← Product Code [LLZ](/productcode/LLZ) · K111694

# ASTRA (K111694)

_Candelis, Inc. · LLZ · Sep 15, 2011 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K111694

## Device Facts

- **Applicant:** Candelis, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Sep 15, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

ASTRA is software imagement intended to receive, process, review, display, print and archive medical images and data from imaging modalities (e.g., CR and DR). Images and data can be stored, communicated, and displayed within the system or across computer systems. ASTRA is comprised with three configurations depending upon the requirements of the user and desired options: ASTRA Plus , ASTRA Lite, and ASTRA Mobile. ASTRA runs on a PC workstation, iPad, or iPhone and may be interfaced with verified and validated image acquisition devices from Candelis or other PACS systems. Diagnosis is not performed by the software but by Radiologists, Clinicians or referring Physicians. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants. ASTRA Plus is used to: - share reports and studies with other ASTRA peers . - review reports and studies . - . download and save reports - send reports to local EMR, EHR, RIS, HIS or PACS systems (HL7 send) . - . route studies to PACS, Workstations, or other ASTRA peers ASTRA Lite is used to: - share reports and studies with other ASTRA peers . - . review reports and studies - download and save reports . - send reports to local EMR, EHR, RIS, HIS or PACS systems (HL7 send) . ASTRA Mobile is used to: - share reports with other ASTRA peers . - review reports . - . download and save reports - . send reports to local EMR, EHR, RIS, HIS or PACS systems (HL7 send) Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA.

## Device Story

ASTRA is a web-enabled software application for medical image management; functions as a Picture Archiving and Communications System (PACS). Inputs include medical images and data from imaging modalities (CR, DR). Software processes, displays, and archives these inputs; enables communication across LAN/WAN/VPN. Used by radiologists, clinicians, and physicians on PC workstations, iPads, or iPhones. Does not perform diagnosis; provides tools for human interpretation. Facilitates report/study sharing, routing to EMR/EHR/RIS/HIS/PACS, and downloading. Benefits include improved access to diagnostic data and workflow efficiency. Operates via DICOM 3.0 standards for interoperability.

## Clinical Evidence

Bench testing only. The system configuration was assessed and tested at the factory; passed all in-house testing criteria without significant failures. Performance verified against functional requirements specified in the SRS and User Manual. No clinical data presented.

## Technological Characteristics

Software-based image management system. Conforms to DICOM 3.0 standard for interoperability. Web-enabled; supports PC, iPad, and iPhone platforms via VPN. Does not contact patients or control life-sustaining equipment. Hazard analysis classified as 'Moderate' level of concern.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- ImageGrid Radiology Viewer ([K092949](/device/K092949.md))
- Mobile MIM ([K103785](/device/K103785.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "510(k) Summary of Safety". The text is in bold and is centered on the image. The text is likely a title or heading for a document or presentation.

K111694

SEP 1 5 2011

This 510(k) summary of safety information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

### Date Prepared:

May 17, 2011

### Submitters Information: 21 CFR 807.92(a)(1)

Mr. Vincent Cipolla 18821 Bardeen Ave. Irvine, CA 92612, USA Tel: 949.852.1000 Fax: 949.752.7317 Email: vincent.cipolla@candelis.com

### Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)

Trade Name: ASTRA Common name: Picture Archiving Communications System . Classification Name: 892.2050 Image Processing System Product Code: LLZ

#### Predicate Device(s): 21 CFR 807.92(a)(3)

| 510(k) Number               | K092949                                   | K103785                                   |
|-----------------------------|-------------------------------------------|-------------------------------------------|
| Manufacturer:               | Candelis, Inc.                            | MIM Software, Inc                         |
| Device Name:                | ImageGrid Radiology<br>Viewer             | Mobile MIM                                |
| Decision Date:              | 10/08/2009                                | 2/4/2011                                  |
| Product Code:               | LLZ                                       | LLZ                                       |
| Device Classification Name: | System, Image Processing,<br>Radiological | System, Image Processing,<br>Radiological |
| Regulation Number:          | Class II - 892.2050                       | Class II - 892.2050                       |
| Reviewed by Third Party:    | Yes                                       | No                                        |

{1}------------------------------------------------

### Device Description: 21 CFR 807.92(a)(4)

ASTRA is a web-enabled software application that provides image processing and viewing tools and access to studies and reports from a Local Area Network. Wide Area Network, remote workstation, personal computer, or an iPhone, or iPad via a Virtual Private Network connection. Diagnosis is not performed by the software but by Radiologists, Clinicians or referring Physicians. The software application conforms to the DICOM 3.0 standard to allow interoperability with other DICOM compliant systems.

