PLANMED NUANCE EXCEL

K111361 · Planmeca USA, Inc. · MUE · Sep 23, 2011 · Radiology

Device Facts

Record IDK111361
Device NamePLANMED NUANCE EXCEL
ApplicantPlanmeca USA, Inc.
Product CodeMUE · Radiology
Decision DateSep 23, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1715
Device ClassClass 2

Indications for Use

Planmed Nuance Excel is a FFDM (Full Field Digital Mammography) X-ray unit, which is intended to be used for screening and diagnosis of breast cancer in the same clinical applications as traditional film/screen mammography X-ray units. The system generates digital mammograms which enable to detect breast cancer at an early stage. The use of Planmed FFDM X-ray unit is allowed only under supervision of a health care professional.

Device Story

Planmed Nuance Excel is a Full Field Digital Mammography (FFDM) system. It utilizes an amorphous selenium-based digital image receptor to convert X-ray photons directly into digital image data. The system is operated by healthcare professionals in clinical settings. Workflow is managed via an acquisition workstation running Planmed Nuance Manager 3 software. Patient data is entered manually or imported via HIS/RIS/MIS modality worklists. The device acquires, processes, and displays images for preview; images are subsequently printed or transferred for soft-copy review. By generating high-quality digital mammograms, the device assists clinicians in the early detection and diagnosis of breast cancer.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and design comparison to the predicate device.

Technological Characteristics

FFDM X-ray system; amorphous selenium-based digital image receptor for direct X-ray to digital conversion; acquisition workstation; Planmed Nuance Manager 3 software; connectivity via HIS/RIS/MIS modality worklists.

Indications for Use

Indicated for screening and diagnosis of breast cancer in patients requiring mammographic imaging. Use restricted to supervision by a healthcare professional.

Regulatory Classification

Identification

A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ **Planned** SEP 2 3 2011 ENCLOSURE 8 P. 8-1 #### KII1361 510K) SUMMARY DATE May 18, 2011 ## PRODUCT, CLASSIFICATION NAME Trade name: Planmed Nuance Excel Common name: Full Field Digital Mammography (FFDM) System Classification: MUE, Class II Regulation number: 21 CFR 892.1715 ## MANUFACTURER Planmed Ov Asentajankatu 6 FI-00880 Helsinki, Finland Phone: +358 20 7795 300 Fax: +358 20 7795 396 Contact person: Lars Moring ## UNITED STATES SALES REPRESENTATIVE (U.S. DESIGNATED AGENT) Planmed USA Inc. 100 North Gary Avenue, Suite A Roselle, IL 60172 Phone: (630) 894 2200 Fax: (630) 894 4271 Contact person : Bob Pienkowski #### INTENDED USE Planmed Nuance Excel is a FFDM (Full Field Digital Mammography) X-ray unit, which is intended to be used for screening and diagnosis of breast cancer in the same clinical applications as traditional film/screen mammography X-ray units. The system generates digital mammograms which enable to detect breast cancer at an early stage. The use of Planmed FFDM X-ray unit is allowed only under supervision of a health care professional. ## PRODUCT DESCRIPTION The Planmed Nuance Excel is a Full Field Digital Mammography (FFDM) system for generating mammographic images that can be used for screening and diagnosis of breast cancer. The Planmed Nuance utilizes an amorphous selenium based digital image receptor to capture digital images. The receptor directly converts the incoming X-ray photons to digital image data. Page 1 of 2 {2}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "Planned" in a bold, sans-serif font. The letters are black and stand out against a white background. The word is horizontally oriented and appears to be a title or heading. ENCLOSURE 8 P. 8-2 The workflow with Planmed Nuance Excel is controlled from the acquisition workstation and Planmed Nuance Manager 3 image acquisition and communications software. The patient information is entered manually or received from the hospital, radiology, or mammography information systems (HIS, RIS, or MIS, respectively), as a format of modality work list. Subsequently, the images are acquired, processed, and displayed for preview. After initial evaluation by the user, the images are either printed or transferred for soft-copy review. ### SUBSTANTIAL EQUIVALENCE We consider this product modification to be similar in design, composition and function to the following device introduced into commercial distribution after May 28, 1976: PMA # 030010 Siemens Mammomat Novation DR The comparison of characteristics supports substantial equivalence. Planmed Nuance Excel is as safe and effective as the predicate device. {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Planmed Ov % Mr. Bob Pienkowski Managing Director, US Agent Planmed USA, Inc. 100 North Gary Avenue, Suite A ROSELLE IL 60172 Re: K111361 Trade/Device Name: Planmed Nuance Excel Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: August 8, 2011 Received: August 10. 2011 SEP 2 3 2011 Dear Mr. Pienkowski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to hittp://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Mary Patel Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: Planmed Nuance Excel KIII361. Indications For Use: Planmed Nuance Excel is a FFDM (Full Field Digital Mammography) X-ray unit, which is intended to be used for screening and diagnosis of breast cancer in the same clinical applications as traditional film/screen mammography X-ray units. The system generates digital mammograms which enable to detect breast cancer at an early stage. The use of Planmed FFDM X-ray unit is allowed only under supervision of a health care professional. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Michael D'Onofrio (Division Sign Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 510K K111361 Page 1 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...