← Product Code [JXN](/productcode/JXN) · K111363

# PPX REAGENT, PPX CALIBRATORS, PPX QC MATERIAL (K111363)

_Medica Corp. · JXN · Dec 6, 2011 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K111363

## Device Facts

- **Applicant:** Medica Corp.
- **Product Code:** [JXN](/productcode/JXN.md)
- **Decision Date:** Dec 6, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3700
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The EasyRA Propoxyphene (PPX) reagent is intended for the qualitative and semi-quantitative measurement of Propoxyphene in human urine, using MEDICA’s EasyRA Chemistry Analyzer in clinical laboratories. The cut-off point is 300ng/ml, and the assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/mass spectrometry (GC/MS) or Liquid Chromatography/mass spectrometry (LC/MS) are the preferred confirmatory method. The semi-quantitative mode is intended to enable laboratories to determine the appropriate dilution of the specimen for confirmation by a reference method such as GCMS or LC/MS and to allow laboratories to establish effective quality control procedures for the PPX assay. Medica’s Propoxyphene (PPX) Calibrators are intended for the calibration of the PPX Enzymatic Immunoassay to estimate propoxyphene in human urine, using Medica’s PPX reagent on the EasyRA clinical chemistry analyzer. Medica’s Propoxyphene (PPX) QC Materials are intended for the validation of the PPX Enzymatic Immunoassay to estimate propoxyphene in human urine, using Medica’s PPX reagent and PPX calibrator on the EasyRA clinical chemistry analyzer.

## Device Story

EasyRA Propoxyphene assay is an enzymatic immunoassay for human urine; utilizes competitive binding between drug in sample and drug-labeled glucose-6-phosphate dehydrogenase (G6PDH) for fixed amount of monoclonal anti-propoxyphene mouse antibody. Enzyme activity inversely proportional to drug concentration; measured at 340 nm. Used in clinical laboratories on Medica EasyRA Chemistry Analyzer. Provides qualitative (positive/negative at 300ng/ml cutoff) or semi-quantitative results. Semi-quantitative mode assists in determining specimen dilution for confirmatory testing (GC/MS or LC/MS) and establishing quality control. Results are preliminary; require confirmation by specific chemical methods. Benefits include rapid screening for propoxyphene presence to guide clinical toxicology workflows.

## Clinical Evidence

No clinical data. Bench testing only. Performance evaluated via precision (within-run/total CVs), linearity (50-800 ng/mL), and analytical specificity (endogenous substances and cross-reactivity). Method comparison against LC/MS showed 95% positive agreement and 100% negative agreement.

## Technological Characteristics

Liquid ready-to-use reagents. R1: monoclonal anti-propoxyphene mouse antibody, G6P, NAD. R2: Propoxyphene-labeled G6PDH. Enzymatic immunoassay principle. Wavelength 340 nm. Storage 2-8°C. Analyzed on Medica EasyRA clinical chemistry analyzer.

## Regulatory Identification

A propoxyphene test system is a device intended to measure propoxyphene, a pain-relieving drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of propoxyphene use or overdose or in monitoring levels of propoxyphene to ensure appropriate therapy.

## Special Controls

*Classification.* Class II (special controls). A propoxyphene test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K111363

B. Purpose for Submission:
New Device

C. Measurand:
Propoxyphene

D. Type of Test:
Qualitative and Semi-quantitative propoxyphene enzyme immunoassay

E. Applicant:
Medica Corp.

F. Proprietary and Established Names:
EasyRA Propoxyphene Reagent, EasyRA Propoxyphene Calibrators, EasyRA Propoxyphene QC Material

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  JXN | Class II | 21 CFR 862.3700
Propoxyphene test system | 91  |
|  DLJ | Class II | 21 CFR 862.3200 Clinical
toxicology calibrator | 91  |
|  LAS | Class I (reserved) | 21 CFR 862.3280 Clinical
toxicology control material | 91  |

H. Intended Use:

1. Intended use(s):

{1}

See indications for use below.

