1/10 | 1/100 | 1/100 | 1/1000 | 1/1000 | 1/10000 | | | | | | | | | | | 1 | 0.21 | 0.75 | 0.20 | 0.83 | 0.20 | 0.79 | 0.12 | | 2 | 0.21 | 0.76 | 0.20 | 0.78 | 0.20 | 0.76 | 0.11 | | 3 | 0.20 | 0.82 | 0.21 | 0.85 | 0.20 | 0.75 | 0.12 | | 4 | 0.21 | 0.75 | 0.19 | 0.82 | 0.20 | 0.74 | 0.11 | | 5 | 0.20 | 0.78 | 0.20 | 0.80 | 0.20 | 0.76 | 0.11 | | 6 | 0.20 | 0.79 | 0.19 | 0.82 | 0.19 | 0.75 | 0.12 | | | | | | | | | | | Mean (μM) | 0.205 | 0.775 | 0.198 | 0.817 | 0.198 | 0.758 | 0.115 | | SD | 0.005 | 0.027 | 0.008 | 0.024 | 0.004 | 0.017 | 0.005 | | Reported (μM) | 2.05 | 7.75 | 19.83 | 81.67 | 198.33 | 758.33 | 1150.00 | Linear regression analysis of this study yielded the following: $$ y = 0.9537x + 2.1285 $$ $$ R^2 = 0.9999 $$ Evaluation of precision in samples with elevated methotrexate levels: Precision was evaluated as described above in 1a. ARK Methotrexate High Range Controls are comprised of three levels (5, 50, and 500 μmol/L). These controls are comprised in a proteinaceous synthetic matrix equivalent to human serum/plasma. High Range controls and patient sample pools with concentrations greater than the calibration range were diluted 10x, 100x or 1000x, respectively in ARK Methotrexate Dilution Buffer prior to testing. High Range controls and samples were diluted once for each run and these diluted samples were tested in quadruplicate Human serum/plasma specimens that contained methotrexate were pooled to create six levels with sufficient volume to complete the 20-day protocol. High patient sample pools (approximately 5, 50 and 500 μmol/L) required that the sample pools were supplemented with methotrexate stock. The stock solution used was prepared by gravimetric addition of pure Methotrexate (USP &gt;99.9 % purity) to dimethylformamide (DMF) and volumetric addition of this stock solution to the specimen pool. {8} 9 | Sample | N | Mean (μmol/L) | Within Run | | Between Day | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | SD | %CV | | Controls | | | | | | | | | | Control 5 | 160 | 4.8 | 0.15 | 3.1 | 0.13 | 2.8 | 0.20 | 4.2 | | Control 50 | 160 | 49 | 1.36 | 2.8 | 2.32 | 4.8 | 2.72 | 5.6 | | Control 500 | 160 | 476 | 15.17 | 3.2 | 30.75 | 6.5 | 34.66 | 7.3 | | Patient Pools | | | | | | | | | | Patient Pool – 5 | 160 | 4.6 | 0.14 | 3.1 | 0.183 | 4.0 | 0.24 | 5.3 | | Patient Pool – 50 | 160 | 45 | 1.33 | 3.0 | 2.63 | 5.9 | 2.93 | 6.6 | | Patient Pool – 500 | 160 | 461 | 11.84 | 2.6 | 27.04 | 5.9 | 29.60 | 6.4 | Evaluation of accuracy in samples with elevated methotrexate levels: A method comparison study was performed to evaluate the agreement between the ARK Methotrexate Assay and predicate device Abbott TDx/TDxFLx Methotrexate II assay (k932615) using samples with elevated methotrexate levels (i.e., samples that require dilution into the assay's measuring range for evaluation). The study was performed using Clinical and Laboratory Standards Institute (CLSI) Protocol EP9-A2 Method Comparison and Bias Using Patient Samples. Forty five patient specimens above the claimed measuring range were evaluated with the proposed and predicate assay; eight samples were spiked with methotrexate. All samples required dilution before testing on both devices (ARK Methotrexate Dilution Buffer was used for the proposed device. The samples ranged from 1.56 – 1440 μmol/L on the predicate. Results of the study are summarized in the table below: | Parameter | Range 1.56 to 1440 μmol/L | | --- | --- | | Number of Samples | 45 | | Slope | 0.96 (0.94 to 0.99) | | y-intercept | 0.11 (-0.52 to 0.33) | | Correlation Coefficient (r²) | 0.997 (0.994 to 0.998) | c. Traceability, Stability, Expected values (controls, calibrators, or methods): ARK Methotrexate Calibrators and controls are prepared by volumetric dilution of high purity, certified Methotrexate solution into a synthetic proteinaceous matrix free of Methotrexate. {9} 10 # ARK Methotrexate Calibrator A master calibrator lot is prepared volumetrically using a certified solution of methotrexate. The ARK Methotrexate Calibrator is traceable to a certified methotrexate solution and the uncertainty in the methotrexate concentration is 5% relative to the