← Product Code [KNW](/productcode/KNW) · K111765

# FULL CORE BIOPSY SYSTEM (K111765)

_Promex Technologies, LLC · KNW · Sep 6, 2011 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K111765

## Device Facts

- **Applicant:** Promex Technologies, LLC
- **Product Code:** [KNW](/productcode/KNW.md)
- **Decision Date:** Sep 6, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Full Core Biopsy System is intended to be used for soft tissue and tumor biopsy of such organs as the liver, spleen, kidney, prostate, lung, breast, and lymph nodes. When used for breast biopsy, the product is for diagnosis only. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal of using standard surgical procedures.

## Device Story

System comprises automatic spring-powered biopsy device and coaxial introducer; used for soft tissue/tumor biopsy. Device features Franseen-style end-cut cannula; introducer features trocar tip. Operation: user positions introducer; advances device; pulls plunger to ready; fires via plunger depression to extend cannula and capture tissue. Tissue retrieved by removing device from introducer and advancing plunger to deliver sample from distal end. Used in clinical settings by physicians. Benefits: provides core tissue samples for histological diagnosis. System provided sterile for single-patient use.

## Clinical Evidence

Bench testing only. Side-by-side comparison against SABD and BioPince predicates. Testing evaluated penetration force, stroke length, and tissue sample retrieval quality. Results confirmed performance is substantially equivalent to predicate devices.

## Technological Characteristics

Spring-powered biopsy device with coaxial introducer. Materials, chemical composition, and manufacturing identical to SABD predicate. Features Franseen-style end-cut cannula and trocar-tip introducer. Available in 16, 18, 20 gauge; 10cm-25cm lengths. Gamma sterilization (25-50kGy) per ISO 11137-2:2006. Mechanical operation via manual plunger.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- SABD™ Automated Core Biopsy Device ([K011270](/device/K011270.md))
- Vibronics Auto Core Biopsy Device (BioPince®) ([K904987](/device/K904987.md))

## Reference Devices

- Coaxial Introducer ([K954265](/device/K954265.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "K111 765 SEP - 6 2011". The text appears to be handwritten, with the date being printed. The text is written in black ink on a white background. The text is slightly slanted to the right.

technoloc les uc

SECTION 5

ィレー

«ﭘ

#### 510(k) SUMMARY:

Date: 21-June-2011

#### 510(k) Owner:

Promex Technologies, LLC 3049 Hudson Street Franklin, Indiana 46131 P: 317-736-0128 F: 317-736-0793

#### Contact:

Allison Scott, Consultant Anson Group P: 317-569-9500 ext. 106 ascott@ansongroup.com

#### Names:

Trade Name - Full Core Biopsy System Common Name – Kit, Needle Biopsy Classification Name - Gastroenterology-Urology Biopsy Instrument 21 CFR 876.1075, Product Code FCG

#### Legally Marketed Devices of Equivalence:

K011270 - SABD™ Automated Core Biopsy Device (Promex Technologies, LLC) (the "SABD"" Predicate") K904987 - Vibronics Auto Core Biopsy Device (also known as BioPince® Full Core Biopsy Instrument - currently sold by Medical Device Technologies, Inc.) (the "BioPince" Predicate)

#### Device Description:

The Full Core Biopsy System (the "System") includes an automatic spring powered Biopsy Device (the "Device") and a Coaxial Introducer (K954265, Promex Technologies, LLC) ("the 

{1}------------------------------------------------

Introducer"). As described more fully in the scientific concepts section below, the Device obtains and delivers a core tissue sample. The end-cut design of the Device allows the entire inner diameter of the cutting cannula to be open to capture tissue.

765

The System (2 components) is comprised of;

- 1. The Device
ン

- 2. The Introducer
and are provided sterile in a pouch and intended for single patient use only.

K 111

The System will be provided in 16, 18, and 20 gage and the working lengths range from 10cm to 25cm. The Device includes a Franseen-style cutting geometry on the end of the cannula and the Introducer includes a standard trocar tip.

#### Indications for Use:

The System is intended to be used for soft tissue and tumor biopsy of such organs as the liver, spleen, kidney, prostate, lung, breast, and lymph nodes. When used for breast biopsy, the product is for diagnosis only. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal of using standard surgical procedures.

