← Product Code [HET](/productcode/HET) · K111788

# ENDOEYE HD II (K111788)

_Olympus America, Inc. · HET · Jan 13, 2012 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K111788

## Device Facts

- **Applicant:** Olympus America, Inc.
- **Product Code:** [HET](/productcode/HET.md)
- **Decision Date:** Jan 13, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1720
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

## Device Story

The Olympus EndoEYE HD II is a high-definition digital video laparoscope used for visualization during endoscopic surgery. It captures optical images of the thoracic and abdominal cavities, including female reproductive organs, and transmits them to a video system center for display on a monitor. The device is operated by physicians in clinical settings (e.g., OR). It features a wider field of view, a 'Fog Free' function, increased depth of field, and utilizes multiple CCDs (in specific models) compared to its predecessors. The output provides real-time visualization to assist the surgeon in performing minimally invasive procedures. The device is used alongside ancillary equipment such as light sources, electrosurgical units, and hand instruments. Benefits include improved image quality and operational features that facilitate surgical navigation and decision-making.

## Clinical Evidence

Bench testing only. Performance testing included validation of the 'Fog Free' function, verification of resolution and color correctness, spectrum analysis, and durability testing after multiple sterilization cycles. No clinical data was required or provided.

## Technological Characteristics

High-definition digital video laparoscope. Features: 5.4mm or 10mm distal end diameter, 0° or 30° direction of view, 20-200mm depth of field. Materials identical to predicate. Connectivity: connects to video system center. Sterilization: autoclavable. Standards applied: IEC 60601-1, IEC 60601-1-1, IEC 60601-2-18, IEC 60601-1-2, ISO 14971, DIN EN ISO 17664, DIN EN ISO 17665-1, DIN EN ISO 14161, ISO 11138-3, DIN EN 556-1, DIN EN 285.

## Regulatory Identification

A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.

## Predicate Devices

- HD EndoEYE ([K090980](/device/K090980.md))
- HD EndoEYE Laparo-Thoraco Videoscope OLYMPUS LTF-VH ([K080948](/device/K080948.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K111788

# JAN 1 3 2012

## January 13, 2012

### General Information ﮩ

- Olympus Winter & Ibe GmbH l Applicant Kuehnstrasse 61 * 22045 Hamburg * Germany Establishment Registration No: 9610773
I Official Correspondent Stacy Abbatiello Kluesner, RAC Regulatory Affairs & Quality Assurance Olympus America Inc. 3500 Corporate Parkway, PO Box 610 Center Valley PA 18034-0610 Phone: (484) 896-5405 Facsimile: (484) 896-7128 Email: Stacy.kluesner @olympus.com Establishment Registration No: 2429304

- Olympus Winter & Ibe GmbH Manufacturer Kuehnstrasse 61 * 22045 Hamburg Germany Establishment Registration Number: 9610773
#### 2. Device Identification

- . Device Name Olympus EndoEYE HD II
- . Common Name High Definition Digital Video Laparoscope
- Regulation Number 21 CFR 876.1500 21 CFR 874.4720 21 CFR 884.1720
- Regulation Name Endoscope and Accessories Mediastinoscope and Accessories Gynecologic Laparoscope and Accessories
- . Regulatory Class ll
- Product Code GCJ /NLM/KOG/ NWB EWY NMH/HET
- I Classification Panel Gastroenterology/Urology Ear Nose& Throat Obstetrics/Gynecology

{1}------------------------------------------------

### Legally Marketed Device to which Substantial Equivalence is claimed 3.

The following table shows the subject and predicate device to which we claim substantial equivalence.

| Subject Device (Part of this submission)                                                                                                                                      | Predicate Device                                               | PD's 510(k) No. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|-----------------|
| HD EndoEYE II<br>- WA50040A, 10 mm, 0°, autoclavable<br>- WA50042A, 10 mm, 30°, autoclavable<br>- WA50050A, 5.4 mm, 0°, autoclavable<br>- WA50052A, 5.4 mm, 30°, autoclavable | HD EndoEYE<br>WA50011A, WA50013A, WA50013L, WA50013T, WA50015L | K090980         |
|                                                                                                                                                                               | HD EndoEYE Laparo-Thoraco<br>Videoscope OLYMPUS LTF-VH         | K080948         |
| Table 1. Subject & Predicate Devices                                                                                                                                          |                                                                |                 |

### Device Description 4 ...

The ENDOEYE HD II - High Definition Digital Video Laparoscope is a video endoscope used for endoscopy and endoscopic surgery within the abdominal cavities, which is basically identical to the predicate devices for the same application areas.