concentration in the certified solution. The concentration of methotrexate in the certified solution is traceable to HPLC; the purity of methotrexate in the certified solution is determined by HPLC and other spectral procedures as performed by the supplier of the certified solution. Bulk solutions of the ARK Methotrexate Calibrator are prepared volumetrically using the certified solution of methotrexate. The concentration of methotrexate in the respective bulk solution must agree within 5% of its corresponding master calibrator. **Value Assignment**: Testing is performed with the ARK Methotrexate Assay on the Roche/Hitachi 917 automated analyzer. Two calibrated runs are performed using the Master Calibrator. In each run, five replicates of Master Lot (reference) and Test Lot are tested as matched pairs for each calibrator level. Mean values for ten replicates are calculated. Test lot mean values are expected to match the Master lot mean values within 5% allowance. Target values for the six levels of calibrator are 0.00 umol/L, 0.05 umol/L, 0.15 umol/L, 0.25 umol/L, 0.50 umol/L and 1.2 umol/L. # ARK Methotrexate Control ARK uses USP methotrexate powder to produce the controls. Six levels are provided: three levels (0.07, 0.40 and 0.80 umol/L) are within the calibration range and three levels above the calibration range (5, 50 and 500 µmol/L) are available for controlling the dilution step. ARK manufactures the controls gravimetrically to contain methotrexate within 10% of the target levels. **Value Assignment**: Testing is performed with the ARK Methotrexate Assay on the Roche/Hitachi 917 automated analyzer, calibrated with the master calibrator lot. Three calibrated runs are performed using five replicates of each level per run. The 5, 50, 500 µmol/L controls are pre-diluted 1:10, 1:100, and 1:1000 respectively with the ARK Methotrexate Dilution Buffer and tested. Results for high range controls are multiplied by the applicable dilution factor. Mean values (15 replicates) for the test lots are expected to result within 10% of the nominal concentration. The control ranges are set to be +/- 25% from the mean values. Each laboratory should establish the mean value for each control level and its own ranges for each new lot of controls. # Calibrator and Control Stability: The calibrators and controls are stable until the expiration date printed on the vial when stored unopened and opened at 2-8 °C. {10} Real time stability studies are ongoing for both unopened and opened calibrators and controls. Stability testing protocols and sponsor’s acceptance criteria were reviewed and found to be acceptable. d. Detection limit: Limit of Detection (LoD) Limit of Quantitation (LoQ) studies were conducted using CLSI Guideline EP 17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation. The study was performed on the Roche/Hitachi 917 automated clinical chemistry analyzer. Samples were prepared by gravimetric addition of pure methotrexate (USP) to dimethylformamide and volumetric addition of this stock solution to human serum negative for methotrexate. Pooled human serum representative of the patient specimen matrix was supplemented with methotrexate to give concentrations of 0.02, 0.03, 0.04, and 0.05 µmol/L. The LoB and LoD were evaluated by testing 60 replicates of pooled human serum (BLANK) and 60 replicates of the first positive level 0.02 µmol/L methotrexate. The analyzer was calibrated and three analytical runs were performed. In each run, 20 replicates of BLANK and 20 replicates of 0.02 µmol/L methotrexate were analyzed. The grand mean and root mean square standard deviation (RMS SD) were calculated. Statistical analyses were performed according to the CLSI Guideline. Since the analyzer did not report values below 0.00 µmol/L, the standard deviation for 0.02 µmol/L methotrexate was used to calculate the LoB based on the equation $$ \text{LoB} = \mu\text{B} + 1.645(\sigma\text{B}) $$$, the standard deviation being relatively constant in this area of measurement. The error of the BLANK (σB) was taken to equal the standard deviation of the sample (SDs) as a fair estimate. The LoQ was determined