← Product Code [GEI](/productcode/GEI) · K111887

# CLOSUREFAST RADIOFREQUENCY CATHETER (K111887)

_Covidien · GEI · Sep 9, 2011 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K111887

## Device Facts

- **Applicant:** Covidien
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Sep 9, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ClosureFAST™ Radiofrequency Catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.

## Device Story

ClosureFAST™ Radiofrequency Catheter is a sterile, single-use, disposable electrosurgical device; features 3cm radiofrequency heating element, flexible shaft, and integrated instrument cable. Used with VNUS RFG2 Generator to deliver radiofrequency energy to distal heating element; positioned at treatment site to coagulate blood vessels. Designed for treating refluxing vein segments less than 7cm in length. Operated by physicians in clinical settings. Output is thermal coagulation of target vessel; enables treatment of shorter vein segments compared to previous iterations. Benefits patients by providing targeted endovascular treatment for superficial vein reflux.

## Clinical Evidence

Bench testing only. Design verification and validation performed on catheter compatibility, design specifications, and performance to compare the modified device to the predicate.

## Technological Characteristics

Electrosurgical catheter; 3cm radiofrequency heating element; flexible shaft; integrated instrument cable. Compatible with VNUS RFG2 Generator. Single-use, disposable, sterile. Materials identical to predicate.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- ClosureFAST™ Catheter ([K061373](/device/K061373.md))

## Reference Devices

- VNUS RFG2 Generator ([K040638](/device/K040638.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# KII1887

# A. Submitter

6.

Covidien 15 Hampshire St. Mansfield, MA 02048 Phone: 508-452-4135

510(k) Summary

# B. Contact Person

Daniel Campion Manager Regulatory Affairs

# C. Date Prepared

June 30, 2011

# D. Trade/Proprietary Name

ClosureFAST™ Radiofrequency Catheter

# E. Common/Usual Name

Electrosurgical Device

# F. Classification Name

Electrosurgical cutting and coagulation device and accessories

### G. Predicate Device(s)

ClosureFAST™ Catheter (K061373)

### H. Device Description

The ClosureFAST™ Radiofrequency Catheter is provided sterile, and is a single-use, disposable device. It has a 3cm radiofrequency heating element, a flexible shaft, and an integrated instrument cable. It is designed for use with the VNUS RFG2 Generator, cleared under 510(k) K040638.

#### -Intended Use

The ClosureFAST™ Radiofrequency Catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.

### J. Technology Characteristics

The ClosureFAST™ Radiofrequency Catheter is identical to the predicate device in terms of its intended use and fundamental scientific technology.

The function of the ClosureFAST™ catheter is to coagulate blood vessels by delivering radiofrequency energy to a heating element on the distal end of the catheter that is positioned at a desired treatment site. The device now will be offered with a 3cm heating element, therefore allowing physicians to treat refluxing vein segments less than 7cm in length.

Image /page/0/Picture/24 description: The image shows the text "SEP - 9 2011". The text appears to be a date, with "SEP" representing the month of September. The number 9 likely represents the day of the month. The year is 2011.

{1}------------------------------------------------

..

### K. Materials

The ClosureFAST™ Radiofrequency Catheter is composed of the same materials as the predicate device.

# L. Performance Data

Testing was performed on the proposed ClosureFast to compare it to the predicate device. Testing included Design Verification and Validation on catheter compatibility, design specifications and performance. Results from the testing demonstrate that the modified device is substantially equivalent to the legally marketed predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, a staff with two snakes entwined around it, which is a common symbol associated with medicine and healthcare.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Covidien % Mr. Daniel Campion Regulatory Affairs Vascular Therapies 15 Hampshire St. Mansfield, MA 02048

SEP - 9 2011

Re: KI11887

Trade/Device Name: ClosureFast Radiofrequency Catheter Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 10, 2011 Received: August 12, 2011

Dear Mr. Campion:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

#### Page 2 - Mr. Daniel Campion

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

E. Keith

Fo Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### 4. Indications for Use Statement

510(k) Number (if known): _KI | | 887

Device Name: ClosureFAST™ Radiofrequency Catheter

Indications for Use: The ClosureFAST™ Radiofrequency Catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.

Prescription use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 CRF 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

---

(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K111887

---

**Source:** [https://fda.innolitics.com/device/K111887](https://fda.innolitics.com/device/K111887)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
