SWEETSPOT DIABETES DATA MANAGEMENT SERVICE

K111509 · Sweetspot Diabetes Care, Inc. · NBW · Nov 9, 2011 · Clinical Chemistry

Device Facts

Record IDK111509
Device NameSWEETSPOT DIABETES DATA MANAGEMENT SERVICE
ApplicantSweetspot Diabetes Care, Inc.
Product CodeNBW · Clinical Chemistry
Decision DateNov 9, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesSoftware as a Medical Device, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K111509 · Nov 9, 2011SWEETSPOT DIABETES DATA MANAGEMENT SERVICESweetspot Diabetes Care, Inc.User evaluations in typical clinical settings; Patient surveys; Healthcare professional (HCP) feedbackThe device was evaluated in real-world clinical settings to assess usability, workflow impact, and the accuracy of the data retrieval (Fetch) process when performed by lay users in a clinical environment.Usability; Workflow; Clinical setting; Lay user evaluation

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Lay User Evaluation; Observational user evaluation in typical clinical settingsTypical outpatients; Sample Size: 88 respondents; Number of Sites: 2Not applicable for this studyEase of use, likelihood to bring meter to appointments, meter download success rate
Back-Office Kiosk HCP Evaluation; Observational user evaluation in a clinical settingHealthcare professionals (HCP); Sample Size: 11 respondents; Number of Sites: 1Not applicable for this studyWorkflow efficiency, ease of use, report consistency

Indications for Use

The SweetSpot Diabetes Data Management Service is intended for use in in clinical settings by both patients and healthcare professionals to assist people with diabetes and their healthcare professionals in the review, analysis and evaluation of historical blood glucose test results to support effective diabetes management. It is intended for use as an accessory to blood glucose meters with data interface capabilities.

Device Story

SweetSpot Diabetes Data Management Service is a web-based SaaS platform for diabetes management. It retrieves historical blood glucose data from various manufacturers' meters via a 'Fetch Utility' and device-specific cables. Input data is processed and stored on remote cloud servers. The system provides three access points: Front-Office Kiosk (patient-operated), Back-Office Kiosk (HCP-operated), and Back-Office Web Application (HCP-operated). The service consolidates data from multiple meters into a standardized report format. HCPs use these reports to monitor patient glucose levels, simplify clinical decision-making, and improve workflow efficiency. The system benefits patients by facilitating data sharing with clinicians and potentially increasing adherence to meter usage.

Clinical Evidence

No clinical trials performed. Evidence consists of bench testing to verify data transfer compatibility and accuracy between meters and the software, and a human factors study (n=21) to verify ease of use, label comprehension, and meter data transfer by lay users.

Technological Characteristics

Web-based SaaS platform utilizing cloud computing for data storage and processing. Interfaces with blood glucose meters via device-specific cables. Compatible with PC and Mac OS. Software-based data retrieval and report generation. Usability testing conducted per ISO 15197:2003.

