K111513 · Philips Ultrasound, Inc. · IYN · Jun 24, 2011 · Radiology
Device Facts
Record ID
K111513
Device Name
CX50 DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
Philips Ultrasound, Inc.
Product Code
IYN · Radiology
Decision Date
Jun 24, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The CX50 Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (or 2-D), M-mode (including Anatomical --mode), Pulse Wave Doppler, continuous Wave Doppler, color Doppler, tissue Doppler Imaging and Harmonics (Tissue and contrast) modes. The device is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Ophthalmic, Fetal, Abdominal, Pediatric, Small Organ, Adult Cephalic, Neonatal Cephalic, Trans-vaginal, Musculo-skeletal, Gynecological, Cardiac Adult, Cardiac pediatric, Trans-Esoph. (Cardiac), Peripheral Vessel, Other (Carotid)
Device Story
The CX50 is a portable, laptop-style, 128-channel diagnostic ultrasound system; AC or battery-powered. It processes ultrasonic signals to generate B-mode, M-mode, Doppler (Pulse Wave, Continuous Wave, Color, Tissue), and Harmonic imaging. Used in point-of-care, clinical, and hospital settings by clinicians for diagnostic imaging and fluid flow analysis. The system utilizes custom digital electronics to transform transducer inputs into diagnostic images/data. Output is displayed for clinician review to support clinical decision-making. This 510(k) adds Cardiac Pediatric and Neonatal Cephalic indications to an existing platform; no new technical features introduced.
Clinical Evidence
No clinical data was generated for this submission. The safety and effectiveness of the CX50 system were established in previous submissions (K091804, K081802). The added Cardiac Pediatric and Neonatal Cephalic indications are well-accepted for ultrasound systems, including the predicate HD11 (K043535).
Technological Characteristics
Portable, laptop-style 128-channel ultrasound system. Employs pulsed echo and pulsed Doppler imaging. Supports various transducers (e.g., S5-1, L12-3, C9-3v, C5-1, D5cwc, D2cwc, X7-2t). Operates under Track 3 FDA limits. Connectivity includes standard ultrasound imaging modes (B, M, Doppler, Harmonics, Tissue Doppler).
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients requiring Ophthalmic, Fetal, Abdominal, Pediatric, Small Organ, Adult/Neonatal Cephalic, Trans-vaginal, Musculo-skeletal, Gynecological, Cardiac (Adult/Pediatric/Trans-Esophageal), and Peripheral Vessel/Carotid examinations.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
K111513
### 510(k) Premarket Notification CX50 Diagnostic Ultrasound System with Additional Indications
JUN 2 4 2011
This summary of safety and effectiveness is provided as part of the Premarket Notification in compliance with 21CFR. Part 807, Subpart E, Section 807.92
1) Submitter's name, address, telephone number, contact person
Penny Greco Philips Ultrasound, Inc. Regulatory Affairs Specialist 3000 Minuteman Road Andover, MA 01810-6302 Tel: (978) 659-4615 Fax (978) 975-7324
Date prepared: May 23, 2011
2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
Common/Usual Name: Diagnostic ultrasound system and transducers
Proprietary Name: CX50 Diagnostic Ultrasound System
Classification Name: Class II
| 21 CFR<br>Section | Classification Name | Product Code |
|-------------------|------------------------------------------------|--------------|
| 892.1550 | System, Imaging, Pulsed Doppler,<br>Ultrasonic | 90 IYN |
| 892.1560 | System, Imaging, Pulsed Echo, Ultrasonic | 90 IYO |
| 892.1570 | Transducer, Ultrasonic, diagnostic | 90 ITX |
3) Substantially Equivalent Devices
| CX50 Diagnostic Ultrasound System | K091804 / K081802 |
|-----------------------------------|-------------------|
| HD11 Diagnostic Ultrasound System | K043535 |
#### 3) Device Description
The CX50 Diagnostic Ultrasound System is a compact, AC or battery powered, 128 channel, cardiac ultrasound imaging. It uses custom digital electronic and fabrication technologies to provide diagnostic ultrasound information and is housed in a portable, laptop-style chassis. The only changes made in this CX50 510(k) are the additional indications of Cardiac Pediatric and Neonatal Cephalic. There are no new or unique features/technical characteristics introduced with the addition of the new indications.
Page 6 of 33
{2}------------------------------------------------
Cardiac Pediatric and Neonatal Cephalic indications have been previously cleared for other Philips Ultrasound systems, including the predicate device, HD11 Ultrasound system (K043535). This modification to the CX50 involves only the two new indications, as reflected in labeling.
