← Product Code [BTR](/productcode/BTR) · K111406

# BEVER REINFORCED ENDOTRACHEAL TUBE WITH CUFF (ORAL/NASAL), BEVER REINFORCED ENDOTRACHEAL TUBE WITHOUT CUFF (ORAL/NASAL) (K111406)

_Hangzhou Bever Medical Devices Co., Ltd. · BTR · Jul 19, 2012 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K111406

## Device Facts

- **Applicant:** Hangzhou Bever Medical Devices Co., Ltd.
- **Product Code:** [BTR](/productcode/BTR.md)
- **Decision Date:** Jul 19, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5730
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The BEVER™ Reinforced Endotracheal Tube is indicated for airway management by oral or nasal intubation of the trachea during anesthesia. The product may be used where the patient's neck is likely to be moved or the patient is in the prone position so that a non-reinforced tracheal tube might become kinked.

## Device Story

Reinforced endotracheal tube; provides kink-resistance via embedded stainless steel wire spiral; maintains lumen patency during neck movement or prone positioning. Available in cuffed (sizes 3.0-10.0mm) and uncuffed (sizes 2.0-7.0mm) variants; includes 15mm connector. Used by clinicians during anesthesia to facilitate breathing; cuffed version seals trachea to prevent gas leakage and aspiration. Single-use; sterile; ethylene oxide sterilized. Device ensures airway patency and ventilation stability in surgical contexts.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, genotoxicity, implantation). Sterilization validated per ISO 11135-1. Shelf life validated to 4 years via accelerated and real-time stability studies. Conforms to ISO 5361:1999(E).

## Technological Characteristics

Materials: PVC and stainless steel wire. Design: Spiral-reinforced tube with 15mm connector. Variants: Cuffed (with inflation system/valve/pilot balloon) and uncuffed. Sterilization: Ethylene oxide. Standards: ISO 5361:1999(E), ISO 10993 series, ISO 11135-1.

## Regulatory Identification

A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

## Predicate Devices

- Well Lead Reinforced Endotracheal Tube ([K073383](/device/K073383.md))
- Portex Reinforced Endotracheal Tube ([K032112](/device/K032112.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Hangzhou Bever dical Devices Co., Ltd.

Add: No. 8-1, Longquan Rd., Canggian Town, Yuhang District, Hangzhou, China Tel: +86-571-8861 6630 Fax: +86-0571-8861 6515

#### ve 111406 510(k) Summary

JUL 19 2012

## 1. Submitter (Owner) of 510 (k):

Hangzhou Bever Medical Devices Co., Ltd. No. 8-1, Longquan Rd., Cangqian Town, Yuhang District 311121, Hangzhou, China Tel: +86-571-8861 6630 Fax: +86-0571-8861 6515 Registration Number: 3008729910

## 2. Contact person:

Allyson Zhou Management Representative Tel: +86-571-8861 6630 Fax: +86-0571-8861 6515 Email: zlshio1224@sina.com

#### 3. Device Name:

Common Name: Reinforced Endotracheal Tube. Reinforced Tracheal Tube Trade Name: BEVER 114 Reinforced Endotracheal Tube Classification name: Tracheal tube(W/Wo Connector) (21 CFR 868.5730. Product Code BTR)

### 4. Predicate Device:

BEVER is claiming substantial equivalence to the following medical devices: Well Lead Reinforced Endotracheal Tube - K073383 Portex Reinforced Endotracheal Tube - K032112

#### 5. Device Description

BEVER™ Reinforced Endotracheal Tube with cuff (Oral/Nasal) is available in sizes 3.0mm~10.0 mm in 0.5 mm increments

BEVER™ Reinforced Endotracheal Tube without cuff (Oral/Nasal) is available in sizes 2.0mm~7.0 mm in 0.5 mm increments

BEVER " Reinforced Endotracheal tube is an Endotracheal tube with additional metal wire spiral reinforcement to provide kink-resistance. This type of product is typically used during operations where a high degree of flexibility is required from the tube, for instance prone position, head and neck surgery, and oral surgery. The plastic material and the spring allow the tube to be easily bent in all directions. The steel reinforcement maintains the patency of the lumen.

{1}------------------------------------------------

The BEVER™ Reinforced Endotracheal tube is sterile, single use device supplied with a standard 15 mm connector. It is available in cuffed and uncuffed variants and is for oral or nasal use. The cuffed tube is composed of main tube, cuff, inflating system (including inflating tube, valve and pilot balloon) and 15 mm connector. The uncuffed tube is composed of main tube and 15 mm connector. The cuff is intended to provide a seal against the trachea, ensuring that inspiratory and expiratory gasses are routed through the tube and not allowed to escape to the patient's upper airway, thus preventing loss of ventilation / anaesthetic and nebulised drugs, and reducing the likelihood of any aspirated stomach contents from entering the lungs. Uncuffed tubes are used mainly for paediatric patients or when patients require less protection from loss of ventilation / anaesthetic and nebulised drugs and or stomach aspiration.

