Paladon ultra is a denture base material for fixed and removable dentures: - Full maxillary and mandibular prostheses . - Implant overdenture .
Device Story
Paladon ultra is a heat-curing, high-impact denture base resin used by dental professionals to fabricate full maxillary/mandibular prostheses and implant overdentures. The system consists of a powder (polymethyl methacrylate) and a liquid (methyl methacrylate) component. The clinician mixes the powder and liquid in a 21g:10ml ratio and processes the material using either an injection procedure or a press-and-pack technique, followed by heat polymerization. The device provides a durable, high-impact resistant base for dental prosthetics. It is intended for clinical use in dental laboratories or offices. The final product serves as the structural foundation for dentures, directly contacting the patient's oral mucosa for long-term use (>30 days).
Clinical Evidence
No clinical trials were conducted. Evidence consists of a clinical evaluation report performed in accordance with MEDDEV 2.7.1 and ISO 14971 risk management processes. The evaluation concluded that the device, as a well-known type of acrylic denture base material, exhibits expected performance and clinical effectiveness. Biocompatibility was documented via testing (cytotoxicity, sensitization, irritation, subacute/subchronic toxicity, genotoxicity) per ISO 10993-1. Physical properties were verified against ISO 20795-1 standards.
Technological Characteristics
Heat-polymerizing high-impact denture base resin. Components: Polymethyl methacrylate powder and methyl methacrylate liquid. Form factor: Powder in HD-PE bottles, liquid in brown glass bottles. Processing: Injection or press-and-pack technique followed by heat curing. Complies with ISO 20795-1:2008 for denture base polymers. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for fabrication of full maxillary and mandibular prostheses and implant overdentures using heat-curing high-impact denture base resin.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
SEP 2 9 2011
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Image /page/0/Picture/1 description: The image contains a sequence of alphanumeric characters. The sequence starts with the letter 'K', followed by two instances of the Roman numeral 'I'. After that, there is the number '832'. The characters are written in a handwritten style, with varying stroke thicknesses.
Image /page/0/Picture/2 description: The image shows the word "Heraeus" in a bold, sans-serif font. The letters are black and appear to have a slightly rough or textured edge. The word is centered and takes up most of the frame.
## 510(k) Summary of Safety and Effectiveness of Paladon ultra
### (1) Submitter's name
Heraeus Kulzer, LLC (for Heraeus Kulzer, GmbH) 300 Heraeus Way South Bend, Indiana 46614
Contact person: Cheryl V. Zimmerman Phone: + 574 299-5444
Date summary was prepared: 2011-05-05
### (2) Name of Device
Trade name - Paladon ultra Classification name - Denture relining, repairing, or rebasing resin (21 CFR 872.3760 Product code EBI) Class - II
## (3) Substantial equilvalence
Paladon ultra, liquid and powders, is substantially equivalent with Palaimpact, liquid and powders, 510(k) No. K043504
Main component of both liquids is methyl methacrylate, of the powders polymethyl methacrylate.
## (4) Description of the device *
Paladon ultra was developed as a heat curing high impact denture base resin similar to the substantially equivalent device Palaimpact.
The product consists of powder in various shades (pink, pink veined, pink live, R50 ыланың жұмысыздың құрылы veined, light pink, light reddish pink, shade 200, dark pink) and liquid. 80 ml or 500 ml of liquid is supplied in brown glass bottles in an outer cardboard box, the 100 g or 1000 g powder is supplied in square HD-PE bottles in an outer cardboard box. 12000 g are available upon request.
{1}------------------------------------------------
## Heraeus
To make a denture, powder and liquid are mixed, in a ratio of 21 g : 10 m!. The material should be polymerised by heating.
The material has a shelf life of five years and fully complies with the requirements of ISO 20795-1:2008 for denture base polymers with improved impact resistance.
#### (5) Intended use
Paladon ultra is a denture base material for fixed and removable dentures:
- Full maxillary and mandibular prostheses .
- Implant overdenture .
## (6) Summaries and Conclusion
#### (a) Technological Characteristics
Paladon ultra was developed as a heat-polymerizing high impact resin with improved impact resistance. In all properties it is comparable with the essentially equivalent heat-polymerizing denture base resin Palaimpact. Paladon ultra is a further development of the earlier product.