after testing samples containing 0.03, 0.04, or 0.05 µmol/L. Eight replicates of each sample were tested in each of five runs (each run performed on a separate day) to yield 40 replicates of each LoQ sample tested. The grand mean and RMS SD were calculated for each sample. The LoQ of the ARK Methotrexate Assay is defined as the lowest concentration for which acceptable inter-assay precision (≤0.01 SD) and recovery (± 0.01 µmol/L) is observed. Where this concentration’s 95% confidence interval exceeds the LoD, the definition of LoQ according to the CLSI Guideline will be met. The criteria of LoQ were met at 0.04 µmol/L. (The same LoQ was established when a similar study was performed as above but also incorporating three lots of reagents, controls and calibrators.) The LoB was determined as 0.01 µmol/L, and the LoD was determined as 0.02 µmol/L using $$ \text{SD} = 0.005 \, \mu\text{mol/L} $$. LoQ was calculated to be 0.04 µmol/L. e. Analytical specificity: Potential interferences of endogenous materials were evaluated according to Clinical and Laboratory Standards Institute (CLSI) Guideline EP7-A2: Interference Testing in Clinical Chemistry. 11 {11} Clinically high concentrations of potentially interfering endogenous substances in serum with known levels of methotrexate (approximately 0.05 and 0.50 μmol/L) were evaluated. Pooled human serum supplemented exogenously with endogenous interfering compounds was prepared prior to addition of methotrexate. Gravimetric addition of methotrexate (USP &gt; 99.9% purity) in dimethylformamide (DMF) and volumetric addition of this stock solution to samples was made to obtain the levels of methotrexate. Six replicates of each sample and their respective serum controls were tested. Respective runs were co-calibrated with the ARK Methotrexate Calibrator. Tri-level calibration range quality controls of the ARK Methotrexate Control were used to qualify runs. The mean results of methotrexate were calculated and the percentage recoveries relative to the respective serum control mean results were determined. Interference was defined as follows: - At methotrexate concentrations near 0.05 μmol/L, the mean concentration of methotrexate in the presence of elevated endogenous substances should fall outside ± 0.02 μmol/L of the mean result for the respective serum control. - At methotrexate concentrations near mid calibration range (0.50 μmol/L), the mean concentration of methotrexate in the presence of elevated endogenous substances should result in &gt; 10% error in detecting methotrexate in comparison to the mean result for the respective serum control. | Interfering Substance | Interferent Concentration | | --- | --- | | Albumin | 12 g/dL | | Bilirubin - conjugated | 70 mg/dL | | Bilirubin - unconjugated | 70 mg/dL | | Cholesterol | 400 mg/dL | | Gamma-Globulin | 12 g/dL | | Hemoglobin | 1000 mg/dL | | Intralipid® | 500 mg/dL | | Rheumatoid Factor | 1100 IU/mL | | Triglycerides | 749 mg/dL | | Uric Acid | 30 mg/dL | {12} Interference with any of these substances was not detected at either level of methotrexate tested. Methotrexate's metabolites, structurally similar compounds, folate derivatives, and potentially co-administered medications were tested to determine whether these compounds affect the quantitation of methotrexate concentrations using the ARK™ Methotrexate Assay. 7-Hydroxymethotrexate (7-OH-MTX) is the main metabolite in serum following high-dose methotrexate (HDMTX) treatment. The concentration of 7-OH-MTX may exceed that of the parent compound by up to 100-fold in plasma shortly after MTX infusion. Methotrexate is also metabolized by intestinal bacteria to the minor, inactive metabolite 2,4-diamino-N¹⁰-methylpteroic acid (DAMPA). Pooled human serum was supplemented with methotrexate prior to addition of potentially cross reacting metabolites (7-OH-MTX and DAMPA) or other compounds with structural similarity. Preparation of serum pools with 0.05 and 0.50 μmol/L methotrexate were identical to the interference study above. Then the potentially cross reactive compounds at their respectively solvated concentrations were added to serum either in the absence