SLIC SCREW REPAIR SYSTEM

K111608 · Acumed, LLC · HWC · Aug 29, 2011 · Orthopedic

Device Facts

Record IDK111608
Device NameSLIC SCREW REPAIR SYSTEM
ApplicantAcumed, LLC
Product CodeHWC · Orthopedic
Decision DateAug 29, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The SLIC Screw Repair System is intended to provide fixation and anatomically reduce two bones or bone portions. Specifically, these indications include scapholunate ligament repair, scapholunate reduction, lunotriquetral ligament repair, lunotriquetral reduction, and carpal instability.

Device Story

SLIC Screw Repair System is a headless, cannulated two-piece screw assembly used for bone fixation and reduction. Device consists of distal and proximal screw components. Used by surgeons in clinical settings to anatomically reduce two bones or bone portions. Modifications include material, interconnection geometry, and size variations. Device provides mechanical stabilization of bone segments to facilitate repair of ligaments and reduction of carpal instability.

Clinical Evidence

Bench testing only. Testing included insertion, pull-apart, and pull-out strength testing to verify performance of modified components against design control acceptance criteria.

Technological Characteristics

Headless, cannulated two-piece screw assembly. Material: 22-13-5 stainless steel conforming to ASTM F1314. Mechanical fixation device.

Indications for Use

Indicated for patients requiring bone fixation and reduction of two bones or bone portions, specifically for scapholunate ligament repair, scapholunate reduction, lunotriquetral ligament repair, lunotriquetral reduction, and carpal instability.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KIII 608 (112) AUG 2 9 2011 #### 5. 510(k) Summary | Device Trade Name: | SLIC Screw Repair System | |--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | Acumed, LLC<br>5885 NW Cornelius Pass Road<br>Hillsboro, OR 97124 | | Contact: | Ms. Lino Tsai<br>Regulatory Specialist<br>Phone: 503.627.9957 | | Prepared by: | Musculoskeletal Clinical Regulatory Advisers, LLC<br>1331 H Street NW, 12th Floor<br>Washington, DC 20005<br>Phone: (202) 552-5800<br>Fax: (202) 552-5798 | | Date Prepared: | June 8, 2011 | | Classification: | 21 CFR 888.3040, Smooth or threaded metallic bone<br>fixation fastener | | Class: | II | | Product Code: | HWC | ### Indications For Use: The SLIC Screw Repair System is intended to provide fixation and anatomically reduce two bones or bone portions. Specifically, these indications include scapholunate ligament repair, scapholunate reduction, lunotriquetral ligament repair, lunotriquetral reduction, and carpal instability. ### Device Description: The SLIC Screw Repair System consists of a headless, cannulated two-piece screw assembly. The two-piece design incorporates the use of distal and proximal screw components to anatomically reduce two bones. The purpose of this Special 510(k) is to modify the material, interconnection geometry and available sizes of the SLIC Screw. All components are made of 22-13-5 stainless steel conforming to ASTM F1314. These modifications do not affect the intended use of the device or alter the fundamental scientific technology of the device. {1}------------------------------------------------ ## Predicate Device: The modified SLIC Screw Repair System is substantially equivalent to the predicate ARC Surgical Bone Reduction System (K063244) with respect to indications, design, function, and materials. # Preclinical Testing: The new components were subjected to insertion testing, pull-apart testing and pull-out strength testing. The results demonstrate that the acceptance criteria defined in the Design Control Activities Summary were met. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper half of the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Acumed, LLC % Ms Lino Tsai Regulatory Specialist 5885 North West Cornelius Pass Road Hillsboro, Oregon 97124 AUG 2 9 2011 Re: K111608 Trade/Device Name: SLIC Screw Repair System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bond fixation fastener Regulatory Class: II Product Code: HWC Dated: August 5, 2011 Received: August 8, 2011 Dear Ms. Tsai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be . found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Ms Lino Tsai CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm. Sincerely yours. Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use 4. 510(k) Number (if known): _ y . Device Name: SLIC Screw Repair System The SLIC Screw Repair System is intended to provide fixation and anatomically reduce two bones or bone portions. Specifically, these indications include scapholunate ligament repair, scapholunate reduction, lunotriquetral ligament repair, lunotriquetral reduction, and carpal instability. Prescription Use ﮯ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) for M.Melleem (Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K111608
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...