K111369 · Helena Laboratories Uk, Ltd. · CFF · Jun 26, 2012 · Immunology
Device Facts
Record ID
K111369
Device Name
THE HELENA V8 IMMUNODISPLACEMENT KIT
Applicant
Helena Laboratories Uk, Ltd.
Product Code
CFF · Immunology
Decision Date
Jun 26, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5510
Device Class
Class 2
Indications for Use
The Helena V8 Immunodisplacement Kit is designed for the detection and the characterization of monoclonal proteins (immunoglobulin’s IgG, IgA, IgM, kappa (bound) and lambda (bound) light chains), in human serum with the Helena V8 Capillary Electrophoresis System. It is used in conjunction with the Helena V8 Serum Protein SPE Kit designed for serum protein separation into 6 major fractions in alkaline buffer. The electrophoretograms of separated proteins mixed with individual specific antisera are evaluated visually to detect the presence of specific reactions with the suspect monoclonal proteins. The test results are to be used in conjunction with clinical findings and other laboratory tests. For In Vitro Diagnostic Use Only.
Device Story
The V8 Immunodisplacement Kit is an in vitro diagnostic reagent system used with the Helena V8 Capillary Electrophoresis System. It detects and characterizes monoclonal proteins in human serum. The process involves mixing serum samples with specific antisera (IgG, IgA, IgM, kappa, lambda) following initial serum protein separation into six fractions via the V8 Serum Protein SPE Kit in an alkaline buffer. The resulting electrophoretograms are evaluated visually by laboratory personnel to identify specific antigen-antibody reactions indicative of monoclonal proteins. The output provides qualitative information regarding the presence and type of monoclonal proteins, which clinicians use alongside other diagnostic data to assess patients for monoclonal gammopathies.
Clinical Evidence
No clinical data provided; substantial equivalence is based on performance characteristics and technological similarities to existing immunological test systems.
Technological Characteristics
Immunological test system for capillary electrophoresis. Utilizes specific antisera (IgG, IgA, IgM, kappa, lambda) for immunodisplacement. Designed for use with the Helena V8 Capillary Electrophoresis System. Operates via electrophoretic separation of serum proteins in alkaline buffer followed by visual evaluation of reaction patterns.
Indications for Use
Indicated for the detection and characterization of monoclonal proteins (IgG, IgA, IgM, kappa/lambda light chains) in human serum. For prescription use in clinical laboratory settings.
Regulatory Classification
Identification
An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k111369
B. Purpose for Submission:
New Device
C. Measurand:
Monoclonal Immunoglobulins (IgG, IgA, IgM) and light chains (kappa, lambda) in serum
D. Type of Test:
Capillary Electrophoresis by Immunodisplacement, Qualitative
E. Applicant:
Helena EU Laboratories
F. Proprietary and Established Names:
Helena V8 Immunodisplacement Kit and Helena V8 Capillary Electrophoresis System
G. Regulatory Information:
1. Regulation section:
21 CFR §866.5510 Immunoglobulins A, G, M, D and E Immunological Test System
21 CFR §866.5550 Immunoglobulin (light chain specific) Immunological Test System
21 CFR §862.1630 Protein (fractionation) Test System
2. Classification:
Class II
3. Product code:
CFF – Immunoelectrophoretic, Immunoglobulins (G, A, M)
DFH – Kappa, Antigen, Antiserum, Control
DEH – Lambda, Antigen, Antiserum, Control
CEF – Electrophoretic, Protein Fractionation
{2}
4. Panel:
Immunology (82)
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
The Helena V8 Immunodisplacement Kit is designed for the detection and the characterization of monoclonal proteins (immunoglobulin’s IgG, IgA, IgM, kappa (bound) and lambda (bound) light chains), in human serum with the Helena V8 Capillary Electrophoresis System. It is used in conjunction with the Helena V8 Serum Protein SPE Kit designed for serum protein separation into 6 major fractions in alkaline buffer. The electrophoretograms of separated proteins mixed with individual specific antisera are evaluated visually to detect the presence of specific reactions with the suspect monoclonal proteins. The test results are to be used in conjunction with clinical findings and other laboratory tests.