of methotrexate or to serum containing methotrexate. Solvent controls for each potentially cross-reacting interferent sample were also prepared. The respective concentrations of cross reacting interferents were as recommended by CLSI EP7-A2, or in the case of trimethoprim and triamterene, above therapeutic levels as tested. Six (6) replicates of each sample and their respective solvent serum controls were tested. The mean results of methotrexate were calculated. The percentage cross reactivity was determined for each compound tested. The ARK Methotrexate Assay did not cross react (≤ 0.01%) with folate analogs or other compounds at ≥ 1000 μmol/L as tested: | Compound | Tested (μmol/L) | Compound | Tested (μmol/L) | | --- | --- | --- | --- | | Adriamycin | 1000 | 5-Methyltetra-hydrofolic acid | 1000 | | Cyclophosphamide | 1500 | Prednisolone | 1000 | | Cytosine | 1000 | Pyrimethamine | 1000 | | Dihydrofolic Acid | 1000 | Sulfamethoxazole | 1600 | | DL-6-Methyl-5,6,7,8-Tetrahydropterine | 1000 | Tetrahydrofolic Acid | 1000 | | Folic Acid | 1000 | Vinblastine | 1000 | | Folinic Acid (leucovorin) | 1000 | Vincristine | 1000 | | 5-Fluorouracil | 3000 | | | | 6-Mercaptopurine | 1000 | | | {13} Cross reactivity to 7-Hydroxymethotrexate, the major metabolite The ARK Methotrexate Assay did not cross react ( $\leq 0.07\%$ ) with the major metabolite 7-OH-MTX. Cross reactivity to the minor, inactive metabolite 2,4-diamino- $\mathbf{N}^{10}$ -methylpteroic acid (DAMPA) The ARK Methotrexate Assay cross reacts substantially with the minor metabolite DAMPA. Tests were performed in the absence of the parent drug methotrexate. Cross reactivity to DAMPA ranged from 64.3 to $100\%$ . The assay should not be used during possible compassionate therapy with glucarpidase (carboxypeptidase G2) that rapidly converts circulating methotrexate to DAMPA. # Drugs that cross react The ARK Methotrexate Assay cross reacts slightly with triamterene and trimethoprim, however these drugs may be contraindicated for MTX cancer treatment due to additional adverse effects if co-administered. The structures of these compounds closely match the pteridine ring moiety of methotrexate. | Compound | Tested (μmol/L) | Methotrexate Absent | | Methotrexate Present 0.05 μmol/L | | Methotrexate Present 0.50 μmol/L | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | MTX (μmol/L) | Cross Reactivity (%) | MTX (μmol/L) | Cross Reactivity (%) | MTX (μmol/L) | Cross Reactivity (%) | | Triamterene | 25 | 0.46 | 1.85 | 0.89 | 3.32 | 1.04 | 2.31 | | Trimethoprim | 100 | 0.17 | 0.17 | 0.16 | 0.12 | 0.99 | 0.54 | f. Assay cut-off: Not applicable. # 2. Comparison studies: a. Method comparison with predicate device: Method comparison studies were performed using Clinical and Laboratory Standards Institute (CLSI) Protocol EP9-A2 Method Comparison and Bias Using Patient Samples. Method comparisons were performed to evaluate the agreement between the ARK Methotrexate Assay and fluorescence polarization immunoassay (FPIA), predicate device Abbott TDx/TDxFLx Methotrexate II assay, k932615. 102 leftover specimens within the claimed measuring range were used for method comparison of the ARK Methotrexate Assay and the FPIA predicate device (TDx Assay). Both serum and plasma specimens were included in the study. Singlicate measurements of all specimens were used for method comparison. {14} Ten specimens were created by pooling or supplementation with methotrexate (USP) to contain methotrexate concentrations useful for the method comparison. Summary of the method comparison data: | Parameter | Range 0.04 to 1.19 μmol/L | | --- | --- | | Number of Samples | 102 | | Slope | 1.00 (1.00 – 1.02) | | y-intercept | 0.01 (0.00 to 0.01) | | Correlation Coefficient (r²) | 0.978 (0.968 to 0.985) | b. Matrix comparison: Anticoagulated plasma and serum were evaluated to demonstrate equivalency of these matrices for measurement of methotrexate with the ARK™ Methotrexate Assay. Matched samples for serum and plasma from eight subjects were evaluated. In this study, blood was separated into plasma or serum prior to supplementation with methotrexate. Each subject donated in