MDRIVE

K111742 · Embla Systems · MNR · Aug 31, 2011 · Anesthesiology

Device Facts

Record IDK111742
Device NameMDRIVE
ApplicantEmbla Systems
Product CodeMNR · Anesthesiology
Decision DateAug 31, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.2375
Device ClassClass 2

Indications for Use

The N7000 system (with MDrive) is intended for clinical sleep applications, with recording and/or amplification capability of numerous types of physiological signals used by a physician or trained technician for the acquisition of respiratory, electroencephalogram (EEG), electrocardiogram (ECG, EKG), positional, user triggered event and Oximetry parameters from a patient connected Headbox and/or Patient Unit during sleep related studies. The general intended environments are hospitals, institutions, sleep centers, and sleep clinics, but the device should be capable of functioning in any environment where patients can sleep reasonably comfortably. The users are the general public, trained physicians, trained sleep technicians (RPGST) or people working under the supervision of one of these professionals. The user may or may not possess knowledge of the physiological signals or test criteria. The N7000 system does not provide any alarms and is not intended to be a monitor. A trained sleep technologist (polysomnographer) and a physician would typically review and analyze the N7000 data when it is communicated and presented on a PC using a separate application software program.

Device Story

Embla N7000 with MDrive is a sleep study data acquisition system. It collects physiological signals (respiratory, EEG, ECG, positional, oximetry) via electrodes and sensors attached to a patient. The system comprises a Headbox, Patient Unit, and the MDrive (an upgraded Communication Unit). The MDrive amplifies signals and transmits them to a host PC via wired or wireless connection. It features a local keypad/display for electrode impedance checking and biocalibration, and an SD card slot for local data recording during PC communication interruptions. Data is reviewed by a physician or sleep technologist on a separate PC application. The device facilitates sleep study recording; it does not provide alarms or real-time monitoring. It benefits patients by enabling comprehensive sleep data collection for clinical analysis.

Clinical Evidence

Bench testing only. Safety and electromagnetic compatibility testing performed per IEC 60601-1/UL60601-1 and IEC 60601-1-2. Performance verification and validation of specifications completed with acceptable results. No clinical data provided.

Technological Characteristics

System components: Communication Unit (MDrive), Headbox, Patient Unit. Connectivity: Wired and wireless to host PC. Local interface: Keypad and display for impedance checking/biocalibration. Storage: SD card for backup recording. Standards: IEC 60601-1, UL 60601-1, IEC 60601-1-2. Signal processing: Amplification and transmission of physiological signals.

Indications for Use

Indicated for use by physicians or trained technicians to acquire EEG and polysomnography (PSG) signals for transmission to a PC during neurophysiologic or sleep examinations in hospitals, sleep centers, or clinics. Not for patient monitoring or diagnosis.