#### Indications for Use ഗ്

This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surqery within the thoracic and abdominal cavities including the female reproductive organs.

## Comparison of Technological Characteristics 6.

The subject EndoEYE HD II is nearly identical to the predicate HD EndoEYE device. It has identical spefications to the predicate device except for; (1) a wider field of view, (2) incorpotion of a Fog Free Function, (3) a greater range in the depth of field and (4) multible CCDs.

| Specifications | Predicate Device   Predicate Device                                                                                                                                            |  |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
|                | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |  |

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the word "OLYMPUS" in bold, black letters. The font is sans-serif and the letters are tightly spaced. The word is centered and takes up most of the frame.

|                                 |                                                          | K090980                                                      | K080948                                    |
|---------------------------------|----------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------|
| Device name                     | EndoEYE HD II                                            | HD EndoEYE                                                   | HD EndoEye<br>Laparo-Thoraco<br>Videoscope |
| 510(k) number                   |                                                          | K955456                                                      | K080948                                    |
| Type                            | WA50040A,<br>WA50042A,<br>WA50050A,<br>WA50052A          | WA50011A,<br>WA50013A,<br>WA50013L,<br>WA50013T,<br>WA50015L | LTF-VH                                     |
| Field of View                   | 90°                                                      | 80°                                                          | 90°                                        |
| Depth of field                  | 20 to 200 mm                                             | 10 to 120 mm                                                 | 15-100 mm                                  |
| Direction of View               | 0°, 30°                                                  | 0°, 30°, 45°                                                 | 0°                                         |
| Outer Diameter<br>of Distal End | 5.4 mm, 10mm                                             | 10 mm                                                        | 10 mm                                      |
| Optical System                  | Color                                                    | Color                                                        | Color                                      |
| Working Length                  | 325 or 300 mm                                            | 250 - 390 mm                                                 | 370 mm                                     |
| Switch for NBI<br>function      | Provided                                                 | Provided                                                     | Provided                                   |
| Number of CCD<br>chips          | 2 (WA50040A &<br>WA50042A)<br>1 (WA50050A &<br>WA50052A) | 1                                                            | 1                                          |

#### 7. Summary of non-clinical testing

Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verifications tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.

Reprocessing validation was carried out in accordance with "Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: FDA Reviewer Guidance - April 1996."

The subject device has identical materials as the predicate device.

The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The device software is considered a "Minor Level of Concern."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "OLYMPUS" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The word is in black color and the background is white. 

The following standards have been applied to the subject EndoEYE HD II devices:

- · IEC 60601-1
- · IEC 60601-1-1
- · IEC 60601-2-18
- · IEC 60601-1-2
- · ISO 14971
- · DIN EN ISO 17664
- · DIN EN ISO 17665-1
- · DIN EN ISO 14161
- · ISO 11138, part 3
- · DIN EN 556-1, Part 1
- · DIN EN 285

Performance testing conducted included validation of the Foq Free function, verification testing for resolution and color correctness, spectrum analysis and testing to confirm the durability of the device after multiple sterilization cycles.

### 8. Conclusion

When compared to the predicate devices, the ENDOEYE HD II does not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device and therefore is substantially equivalent to the idenfied predicate devices.

{4}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three tail feathers, representing the department's commitment to health, services, and people. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Olympus Winter & Ibe GmbH % Ms. Stacy Abbatiello Kluesner, M.S., RAC Project Manager Olympus America Inc. 3500 Corporate Parkway CENTER VALLEY PA 18034

JAN 1 3 2012

Re: K111788

Trade/Device Name: ENDOEYE HD II Type WA50040A, WA50042A, WA50050A, WA50052A

Regulation Number: 21 CFR§ 884.1720 Regulation Name: Gynecologic laparoscope and accessories Regulatory Class: II Product Code: HET Dated: January 10, 2012 Received: January 11, 2012

Dear Ms. Kluesner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{5}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,
. Benjamin K. twka

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known): ______ K111788

ENDOEYE HD II Device Name: Type WA50040A, WA50042A, WA50050A, WA50052A

Indications For Use:

This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concerrence of CDRH, Office of Device Evaluation (ODE)

Ma

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K111788

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/device/K111788](https://fda.innolitics.com/device/K111788)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