2. **Indication(s) for use:**

The EasyRA Propoxyphene (PPX) reagent is intended for the qualitative and semi-quantitative measurement of Propoxyphene in human urine, using MEDICA’s EasyRA Chemistry Analyzer in clinical laboratories. The cut-off point is 300ng/ml, and the assay provides only a preliminary analytical result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/mass spectrometry (GC/MS) or Liquid Chromatography/mass spectrometry (LC/MS) are the preferred confirmatory method.

**The semi-quantitative mode is intended to enable laboratories to determine the appropriate dilution of the specimen for confirmation by a reference method such as GCMS or LC/MS and to allow laboratories to establish effective quality control procedures for the PPX assay.**

Medica’s Propoxyphene (PPX) Calibrators are intended for the calibration of the PPX Enzymatic Immunoassay to estimate propoxyphene in human urine, using Medica’s PPX reagent on the EasyRA clinical chemistry analyzer.

Medica’s Propoxyphene (PPX) QC Materials are intended for the validation of the PPX Enzymatic Immunoassay to estimate propoxyphene in human urine, using Medica’s PPX reagent and PPX calibrator on the EasyRA clinical chemistry analyzer.

3. **Special conditions for use statement(s):**

For prescription use only

4. **Special instrument requirements:**

Medica EasyRA analyzer

I. **Device Description:**

The reagents are liquid ready to use. R1 contains a monoclonal anti-propoxyphene mouse antibody, glucose-6-phosphate (G6P), Nicotinamide adenine dinucleotide (NAD), stabilizers and preservatives. R2 contains Propoxyphene-labeled glucose-6-phosphate dehydrogenase (G6PDH), stabilizers and preservatives.

Controls and calibrators are human urine based, liquid ready to use, spiked with various concentrations of propoxyphene.

{2}

J. Substantial Equivalence Information:

1. Predicate device name(s):

Lin-Zhi International, Inc. Simultaneous Barbiturate-Methadone-Benzodiazepine-Propoxyphene (BMBP) Multiple Analyte Enzyme Immunoassay

Lin-Zhi International, Inc. Propoxyphene Calibrators

Lin-Zhi International, Inc. Propoxyphene Controls

2. Predicate 510(k) number(s):

k033885, k023316

3. Comparison with predicate:

|  Reagent Similarities and Differences  |   |   |
| --- | --- | --- |
|  Item | Submitted device: Easy RA PPX Reagent | Predicate Device: Propoxyphene reagent  |
|  Intended Use and indications for use | The Propoxyphene Enzyme Immunoassay is intended for the qualitative and semi-quantitative measurement of Propoxyphene, a pain-relieving drug, in human urine, using the EasyRA clinical chemistry analyzer. | Same  |
|  Sample type | Urine | Urine  |
|  Wavelength | 340 nm | 340 nm  |
|  Reaction type | Enzymatic Immunoassay | Enzymatic Immunoassay  |
|  Reagent storage | 2 – 8 °C | 2 – 8 °C  |
|  Test Methodology | The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. | The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity.  |
|  Reagent Similarities and Differences  |   |   |

{3}

|  Item | Submitted device: Easy RA PPX Calibrator and Control | Predicate Device:  |
| --- | --- | --- |
|  Intended Use | The propoxyphene Calibrators are intended for the calibration of the PPX Enzymatic Immunoassay to quantitate propoxyphene in human urine, using Medica's PPX reagent on the EasyRA chemistry analyzer. The Propoxyphene QC controls are intended for the validation of the PPX Enzymatic Immunoassay to determine Propoxyphene in human urine, using Medica's PPX reagent on the EasyRA chemistry analyzer. | Same  |
|  Matrix | Urine | Urine  |
|  Reagent storage | 2 – 8 °C | 2 – 8 °C  |

# K. Standard/Guidance Document Referenced (if applicable):

None were referenced.

# L. Test Principle:

The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of mouse antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity.