### Indications for Use: 21 CFR 807.92(a)(5)

ASTRA is software image management intended to receive, process, review, display, print and archive medical images and data from imaging modalities (e.g., CR and DR), Images and data can be stored, communicated, and displayed within the system or across computer systems. ASTRA is comprised with three configurations depending upon the requirements of the user and desired options: ASTRA Plus , ASTRA Lite, and ASTRA Mobile. ASTRA runs on a PC workstation, iPad, or iPhone and may be interfaced with verified and validated image acquisition devices from Candelis or other PACS systems. Diagnosis is not performed by the software but by Radiologists, Clinicians or referring Physicians. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants.

ASTRA Plus is used to:

- share reports and studies with other ASTRA peers .
- review reports and studies .
- . download and save reports
- send reports to local EMR, EHR, RIS, HIS or PACS systems (HL7 send) .
- . route studies to PACS, Workstations, or other ASTRA peers

ASTRA Lite is used to:

- share reports and studies with other ASTRA peers .
- . review reports and studies
- download and save reports .
- send reports to local EMR, EHR, RIS, HIS or PACS systems (HL7 send) .

ASTRA Mobile is used to:

- share reports with other ASTRA peers .
- review reports .
- . download and save reports
- . send reports to local EMR, EHR, RIS, HIS or PACS systems (HL7 send)

Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA.

{2}------------------------------------------------

# 510(k) Summary of Safety

## Technological Characteristics: 21 CFR 807.92(a)(6)

ASTRA is a software product that handles digital medical images and reports. The device does not contact the patient, nor does it control any life sustaining devices. A physician. providing ample opportunity for competent human intervention interprets images and information being displayed and printed.

### Testing:

The complete system configuration has been assessed and tested at the factory and the device has passed all in-house testing criteria without significant failures. The data presented in the submission demonstrates that the ASTRA device performs all required actions according to the functional requirements specified in the SRS and User Manual with no errors that had an impact on safety or efficacy.

### Conclusion: 21 CFR 807.92(b)(1)

The 510 (k) Pre-Market Notification for Candelis ASTRA contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device. ASTRA has been and will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the "Level of Concern for potential hazards has been classified as "Moderate".

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features three diagonal lines representing the branches of government. The caduceus is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" in a circular arrangement.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Vincent Cipolla Manager Quality/Regulatory Compliance Candelis, Inc. 18821 Bardeen Avenue IRVINE CA 92612

Re: K111694

SEP 15 2015

Trade/Device Name: ASTRA Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 31. 2011 Received: August 31. 2011

Dear Mr. Cipolla:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

ff you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Mary Patel

Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a black and white photograph. The image is mostly white with a few black spots scattered throughout. The black spots are small and irregular in shape. 

Image /page/5/Picture/1 description: The image shows the word "CANDELIS" in a stylized font. The letters are bold and have a slightly distressed appearance. There is a curved line above the word, possibly indicating an arc or a design element.

## Indications for Use Statement

510(k) Number (if known):

Device Name: ASTRA

Indications for Use:

ASTRA is software imagement intended to receive, process, review, display, print and archive medical images and data from imaging modalities (e.g., CR and DR). Images and data can be stored, communicated, and displayed within the system or across computer systems. ASTRA is comprised with three configurations depending upon the requirements of the user and desired options: ASTRA Plus , ASTRA Lite, and ASTRA Mobile. ASTRA runs on a PC workstation, iPad, or iPhone and may be interfaced with verified and validated image acquisition devices from Candelis or other PACS systems. Diagnosis is not performed by the software but by Radiologists, Clinicians or referring Physicians. Typical users of this system are trained professionals, e.g. physicians, radiologists, nurses, medical technicians, and assistants.

ASTRA Plus is used to:

- share reports and studies with other ASTRA peers �
- review reports and studies .
- download and save reports .
- send reports to local EMR, EHR, RIS, HIS or PACS systems (HL7 send) .
- route studies to PACS, Workstations, or other ASTRA peers .

ASTRA Lite is used to:

- . share reports and studies with other ASTRA peers
- review reports and studies .
- download and save reports
- send reports to local EMR, EHR, RIS, HIS or PACS systems (HL7 send)

ASTRA Mobile is used to:

- share reports with other ASTRA peers .
- review reports .
- download and save reports .
- send reports to local EMR, EHR, RIS, HIS or PACS systems (HL7 send)

Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA.

Prescription Use × Over-The-Counter Use (21 AND/OR (Part 21 CFR 801 Subpart D) CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muri S. Pastel

510K K11694

---

**Source:** [https://fda.innolitics.com/device/K111694](https://fda.innolitics.com/device/K111694)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
