The SYNSTER CERVICAL CAGE is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage. The SYNSTER (ALIF, PLIF, PTLIF and TLIF) CAGE is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at one level or two continuous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
Device Story
Intervertebral body fusion device; provides structural support to spinal column during fusion process. Configurations include cervical (anterior approach) and lumbar (ALIF, PLIF, PTLIF, TLIF approaches). Implanted by surgeons in clinical settings. Device consists of PEEK Optima LT1 cage with titanium markers for radiographic visualization. Used with supplemental fixation and autograft. Benefits patient by stabilizing spinal segment, facilitating fusion, and alleviating discogenic pain. Mechanical performance validated via static/dynamic compression, torsion, subsidence, and expulsion testing.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing included static and dynamic compression, static compression shear, static and dynamic torsion (ASTM F2077-03), subsidence (ASTM F2267-04), and expulsion testing (ASTM Draft Standard F 04.25.02.02).
Technological Characteristics
Materials: PEEK Optima LT1 (ASTM F2026) and titanium markers (ASTM F136). Instruments: stainless steel (ASTM F899). Configurations: Cervical and Lumbar (ALIF, PLIF, PTLIF, TLIF). Mechanical testing per ASTM F2077-03 and F2267-04.
Indications for Use
Indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD). Cervical: C2-T1, one level, radicular symptoms, 6 weeks non-operative treatment. Lumbar: L2-S1, one or two levels, up to Grade I spondylolisthesis, 6 months non-operative treatment. Requires supplemental fixation and autograft.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for BMK Global Medical Company. The logo consists of two parts: a circular emblem on the left and the company name on the right. The emblem features a stylized human figure with outstretched arms inside a circle, surrounded by text. The text "BMK Global Medical Company" is displayed in bold, with the "R" in a circle indicating a registered trademark.
MAR 1 4 2012
# 510(k) Summary
The following 510(k) summary is submitted as required by 21 CFR Part 807.92:
Date Prepared: Feb 21, 2012
### 1. Submission information:
| a) Submitter | |
|--------------|----------------------------------------------------------------------|
| Name | BM KOREA CO., LTD. |
| Address | 325-26 Dangjeong-dong, Gunpo-si, Gyeonggi-do-<br>435-832 South Korea |
| Phone | 82-31-451-9294~5 |
| Fax | 82-31-451-9248 |
| Contact | Ju Yun |
#### b) U.S Agent
| Name | LK Consulting Group |
|---------|-------------------------|
| Address | 951 Starbuck St. Unit J |
| | Fullerton, CA 92833 |
| Phone | 714-869-3080 |
| Fax | 714-409-3357 |
| Contact | Priscilla Chung |
## 2. Device Identification:
Trade Name: SYNSTER® CERVICAL CAGE
SYNSTER® ALIF CAGE SYNSTER® PLIF CAGE SYNSTER® PTLIF CAGE SYNSTER® TLIF CAGE
Common Name: Intervertebral Body Fusion Device
Classification Name: Intervertebral Fusion Device with Bone Graft, Cervical
(21 CFR 880.3080, Product Code ODP)
Intervertebral Fusion Device with Bone Graft, Lumbar
(21 CFR 880.3080, Product Code MAX)
- 3. Identification of the Legally Marketed Devices (Predicate):
Substantial Equivalence for the SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE is based on its similarities in indications for use, design features, operational principle and material composition when compared to the predicate devices cleared under the following:
Eminent Spine Interbody Fusion System (K090064) Genesys Spine Interbody Fusion System (K103034)
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image contains the logo for BMK Global Medical Company. On the left side of the logo is a circular emblem with a stylized human figure inside a globe. To the right of the emblem is the text "BMK" in large, bold letters, with the words "Global Medical Company" underneath in a smaller font.
BAK/Cervical Interbody Fusion System (P980048) Affinity Anterior Cervical Cage System (P000028) RAY Threaded Fusion Cage (P950019) Lumbar I/F Cage (P960025) INTERFIX Threaded Fusion Device (P970015) RABEATM Spinal Implant (K082848) Aesculap PEEK Spinal Implant System (K071983) Stryker Spine AVS® TL PEEK Spacers (K083661) Synthes T-PAL Spacer (K100089) AVS® PL PEEK Spacers (K093704) AnyPlus ALIF, PLIF, TLIF PEEK Lumbar Cage (K100516)
# 4. Device Description:
The SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE will be offered in various device configurations based on surgical approach and patient anatomy, and consist of:
- 1) Cervical Interbody Fusion Device (SYNSTER® CERVICAL CAGE), which may be implanted as a single device via an anterior approach.