In the absence of drug in the sample, propoxyphene-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when free drug is present in the sample, antibody will bind to free drug, and the unbound propoxyphene-labeled G6PDH then exhibits its maximal enzyme activity.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

Within-run and total precision were performed by analyzing nine levels of urine samples spiked with propoxyphene to 75, 150, 225, 300, 375, 450, 525, and  $600\mathrm{ng / mL}$  on a Medica EasyRA analyzer. Full calibration with five level calibrators was performed to generate calibration curves for the semiquantitative samples. The same samples were tested qualitatively using a one level  $(300\mathrm{ng / mL})$  calibrator. Each sample was tested twice each morning and twice each afternoon for 20 days. The results are summarized below.

{4}

Qualitative

|  Target conc. (ng/ml) | 0 | 75 | 150 | 225 | 300 | 375 | 450 | 525 | 600  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Pos/Neg | 0/80 | 0/80 | 0/80 | 0/80 | 1/79 | 80/0 | 80/0 | 80/0 | 80/0  |

Semi-quantitative

|  Target conc. (ng/ml) | 0 | 75 | 150 | 225 | 300 | 375 | 450 | 525 | 600  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Pos/Neg | 0/80 | 0/80 | 0/80 | 0/80 | 20/60 | 80/0 | 80/0 | 80/0 | 80/0  |
|  Within run %CV | N/A | 5.10 | 2.44 | 1.38 | 1.65 | 1.45 | 1.69 | 1.65 | 2.60  |
|  Total %CV | N/A | 5.80 | 2.86 | 2.22 | 2.16 | 2.13 | 2.16 | 2.37 | 3.45  |

b. Linearity/assay reportable range:

Recovery was evaluated by testing nine urine samples spiked with propoxyphene to concentrations ranging from 50 to 800 ng/mL. Each sample was tested n=10 and the average value was calculated and compared to the actual value of the spiked sample. The results are summarized below.

|   | Sample # 1 | Sample # 2 | Sample # 3 | Sample # 4 | Sample # 5 | Sample # 6 | Sample # 7 | Sample # 8 | Sample # 9  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Target conc. | 50ng/ml | 100ng/ml | 150ng/ml | 200ng/ml | 300ng/ml | 400ng/ml | 500ng/ml | 600ng/ml | 800ng/ml  |
|  Mean | 56.3 | 98.0 | 143.8 | 210.1 | 324.6 | 429.7 | 522.8 | 633.0 | 868.2  |
|  SD | 5.6 | 3.6 | 4.3 | 3.5 | 4.4 | 13.7 | 16.8 | 28.2 | 63.2  |
|  CV | 10.0 | 3.7 | 3.0 | 1.7 | 1.3 | 3.2 | 3.2 | 4.5 | 7.3  |
|  % Recovery | 112.6% | 98.0% | 95.9% | 105.1% | 108.2% | 107.4% | 104.6% | 105.5% | 108.5%  |

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Calibrators and controls were previously cleared under k023316. Medica is re-labeling these products for use on the EasyRA analyzer and verifying the calibrator and control values of the previously cleared products on the EasyRA analyzer to ensure traceability of the calibrators and controls.

Accelerated and real time stability studies were performed and support Medica's labeled storage conditions and expiration dating.

d. Detection limit:

Performance at low drug concentrations in the semi-quantitative assay was

{5}

characterized by determination of recovery (see section b above).

# e. Analytical specificity:

Medica performed two interference studies for the PPX assay. The first coved interference from ten endogenous substances. The second investigated the potential cross-reactivity of thirty drugs with the PPX reagent.

# Interference from urine endogenous substances:

Ten endogenous substances, seven pH levels, and seven specific gravity levels were evaluated as potential interferants of the propoxyphene (PPX) semi-Quantitative Assay on the EasyRA analyzer. Medica evaluated the potential interference at urine propoxyphene levels of  $150\mathrm{ng / ml}$ ,  $300\mathrm{ng / ml}$ , and  $450\mathrm{ng / ml}$ . The pH and specific gravity interference was evaluated in urine samples containing  $225\mathrm{ng / ml}$ ,  $300\mathrm{ng / ml}$ , and  $375\mathrm{ng / ml}$  propoxyphene. The following substances did not demonstrate significant interferences at the concentrations shown.