Indications for Use

Indicated for patients with diabetes and healthcare professionals to assist in the review, analysis, and evaluation of historical blood glucose test results to support diabetes management. Used as an accessory to blood glucose meters with data interface capabilities.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE A. 510(k) Number: k111509 B. Purpose for Submission: New submission for an accessory data management software application for cleared glucose meters C. Manufacturer and Instrument Name: SweetSpot Diabetes Care, Inc. SweetSpot Diabetes Data Management Service D. Type of Test or Tests performed: Diabetes data management system E. System Descriptions: 1. Device Description: The SweetSpot Service allows patients or healthcare professionals to download data from blood glucose meters (BGM) and generates a report from the downloaded data, which is delivered to the healthcare professional for use in patient management. The Service is comprised of three different types of data retrieval stations, a Fetch Utility, a data processing and storage platform, report generation software, and an information delivery service. 2. Principles of Operation: The Service is Microsoft Windows and Apple OSX compatible, and is delivered as Software as a Service (SaaS) over the internet. The Service centrally manages data and activities through multiple, loosely-coupled units of functionality. The SweetSpot Service is comprised of 3 subsystems: Kiosk, Fetch Utility, and SweetSpot Platform. The Fetch Utility resides on a Kiosk and is updated from the Platform as needed. If the Back-Office Web Application is used, the Fetch Utility is called from the Platform for each instance of use (it does not reside on a standalone computer). The Platform resides on web. 3. Modes of Operation: Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device? Yes ☐ X or No ☐ {1} Page 2 of 4 Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? Yes ☐ or No ☑ If a BGM uses wireless or Bluetooth® communication, the Kiosk is capable of detecting these signals and downloading the BGM data using 802.11(g) or Bluetooth® 2.1 protocols. The SweetSpot Service operates in the following manner: - The patient or clinician “checks-in” (downloads) BGM data - The BGM data are sent to the SweetSpot Platform on the Web - The SweetSpot Platform performs various data processing steps on the BGM data - The SweetSpot Platform generates a report from the BGM data - The SweetSpot Service delivers the report to the clinician, and to the patient if requested 4. Specimen Identification: Meter controlled download to data manager 5. Specimen Sampling and Handling: Not Applicable 6. Calibration: Not Applicable 7. Quality Control: Not Applicable 8. Software: FDA has reviewed the applicant’s Hazard Analysis and software Documentation: Yes ☑ or No ☐ F. Regulatory Information: 1. Regulation Section: 21CFR §862.1345 - Glucose test system. 21CFR §862.2100 - Calculator/data processing module for clinical use. 2. Classification: Class II and I respectively 3. Product Code: NBW - System, Test, Blood Glucose, Over the Counter JQP - Calculator/Data Processing Module, for Clinical Use 4. Panel: Chemistry (75) G. Intended Use: 1. Indication(s) for Use: {2} Page 3 of 4 The SweetSpot Diabetes Data Management Service is intended for use in clinical settings by both patients and healthcare professionals, to assist in the review, analysis, and evaluation of blood glucose test results by the clinician to support effective diabetes management. It is intended for use as an accessory to blood glucose meters with data interface capabilities. 2. Special Condition for use Statement(s): Prescription Use H. Substantial Equivalence Information: 1. Predicate device name(s) and 510(k) numbers: MyCareTeam-Diabetes (MCT-Diabetes) data management software k073699 2. Comparison with Predicate Device: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Indication for use | Service is intended for use in clinical settings by both patients and healthcare professionals, to assist in the review, analysis, and evaluation of blood glucose test results by the clinician to support effective diabetes management. It is intended for use as an accessory to blood glucose meters with data interface capabilities. | same | | User | patients and healthcare professionals | same | | Data access | Accessed from a typical PC | same | | Interface | Icon-based user interfaces | same | | Meter | support commonly available blood glucose monitors | same | | Reports | Automated reports to aid in diabetes management | same | | Differences | | | | Item | Device | Predicate | | Data Upload | Patient or Clinician at the clinic | Patient from home | | Report | Reports are automatically sent to the clinician via a predefined workflow | reports to the clinician only after the patient has granted permission | | Computer | Windows or Macintosh | Windows | {3} Page 4 of 4 I. Standard/Guidance Document Referenced (if applicable): - ISO 14971:2007, Medical devices - Application of risk management to medical devices - ISO15197:2003 - In vitro diagnostic test systems - Requirements for blood - glucose monitoring systems for self-testing in managing diabetes mellitus J. Performance Characteristics: 1. Analytical Performance: a. Accuracy: Not Applicable b. Precision/Reproducibility: Not Applicable c. Linearity: Not Applicable d. Carryover: Not Applicable e. Interfering Substances: Not Applicable 2. Other Supportive Instrument Performance Data Not Covered Above: A Human Factors study of 21 participants was performed to verify ease of use, label comprehension, meter data transfer in the hands of lay users. Additionally, meters are bench tested to verify data transfer compatibly and accuracy. K. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. L. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...