OVER-THE-WIRE EXPANDABLE LEMAITRE VALVULOTOME

K111884 · LeMaitre Vascular, Inc. · MGZ · Jul 29, 2011 · Cardiovascular

Device Facts

Record IDK111884
Device NameOVER-THE-WIRE EXPANDABLE LEMAITRE VALVULOTOME
ApplicantLeMaitre Vascular, Inc.
Product CodeMGZ · Cardiovascular
Decision DateJul 29, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4885
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Over-The-Wire Expandable LeMaitre Valvulotome is indicated for cutting saphenous vein valves during In Situ bypass.

Device Story

The OTW-Expandable LeMaitre Valvulotome is a self-centering, self-sizing catheter designed for use during in-situ bypass surgery. The device is navigated through the saphenous vein over a guidewire. Upon deployment, integrated centering hoops expand to contact the vessel walls. As the clinician retracts the catheter, the hoops maintain a safe distance between the vessel wall and the cutting blades, ensuring the blades engage and cut the valves as they are encountered. The device is used by surgeons in a clinical setting to facilitate the bypass procedure. By effectively cutting the valves, the device allows for the proper flow of blood through the vein graft, potentially improving patient outcomes in vascular bypass surgery.

Clinical Evidence

No clinical data. Evidence consists of in-vitro bench-top testing and in-vitro cadaver testing. Bench testing included tensile testing of bond joints, dimensional verification, guidewire compatibility, trackability (simulated use), efficacy of valve cutting (simulated use), hemostasis, and fatigue testing. Biocompatibility was verified per AAMI/ANSI/ISO 10993-1:2003, including MEM elution, complement activation, platelet/leukocyte counts, hemolysis, PTT, intracutaneous reactivity, and murine local lymph node assays.

Technological Characteristics

Self-centering, self-sizing catheter with integrated hoops and cutting blades. Navigates via guidewire. Materials verified for biocompatibility per ISO 10993-1:2003. Mechanical operation; no energy source or software components.

Indications for Use

Indicated for patients undergoing in-situ bypass procedures requiring the cutting of saphenous vein valves.

Regulatory Classification

Identification

An external vein stripper is an extravascular device used to remove a section of a vein.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ #### JUL 2 9 2011 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Section 5.0 #### A. Submitter Information | Submitter's Name: | LeMaitre Vascular Inc. | |----------------------|------------------------------------------------------------------------------| | Address: | 63 Second Avenue<br>Burlington, MA 01803 | | Telephone: | 781-221-2266 | | Fax: | 781-425-5049 | | Contact Person: | Andrew Hodgkinson, Vice President Clinical<br>Regulatory and Quality Affairs | | Date of Preparation: | July 28, 2011 | #### B. Subject Device Trade Name: Common/Usual Name: Classification Name: OTW-Expandable LeMaitre Valvulotome Valvulotome Valvulotome (870.4885, Product Code MGZ) # C. Predicate Device Name: Trade Name: Expandable LeMaitre Valvulotome, (K980723) #### D. Device Description: The OTW ELV is a self-centering and self-sizing catheter used for cutting valves in the saphenous vein while navigating over a guidewire. #### E. Intended Use: The Over-The-Wire Expandable LeMaitre Valvulotome is indicated for cutting saphenous vein valves during in-situ bypass. #### F. Summary of Technological Characteristics The OTW ELV navigates the vein via passage over a guidewire. Upon deployment, the centering hoops expand to the walls of the vessel. As the device is retracted, the hoops keep the blades away from the vessel while allowing them to cut the valves as they are encountered. #### G. Performance Data: Biocompatibility testing was conducted in accordance with AAMI/ANSI/ ISO 10993-1: 2003 Biological Evaluation of Medical Devices- Part 1 Evaluation and Testing. The following biocompatibility tests were conducted with results that demonstrated that the proposed device is non-toxic and non-sensitizing to biological tissues consistent with its intended use: - . ISO MEM Elution Assay - Complement Activation C3a and SC5b-9 Assay . - Platelet & Leukocyte Count . {2}------------------------------------------------ - ASTM Hemolysis Assay ● - Partial Thromboplastin Time ● - ISO Intracutaneous Reactivity . - Murine Local Lymph Node Assay . - ISO Acute Systemic Injection Test in the Mouse t The proposed OTW-ELV was tested through in-vitro bench-top testing as well as in-vitro cadaver testing. The following performance bench testing was completed for the OTW-ELV: 111 284 2.22 - Tensile testing of all bond joints . - Dimensional verification against product design specifications t - Guidewire compatibility . - Trackability through in-vitro models (simulated use) . - Trackability through cadaver models (simulated use) . - Efficacy cutting of in-vitro vein valves (simulated use) . - Efficacy cutting of cadaver vein valves (simulated use) t - Hemostasis . - . Fatigue testing All test results demonstrated that the materials chosen, the manufacturing process, and the design utilized for OTW-Expandable LeMaitre Valvulotome met the established specifications necessary for consistent performance during its intended use as well as substantial equivalence to its predicate. ### H. Conclusion: LeMaitre Vascular has demonstrated that the OTW-Expandable LeMaitre Valvulotomc is substantially equivalent to the predicate device based on its indications for use and fundamental scientific technology. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUL 2 9 2011 LeMaitre Vasculature c/o Andrew Hodgkinson Vice President, Clinical, Regulatory, & Quality Affairs . 63 Second Avenue Burlington, MA 01803 Re: K111884 Trade/Device Name: OTW-Expandable LeMaitre Valvulotome Regulation Number: 21 CFR 870.4885 Regulation Name: External Vein Stripper Regulatory Class: Class II Product Code: MGZ Dated: June 30, 2011 Received: July 1, 2011 Dear Mr. Hodgkinson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 – Mr. Andrew Hodgkinson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram Zuckerman, M Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### SECTION 4: INDICATION FOR USE STATEMENT 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Over-The-Wire Expandable LeMaitre Valvulotome #### Indications for Use: The Over-The-Wire Expandable LeMaitre Valvulotome is indicated for cutting saphenous vein valves during In Situ bypass. Prescription Use______________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE ID NEEDED) Concurrence of CDRH, Office of Device. Evaluation (ODE) LL ovascular Devices 510(k) Number K111884
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...