WONDFO AMPHETAMINE URINE TEST, WONDFO BARBITURATES URINE TEST, WONDFO BENZODIAZEPINES URINE TEST

K111560 · Guangzhou Wondfo Biotech Co., Ltd. · DKZ · Jul 15, 2011 · Clinical Toxicology

Device Facts

Record IDK111560
Device NameWONDFO AMPHETAMINE URINE TEST, WONDFO BARBITURATES URINE TEST, WONDFO BENZODIAZEPINES URINE TEST
ApplicantGuangzhou Wondfo Biotech Co., Ltd.
Product CodeDKZ · Clinical Toxicology
Decision DateJul 15, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3100
Device ClassClass 2

Intended Use

Wondfo Amphetamine Urine Test: Wondfo Amphetamine Urine Test is an immunochromatographic assay for the qualitative determination of d-amphetamine in human urine at a cutoff concentration of 1000 ng/mL. The test is available in a cassette format and a strip format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. Wondfo Secobarbital Urine Test: Wondfo Secobarbital Urine Test is an immunochromatographic assay for the qualitative determination of secobarbital in human urine at a cutoff concentration of 300 ng/mL. The test is available in a cassette format and a strip format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. Wondfo Oxazepam Urine Test: Wondfo Oxazepam Urine Test is an immunochromatographic assay for the qualitative determination of oxazepam in human urine at a cutoff concentration of 300 ng/mL. The test is available in a cassette format and a strip format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Device Story

Lateral flow immunochromatographic assay for qualitative detection of Amphetamine, Secobarbital, and Oxazepam in human urine; utilizes monoclonal antibody-dye conjugate against drug, gold chloride, fixed drug-protein conjugate, and anti-mouse IgG polyclonal antibody on membrane. Available in strip or cassette formats. Used by healthcare professionals or patients (OTC) for preliminary drug screening. Results interpreted visually; positive results require confirmatory testing via GC/MS. Assists in clinical decision-making regarding drug use status.

Clinical Evidence

No clinical diagnostic data. Performance established via analytical bench testing and a lay-user study. Lay-user study (n=140 per analyte/format) compared device results to GC/MS across various drug concentrations. Overall agreement with GC/MS ranged from 96.4% to 97.8%. Precision studies performed across multiple lots and operators. Specificity and cross-reactivity evaluated against numerous compounds.

Technological Characteristics

Lateral flow immunochromatographic assay; competitive binding principle. Components include monoclonal antibody-dye conjugate, gold chloride, fixed drug-protein conjugate, and anti-mouse IgG polyclonal antibody. Form factors: strip or cassette. No electronic components, software, or external energy source required.

Indications for Use

Indicated for qualitative detection of d-amphetamine (1000 ng/mL), secobarbital (300 ng/mL), or oxazepam (300 ng/mL) in human urine. Intended for prescription and over-the-counter use as a preliminary screening test.