### (b) Nonclinical tests and clinical tests/evaluations
(1) Nonclinical tests: In accordance with the Medical Device Directive 93/42/EEC and National European medical device legislation, any medical device, must be evaluated by the legal medical device manufacturer regarding its clinical performance and safety. This includes an evaluation of biocompatibility in accordance with ISO 10993-1. As Paladon ultra is exclusively used for the fabrication denture bases as well as implant-supported prostheses, it can be concluded that there is only contact with the mucosa. The duration of contact of Paladon ultra is > 30 days. According to this classification, (DIN EN ISO 10993-1, Table 1) the following tests must be considered: Cytotoxicity, Sensitization, Irritation / intracutaneous reactivity, Subacute / Subchronic toxicity, Genotoxicity. On the basis of the test results, the
biocompatibility of Paladon ultra in the aforementioned indication was documented in a biocompatibility evaluation report and the benefit / riskrelation has been judged as positive.
- (2) Clinical evaluation
In accordance with the Medical Device Directive 93/24/EEC and National European medical device evaluation, any medical device is requested to be
{2}------------------------------------------------
# Heraeus
evaluated by the legal medical device manufacturer regarding its clinical performance and safety. This includes a clinical evaluation in accordance with MEDDEV 2.7.1, which is intended to critically evaluate the clinical benefits of the medical device in comparison to its potential risks. Therefore, a clinical evaluation is part of the compulsory risk management process according to ISO 14971, and critical findings must further be considered in the current risk management process of the medical device manufacturer responsible for the evaluated device.
On this background, the clinical evaluation was performed in order to comply with the current European medical device legislation, in particular with MEDDEV 2.7.1. This critical evaluation followed the procedures outlined in the corresponding clinical evaluation plan.
Paladon ultra is classified according to annex IX of council directive 93/42/EEC as a class Ila medical device.
Paladon ultra represents as well-known type of acrylic denture base material which has proven to exhibit the expected performance and clinical effectiveness. There is no hint for undesirable effects and potential risks when Paladon ultra is applied according to the instructions for use. Considering the evaluated data and technical results for Paladon ultra, it is concluded that the product will exhibit that claimed clinical and technical performance and that potential undesirable clinical effects and risks seem well controlled and acceptable when weighed against their benefits in dentistry. Therefore, the benefits versus risks ratio was stated to be positive for Paladon ultra, provided that the product is applied in accordance with its intended use according to the manufacturer's information for use. Nevertheless, a risk for irritation or sensitization in susceptible patients or users due to the contact with Paladon ultra cannot be excluded.
The clinical report was carried out in accordance with MEDDEV 2.7.1 and followed the provisions of the corresponding clinical evaluation plan.
(c) Conclusion
The risk potential of the denture base resin Paladon ultra was proven. All properties.of.the product were verified.successfully ... .....................................................................................................................
The biological compatibility of the denture base resin was investigated to evaluate the toxicological risk. A biological evaluation report has confirmed that the product Paladon ultra meets the requirements of ISO 10993 standard. The results were discussed in the biological evaluation report and the benefit / risk relation has been judge as positive.
- -
{3}------------------------------------------------
# Heraeus
The physical properties of Paladon ultra were determined in accordance with ISO 20795-1 for denture base polymers. All properties comply with and exceed the requirements of the standard. This is stated in section (4).
Based on the results of the clinical evaluation report it is concluded that the product can be expected to exhibit the claimed technical performance and that potential undesirable clinical effects and risks seem well controlled and acceptable when weighed against their benefits in dentistry. The risk analysis according to ISO 14971 was carried out for Paladon ultra and showed that the application of the product according to the manufacturer's instruction for use can be considered as safe.
Paladon ultra meets all relevant requirements for denture base resins in accordance with the Medical Device Directive 93/42/EEC and National European medical device legislation. Based on the actual facts Paladon ultra is considered to be effective and safe when used in accordance with the manufacturer's instructions for use.
Webrheim, September 26, 2011
Wehrheim, Septemb
Dr. K. Ruppert
U. Lialich
A. Keishold
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure, with its wings spread and feathers represented by curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MID 20993-0002
Ms. Cheryl Zimmerman Director, Quality Assurance & Regulatory Affairs Heraeus Kulzer. LLC 300 Heracus Way South Bend, Indiana 46614
2 9 2311
Re: K111832
Trade/Device Name: Paladon Ultra Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: II Product Code: EBI Dated: August 24, 2011 Received: August 26, 2011
Dear Ms. Zimmerman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Zimmerman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Reportal/roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \ \
Device Name:
Indications for use:
Paladon ultra is a heat curing high impact denture base resin designed for injection procedure and press and pack technique for:
- . maxillary and mandibular prostheses
- implant over-denture .
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
REBER DDS Yov Dr Siremes
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: K111832
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.