#### 4) Intended Use
The CX50 Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (or 2-D), M-mode (including Anatomical --mode), Pulse Wave Doppler, continuous Wave Doppler, color Doppler, tissue Doppler Imaging and Harmonics (Tissue and contrast) modes. The device is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Ophthalmic, Fetal, Abdominal, Pediatric, Small Organ, Adult Cephalic, Neonatal Cephalic, Trans-vaginal, Musculo-skeletal, Gynecological, Cardiac Adult, Cardiac pediatric, Trans-Esoph. (Cardiac), Peripheral Vessel, Other (Carotid)
#### 5) Technological comparison to predicate devices
Philips CX50 and HD11 Diagnostic Ultrasound are Track 3 systems that employ the same fundamental scientific technology.
#### 6) Determination of Substantial Equivalence
### Non-clinical performance data
No new hazards were identified with the addition of cardiac pediatric and neonatal cephalic indications. No new testing was required to determine safety and efficacy of the CX50 with the new indications.
#### Summary of Clinical Tests
The CX50 required no modifications to support the neonatal cephalic and cardiac pediatric indications. The clinical safety and effectiveness of the system and transducers were identified in previous CX50 submissions (K091804 and K081802). The clinical safety and effectiveness of the Neonatal Cephalic and Cardiac Pediatric indications are well accepted for use with ultrasound systems including the predicate device, Philips HD11 (K043535).
{3}------------------------------------------------
### 7) Conclusions
CX50 with additional indications is substantially equivalent in safety and effectiveness to the predicate identified above:
- . The predicate devices and CX50 with additional indications are indicated for the diagnostic ultrasonic imaging and fluid flow analysis.
- The predicate devices and CX50 with additional indications have the same gray-scale and . Doppler capabilities.
- The predicate devices and CX50 with additional indications use essentially the same . technologies for imaging, Doppler functions and signal processing.
- The predicate devices and CX50 with additional indications have acoustic output levels . below the Track 3 FDA limits.
- . The predicate devices and CX50 with additional indications are manufactured under equivalent quality systems.
- The predicate devices and CX50 with additional indications are manufactured of . materials with equivalent bio safety. The materials have been evaluated and found to be safe for this application.
- . The predicate devices and CX50 with additional indications are designed and manufactured to the same electrical and physical safety standards.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of a caduceus, a symbol often associated with medicine and healthcare.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Philips Ultrasound, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO NY 55313
JUN 2 4 2011
Re: K111513
Trade/Device Name: Philips CX50 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: May 31, 2011 Received: June 1, 2011
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Philips CX50 Diagnostic Ultrasound System, as described in your premarket notification:
> Transducer Model Number ટર-I L12-3
C9-3v C2-1 D5-cwc D2cwc X7-2t
{5}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Joshua Nipper at (301) 796-6524.
Sincerely Yours,
Mary S Patil
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{6}------------------------------------------------
### Indications for Use
510(k) Number (if known): KIJ1513
Device Name: Philips CX50 Diagnostic Ultrasound System
Philips CX50 Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (or 2-D), M-mode (including Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Tissue Doppler Imaging and Harmonics (Tissue and Contrast) modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:
- Ophthalmic Fetal Abdominal Pediatric Small Organ Adult Cephalic Neonatal Cephalic Trans-vaginal Musculo-skeletal Gynecological Cardiac Adult Cardiac Pediatric Trans-Esoph. (Cardiac) Peripheral Vessel Other (Carotid)
The clinical environments where the CX50 Diagnostic Ultrasound System can be used include point-of-care areas in offices, clinical and hospital settings for diagnosis of patients.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K111513
Page 1 of 9
ﺳﺮﺓ ﺍﻟﺘﻮﺩﻳ
24 11:52 :
{7}------------------------------------------------
મારા ક
CX50 Diagnostic Ultrasound System Device name:
510(k) Number: _
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | |
|----------------------------------------------------|----------------------------------------------------------------------------|---|-------------------|-----|-----|-----------------------------------------------|------------------------------------|---------------------|
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>See below | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | P | P | P | | P | P | P (1,4,6,7) |
| | Fetal/Obstetric | P | P | P | P | P | P | P (1,3-8) |
