Covidien, Formerly US Surgical A Divison of Tyco H
Product Code
GAM · General, Plastic Surgery
Decision Date
Jun 3, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
V-Loc™ 180 Absorbable Reload: The V-Loc™ 180 absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ 180 absorbable reloads are indicated for soft tissue approximation where use of an absorbable suture is appropriate. V-Loc™ PBT Non-Absorbable Reload: The V-Loc™ PBT non-absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ PBT non-absorbable reloads are indicated for soft tissue approximation.
Device Story
V-Loc 180 and V-Loc PBT reloads are synthetic suture components designed for use with Endo Stitch and SILS Stitch endoscopic suturing devices. These reloads feature a braided leader and ferrule modification to enable attachment to needles compatible with existing endoscopic suturing hardware. Used in endoscopic surgery by surgeons to place running stitches in soft tissue. The device facilitates tissue approximation; the V-Loc 180 provides absorbable support, while the V-Loc PBT provides non-absorbable support. The system maintains the functionality of the predicate suturing devices while allowing for the use of barbed suture technology in an endoscopic delivery format.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included USP Physical Tests <861> (Diameter) and <871> (Needle Attachment), plus bench and in vivo evaluations of insertion/removal forces, tensile strength, barbed suture holding strength, and tissue passage.
Technological Characteristics
Implantable synthetic sutures (absorbable poly(glycolide/l-lactide) or non-absorbable poly(ethylene terephthalate)). Features braided leader and ferrule for needle attachment. Evaluated per ISO 10993-1. Compatible with existing Endo Stitch/SILS Stitch endoscopic suturing devices. Sterilization method not specified.
Indications for Use
Indicated for soft tissue approximation in patients undergoing endoscopic surgery requiring running stitches; V-Loc 180 for cases where absorbable suture is appropriate; V-Loc PBT for general soft tissue approximation.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
V-Loc™ 180 Absorbable Wound Closure Device, Covidien (K091087 and K082662)
{1}------------------------------------------------
| 510(k) Summary of Safety and Effectiveness | |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER: | Covidien llc (formerly registered as Tyco Healthcare, LP)<br>60 Middletown Avenue<br>North Haven, CT 06473<br>Tel. No .: (203) 492-5000 |
| CONTACT PERSON: | Robert Zott<br>Program Director, Regulatory Affairs<br>60 Middletown Avenue<br>North Haven, CT 06473 USA<br>Phone: (203) 492-5493<br>Fax: (203) 492-6602 |
| DATE PREPARED: | May 17, 2011 |
| TRADE/PROPRIETARY NAME: | V-Loc™ 180 Absorbable Reload |
| | V-Loc™ PBT Non-Absorbable Reload |
| | Endo Stitch™ Endoscopic Suturing Device |
| | SILS™ Stitch Endoscopic Suturing Device |
| COMMON/USUAL NAME: | Synthetic Absorbable and Nonabsorbable Sutures |
| | Endoscopic Suturing Devices |
| CLASSIFICATION NAME: | V-Loc™ 180 Absorbable Reload:<br>Absorbable poly(glycolide/l-lactide) surgical suture<br>21 CFR 878.4493 |
| | V-Loc™ PBT Non-Absorbable Reload:<br>Nonabsorbable poly(ethylene terephthalate) surgical suture<br>21 CFR 878.5000 |
| | Endo Stitch™ Endoscopic Suturing Device<br>Endoscopic Tissue Approximation Device<br>21 CFR 876.1500 |
| | SILS™ Stitch Endoscopic Suturing Device<br>Endoscopic Tissue Approximation Device<br>21 CFR 876.1500 |
| PREDICATE DEVICE(S): | V-Loc™ 180 Absorbable Wound Closure Device,<br>Covidien (K091087 and K082662) |
| | V-Loc™ Non-Absorbable Wound Closure Device (PBT),<br>Covidien (K103052) |
| | Endo Stitch™ Endoscopic Suturing Device, Covidien<br>(K934738) |
| | SILS™ Stitch, Covidien (K090419) |
Addendum to Special 510(k) Premarket Notification for V-Loc™ 180 and V-Loc™ PBT Reloads - 06/01/2011
™ Trademark Page 1 of 26
.