#### 6. Indications for Use

The BEVER " Reinforced Endotracheal Tube is indicated for airway management by oral or nasal intubation of the trachea during anesthesia. The product may be used where the patient's neck is likely to be moved or the patient is in the prone position so that a non-reinforced tracheal tube might become kinked.

## 7. Substantial Equivalence

The BEVER™ Reinforced Endotracheal Tube maintains the same intended use as the predicate devices. It is a device inserted into the trachea through the mouth or nose to facilitate breathing and it may be used where the patient's neck is likely to be moved or flexed or the patient is in the prone position so that a non-reinforced tracheal tube might become kinked.

The BEVER™ Reinforced Endotracheal Tube is composed essentially of the same materials as the predicate devices. The material for both devices is PVC and stainless steel wire.

The BEVERTM Reinforced Endotracheal tube has the same design and performance characteristic as the predicate devices.

#### 8. Device Performance

The dimension, design, material, sterility, packaging and labeling of BEVER™ Reinforced Endotracheal tubes are conformed with ISO 5361:1999 (E).

## 9. Summary of Testing

Biocompatibility testing was performed based on ISO 10993 standards. The subject device passed the following biocompatibility testings:

- Cytotoxicity �
- Sensitization .

Section 2, Page 2 of 3 Pages

{2}------------------------------------------------

- . Irritation
- Genotoxicity .
- Implantation .

And the subject device is compliance with the following Biocompatibility standards:

- AAMI / ANSI / ISO 10993-5:2009 Biological evaluation of medical devices --. Part 5: Tests for In Vitro cytotoxicity
- AAMI / ANSI / ISO 10993-10: 2002/Amd. 1:2006(E) Biological evaluation of . medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity AMENDMENT 1
- AAMI / ANSI / ISO 10993-6:2007 Biological evaluation of medical devices --� Part 6: Tests for local effects after implantation
- AAMI / ANSI / ISO 10993-3:2009 Biological evaluation of medical devices -. Part 3: Tests for genotoxicity, carcinogenicity, and reproductive toxicity

## 10. Sterility and shelf life

The device is sterilized by ethylene oxide. The sterilization process has been validated to be compliance with

- AAMI / ANSI / ISO ISO11135-1: 2007 Sterilization of health care products . - Ethylene oxide - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
- AAMI / ANSI / ISO 10993-7: 2008 Biological Evaluation of Medical . Devices - Part 7: Ethylene Oxide Sterilization Residuals

Regarding the accelerated aging testing result, after 183 days accelerated aging, BEVER™ Reinforced Endotracheal Tubes are still compliance the requirements of device specification. The shelf life of device could be considered as 4 years. And, according to the real time stability study, the 4 years shelf life of device has been validated. So the shelf life of subject device is 4 years.

### 11. Conclusions

These proposed devices have the same intended use and technological characteristics to the currently-marketed predicate devices. No new issues of safety or effectiveness are introduced by using these devices. Therefore we believe the proposed devices are substantially equivalent to the currently-marketed predicate devices.

#### 12. Date of submission

January 13, 2012

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Allyson Zhou Management Representative Hangzhou Bever Medical Devices Company, Limited No. 8-1 Longquan Road Cangqian Town, Yuhang District Hangzhou, Zhejiang China 311121

JUL 1 9 2012

Re: K111406

Trade/Device Name: BEVER™ Reinforced Endotracheal Tube Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: BTR Dated: July 12, 2012 Received: July 12, 2012

Dear Ms. Zhou:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

#### Page 2- Ms. Zhou

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Rh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): Device Name: BEVER™ Reinforced Endotracheal Tube Indications For Use:

The BEVER™ Reinforced Endotracheal Tube is indicated for airway management by oral or nasal intubation of the trachea during anesthesia. The product may be used where the patient's neck is likely to be moved or the patient is in the prone position so that a non-reinforced tracheal tube might become kinked.

Over-The-Counter Use No AND/OR . Prescription Use _ Yes (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

y Schutts

(Division Sign-Off) Division of Anasthesiology, General Hospital Intection Control. Dental Devices

510(k) Number よう

Section 1, Page 1 of 1 Page

---

**Source:** [https://fda.innolitics.com/device/K111406](https://fda.innolitics.com/device/K111406)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