| | Abdominal | P | P | P | P | P | P | P (1,3-9) |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | P | P | P | | P | P | P (1,3-8) |
| | Small Organ (thyroid,<br>scrotum, prostate, breast) | P | P | P | | P | P | P (1,3-8) |
| | Neonatal Cephalic | N | N | N | N | N | N | N(1-8) |
| | Adult Cephalic | P | P | P | P | P | P | P (1,3-7) |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | P | P | P | | P | P | P (1,3-8) |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Musculo-skel<br>(conventional) | P | P | P | | P | P | P (1,3-8) |
| | Musculo-skel (superficial) | P | P | P | | P | P | P (1,3-8) |
| | Other (Gynecological) | P | P | P | | P | P | P (1,3-9) |
| Cardiac | Cardiac Adult | P | P | P | P | P | P | P(1-4) |
| | Cardiac Pediatric | N | N | N | N | N | N | N (1-7) |
| | Trans-esoph. (Cardiac) | P | P | P | P | P | P | P(1-4) |
| | Other (Fetal) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | P | P | P (1,3-8) |
| | Other (Carotid) | P | P | P | | P | P | P (1,3-8) |
| | N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | |
| *Other modes:<br>1. Harmonics (Tissue or Contrast) | | | | | | 5. Angio Imaging<br>6. 3D (Freedhand) Imaging | | |
| *Other modes: | 5. Angio Imaging |
|---------------------------------------------------------------------------------|---------------------------------------------------|
| 1. Harmonics (Tissue or Contrast) | 6. 3D (Freedhand) Imaging |
| 2. Tissue Doppler Imaging | 7. SonoCT |
| 3. iSCAN | 8. Biopsy guidance |
| 4. X-Res | 9. Infertility monitoring of follicle development |
| Combined modes: B+PWD, B+Color, B+M, B+M+Color, B+Color+PWD, B+CWD, B+Color+CWD | |
| Previous submission: K091804 for 2.0 release of CS50 | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of In Vitro Diagnostics
Prescription Use (Per 21 CFR 801.109)
Page 2 of 9
{8}------------------------------------------------
1111513 510(k) Number: _
S5-1 transducer for use with CX50 Diagnostic Ultrasound System Device name:
Intended Use: Diagnostic ultrasound imaging or tluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|----------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------|---|-----|-----|------------------|------------------------------------|---------------------|
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>See below | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | P | P | P | | P | P | P<br>(1.4.6,7) |
| | Fetal/Obstetric | | | | | | | |
| | Abdominal | P | P | P | | P | P | P (1.3-8) |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | | | | | | | |
| | Neonatal Cephalic | N | N | N | N | N | N | N (1-8) |
| | Adult Cephalic | P | P | P | P | P | P | P (1,3-7) |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Musculo-skel (conventional) | | | | | | | |
| | Musculo-skel (superficial) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | P | P |
| Cardiac | Cardiac Pediatric | N | N | N | N | N | N | N (1-7) |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
| | N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | |
| *Other modes: | | | | | | 5. Angio Imaging | | |
| 1. Harmonics (Tissue & Contrast) | | 6. 3D (Freedhand) Imaging | | | | | | |
| 2. Tissue Doppler Imaging | | 7. SonoCT | | | | | | |
| 3. iSCAN | | 8. Biopsy guidance | | | | | | |
| 4. X-Res | | 9. Infertility monitoring of follicle development | | | | | | |
| | Combined modes: B+PWD, B+Color, B+M, B+M+Color, B+Color+PWD, B+CWD, B+Color+CWD | | | | | | | |
Previous submission: K091804 – use of S5-1 transducer with the 2.0 release of CX50
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of In Vitro Diagnostics
Prescription Use (Per 21 CFR 801.109)
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
| 510K | K111513 |
|------|---------|
|------|---------|
Page 3 of 9
{9}------------------------------------------------
1111513 510(k) Number:
L12-3 transducer for use with CX50 Diagnostic Ultrasound System Device name:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|----------------------------------------------------------------------------|---------------------------|--------------------|-----------|-----|------------------|------------------------------------|---------------------|--|--|
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>See below | Other*<br>(Specify) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal/Obstetric | | | | | | | | | |
| | Abdominal | P | P | P | | P | P | P (1,3-8) | | |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | P | P | P | | P | P | P (1,3-8) | | |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | P | P | P | | P | P | P (1,3-8) | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Intra-luminal | | | | | | | | | |
| | Musculo-skel (conventional) | P | P | P | | P | P | P (1,3-8) | | |
| | Musculo-skel (superficial) | P | P | P | | P | P | P (1,3-8) | | |
| | Other (Gynecological) | | | | | | | | | |
| | Cardiac Adult | | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Other (Fetal) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | P | P (1,3-8) | | |