three different anticoagulant tubes and a serum tube to produce a matched set. A methotrexate stock solution was prepared by gravimetric addition of methotrexate (USP &gt; 99% purity) to dimethylformamide (DMF). For each matrix and subject, an aliquot was supplemented by volumetric addition of methotrexate stock solution. For eight subjects, methotrexate was added to each matrix to give 1.00 μmol/L, and then serial two-fold dilutions were made with its own matrix/subject to give 0.50 and 0.25 μmol/L, respectively. For four subjects, methotrexate was added to each matrix to give 0.05 μmol/L. Since the calibration range spans 0.00 to 1.20 μmol/L, equivalency of serum and plasma matrices for measurement of methotrexate with the ARK Methotrexate Assay was evaluated from near the lower limit of quantitation (0.04 μmol/L) to the upper calibration range. Compared to the mean methotrexate level in the individual’s matched serum control, the mean methotrexate concentration in the individual’s corresponding plasma was acceptable if the concentration measured was within ±10% of the serum control for concentrations at &gt; 0.1 μmol/L and within ±0.02 μmol/L at methotrexate concentrations ≤ 0.1 μmol/L. The overall percentage of serum levels of spiked methotrexate above 0.10 μmol/L in plasma ranged 93.7 to 107.4%. At 0.05 μmol/L methotrexate, the difference in measurement of methotrexate between the plasma samples and serum controls was ≤0.01 μmol/L. 15 {15} 16 3. Clinical studies: a. Clinical Sensitivity: Not applicable. Not typical for this type of assay. b. Clinical specificity: Not applicable. Not typical for this type of assay. c. Other clinical supportive data (when a. and b. are not applicable): The sponsor provided a discussion with balanced and representative literature discussing clinical use of methotrexate measurements. 4. Clinical cut-off: See expected values below. 5. Expected values/Reference range: The following is included in the package insert: Methotrexate serum levels depend on indication for use, dosage, mode of administration, treatment regimen, individual pharmacokinetics, metabolism and other clinical factors.¹,² While the serum level may typically reach approximately 10 to 100 μmol/L in treatment of breast cancer (for example),⁶ concentrations may exceed 1000 μmol/L with high dose therapy for osteosarcoma, and up to 3100 μmol/L methotrexate was reached following a 4-hour infusion in pediatric patients with osteosarcoma.⁸ For treatment of osteosarcoma,⁷ the methotrexate decay curve has wide variability: 24 hours, 30 to 300 μmol/L; 48 hours, 3 to 30 μmol/L; and 72 hours, less than 0.3 μmol/L. A dose of 10 mg of leucovorin is usually administered intravenously 24 hours after initiation of the MTX infusion. Subsequent doses are adjusted and administered according to the MTX levels obtained at 24, 48, and 72 hours. Methotrexate levels in excess of 50 μmol/L at 24 hours, 10 μmol/L at 48 hours, and 0.5 μmol/L at 72 hours portend potential toxicity and are usually treated with an increase in the dose of leucovorin in accordance with algorithms until the MTX level is &lt;0.1 μmol/L. Guidelines for methotrexate therapy with leucovorin rescue usually recommend continuance of leucovorin until the methotrexate level falls below 0.05 μmol/L.¹,³ Some centers follow ≤ 0.10 μmol/L.⁷,⁹ From prescribing and other information: Laboratory Indicators of Toxicity Following Leucovorin Rescue Schedules with High Dose Methotrexate.¹,³,¹⁰ {16} | Clinical Situation | Laboratory Findings | | | --- | --- | --- | | | Methotrexate Level (μmol/L) | Hours after administration | | Normal Methotrexate Elimination | ~10 | 24 | | | ~1 | 48 | | | <0.2 | 72 | | Delayed Late Methotrexate Elimination | >0.2 | 72 | | | >0.05 | 96 | | Delayed Early Methotrexate Elimination and/or | ≥50 | 24 | | | ≥5 | 48 | | | OR | | | Evidence of Acute Renal Injury | ≥100% increase in serum creatinine | 24 | Renal toxicity is a significant risk and may be exacerbated by coadministration of other drugs, $^{5,10}$ for example vancomycin. $^{11}$ Other forms of toxicity can occur, including digestive disorders (e.g., nausea, vomiting, abdominal