|  Substance | Concentration Tested without significant interference  |
| --- | --- |
|  Ascorbic Acid | 2.5 mg/dl  |
|  Creatinine | 500 mg/dl  |
|  Glucose | 1200 mg/dl  |
|  Hemoglobin | 100 mg/dl  |
|  Albumin (HSA) | 300 mg/dl  |
|  Oxalic Acid | 0.5 mg/dl  |
|  Sodium Chloride | 2300 mg/dl  |
|  Acetone | 790 mg/dl  |
|  Ethanol | 790 mg/dl  |

The pH interference testing showed not significant interference due to pH ranging from 4.5 to 8.0.

The specific gravity interference testing showed not significant interference due to specific gravities ranging from 1.005 to 1.031.

# Interference from cross reactivity

Thirty substances were evaluated as potential interferants of the propoxyphene (PPX) assay on the EasyRA analyzer.

The following compounds had a cross reactivity of less than  $0.01\%$  when spiked into propoxyphene free urine and urine with propoxyphene levels of 225 and  $375~\mathrm{ng / mL}$ : Amitriptyline, Amphetamine, Bupropion, Ecgonine, Ephedrine, Codeine, Caffeine, Chlorpheniramine, Cocaine, Lidocaine,

{6}

Ranitidine, Meperidine, Valproic Acid, Methaqualonel, Nortriptyline, Phenobarbital, Imipramine, Promethazine, Acetaminophen, Chlorpromazine, Dextromethorphan, Acetylsalicylic acid, Benzoylecgonine, Methadone, Methamphetamine, Morphine, Oxazepam, PCP, Secobarbital.

Nor-propoxyphene cross reactivity is summarized below.

|  Substance | Nominal PPX value 0.0ng/ml  |   |   |
| --- | --- | --- | --- |
|   | True PPX (ng/ml) | Read PPX (ng/ml) | % Cross Reactivity  |
|  Norproxyphene 620 ng/ml | 0 | 279 | 45.06%  |

f. Assay cut-off:

Characterization of how the device performs analytically around the claimed cutoff concentration appears in the precision section, 1.a, above.

2. Comparison studies:

a. Method comparison with predicate device:

For the method comparison study, human urine samples with LC/MS propoxyphene and norpropoxyphene values were obtained from a reference laboratory. The samples were tested using the EasyRA propoxyphene assay in both qualitative and semi-quantitative modes of operation and compared to the GC/MS results. The results are summarized below.

Semi-quantitative mode

|   | Low Neg by LC/MS (less than -50%) | Near Cutoff Negative by LC/MS (Between -50% and cutoff) | Near Cutoff Positive by LC/MS (Between cutoff and +50%) | High Positive by LC/MS (greater than +50%) | Percent Agreement with LC/MS  |
| --- | --- | --- | --- | --- | --- |
|  Pos | 0 | 1 | 11 | 45 | 95%  |
|  Neg | 54 | 7 | 0 | 0 | 100%  |

Qualitative results

|   | Low Neg by LC/MS (less than -50%) | Near Cutoff Negative by LC/MS (Between -50% and | Near Cutoff Positive by LC/MS (Between cutoff and | High Positive by LC/MS (greater than +50%) | Percent Agreement with LC/MS  |
| --- | --- | --- | --- | --- | --- |
|   |  | cutoff) | +50%) |  |   |

{7}

8

|   |  | cutoff) | +50%) |  |   |
| --- | --- | --- | --- | --- | --- |
|  Pos | 0 | 1 | 11 | 45 | 95%  |
|  Neg | 54 | 7 | 0 | 0 | 100%  |

Discrepant Sample Results

|  Total LC/MS Value (ng/ml) | EasyRA Semi-Quantitative | EasyRA Qualitative  |
| --- | --- | --- |
|  220 | POS | POS  |

b. Matrix comparison:

Not applicable

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

Not applicable

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K111363](https://fda.innolitics.com/device/K111363)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