| | Other (Carotid) | P | P | P | | P | P | P (1,3-8) | | |
| | N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | |
| *Other modes: | | | 5. Angio Imaging | | | | | | | |
| 1. Harmonics (Tissue & Contrast) | | 6. 3D (Freedhand) Imaging | | | | | | | | |
| | 2. Tissue Doppler Imaging | | | 7. SonoCT | | | | | | |
| 3. iSCAN | | | 8. Biopsy guidance | | | | | | | |
Previous submission: K091804- use of L12-3 transducer with the 2.0 release of CX50
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of In Vitro Diagnostics
Combined modes: B+PWD, B+Color, B+M+Color, B+Color+PWD, B+CWD, B+CWD, B+Color+CWD
Prescription Use (Per 21 CFR 801.109)
4. X-Res ·
| (Division Sign-Off) | |
|------------------------------------------------------------|---------|
| Division of Radiological Devices | |
| Office of In Vitro Diagnostic Device Evaluation and Safety | |
| 510K | K111513 |
9. Infertility monitoring of follicle development
Page 4 of 9
{10}------------------------------------------------
K111513 510(k) Number: _
C9-3v transducer for use with CX50 Diagnostic Ultrasound System Device name:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | | Mode of Operation | | | | | | | |
|------------------------------|-----------------------------------------------------|---|---|-------------------|-----|------------------|------------------------------------|---------------------|--|--|--|
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>See below | Other*<br>(Specify) | | | |
| Ophthalmic | Ophthalmic | | | | | | | | | | |
| | Fetal/Obstetric | P | P | P | | P | P | P (1,3-8) | | | |
| | Abdominal | P | P | P | | P | P | P (1,3-9) | | | |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | | |
| | Laparoscopic | | | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | | | | |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | | |
| | Trans-rectal | | | | | | | | | | |
| | Trans-vaginal | P | P | P | | P | P | P (1,3-9) | | | |
| | Trans-urethral | | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | | |
| | Intra-luminal | | | | | | | | | | |
| | Musculo-skel (conventional) | | | | | | | | | | |
| | Musculo-skel (superficial) | | | | | | | | | | |
| | Other (Gynecological) | P | P | P | | P | P | P (1,3-9) | | | |
| | Cardiac Adult | | | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | | |
| | Other (Fetal) | | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | |
| | Other (Specify) | | | | | | | | | | |
Previous submission: K091804 - use of C9-3v transducer with the 2.0 release of CX50
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of In Vitro Diagnostics
Prescription Use (Per 21 CFR 801.109)
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
| 510K | K111513 |
|------|---------|
|------|---------|
{11}------------------------------------------------
111513 510(k) Number:
Device name: C5-1 transducer for use with CX50 Diagnostic Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------|---------------------------|-----|-----|------------------|---------------------------------------------------|---------------------|--|
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>See below | Other*<br>(Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal/Obstetric | P | P | P | | P | P | P (1,3-8) | |
| | Abdominal | P | P | P | | P | P | P (1,3-9) | |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | P | P | P | | P | P | P (1,3-9) | |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Musculo-skel (conventional) | | | | | | | | |
| | Musculo-skel (superficial) | P | P | P | | P | P | P (1,3-8) | |
| | Other (Gynecological) | P | P | P | | P | P | P (1,3-9) | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| Peripheral<br>Vessel | Other (Fetal) | | | | | | | | |
| | Peripheral vessel | P | P | P | | P | P | P (1,3-8) | |
| | Other (Specify) | | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | |
| *Other modes: | | | | | | 5. Angio Imaging | | | |
| 1. Harmonics (Tissue & Contrast) | | | 6. 3D (Freedhand) Imaging | | | | | | |
| | 2. Tissue Doppler Imaging | | 7. SonoCT | | | | | | |
| | 3. ISCAN | | 8. Biopsy guidance | | | | | | |
| | | | | | | | 9. Infertility monitoring of follicle development | | |
| | Combined modes: B+PWD, B+Color, B+M, B+M+Color, B+Color+PWD, B+CWD, B+Color+CWD | | | | | | | | |
| | Previous submission: K091804- use of C5-1 transducer with the 2.0 release of CX50 | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of In Vitro Diagnostics
Prescription Use (Per 21 CFR 801.109)
| (Division Sign-Off) | |
|------------------------------------------------------------|---------|
| Division of Radiological Devices | |
| Office of In Vitro Diagnostic Device Evaluation and Safety | |
| 510K | K111513 |
Page 6 of 9
{12}------------------------------------------------
1111513 510(k) Number: J
D5cwc transducer for use with CX50 Diagnostic Ultrasound System Device name:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | | |
