LIFEBRIDGE

K111384 · Lifebridge Medizintechnik AG · KFM · May 24, 2011 · Cardiovascular

Device Facts

Record IDK111384
Device NameLIFEBRIDGE
ApplicantLifebridge Medizintechnik AG
Product CodeKFM · Cardiovascular
Decision DateMay 24, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4360
Device ClassClass 2
AttributesTherapeutic

Intended Use

The LIFEBRIDGE SYSTEM is indicated for use as an extracorporeal blood oxygenation system for patients needing short term, 6 hours or less, cardiac and/or pulmonary support.

Device Story

LIFEBRIDGE System is a compact, modular extracorporeal blood oxygenation system. Input: patient blood flow. Transformation: blood is processed through a circuit containing a centrifugal pump, oxygenator, and arterial filter; system monitors flow, pressure, level, and bubbles via integrated sensors. Output: oxygenated blood returned to patient. Used in clinical settings for short-term (≤ 6 hours) cardiac/pulmonary support. Operated by trained medical personnel. Control module with touch screen interface manages system parameters and provides real-time monitoring. Modification involves replacing the BioCor 200 oxygenator with the Medtronic Affinity NT oxygenator featuring Trillium Biopassive surface coating to improve biocompatibility.

Clinical Evidence

No clinical data. Performance testing was conducted to validate the modification and ensure equivalence to the original system.

Technological Characteristics

Modular extracorporeal circuit including centrifugal pump, hollow fiber membrane oxygenator (Medtronic Affinity NT with Trillium Biopassive surface coating), arterial filter, and reservoir. Integrated sensors for flow, pressure, level, and bubble detection. Control module with touch screen interface. System is pre-assembled.

Indications for Use

Indicated for patients requiring short-term (≤ 6 hours) cardiac and/or pulmonary support via extracorporeal blood oxygenation.

Regulatory Classification

Identification

A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.

Special Controls

*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability; (ii) The patient-contacting components of the device must be demonstrated to be biocompatible; (iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and (iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device. (b) *Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2) *Classification.* Class III (premarket approval).(c) *Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ **LIFEBRIDGE**® K11384 LIFEBRIDGE Affinity Special 510(k) MAY 2 4 2011 ## Special 510(k) Summary SUBMITTER: LIFEBRIDGE® Medizintechnik AG Simon-Ohm-Str 1 84539 Ampfing Germany CONTACT PERSON: Kathleen Johnson Medical Device Approvals, Inc. Phone: (641)472-0775 Fax: (641)472-0241 Mobile: (302) 521-9496 Class III per 21 CFR 870.4360 DATE PREPARED: May 3, 2011 DEVICE TRADE NAME: LIFEBRIDGE SYSTEM COMMON/USUAL NAME: Cardiopulmonary Support System CLASSIFICATION: PRODUCT CODE: ### CLASSIFICATION NAMES: Cardiopulmonary bypass heart/lung machine console (21 CFR 870.4220) Non-roller type cardiopulmonary bypass blood pump (21 CFR 870.4360) Cardiopulmonary bypass vascular catheter, cannulae, or tubing (21 CFR 870.4210) Cardiopulmonary bypass adaptor, stopcock, manifold, or fitting (21 CFR 870.4290) Cardiopulmonary bypass blood reservoir (21 CFR 870.4400) Cardiopulmonary bypass pump speed control (21 CFR 870.4380) Cardiopulmonary bypass oxygenator (21 CFR 870.4350) Cardiopulmonary bypass heat exchanger (21 CFR 870.4240) Cardiopulmonary bypass arterial line blood filter (21 CFR 870.4260) Cardiopulmonary bypass level sensing monitor and/or control (21 CFR 870.4340) Cardiopulmonary bypass bubble detector (21 CFR 870.4205) KFM # PREDICATE DEVICES: LIFEBRIDGE B2T System K090006 Avecor Affinity Hollow Fiber oxygenator with Trillium Biopassive Surface K973760 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "LIFEBRIDGE" in bold, black letters. Above the word is a curved line that resembles a bridge. The word is in all caps and has a registered trademark symbol on the upper right. ## DEVICE DESCRIPTION: The LIFEBRIDGE SYSTEM is a compact, pre-assembled, modular system consisting of: 1. Patient module housing an extracorporeal circuit comprised of several previously 510k-cleared devices. The circuit includes a rigid reservoir bag, a centrifugal pump, oxygenator, arterial filter, active air management system, tubing and connectors. 2. Sensors, including flow, pressure, level and bubble to read system parameters. 3. Control module that contains the electronics and user interface. 4. Base module that contains a touch screen, the main power connection and acts as a stable frame for the svstem. The modification to the current Lifebridge System is to exchange the current oxygenator, BioCor 200, used in the patient module, for the Medtronic Affinity NT oxygenator with Trillium Biopassive surface coating. #### INDICATIONS FOR USE: The LIFEBRIDGE SYSTEM is indicated for use as an extracorporeal blood oxygenation system for patients needing short term, 6 hours or less, cardiac and/or pulmonary support. #### STATEMENT OF TECHNICAL CHARACTERISTICS COMPARISON: This "Special 510(k)" is being submitted for a modification to the patient module of the Lifebridge System. The modification is to exchange one 510(k) cleared oxygenator for another. Both are hollow fiber membrane oxygenators cleared for short-term cardio-pulmonary support up to 6 hrs. The major differences between the oxygenators are availability and a surface coating. The Medtronic Affinity NT oxygenator provides a surface coating to the large surface area of the oxygenator and the BioCor oxygenator does not. The coating isTrillium Biopassive Surface coating. The modification has been implemented through adherence to Design Controls. Risks posed by the change have been identified and analyzed. Measures to reduce any possible risks have been identified. Performance testing has been carried out to validate the mitigations and to ensure that the change to the patient module is safe, does not create any new risks and that performance of the modified module is equivalent to that of the original. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing and tail. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Lifebridge Medizintechnik AG c/o Ms. Kathleen Johnson Medical Device Approvals, Inc. 1282 Round Hill Rd. Bryn Mawr, PA 19010 2 4 2011 NAY Re: K111384 LIFEBRIDGE System Regulation Number: 21 CFR 870.4360 Regulation Name: Non-roller type CPB Pump Regulatory Class: Class III (three) Product Code: KFM Dated: May 3, 2011 Received: May 17, 2011 Dear Ms. Kathleen Johnson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Kathleen Johnson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Lifebridge System Indications For Use: The Lifebridge System is indicated for use as an extracorporeal blood oxygenation system for patients needing short term, 6 hours or less, cardiac and / or pulmonary support. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of (Division Sign-Off) Divis Irdiovascular Devices 510(lk) Number
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%