#### Scientific Concepts that Form the Basis for the Device:

The Device has an outer Cutting Cannula having a sharpened tip and an inner stylet. The Cutting Cannula and the inner stylet are both operated manually by the user by a Plunger. In operation, the Introducer is positioned within tissue in a standard fashion. As with the SABD" Predicate device, the Device is first placed in a Ready condition (prior to introduction to the patient) by pulling back on a Plunger, and then the Device is advanced within the Introducer. The Plunger is first advanced to the first stop. To fire (advance the outer cutting cannula to obtain a biopsy tissue sample), the Plunger is fully depressed. When the Device is fired, the outer Cutting Cannula extends forward and cuts tissue. A tissue specimen is retained within the Cutting Cannula. As with both Predicate devices, the Device is then removed from the Introducer to retrieve the tissue specimen from the Device. From the user perspective, the tissue specimen is delivered from the Device the same as the SABD™ Predicate, by pulling back on the Plunger and then advancing the Plunger forward to the first stop. In a device such as the SABD" Predicate, this action results in exposure of the sampling notch so that the specimen can be manually removed. With the Device, this action results in the delivery of the tissue specimen from the distal end of the outer Cutting Cannula of the Device, as occurs with the BioPince® Predicate. If it is desired to obtain another tissue specimen, the Device is again placed in the Ready condition by pulling back on the Plunger. From the user perspective, this is identical to how the SABD" Predicate is used.

#### Predicate Device Comparison:

Full Core Biopsy System 510(k) Section 5 - Summary Section 5/Page 2 of 4

{2}------------------------------------------------

K 11 765

Page 3 of 4

### The SABD™ Predicate

్రా

ェ・ン

The Full Core Biopsy System has the same intended use as the SABD™ Predicate. All materials, manufacturing methods, chemical composition of components, and energy source are the same as the predicate SABD" Predicate. The design is the same as the SABD" Predicate, with the exception of the cutting feature.

The System uses an end-cut technology on the Device rather than the side-cut technology found in the SABD" Predicate. The difference in the cutting feature poses no concerns regarding safety and effectiveness, as it is equivalent to the cutting feature of the Biopince® Predicate.

The SABD" Predicate has a fifty-nine month shelf life. The subject device will have a thirty-six month shelf life. The shorter shelf-life for the subject device is due to marketing purposes. The difference in shelf-life poses no concerns regarding safety and effectiveness.

#### The Biopince® Predicate

The Full Core Biopsy System has equivalent intended use as the Biopince® Predicate". The end-cut technology (mechanism of cutting tissue) of the Device and delivery of tissue sample (as described above) is equivalent to that of the Biopince® Predicate. The Device has a fixed stroke, which is a simpler construction from the Biopince® Predicate, which has a variable stroke.

#### Performance Testing:

#### Bench Testing:

The Full Core Biopsy System was tested in a side-by-side comparison against the SABD™ Predicate and the Biopince® Predicate.

Penetration testing confirms that the penetration force of the System falls between the penetration forces of the SABD™ Predicate and the Biopince® Predicate.

Testing of stroke lengths confirms that the stroke length of the Device is the stroke length of the SABD" Predicate and has a stroke length that is equivalent to one of the settings of the BioPince® Predicate.

The Full Core Biopsy System was tested in a side-by-side comparison against the SABD™ Predicate which was used in conjunction with a Coaxial Introducer. Testing against the Biopince® Predicate with Coaxial Introducer was also conducted. Test results confirmed that the Full Core Biopsy System would retrieve a tissue sample that was substantially equivalent to the tissue sample retrieved by the predicate devices.

#### Mechanical Durability Testing:

The Device utilizes the same components, springs, manufacturing processes, and tests as used in the assembly of the SABD™ Predicate. Mechanical Durability testing of the Device will be a part

{3}------------------------------------------------

K 1 1 765

Page 4 of 4

of the normal inspection process, which is the same as the inspection process for the 510k Owner's SABD™ Predicate.

## Sterilization:

The sterilization process will be identical for the System as the SABD™ Predicate. The SABD™ Predicate and the Introducer have been validated for gamma sterilization at a minimum level of 25kGy and a maximum of 50kGy. The process is validated and monitored to a 10° SAL per ISO11137-2:2006, method VDmax25

Aging Tests:
The SABD" Predicate has been evaluated for shelf-life stability using accelerated aging as well as real-time shelf exposure under normal storage conditions. All materials, manufacturing methods, and packaging are the same for the subject Device and the SABD™ Predicate; therefore no further testing of the System is necessary.

# Conclusion:

The Full Core Biopsy System is safe and effective and performs substantially equivalent to the Predicates.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MID 20993-0002

Promex Technologies. LLC % Anson Group Ms. Allison Scott 9001 Wesleyan Road Indianapolis, Indiana 46268

SEP - 6 2011

Re: K111765

Trade/Device Name: Full Core Biopsy System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW, FCG Dated: June 22, 2011 Received: June 23, 2011

Dear Ms. Scott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{5}------------------------------------------------

Page 2 - Ms. Allison Scott

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (2) CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Eric Keith

Sor Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known): K111765

Device Name: Full Core Biopsy System

Indications for Use:

The Full Core Biopsy System is intended to be used for soft tissue and tumor biopsy of such organs as the liver, spleen, kidney, prostate, lung, breast, and lymph nodes. When used for breast biopsy, the product is for diagnosis only. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal of using standard surgical procedures.

Prescription Use Yes (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use No (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

for mkn

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K11765

---

**Source:** [https://fda.innolitics.com/device/K111765](https://fda.innolitics.com/device/K111765)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