|------------------------------|----------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|------------------------------------|---------------------|--|--|
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>See below | Other*<br>(Specify) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal/Obstetric | | | | | | | | | |
| | Abdominal | | | | | | | | | |
| | Intra-operative | | | | | | | | | |
| | (vascular/epicardial) | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Laparoscopic | | | | | | | | | |
| | Pediatric | | | | | | | | | |
| | Small Organ (thyroid, scrotum, | | | | | | | | | |
| | prostate, breast) | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Intra-luminal | | | | | | | | | |
| | Musculo-skel (conventional) | | | | | | | | | |
| | Musculo-skel (superficial) | | | | | | | | | |
| | Other (Gynecological) | | | | | | | | | |
| | Cardiac Adult | | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | N | | | N (1-7) | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Other (Fetal) | | | | | | | | | |
| Peripheral | Peripheral vessel | | | | P | | | | | |
| Vessel | Other (Carotid) | | | | P | | | | | |
| | N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | |
| *Other modes: | | | | | | | | | | |
| | 1. Harmonics (Tissue & Contrast) | | | | | | | | | |
| | 2. Tissue Doppler Imaging | | | | | | | | | |
| 3. iSCAN | | | | | | | | | | |
| 4. X-Res | | | | | | | | | | |
Combined modes: B+PWD, B+Color, B+M+Color, B+Color+PWD, B+CWD, B+Color+CWD Previous submission: K091804- use of D5cwc transducer with the 2.0 release of CX50
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of In Vitro Diagnostics
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
| 510K | K111513 |
|------|---------|
|------|---------|
{13}------------------------------------------------
| 510(k) Number: | K111573 |
|----------------|------------------------------------------------------------------------------------|
| Device name: | D2cwc transducer for use with CX50 Diagnostic Ultrasound System |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: |
| Clinical Application | | | Mode of Operation | | | | | |
|---------------------------|--------------------------------------------------|---|-------------------|-----|-----|---------------|--------------------|------------------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging & Other | Fetal/Obstetric | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (thyroid, scrotum, prostate, breast) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Musculo-skel (conventional) | | | | | | | |
| | Musculo-skel (superficial) | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| Cardiac | Cardiac Adult | | | | P | | | |
| | Cardiac Pediatric | | | | N | | | N (1-7) |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E |
|----------------------------------------------------------------------------|
| * Other modes: |
| Combined modes: |
Previous submission: K081802- use of D2cwc transducer with the first release of CX50
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of In Vitro Diagnostics
Prescription Use (Per 21 CFR 801.109)
| (Division Sign-Off) |
|------------------------------------------------------------|
| Division of Radiological Devices |
| Office of In Vitro Diagnostic Device Evaluation and Safety |
| 510K | K111513 |
|------|---------|
|------|---------|
Page 8 of 9
{14}------------------------------------------------
KII 213 510(k) Number:
Device name: X7-2t transducer for use with CX50 Diagnostic Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Property 1.7 Mode of Operation
| Clinical Application | | Mode of Operation | | | | | | | |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|--|
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal/Obstetric<br>Abdominal<br>Intra-operative<br>(vascular/epicardial)<br>Intra-operative (Neuro)<br>Laparoscopic<br>Pediatric<br>Small Organ (thyroid, scrotum,<br>prostate, breast)<br>Neonatal Cephalic<br>Adult Cephalic<br>Trans-rectal<br>Trans-vaginal<br>Trans-urethral<br>Trans-esoph. (non-Card.)<br>Intra-luminal<br>Musculo-skel (conventional)<br>Musculo-skel (superficial)<br>Other (Gynecological) | | | | | | | | |
| Cardiac | Cardiac Adult<br>Cardiac Pediatric<br>Trans-esoph. (Cardiac)<br>Other (Fetal) | P | P | P | P | P | P | P | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| | N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | |
* Other modes: Harmonics (Tissue & Contrast), Tissue Doppler Imaging
| Combined modes: | B+PWD, B+Color, B+M, B+M+Color, B+Color+PWD, B+CWD, B+Color+CWD |
|----------------------|-----------------------------------------------------------------|
| Previous submission: | K081802- use of X7-2t transducer with the first release of CX50 |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of In Vitro Diagnostics
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
| 510K | K111513 |
|------|---------|
|------|---------|
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.