Regulatory Classification

Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | TABLE 4 | SUMMARY | |-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------| | 1. Date the summary was prepared: | May 03, 2011 | | 2. Submitter's name: | Guangzhou Wondfo Biotech Co., Ltd. | | Address: | South China University of Technology<br>Guangzhou, P.R. China 510641 | | Phone: | 012-86-20-32296069 | | Name of contact person: | Joe Shia<br>LSI International Inc.<br>504 East Diamond Ave.,<br>Suite F Gaithersburg, MD 20878<br>Telephone: 240-505-7880<br>Fax: 301-916-6231 | | 3. Name of the device | | | Common or usual name: | Amphetamine Urine Test<br>Secobarbital Urine Test<br>Oxazepam Urine Test | | Trade or proprietary name: | Wondfo Amphetamine Urine Test | Classification: All are Class II medical devices with the following various product codes with Code of Federal Regulation references: Wondfo Secobarbital Urine Test Wondfo Oxazepam Urine Test | Product Code | CFR # | |--------------|----------------| | DKZ | 21CFR 862.3100 | | DIS | 21CFR 862.3150 | | JXM | 21CFR 862.3170 | - 4. The legally marketed device to which we are claiming equivalence [807.92(a)(3)]: Acon Laboratories, Inc. One Step Drug Screen Test Card, K020771. - 5. Description of the device: Assay Principle: Immunochromatograph assay for Amphetamine, Secobarbital, and Oxazepam Urine Test using a lateral flow, one step system for the qualitative detection of Amphetamine, Secobarbital, and Oxazepam in human urine. Each assay uses a monoclonal antibody-dve conjuqate from mouse aqainst drug with gold chloride and fixed drug-protein conjugate and anti-mouse IgG polyclonal antibody in membrane. - 6. Intended use of the device: Wondfo Amphetamine Urine Test, Wondfo Secobarbital Urine Test, and Wondfo Oxazepam Urine Test are intended for the qualitative determination of Amphetamine, Secobarbital, Oxazepam at the specific cut-off concentration in human urine. They are intended for healthcare professional use and over the counter use. 7. Comparison to the predicate device A summary comparison of the features of the Wondfo Amphetamine Urine Test, Wondfo Secobarbital Urine Test, Wondfo Oxazepam Urine Test and the predicate devices is provided in the Table 1. {1}------------------------------------------------ | Item | Device | Predicate | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------| | Indication(s) for<br>use | For the qualitative determination of<br>Amphetamine, Secobarbital, Oxazepam<br>individual in human urine. | Same (but the number<br>of drugs detected<br>different) | | Methodology | Competitive binding, lateral flow<br>immunochromatographic assays based<br>on the principle of antigen antibody<br>immunochemistry. | Same | | Type Of Test | Immunoassay principles that rely on<br>antigen-antibody interactions to<br>indicate positive or negative result | Same | | Results | Qualitative | Same | | Specimen Type | Human urine | Same | | Cut Off Values | Amphetamine: 1000ng/ml<br>Secobarbital: 300 ng/ml<br>Oxazepam: 300ng/ml | Same (but the number<br>of drugs detected<br>different) | | Configurations | Strip, cassette | Card | | Intended Use | OTC Use & Prescription Use | Prescription Use | | Table 1: Features comparison of Wondfo assays and the predicate devices | | |-------------------------------------------------------------------------|--| |-------------------------------------------------------------------------|--| The Wondfo Amphetamine Urine Test, Wondfo Secobarbital Urine Test, and Wondfo Oxazepam Urine Test have similar technological characteristics and performance to the predicate and are equivalent. . {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. To the right of the text is a stylized symbol that resembles a bird or a series of flowing lines. The symbol is black and appears to be in motion. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 Guangzhou Wondfo Biotech Co., LTD. c/o Joe Shia, LSI International Inc. 504 East Diamond Ave. Suite F Gaithersburg, MD 20877 k111560 Re: Trade Name: Wondfo Amphetamine Urine Test, Wondfo Secobarbital Urine Test, and Wondfo Oxazepam Urine Test Regulation Number: 21 CFR §862.3100 Regulation Name: Amphetamine Test System Regulatory Class: Class II Product Codes: DKZ, DIX, and JXM Dated: July6, 2011 Received: July 8, 2011 Dear Mr. Shia: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollefree no mber (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, C.C. Courtney Harper, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health : تركة : Enclosure {4}------------------------------------------------ ## Indications for Use Form 510(k) Number (if known): _ K111560 Device Name: _Wondfo Amphetamine Urine Test Indications for Use: Wondfo Amphetamine Urine Test is an immunochromatographic assay for the qualitative determination of d-Amphetamine in human urine at a cutoff concentration of 1000ng/mL. The test is available in a cassette format and a strip format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must be to obtain a conformed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. Prescription Use __ X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use_ X (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Size Off Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) 111560 Page 1 of 3 {5}------------------------------------------------ ## Indications for Use Form 510(k) Number (if known): _ K111560 Device Name: Wondfo Secobarbital Urine Test Indications for Use: Wondfo Secobarbital Urine Test is an immunochromatographic assay for the qualitalive determination of Secobarbital (major metabolite of Barbiturates) in human urine at a cutoff concentration of 300ng/mL. The test is available in a cassette format and a strip format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a conformed analytical result. GCMS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. Prescription Use __ × (21 CFR Part 801 Subpart D) And/Or Over the Counter Use_ × (21 CFR Part 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign Off Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K111560 Page 2 of 3 {6}------------------------------------------------ ## Indications for Use Form 510(k) Number (if known): _ K111560 Device Name: Wondfo Oxazepam Urine Test Indications for Use: Wondfo Oxazepam Urine Test is an immunochromatographic assay for the qualitative determination of Oxazepam (major metabolite of Benzodiazepines) in human urine at a cutoff concentration of 300ng/mL. The test is available in a cassette format and a strip format. It is intended for prescription use and over the counter use. The test provides only preliminary test results. A more specific alternative chemical method must to obtain a conformed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (21 CFR Part 801 Subpart D) And/Or Over the Counter Use_ X (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign Off Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K111360 Page 3 of_ 3
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...