Regulatory Classification

Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # K111742 # ATTACHMENT 'C' AUG 3 1 2011 ## 510(K) SUMMARY #### SUBMITTER INFORMATION · | A. Company Name: | Embla Systems | |-----------------------|------------------------------------------------------------------------------------------| | B. Company Address: | 9351 Grant St.., Suite 600<br>Thornton, Colorado 80229 | | C. Company Phone: | (303) 790-1801 | | D. Company Facsimile: | (303) 790-1810 | | E. Company Contact: | Robert G. Schueppert<br>Director, Regulatory Affairs<br>E-mail: bob.schueppert@embla.com | #### PREPARATION DATE June 16, 2011 #### DEVICE IDENTIFICATION | A. Device Trade Name: | Embla N7000 with MDrive | |-------------------------|-----------------------------| | B. Device Common Name: | Embla N7000 | | C. Classification Name: | Breathing Frequency Monitor | | D. Regulation Number: | 21 CFR 868.2375 | | E. Product Code: | MNR | | F. Device Class: | Class II | #### PREDICATE DEVICES A. Trade Name: Embla N7000, 510(k) Number: K024322 {2}------------------------------------------------ #### DEVICE DESCRIPTION The original cleared N7000 system consists of 3 separate components; a mains connected Communication Unit (CU) for signal amplification and transmission to a host PC, a Headbox that is cable connected to the CU, and a Patient unit that is cable connected to the Headbox. Patient electrodes are connected to the Headbox and the Patient Unit with several electrodes and sensors non-invasively attached to a patient during sleep. The MDrive device is intended as an alternate upgrade for the current CU component and provides all original functionality and system connectivity that exists with the CU. The major additions that the MDrive provides over the existing CU component in the N7000 svstem are: - A wireless connection option to the host PC, in addition to a wired connection, . 1. - a local keypad and display for the user to perform electrode impedance checking 2. and biocalibration locally rather that from a remote PC, and - study recording capability using an SD card in the event of PC host 3. communication interruption. The SD card study can then be uploaded to the host PC. #### INTENDED USE The N7000 system (with MDrive) is intended for clinical sleep applications, with recording and/or amplification capability of numerous types of physiological signals used by a physician or trained technician for the acquisition of respiratory, electroencephalogram (EEG), electrocardiogram (ECG, EKG), positional, user triggered event and Oximetry parameters from a patient connected Headbox and/or Patient Unit during sleep related studies. The general intended environments are hospitals, institutions, sleep centers, and sleep clinics, but the device should be capable of functioning in any environment where patients can sleep reasonably comfortably. The users are the general public, trained physicians, trained sleep technicians (RPGST) or people working under the supervision of one of these professionals. The user may or may not possess knowledge of the physiological signals or test criteria. The N7000 system does not provide any alarms and is not intended to be a monitor. A trained sleep technologist (polysomnographer) and a physician would typically review and analyze the N7000 data when it is communicated and presented on a PC using a separate application software program. {3}------------------------------------------------ #### INDICATIONS FOR USE The Embla N7000 is intended for use by a physician or trained technician for the acquisition of EEG and polysomnography (PSG) signals and transmission of these signals to a PC during neurophysiologic or sleep examinations. The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments. The use of the Embla N7000 system does not involve any patient monitoring or diagnosis. #### COMPARISON TO PREDICATE DEVICES The Embla MDrive is intended as an upgrade and direct replacement for the Communications Unit in the N7000 predicate device. It is substantially equivalent to the Communications Unit in the following technological ways to the intended use and application in the N7000 system; - Indications for Use . - Target population ● - Basic design and architecture - . Where used - Standards met . - Signals recorded and transmitted . #### TESTING AND PERFORMANCE DATA Safety tests on the N7000 with the MDrive have been performed to verify compliance with IEC 60601-1/UL60601-1 and any applicable particular standards in this family of international safety standards to ensure that there are no potential hazards on patients, operators, or the surroundings. Electromagnetic Compatibility tests according to IEC 60601-1-2 have been performed to ensure that no intolerable electro-magnetic disturbances are introduced. Immunity tests to IEC 60601-1-2 have been performed to ensure that the device operates satisfactorily in an electromagnetic environment. The internal testing, verification in various design phases, and validation of performance specifications have been completed with acceptable results. #### BIOCOMPATIBILITY, STERILIZATION, PACKAGING, AND BENCH TESTING Biocompatibility and Sterilization do not apply. {4}------------------------------------------------ #### CONCLUSION The signals received, amplified, recorded and transmitted by the MDrive in the Embla N7000 system were compared to signals received, amplified and transmitted by the Communications Unit in the N7000 predicate system. The result of the comparison is that all signals received, amplified, recorded and transmitted by the MDrive were identical to signals received, amplified and transmitted by the predicate devices. It is therefore the conclusion of Embla Systems that the Embla MDrive is substantially equivalent to devices already on the market (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness. {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MI) 20993-0002 Mr. Robert G. Schueppert Director, Regulatory Affairs Embla Systems 9351 Grant Street, Suite 600 Thornton, Colorado 80229 AUG 3 1 2011 Re: K111742 Trade/Device Name: Embla N7000 with MDrive Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: July 29, 2011 Received: August 2, 2011 Dear Mr. Schueppert: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Mr. Schueppert Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies, You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. for Anthony D. Watson, B.S., M.S.. M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {7}------------------------------------------------ ## Special 510(k) Notification for Embla N7000 #### SECTION 9 ## STATEMENT OF INDICATIONS OF USE 510(k) Number (if known): Unknown Device Name: Embla N7000 with MDrive Indications For Use: The Embla N7000 is intended for use by a physician or trained technician for the acquisition of EEG and polysomnography (PSG) signals and transmission of these signals to a PC during neurophysiologic or sleep examinations. The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments. The use of the Embla N7000 system does not involve any patient monitoring or diagnosis. X = Prescription Use AND/OR 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (Part (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schultheis (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices KILTY2 510(k) Number:
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...