For In Vitro Diagnostic Use Only.
2. Indication(s) for use:
Same as Intended use.
3. Special conditions for use statement(s):
For prescription only.
4. Special instrument requirements:
The Helena V8 Capillary Electrophoresis System
I. Device Description:
The Helena V8 Immunodisplacement Kit contains 5 antisera bottles (1x1.3 mLx50 tests) ready for use. The antisera bottles have monospecific antibodies against heavy chains (IgG, IgA, IgM); Kappa (bound) and Lambda (bound) light chains. The Helena V8 Immunodisplacement Kit is to be used on the Helena V8 Capillary Electrophoresis System.
J. Substantial Equivalence Information:
1. Predicate device name(s) and 510(k) number(s):
Helena SPIFE Kit, k973040
{3}
2. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | For detection and the characterization of monoclonal proteins | Same |
| Antisera Specificity | Antibody specificity to heavy chains (IgG, IgA, IgM) and to kappa (bound) and lambda (bound) light chains . | Same |
| Antisera Storage | 2 – 8°C | Same |
| Results Interpretation | Qualitative | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Immunoglobulin detection | IgG, IgA, IgM, kappa (bound) and lambda (bound) light chains | IgG, IgA, IgM, kappa: (bound & free) and lambda (bound & free) light chains |
| Sample Type | Serum | Serum, CSF and urine |
| Technology | Capillary Electrophoretic migration with Immunofixation by Immunodisplacement | Agarose Gel Electrophoretic Migration by Immunofixation |
| Instrument | Helena V8 Capillary Electrophoresis System | Helena SPIFE 2000 or SPIFE 3000 electrophoresis analyzer systems |
| Method of monoclonal protein identification | Removal of target protein | Isolation and fixation of target protein |
| Visualisation of target protein | UV absorbance scan on CE Electrophoretogram | Acid Violet Gel Staining |
| Lowest Detectable Limit | Serum: 0.25 g/L | Serum, CSF and Urine: Not available |
K. Standard/Guidance Document Referenced (if applicable):
None.
L. Test Principle:
The V8 Immunodisplacement Kit is intended for the separation and determination of monoclonal gammopathies by capillary zone electrophoresis. This technique separates
{4}
proteins on the basis of their net charge in an alkaline buffer solution in combination with their differing interaction with the wall of the silica capillary. Immunotyping of gammaglobulins is achieved by testing aliquots of sample with a panel of monospecific antisera. The complex formed by the test antisera and their target proteins has a modified migration profile and is therefore displaced from the standard serum protein electropherotogram. By comparing the results from the test panel with a reference analysis the immunoglobulin type present can be determined by the specific removal or reduction of the abnormal spike from the CE electrophoretogram.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Eight samples were used for reproducibility study, including one normal sample and seven pathological samples: monoclonal subtypes IgGκ (n=1), IgGλ (n=1), IgAκ (n=1), IgAλ (n=1), IgMκ (n=2), IgMλ (n=1). The total immunoglobulin levels were between 2.0 and 22.5 g/L. Each sample was analyzed in six replicates and the run was repeated with three different lots of antisera. According to the identified monoclonal protein characterization, replicates of each sample showed 100% concordant and reproducible results within each run and between different lots of antisera.
#### b. Linearity/assay reportable range:
Not applicable.
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Stability data from 3 lots support the assay shelf life claims of 24 months for unopened antisera vials and four weeks for opened vials at 2-8°C. Testings were performed at 0, 1, 2, 4, 6 week and at 12, 26, 28 months intervals.
#### d. Detection limit:
Six pathological samples comprising each monoclonal isotype IgGκ (51 g/L), IgGλ (37 g/L), IgAκ (69 g/L), IgAλ (61 g/L), IgMκ (42 g/L), IgMλ (69 g/L) were diluted with a normal sample until a sample was produced with a monoclonal peak of 0.75 g/L. These initial diluted samples were further serially diluted to produce a range of monoclonal concentrations of 0.45, 0.35, 0.25, 0.20 and 0.15 g/L. These dilutions were analyzed by the V8 System to show the decreasing appearance of the monoclonal peak. The data demonstrated the claimed detection limit of 0.25 g/L.