{2}------------------------------------------------
| | USSC Non-absorbable Sutures, Covidien (K955747) |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DEVICE DESCRIPTION: | Absorbable and non-absorbable synthetic suture reloads for endoscopic suturing devices |
| INTENDED USE: | V-Loc™ 180 Absorbable Reload:<br>The V-Loc™ 180 absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ 180 absorbable reloads are indicated for soft tissue approximation where use of an absorbable suture is appropriate.<br><br>V-Loc™ PBT Non-Absorbable Reload:<br>The V-Loc™ PBT non-absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ PBT non-absorbable reloads are indicated for soft tissue approximation. |
| TECHNOLOGICAL CHARACTERISTICS: | The implantable portions of the V-Loc™ Reloads are identical to the predicate V-Loc™ Wound Closure Devices in terms of indications, labeling, materials, performance, and technology. The only differences are the manner in which they are attached to their respective needles and the lengths in which they are supplied. The modifications include the addition of a braided leader and ferrule in order to connect the V-Loc™ suture to a needle specifically designed to be captured and passed between the jaws of an endoscopic suturing device. The needles and materials used for the leader and ferrule have been previously-cleared in the cited predicate devices. The Endo Stitch™ and SILS Stitch™ Endoscopic Suturing Devices have not been changed. |
| MATERIALS: | All component materials of the V-Loc™ 180 Absorbable Reload and the V-Loc™ PBT Non-Absorbable Reload have been evaluated in accordance with ISO 10993-1: Biological Evaluation of Medical Devices, according to their intended use. These materials have also been previously cleared in predicate devices for the identical intended use. |
| PERFORMANCE DATA: | Performance testing has been performed in support of the intended use of these devices. This testing verifies that the V-Loc™ 180 Absorbable Reload and the V-Loc™ PBT Non-Absorbable Reload are substantially equivalent to the predicate devices. The following performance standards were used: |
{3}------------------------------------------------
- · USP Monograph for Sutures (33, NF 28, 2010) Physical Test <861> Diameter
- · USP Monograph for Sutures (33, NF 28, 2010) Physical Test <871> Needle Attachment
The subject devices were evaluated in the following bench top and in vivo tests:
- Needle Attachment
- · Insertion / Removal Forces through Port Cannula
- · Tensile Strength
- · Barbed Suture Holding Strength
- · Tissue Passage
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like strokes. The bird is positioned to the right of the text, which is arranged in a circular pattern around the bird. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Covidien % Mr. Robert Zott Program Director, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473
JUN - 3 2011
## Re: K111442
Trade/Device Name: V-Loc™ 180 Absorbable Reload, V-Loc™ PBT Non-Absorbable Reload Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: II Product Code: GAM, GAT, OCW Dated: May 17, 2011 Received: May 24, 2011
Dear Mr. Zott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
## Page 2 - Mr. Robert Zott
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.him for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
fn Dothan,
Oct.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known).
V-Loc™ 180 Absorbable Reload Device Name:
Indications for Use:
The V-Loc™ 180 absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ 180 absorbable reloads are indicated for soft tissue approximation where use of an absorbable suture is appropriate.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111442
{7}------------------------------------------------
Covidien IIc (formerly registered as Tyco Healthcare, LP)
## Indications for Use
510(k) Number (if known):
Device Name: V-Loc™ PBT Non-Absorbable Reload
Indications for Use:
The V-Loc™ PBT non-absorbable reloads are intended for use with the Endo Stitch™ suturing device and SILS™ Stitch suturing device. The Endo Stitch™ suturing device and SILS™ Stitch suturing device, when used with the reloads, have application in endoscopic surgery for the placement of running stitches in soft tissue. The V-Loc™ PBT non-absorbable reloads are indicated for soft tissue approximation.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Kuneber Mxm
rgical, Orthopedic, and Restorative Devices
510(k) Number K111442
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.