pain), cutaneous-mucous disorders (especially mucositis), hematological abnormalities (e.g., neutropenia and thrombocytopenia), liver function test disturbances, and neurotoxicity. $^{12-19}$ Given the profile of the appearance of the 7-hydroxymethotrexate metabolite, $^{6,18}$ its molar ratio to methotrexate of up to approximately 100-fold, $^{20}$ and relative insolubility versus the parent drug, $^{5,10}$ possible nephrotoxicity due to precipitation of the metabolite in renal tubules $^{20}$ may delay elimination of methotrexate itself. Glucarpidase therapy (available for compassionate use) reduces the circulating level of methotrexate rapidly, not intracellular drug. A rebound effect in the serum level of methotrexate following glucarpidase therapy has been observed. $^{5}$ Elimination of DAMPA may take several days before it no longer interferes with the monitoring of methotrexate by immunoassay. $^{4}$ 1. Prescribing information. 2008. Methotrexate Injection, USP. Hospira, Inc. Lake Forest, IL. 2. Bleyer, W. A. 1978. The clinical pharmacology of methotrexate: New applications of an old drug. Cancer 41:36 - 51. 3. Leucovorin (Fusilev) Prescribing Information. 2008. Spectrum Pharmaceuticals, Inc. Irvine, CA. 4. Al-Turkmani, M. R. et al., 2010. Difficulty Measuring Methotrexate in a Patient with High-Dose Methotrexate-Induced Nephrotoxicity. Clin Chem 56:1792 - 1796. 5. Schwarz, S. et al. 2007. Glucarpidase (Carboxypeptidase G2) intervention in adult and elderly cancer patients with renal dysfunction and delayed methotrexate elimination after high-dose methotrexate therapy. The Oncologist 12:1299 - 1308. 6. Bore, P. et al. 1987. Pharmacokinetics of Methotrexate and 7-Hydroxy-Methotrexate After Methotrexate Infusions. Cancer Drug Delivery 4:177 - 183. 7. Jaffe, N. and Gorlick, R. 2008. High-Dose Methotrexate in Osteosarcoma: Let the Questions Surcease—Time for Final Acceptance. J Clin Oncol 26:4365 - 4366. {17} 8. Colom, H. et al. 2009. Population Pharmacokinetics of High-Dose Methotrexate After Intravenous Administration in Pediatric Patients With Osteosarcoma. *Ther Drug Monit* 31:76–85. 9. Dombrowsky, E. et al. 2011. Evaluating performance of a decision support system to improve methotrexate pharmacotherapy in children and young adults with cancer. *Ther Drug Monit* 33:99–107. 10. Widemann, B. C. and Adamson, P. C. 2006. Understanding and managing methotrexate nephrotoxicity. *Oncologist* 11:694–703. 11. Blum, R. et al. 2002. Significant impairment of high-dose methotrexate clearance following vancomycin administration in the absence of overt renal impairment. *Annals of Oncology* 13:327–330. 12. Martelli, N. et al. 2011. Methotrexate pharmacokinetics in childhood acute lymphoblastic leukaemia: a prognostic value? *J Clin Pharm Ther* 36:237–245. 13. Mazanec, D. J. and Grisanti, J. M. 1989. Drug-induced osteoporosis. *Cleve Clin J of Med* 56:297–303. 14. Chessells, J. M. et al. 1990. Neurotoxicity in lymphoblastic leukaemia: Comparison of oral and intramuscular methotrexate and two doses of radiation. *Archives of Disease in Childhood* 65:416–422. 15. Allen, J. C. et al. 1980. Leukoencephalopathy following high-dose IV methotrexate chemotherapy with leucovorin rescue. *Cancer Treat Rep* 64:1261–1273. 16. Jacobs, P. et al. 1991. Methotrexate encephalopathy. *Eur J Cancer* 27:1061–1062. 17. Flombaum, C. D. and Meyers, P. A. 1999. High-Dose Leucovorin as Sole Therapy for Methotrexate Toxicity. *J Clin Oncol* 17:1589–1594. 18. Collier, C. P. et al. 1982. Analysis of methotrexate and 7-hydroxymethotrexate by high-performance liquid chromatography and preliminary clinical studies. *Ther Drug Monit* 4:371–380. 19. Widemann, B. C. et al. 2010. Glucarpidase, leucovorin, and thymidine for high-dose methotrexate-induced renal dysfunction: clinical and pharmacologic factors affecting outcome. *J Clin Oncol* 28:3979–3986. 20. Erttmann, R. et al. 1985. 7-Hydroxy-Methotrexate and Clinical Toxicity Following High-Dose Methotrexate Therapy. *J Cancer Res Clin Oncol* 109:86–88. ## N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. ## O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...