#### e. Analytical specificity:
Interference by endogenous and other substances: No interference was observed in
4
{5}
six pathological samples (IgGκ, IgGλ, IgAκ, IgAλ, IgMκ, IgMλ ranging from 1.3 g/L to 43.0 g/L monoclonal protein concentrations) spiked with hemoglobin up to 0.17 g/dL; 16.05 mg/L indirect bilirubin; 15.9 mg/dL direct bilirubin; triglycerides at 386, 669 or 1991 mg/dL; 317 IU/mL RF; and up to 16.28 mmol/L cholesterol.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
A total of 130 serum samples (108 pathological and 22 normal) were performed on Helena V8 Immunodisplacement (ID) kit using the Helena V8 System and Helena SPIFE/IFE Gel electrophoresis using Helena SPIFE System. The study was performed at three different sites and demonstrated 99% agreement between the two methods. The numbers of samples for different subtypes in the study are listed below.
| Qualitative Results | Total Number | Complete Agreement | Partial Agreement | Monoclonal Concentration Ranges (g/L) |
| --- | --- | --- | --- | --- |
| Normal | 22 | 22 | 0 | N/A |
| IgGκ | 36 | 36 | 0 | 0.35 – 4.78 |
| IgGλ | 23 | 23 | 0 | 0.47 – 4.39 |
| IgAκ | 9 | 8 | 1 | 0.27 – 4.09 |
| IgAλ | 11 | 11 | 0 | 0.25 – 4.09 |
| IgMκ | 18 | 18 | 0 | 0.30 – 3.10 |
| IgMλ | 5 | 5 | 0 | 0.80 – 2.80 |
| Biclonal | 2 | 2 | 0 | 0.40 – 0.70 |
| Oligoclonal | 1 | 1 | 0 | 1.50 |
| Other | 4 | 4 | 0 | - |
| Grand Total | 131 | 130 | 1 | N/A |
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
{6}
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Same as Expected values/ Reference range
5. Expected values/Reference range:
Absence of monoclonal immunoglobulins
N. Instrument Name:
The Helena V8 Capillary Electrophoresis System
O. System Descriptions:
1. Modes of Operation:
V8 is designed as a fully automated Clinical Capillary Electrophoresis System, providing automation from primary sample tube handling to analyte profiling and abnormal data detection. Sample identification, preparation, analysis, capillary cleaning, instrument maintenance and data transfer are fully automated. The V8 system can be integrated to all host networks for bi-directional interfacing. V8 is designed for professional laboratory conditions with minor installation and bench-space requirements. Operators load the V8 instrument with primary sample tubes and monitor the process. Trained clinicians and scientists diagnose the result for proper patient treatment.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☑ or No ☐
3. Specimen Identification:
The instrument is a fully automated 8 channel capillary electrophoresis instrument. The system can hold up to 112 blood samples or urine samples in bar-coded primary sample tubes. V8 employs a barcode scanner to detect the barcodes on the sample tubes and the racks containing the sample tubes.
6
{7}
4. Specimen Sampling and Handling:
The system can hold up to 112 blood samples or urine samples in bar-coded primary sample tubes and is able to identify, prepare, separate by CE, and transfer the sample data to a PC. The two main functions of the V8, sample preparation and sample separation, are conducted concurrently. Prepared samples are introduced to the CE where an 8-channel detector measures the samples using a UV-sensitive detector and sends results to the PC. The V8 also prepares samples and loads a sample tray to be used for gel electrophoresis.
5. Calibration:
All V8 instrument calibration should be carried out by a Helena Bioscience’s trained and certified engineer; this will be carried out during scheduled site service visits.
6. Quality Control:
It is recommended that Helena Bioscience’s control ranges (Catalogue number: 802400, 802500) are used to perform daily calibration check on the performance of the V8 instrument. All control kits are supplied with assay sheets which provide details of these normal ranges. It is recommended that each lab configures its own normal range.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.