- 2) Lumbar Interbody Fusion Device [SYNSTER® (ALIF, PLIF, PTLIF and TLIF) CAGE], which may be implanted
- As a single device via an Anterior or Anterolateral or Lateral (ALIF) approach; ।
- Bi-laterally via a posterior (PLIF) approach; ।
- As a single device via a posterior transforaminal (PTLIF) approach; -
- As a single device via a transforaminal (TLIF) approach. -
The SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE components are made of polyether ether ketone (PEEK Optima LT1) that conforms to ASTM F2026. Additionally, the devices contain titanium makers (ASTM F136) to assist the surgeon with proper placement of the device.
The SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE is implanted by using the (SCCI, SACI, SPCI, SPTCI and STCI) instruments manufactured from stainless steel materials that conform to ASTM F899.
## 5. Indications for Use:
The SYNSTER CERVICAL CAGE is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.
The SYNSTER (ALIF, PLIF, PTLIF and TLIF) CAGE is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at one level or two continuous levels for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as back pain of discogenic
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for BMK Global Medical Company. The logo consists of two parts: a circular emblem on the left and the company name on the right. The emblem features a stylized human figure with arms outstretched inside a globe. The text "BMK Global Medical Company" is written in bold, black letters, with the letters "BMK" being larger than the rest of the text.
origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
## 6. Summary of Technology Characteristics:
The purpose of this premarket notification is to obtain clearance to market the SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE. The SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE is comprised of various device configurations designed to accommodate patient anatomy and provide the surgeon with different surgical approach options.
The SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE Implant components are made of polyether ether ketone (PEEK Optima LT1) that conforms to ASTM F2026. Additionally, the devices contain titanium makers (ASTM F136) to assist the surgeon with proper placement of the device.
The subject system has similar technological characteristics as the predicate devices identified above. Specifically, the following characteristics support this conclusion;
- Intended for use at one level from the C2-C3 disc to the C7-T1 disc for the । treatment degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at the involved level.
- Intended for use at either one level or two contiguous levels from L2-S1 for । the treatment of degenerative disc disease (DDD) with up Grade I spondylolisthesis or retrolisthesis at the involved level(s).
- Substantially equivalent result of non-clinical testing relative static and dynamic testing (per ASTM F2077-03), subsidence testing (per ASTM F2267-04), and expulsion testing (per ASTM Draft Standard F 04.25.02.02)
#### 7. Discussion of Non-clinical Testing
The following non-clinical tests were conducted:
- Static and dynamic compression testing, static compression shear testing, conducted in accordance with ASTM F2077-03
- Static and dynamic torsion testing, conducted in accordance with ASTM -F2077-03
- Subsidence testing, conducted in accordance with ASTM F2267-04 -
- Expulsion testing (per ASTM Draft Standard F 04.25.02.02) -
#### 8. Conclusions
The subject and predicate device(s) share the same intended use, primary implant design and equivalent material of manufacture. Tests performed according to ASTM F2077/F2267 indicate that The SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE meet required mechanical strengths. Some of the predicate devices have a different geometry than The SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE. But the non-clinical mechanical test results demonstrate that any minor differences do not impact performance as compared to the predicates and demonstrate that The SYNSTER® (CERVICAL, ALIF, PLIF, PTLIF and TLIF) CAGE is substantially equivalent to the predicate device.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with outstretched wings, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
BM KOREA CO., LTD. % LK Consulting Group Ms. Priscilla Chung 951 Starbuck Street, Unit J Fullerton, California 92833
Re: K111820
Trade/Device Name: SYNSTER® CERVICAL CAGE SYNSTER® ALIF CAGE SYNSTER® PLIF CAGE SYNSTER® PTLIF CAGE SYNSTER® TLIF CAGE Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, ODP Dated: March 1, 2012 Received: March 8, 2012
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I va may, more of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r touse be advised a determination that your device complies with other requirements of the Act that I Dri has intact a and regulations administered by other Federal agencies. You must
MAR 1 4 2012
{4}------------------------------------------------
# Page 2 -- Ms. Priscilla Chung
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I ou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
K111820 510(k) Number (if known)
Device Name: SYNSTER® CERVICAL CAGE SYNSTER® ALIF CAGE SYNSTER® PLIF CAGE SYNSTER® PTLIF CAGE SYNSTER® TLIF CAGE .
#### Indications for Use:
The SYNSTER CERVICAL CAGE is indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level from the C2-C3 disc to the C7-T1 disc. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.
The SYNSTER (ALIF, PLIF, PTLIF and TLIF) CAGE is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at one level or two continuous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.
Prescription Use >< (Part 21CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
KIII820 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.